Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    DEN160003

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-07-22

    (185 days)

    Product Code
    Regulation Number
    866.6060
    Type
    Direct
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The QuantideX qPCR BCR-ABL IS Kit is an in vitro nucleic acid amplification test for the quantitation of BCR-ABL1 and ABL1 transcripts in total RNA from whole blood of diagnosed t(9,22) positive Chronic Myeloid Leukemia (CML) patients expressing BCR-ABL1 fusion transcripts type e13a2 and/or e14a2. The QuantideX qPCR BCR-ABL IS Kit is a reverse transcription-quantitative PCR performed on the Applied Biosystems 7500 Fast Dx Real-Time PCR Instrument and is intended to measure BCR-ABL1 to ABL1, expressed as a log molecular reduction (MR value) from a baseline of 100% on the International Scale, in t(9;22) positive CML patients during monitoring of treatment with Tyrosine Kinase Inhibitors (TKIs).

    The test does not differentiate between e13a2 or e14a2 fusion transcripts and does not monitor other rare fusion transcripts resulting from t(9:22). This test is not intended for the diagnosis of CML.

    Device Description

    The QuantideX qPCR BCR-ABL IS Kit reagents are adapted for use on the ABI 7500 Fast Dx Real-Time PCR Instrument. The assay includes reagents sufficient for 60 reactions. A description of the reagents provided is described below in Table 1.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document describes both analytical and clinical performance acceptance criteria.

    Analytical Performance - Precision/Reproducibility

    Performance MetricAcceptance Criteria (SD)Reported Device Performance (Max Total SD)Met?
    Within-lab Precision
    MR Value < 3.5≤ 0.210.128 (at MR 1)Yes
    MR Value 3.5-4.25≤ 0.290.111 (at MR 3.5)Yes
    MR Value > 4.25≤ 0.360.134 (at MR 4)Yes
    Site-to-Site Reproducibility
    MR Value < 3.5≤ 0.210.117 (at MR 3)Yes
    MR Value 3.5-4.25≤ 0.290.169 (at MR 3.5)Yes
    MR Value > 4.25≤ 0.360.167 (at MR 4)Yes

    Analytical Performance - Linearity

    Performance MetricAcceptance Criteria (Slope/Intercept for Regression)Reported Device Performance (Slope/Intercept/Max SD)Met?
    Linear RegressionSlope close to 1 (not explicitly defined but implied typical regression expectations), intercept close to 0 (implied)e13a2: Slope 1.01, Intercept -0.11, Max SD 0.17 e14a2: Slope 1.01, Intercept -0.05, Max SD 0.17Yes
    Range of Linearity(Implied based on clinical use range)MR 0.3 (50% IS) to MR 4.7 (0.002% IS) for both transcriptsYes

    Analytical Performance - Limit of Quantitation (LoQ)

    Performance MetricAcceptance Criteria (SD)Reported Device Performance (MR Range/SD Range)Met?
    LoQ≤ 0.36 at MR 4.5 or greaterMR values 4.6 to 4.87, SD values 0.23 to 0.34Yes

    Analytical Performance - Interfering Substances

    Performance MetricAcceptance Criteria (Mean difference from control)Reported Device PerformanceMet?
    InterferenceMean of test samples ± 0.5 MR of the controlAll met acceptance criteriaYes

    Analytical Performance - Primer Specificity

    Performance MetricAcceptance CriteriaReported Device PerformanceMet?
    SpecificitySpecimens without a major BCR-ABL1 breakpoint reported as negative or below LoD in > 8/9 replicatesMet acceptance criteria (e.g., cell lines and IVT controls without e13a2/e14a2 were <LOD in 8/9 or 9/9 replicates)Yes

    Analytical Performance - Specimen Carryover Contamination

    Performance MetricAcceptance CriteriaReported Device PerformanceMet?
    Carryover> 95% of CML-negative samples tested negative or below LoDAll negative specimens (25/25) reported as "Negative (sufficient ABL1)"Yes

    Analytical Performance - RNA Input

    Performance MetricAcceptance Criteria (SD criteria from Table 3)Reported Device Performance (RNA input range with met criteria)Met?
    RNA Input RangeSame as Table 3 for corresponding MR values750 to 6000 ng from MR 1 to MR 4.5Yes

    Analytical Performance - Traceability

    Performance MetricAcceptance Criteria (Regression to WHO panel)Reported Device Performance (Slope/Intercepts for 3 lots)Met?
    TraceabilitySlope 1.0 to 1.1, Intercept 0.02 - 0.11Lot1: y=-0.11+1.1x; Lot2: y=0.02+1x; Lot3: y=-0.059+1.1xYes

    Analytical Performance - Specimen Stability (Whole Blood)

    Performance MetricAcceptance Criteria (MR unit difference from baseline)Reported Device Performance (Duration and storage conditions)Met?
    Whole Blood StabilityAll results within 0.5 MR units from baselineUp to 72 hours at 2-8°CYes

    Clinical Performance

    Performance MetricAcceptance CriteriaReported Device PerformanceMet?
    Event-Free Survival (EFS) Difference at 36 monthsStatistically significantly different AND point estimates differing by at least 10 percentage points (for MR < 3 vs MR ≥ 3 groups)EFS difference: 22.2% (95% CI: 2.0%-42.4%), p = 0.0279Yes
    Event Status Known at 36 months95% Wilson score confidence interval for the difference > 0MetYes

    2. Sample Size Used for the Test Set and Data Provenance

    The primary test set for clinical performance involved a retrospective study with:

    • Sample Size: 137 evaluable samples from 96 subjects (out of 139 samples from 98 patients initially collected).
    • Data Provenance: Retrospective, collected at 2 clinical sites in the US (OHSU and Hospital of the University of Pennsylvania).

    For analytical performance studies, various sample compositions and quantities were used:

    • Precision (within-lab & site-to-site): 25 samples formulated from 5 human RNA specimens positive for BCR-ABL1 diluted into RNAs from CML-negative blood. The site-to-site reproducibility study involved 1200 measurements from this 25-sample pool.
    • Linearity/Reportable Range: 2 separate RNA specimens (one e13a2, one e14a2) diluted into RNAs from CML-negative whole blood (ranging from MR 0.1 to MR 4.8).
    • Limit of Blank (LoB): 30 non-leukemic human RNA specimens.
    • Limit of Detection (LoD): 4 separate human RNA specimens positive for BCR-ABL1, serially diluted to 28 levels.
    • Limit of Quantitation (LoQ): 6 specimens derived from 6 human RNA positive for BCR-ABL (diluted to target MR 4.7).
    • Interfering Substances: Residual CML-positive blood diluted to approximately MR 4.0 into RNA from CML-negative whole blood (tested in 9 replicates).
    • Primer Specificity: 11 leukemic specimens (CML, AML, ALL) and 2 non-leukemic RNA specimens.
    • Specimen Carryover Contamination: Serially diluted total RNA from residual clinical CML-positive blood into RNA from CML-negative whole blood in a checkerboard pattern (25 high positive, 25 negative).
    • RNA Isolation: CML-positive white blood cells serially diluted across 4 levels (MR 1-4) into CML-negative human anti-coagulated whole blood.
    • RNA Input: 30 samples derived from 2 primary RNA samples diluted into non-leukemic human RNA (total 216 evaluable observations).
    • Traceability: WHO Reference Panel (4 panel members tested in duplicate).
    • Reagent Stability: 5 samples diluted to approximately MR 1, 2, 3, 4, and < LoD (15 replicates per time point).
    • Specimen Stability (Whole Blood): 13 CML positive specimens and 3 CML negative specimens.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not explicitly state the number of experts or their specific qualifications (e.g., radiologist with X years of experience) used to establish the ground truth for the test set.

    For the clinical study, the "ground truth" for event-free survival (EFS) was based on clinical outcomes defined by specific criteria:

    • Death by any cause
    • Development of accelerated-phase or blast crisis CML
    • Loss of complete hematologic response
    • Loss of complete cytogenetic response
    • Appearance of mutation
    • Change in TKI treatment not due to toxicity and not due to the results of another BCR-ABL assay

    These outcomes would typically be determined by treating physicians and medical records, often relying on established clinical guidelines and diagnostic criteria. While not explicitly stated as "experts establishing ground truth," the medical documentation and clinical assessments would serve this role.

    4. Adjudication Method for the Test Set

    The document does not describe an explicit adjudication method (e.g., 2+1, 3+1, none) for a test set, particularly in the context of expert review. The ground truth for the clinical study was based on observed clinical outcomes and treatment changes, which are derived from patient records and physician decisions rather than expert consensus on image or assay interpretation.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This device is a quantitative in vitro diagnostic assay, not an imaging device or a device requiring human interpretation in the loop in the same way an MRMC study evaluates. The clinical study assessed the device's ability to predict clinical outcomes (EFS) based on its quantitative output (MR value), rather than comparing human reader performance with and without AI assistance.

    6. If a Standalone (Algorithm Only Without Human-in-the-loop Performance) Was Done

    Yes, the primary clinical performance and all analytical performance studies represent a standalone assessment of the device. The QuantideX qPCR BCR-ABL IS Kit is an in vitro diagnostic test that provides quantitative results (MR values) directly. There is no human-in-the-loop interpretation of the assay's raw output described; the device itself calculates the BCR-ABL1 to ABL1 ratio and MR value. The clinical study then correlated these device-generated MR values with patient outcomes.

    7. The Type of Ground Truth Used

    • Analytical Ground Truth: For many analytical performance studies (precision, linearity, LoD, LoQ, interference), the ground truth was established by known concentrations or dilutions of BCR-ABL1 and ABL1 transcripts, or by using characterized RNA specimens (e.g., CML-positive or CML-negative clinical samples, or cell lines with specific translocations) to create spiked samples or panels. Traceability was established against the First (1st) WHO International Genetic Reference Panel.
    • Clinical Ground Truth: For the clinical performance study, the ground truth was patient outcomes data, specifically "event-free survival" at 32-40 months after TKI treatment. "Events" were defined by specific clinical criteria related to disease progression, death, or changes in therapy.

    8. The Sample Size for the Training Set

    The document does not specify a separate training set size. As an in vitro diagnostic kit, the "training" analogous to machine learning would be the extensive analytical validation and optimization performed by the manufacturer during product development, which informs the assay's design, protocols, and reagent formulations. The clinical and analytical studies described are primarily for validation and verification of the final device, not for training a specific algorithm.

    9. How the Ground Truth for the Training Set Was Established

    Since a distinct "training set" in the machine learning sense is not explicitly mentioned or applicable for this type of IVD kit, the concept of establishing ground truth for a training set does not directly apply. Instead, the analytical characteristics (like limit of detection, linearity, precision) are established using controlled experiments with known concentrations and reference materials, and clinical utility is demonstrated by correlating the device's measurements with observed patient clinical outcomes. These outcomes in the clinical study are determined by medical definitions and patient records, as described in point 3.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1