Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K241366

    Validate with FDA (Live)

    Date Cleared
    2024-10-29

    (168 days)

    Product Code
    Regulation Number
    868.5870
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The demand valve is used for the application of 100 vol% oxygen during manual ventilation with the resuscitator bag, as well as during direct non-invasive inhalation with the resuscitator mask in spontaneously breathing patients.

    Device Description

    The valve, non - rebreathing QualityFlow O2 demand valve is a medical device to deliver medical oxygen or a medical air - oxygen mixture to a patient on demand. Unlike continuous oxygen systems, the demand valve only supplies medical gas when the patient inhales, reducing oxygen consumption and ensuring the patient receives gas only when needed. The QualityFlow O2 Demand Valve from DEHAS consists of a membrane that is activated by negative pressure during inspiration and a valve mechanism that controls gas flow. During inspiration, the valve opens due to the movement of the membrane, while during expiration it is closed by a return spring. The function of the demand valve is based on providing gas only when a negative pressure is applied to the membrane, e.g. by the patient inhaling with a connected mask or suction of the resuscitator when connected (expansion after previous compression of the resuscitator). The valve therefore ensures that gas only flows when it is needed.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for a medical device called "QualityFlow O2 Demand Valve". This type of submission aims to demonstrate that a new device is substantially equivalent to a legally marketed predicate device, and thus does not require clinical studies in many cases.

    Based on the provided text, the device in question (QualityFlow O2 Demand Valve) is a mechanical device, not an AI/software-driven one. Therefore, the questions related to AI/MRMC studies, training/test sets, ground truth establishment by experts, and adjudication methods are not applicable to this submission. The "acceptance criteria" discussed in this document refer to the device's ability to meet various engineering, safety, and performance standards, rather than diagnostic accuracy metrics common for AI/ML devices.

    Here's an analysis based on the information relevant to this type of medical device submission:

    1. Table of Acceptance Criteria (Relevant Standards) and Reported Device Performance:

    The document summarizes non-clinical tests performed according to international standards to demonstrate substantial equivalence and that the device performs as intended.

    Acceptance Criteria (Standard Used)Description of TestReported Device Performance
    ISO 5356-1 Third edition 2004-05-15: Anaesthetic and respiratory equipment - Conical connectors: Part 1: Cones and socketsSpecifies the dimension and dimensional requirements for cones and bushings intended for the connection of anesthesia and ventilator equipment (e.g., in ventilation systems, anesthetic gas transport systems and vaporizers).Result: pass
    IEC 62366-1 Edition 1.1 2020-06 CONSOLIDATED VERSION: Medical devices - Part 1: Application of usability engineering to medical devicesUsed to verify and validate the usability of the QualityFlow O2 demand valve.Result: pass
    ISO 14971 Third Edition 2019-12: Medical devices - Application of risk management to medical devicesProcess for identification and assessment of hazards and associated risks, control of these risks, and monitoring effectiveness of risk control measures.Result: pass
    ISO 15001:2010: Anesthetic and respiratory equipment - Compatibility with oxygenUsed to ensure the oxygen compatibility of the used materials.Result: pass
    ISO 10993-1 Fifth edition 2018-08: Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management processUsed for biological evaluation of medical devices and testing within a risk management process.Result: pass
    ISO 18562-1 First edition 2017-03: Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 1: Evaluation and testing within a risk management processEvaluated biocompatibility of breathing gas pathways, specifically for indirect patient contact via medical oxygen (gas pathway contact, limited exposure < 24 hours).Result: pass
    ISO 18562-3: Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 3: Tests for emissions of volatile organic compoundsAssessed gas emission of VOCs with a risk assessment of any identified chemical.Result: pass

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size for Test Set: The document does not specify a "sample size" in terms of patient data or images, as this is a non-AI/ML mechanical device. The tests performed are laboratory/bench tests on the device itself and its components, according to the listed international standards.
    • Data Provenance: Not applicable in the context of patient data. The provenance for the testing refers to the execution of tests in a laboratory setting by the manufacturer (DEHAS Medical Systems GmbH) or their designated testing facilities, adhering to international standards. The tests are non-clinical.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

    • Not applicable. "Ground truth" in the context of device performance is established by the detailed specifications of the international standards themselves and the objective measurements derived from testing the physical device against those specifications, using calibrated equipment. There is no subjective human interpretation of patient data here that requires expert consensus for ground truth.

    4. Adjudication Method for the Test Set:

    • Not applicable. Adjudication methods (e.g., 2+1, 3+1) are used for resolving disagreements among human readers or expert reviewers, typically in diagnostic studies with subjective image interpretation. This is a mechanical device tested against objective engineering standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    • No. MRMC studies are specific to evaluating the diagnostic performance of AI or imaging devices with human readers interpreting cases. This device is a mechanical demand valve, not an imaging or AI diagnostic device.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study Was Done:

    • Not applicable. This is not an algorithm. The "standalone" performance here refers to the device's inherent mechanical functionalities (e.g., flow rates, pressure triggers, material compatibility) as measured by laboratory testing according to the specified standards. All tests listed in the "Summary of non-clinical tests" (Table 3 on page 10-11) evaluate the device's technical performance.

    7. The Type of Ground Truth Used:

    • Ground Truth: The "ground truth" for this type of device is defined by the objective, quantifiable specifications and pass/fail criteria outlined in the referenced international standards (e.g., ISO 5356-1 for connector dimensions, ISO 15001 for oxygen compatibility, ISO 18562-3 for VOC emissions). These are engineering and safety standards, not medical diagnostic "ground truth" derived from patient outcomes or expert consensus.

    8. The Sample Size for the Training Set:

    • Not applicable. This is a mechanical device, not an AI/ML model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable. As there is no training set for an AI/ML model, no ground truth was established for it.

    Summary of Study Proving Device Meets Acceptance Criteria:

    The study proving the device meets the acceptance criteria is a series of non-clinical bench tests and evaluations conducted according to recognized international standards. These tests assess various aspects of the device's design, materials, usability, and performance to ensure it is safe and effective and substantially equivalent to its predicate device. The results for all listed standards were "pass", indicating the device met the pre-defined criteria for each test. The manufacturer concluded that there are "no differences which raise any different questions of safety and effectiveness" compared to the predicate device, and that its performance is "better or identical" to the predicate. No animal or clinical testing was performed for this 510(k) submission, as it was deemed unnecessary to support substantial equivalence.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1