Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K974370

    Validate with FDA (Live)

    Device Name
    RAPET RF
    Manufacturer
    Date Cleared
    1998-02-17

    (89 days)

    Product Code
    Regulation Number
    866.5775
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    RaPET RF (Class II) is intended for the qualitative and semi-quantitative detection of rheumatoid factor in human serum. The latex slide test is intended to be used as an aid in the diagnosis of rheumatoid arthritis.

    Device Description

    The device test kit is comprised of RF Latex Reagent, RF Positive Control. Negative Control and Glycine/Saline Buffer for performing semi-quantitative testing.

    AI/ML Overview

    Here's an analysis of the provided information regarding the RaPET® RF device, focusing on its acceptance criteria and the study used to demonstrate it:

    Acceptance Criteria and Device Performance for RaPET® RF

    The RaPET® RF device is a latex slide test intended for the qualitative and semi-quantitative detection of rheumatoid factor in human serum, to be used as an aid in the diagnosis of rheumatoid arthritis.

    1. Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Method ComparisonCorrelation coefficient of 0.998 with a regression equation of Y=1.003X - 2.82 when compared to an unspecified predicate RF latex method.
    PrecisionResults found to be acceptable. (Specific thresholds not provided)
    SensitivityResults found to be acceptable. (Specific thresholds not provided)
    SpecificityResults found to be acceptable. (Specific thresholds not provided)
    InterferenceResults found to be acceptable. (Specific thresholds not provided)

    Note: The specific numerical acceptance criteria (e.g., minimum correlation coefficient, ranges for precision, sensitivity, and specificity) are not explicitly stated in the provided summary. The statement "Results of these tests were found to be acceptable" indicates that the device met internal or regulatory predefined criteria, but the criteria themselves are not detailed.

    2. Sample Size and Data Provenance for the Test Set

    • Sample Size: The sample size used for the method comparison and other studies (precision, sensitivity, specificity, interference) is not specified in the provided document.
    • Data Provenance: The document does not specify the country of origin of the data or whether the studies were retrospective or prospective.

    3. Number of Experts and Qualifications for Ground Truth Establishment (Test Set)

    • Number of Experts: The document does not specify the number of experts used to establish ground truth for the test set.
    • Qualifications of Experts: The document does not specify the qualifications of any experts involved in establishing ground truth. Since this is an in-vitro diagnostic device, ground truth often refers to a reference method or clinical diagnosis rather than expert interpretation of images. However, if any such interpretation was part of the study, it's not detailed.

    4. Adjudication Method for the Test Set

    The document does not specify any adjudication method. For in-vitro diagnostic devices like this, an adjudication method (like 2+1, 3+1) is typically associated with studies involving human interpretation or subjective assessments, which are not the primary focus of an RF latex test validation. The "ground truth" would likely be established by a reference method or clinical diagnosis itself.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    An MRMC comparative effectiveness study was not explicitly mentioned or performed as described for this device. This type of study focuses on how human readers' performance (e.g., diagnostic accuracy) is affected by AI assistance, which is not applicable to a manual latex agglutination test that does not involve "readers" in the same sense as an imaging AI.

    6. Standalone Performance Study

    A standalone study was performed in the sense that the device's performance characteristics (precision, sensitivity, specificity, interference) were evaluated as an algorithm/device only, without human-in-the-loop performance being a primary metric. The method comparison also evaluates the device's standalone output against a predicate device's output.

    7. Type of Ground Truth Used

    The primary ground truth for the method comparison appears to be the results obtained from a predicate RF latex method.
    For precision, sensitivity, specificity, and interference studies, the ground truth would typically be established based on:

    • Reference method/clinical diagnosis: For sensitivity and specificity, indicating the true presence or absence of rheumatoid factor/rheumatoid arthritis.
    • Known concentrations: For precision studies using controls with known analyte concentrations.
    • Known interferents: For interference studies, using samples spiked with known interfering substances.
      The document provides no further specifics on these ground truth methods.

    8. Sample Size for the Training Set

    • Training Set Sample Size: The document does not specify a training set sample size. For traditional in-vitro diagnostic kits like RaPET® RF, which are based on biochemical reactions (latex agglutination), there isn't typically a "training set" in the machine learning sense. The device itself is designed and optimized, and its performance is then validated.

    9. How Ground Truth for the Training Set Was Established

    As there is no explicit "training set" in the machine learning context for this device, a method for establishing its ground truth is not applicable or described in the provided document. The development of such a test relies on established biochemical principles and experimental validation, rather than algorithmic training on a dataset.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1