Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K163013

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-01-06

    (70 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Quantib™ Brain is a non-invasive medical imaging processing application that is intended for automatic labeling, visualization, and volumetric quantification of segmentable brain structures from a set of magnetic resonance (MR) images. The Quantib™ Brain output consists of segmentations, visualizations and volumetric measurements of grev matter (GM), white matter (WM), and cerebrospinal fluid (CSF). The output also visualizes white matter hyperintensity (WMH) candidates. Users need to review and if necessary, edit WMH candidates using the provided tools, before validation of the WMHs. It is intended to provide the trained medical professional with complementary information for the evaluation and assessment of MR brain images and to aid the trained medical professional in quantitative reporting. Quantib™ Brain is a post-processing plugin for the GE Advantage Workstation (AW 4.7) or AW Server (AWS 3.2) platforms.

    Device Description

    Quantib™ Brain is post-processing analysis software for the GE Advantage Workstation (AW 4.7) and AW Server (AWS 3.2) platforms using Volume Viewer Apps. 12.3 Ext 8 (or higher). It is intended for automatic labeling, visualization, and volumetric quantification of identifiable brain structures from magnetic resonance images (a 3D T1-weighted MR image, with an additional T2-weighted FLAIR MR image for white matter hyperintensities (WMH) segmentation system relies on a number of atlases each consisting of a 3D T1-weighted MR image and a label map dividing the MR image into different tissue segments. Quantib™ Brain provides quantitative information on both the absolute and relative volume of the segmented regions. The automatic WMH segmentation is to be reviewed and if necessary, edited by the user before validation of the segmentation, after which volumetric information is accessible. Longitudinal analysis can be performed for the brain tissue segmentation and WMH segmentation in order to compare multiple exams of an individual patient. Quantib Brain is intended to provide the trained medical professional with complementary information for the evaluation and assessment of MR brain images and to aid the radiology specialist in quantitative reporting.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for Quantib™ Brain 1.2, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document primarily focuses on the validation of a newly added algorithm for classifying White Matter Hyperintensities (WMH) as consistent, new, or disappearing between longitudinal scans. The overall performance of core segmentation algorithms (brain volumetry and WMH) is stated as unchanged from the predicate device.

    Acceptance Criteria (Implicit for new WMH labeling algorithm)Reported Device Performance (Quantib™ Brain 1.2)
    Accurate labeling of WMHs as consistent, new, or disappearing in longitudinal comparison.The automatic labeling of WMHs was 99.6% identical to manual labeling of these WMHs for WMH volume.
    No impact on the safety of the device."The changes made in Quantib™ Brain 1.2 do not affect the safety of the device."
    Continued performance of existing algorithms."The performance of the already existing algorithms did not change."

    2. Sample Size Used for the Test Set and Data Provenance:

    • Test Set Sample Size: 12 datasets.
    • Data Provenance: The document states "12 datasets of different subjects, each consisting of a baseline exam and 1 to 3 follow-up exams." It does not explicitly state the country of origin or if the data was retrospective or prospective. However, given the nature of longitudinal studies, it's highly likely to be retrospective data collected over time from individual patients.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    The document states "manual labeling of these WMHs" was used for comparison. It does not specify the number of experts or their qualifications (e.g., radiologist with X years of experience).

    4. Adjudication Method for the Test Set:

    The document mentions "manual labeling" for comparison but does not detail an adjudication method (e.g., 2+1, 3+1, none) among multiple experts, suggesting the ground truth was established by a single manual labeling process, or the details of such a process are not provided.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:

    No, an MRMC comparative effectiveness study is not mentioned in the provided text. The evaluation focuses on the algorithm's standalone performance compared to manual labeling, not on human reader performance with or without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Yes, a standalone performance evaluation of the new WMH labeling algorithm was done. The study assessed the automatic labeling of WMHs against manual labeling, without a human-in-the-loop component.

    7. The Type of Ground Truth Used:

    The ground truth used for the WMH labeling algorithm was expert manual labeling. The document states, "The automatic labeling of WMHs was for 99.6% of the WMH volume identical to manual labeling of these WMHs."

    8. The Sample Size for the Training Set:

    The document does not provide any information regarding the training set sample size for the new WMH labeling algorithm or for the existing core algorithms.

    9. How the Ground Truth for the Training Set Was Established:

    The document does not provide any information on how the ground truth for the training set was established, as details about the training set itself are absent.

    Ask a Question

    Ask a specific question about this device

    K Number
    K153351

    Validate with FDA (Live)

    Device Name
    Quantib Brain 1
    Manufacturer
    Date Cleared
    2016-06-17

    (210 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Quantib™ Brain is a non-invasive medical imaging processing application that is intended for automatic labeling, visualization, and volumetric quantification of segmentable brain structures from a set of magnetic resonance (MR) images. The Quantib™ Brain output consists of segmentations, visualizations and volumetric measurements of grey matter (GM), white matter (WM), and cerebrospinal fluid (CSF). The output also visualizes and quantifies white matter hyperintensity (WMH) candidates. Users need to review and if necessary, edit WMH candidates using the provided tools, before validation of the WMHs. It is intended to provide the trained medical professional with complementary information for the evaluation and assessment of MR brain images and to aid the trained medical professional in quantitative reporting. Quantib™ Brain is a post-processing plugin for the GE Advantage Workstation (AW 4.7) or AW Server (AWS 3.2) platforms.

    Device Description

    Quantib™ Brain is post-processing analysis software for the GE Advantage (AW 4.7) or AW Server (AWS 3.2) platforms using Volume Viewer Apps. 12.3 Ext 6. It is intended for automatic labeling, visualization, and volumetric quantification of identifiable brain structures from magnetic resonance images (a 3D T1-weighted MR image, with an additional T2-weighted FLAIR MR image for white matter hyperintensities (WMH) segmentation). The segmentation system relies on a number of atlases each consisting of a 3D T1-weighted MR image and a label map dividing the MR image into different tissue segments. Quantib™ Brain provides quantitative information on both the absolute and relative volume of the segmented regions. The automatic WMH segmentation is to be reviewed and if necessary, edited by the user before validation of the segmentation, after which volumetric information is accessible. It is intended to provide the trained medical professional with complementary information for the evaluation and assessment of MR brain images and to aid the radiology specialist in quantitative reporting.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for Quantib™ Brain 1, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" but rather presents a performance study comparing the device's output to manual segmentations. Therefore, the "acceptance criteria" are inferred from the reported performance values that were deemed acceptable for market clearance.

    Metric (Inferred Acceptance Criterion)Reported Device Performance (Mean ± Standard Deviation)
    Brain Tissue Segmentation (GM, WM, CSF)
    Dice Index for CSF0.78 ± 0.05
    Dice Index for GM0.83 ± 0.02
    Dice Index for WM0.86 ± 0.02
    Absolute Difference of Relative CSF Volume (pp)1.6 ± 1.0
    Absolute Difference of Relative GM Volume (pp)2.8 ± 1.9
    Absolute Difference of Relative WM Volume (pp)2.6 ± 1.6
    ICV Segmentation
    Dice Index for ICV0.97 ± 0.01
    White Matter Hyperintensity (WMH) Segmentation
    Dice Index for WMH0.61 ± 0.13
    Absolute Difference of Relative WMH Volume (pp)0.6 ± 0.7

    Note: The document states that the performance data "shows that Quantib™ Brain is as safe and effective as the predicate device," implying these performance metrics were sufficient to demonstrate substantial equivalence.


    2. Sample Size Used for the Test Set and Data Provenance

    • Brain Tissue Segmentation (GM, WM, CSF, ICV):

      • Sample Size: 33 T1w MR images with 6 selected slices per scan for comparison.
      • Data Provenance: The set was "carefully selected to include data from multiple vendors and a series of representable scan settings." The document does not specify the country of origin or whether the data was retrospective or prospective.
    • White Matter Hyperintensity (WMH) Segmentation:

      • Sample Size: 30 3D T1w images with corresponding T2w FLAIR images.
      • Data Provenance: This set also "represented various scan settings." The document does not specify the country of origin or whether the data was retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    The document does not explicitly state the number of experts or their qualifications for establishing the ground truth. It simply refers to "manual segmentations" for brain tissues and "manually segmented" WMHs.


    4. Adjudication Method for the Test Set

    The document does not specify any adjudication method (e.g., 2+1, 3+1) for establishing the ground truth or resolving discrepancies in manual segmentations. It merely states "manual segmentations."


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, the document does not indicate that a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done. The study focuses on comparing the algorithm's performance against manual segmentations, not on how human readers improve with or without AI assistance.


    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, a standalone performance study was done. The reported "Algorithm performance" (Section VII.2) directly compares the Quantib™ Brain's automatic segmentations and volume measurements to manual segmentations for the same scans, without human intervention in the device's output during the test. The "Users need to review and if necessary, edit WMH candidates using the provided tools, before validation of the WMHs" statement in the Indications for Use refers to the intended use case, not the performance validation study's methodology. The study itself assesses the raw algorithmic output.


    7. The Type of Ground Truth Used

    The ground truth used was expert manual segmentation.

    • For brain tissue volumetry (GM, WM, CSF, ICV), the device's relative brain tissue volumes were compared to "relative volumes derived from manual segmentations."
    • For WMH, "WMHs were manually segmented on the T2w FLAR images" and compared to the device's automatic segmentation.

    8. The Sample Size for the Training Set

    The document does not specify the sample size for the training set. It mentions that "The segmentation system relies on a number of atlases each consisting of a 3D T1-weighted MR image and a label map dividing the MR image into different tissue segments," but does not give a number for these atlases or the data used to create them or train the system.


    9. How the Ground Truth for the Training Set Was Established

    The document does not explicitly state how the ground truth for the training set was established. It mentions that the system "relies on a number of atlases each consisting of a 3D T1-weighted MR image and a label map dividing the MR image into different tissue segments." It can be inferred that these atlases contain expert-defined segmentations (label maps), but the method of their creation (e.g., by how many experts, what qualifications, adjudication) is not detailed.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1