510(k) Data Aggregation

    K Number
    K242255

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-04-29

    (272 days)

    Product Code
    Regulation Number
    880.5440
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Qitexio® 4-Way Stopcock is intended to serve as a flow control and delivery device for I.V. fluids injection to the patient's vascular system. The device is indicated for medium pressure injection of fluids. It is also used for delivery of Lipiodol® (Ethiodized Oil) injection. It is intended for single use only. The device is not indicated for use in infusion.

    Device Description

    Qitexio® 4-Way Stopcock is intended to serve as a flow control and delivery device for I.V. fluids injection to the patient's vascular system. Qitexio® 4-Way Stopcock is not intended to facilitate bi-directional flows, thus is not designed to be used in infusion. Qitexio® 4-Way Stopcock is designated for single use only (disposable) and is MR safe. Qitexio® 4-Way Stopcock is robust under medium pressure and is verified to be used for Lipiodol (Ethiodized oil) delivery.

    Qitexio® 4-Way Stopcock is made of a body with four ports (3 female and one male) and one handle with a L-Shape diverter. The L-shaped diverter opens two neighboring ports at the same time, creating a L-shaped communication channel. The male connector includes a mobile ring assembled on the end of male port, and it enables connection locking over female Luer. A small amount of lubricant is applied on the handle stem before placing it in the Stopcock body. The neighboring female port(s) is connected to the I.V. set to accept the delivered drug or fluid. One female port, that is on the opposite side of the male port, does not directly communicate with the patient (male port). This extra port is designed to enable users to inject multiple I.V. fluids, if needed.

    The operating mechanism is manual and with a simple twisting handle position to determine direction of the fluids. Qitexio® 4-Way Stopcock has only two ports communicating at a time.

    AI/ML Overview

    This document is a 510(k) clearance letter for the Qitexio® 4-Way Stopcock, a physical medical device (an intravascular administration set), not an AI/Software as a Medical Device (SaMD).

    Therefore, many of the requested criteria related to AI/SaMD performance studies (e.g., sample size for test/training sets, ground truth establishment for AI, expert adjudication, MRMC studies, standalone performance) are not applicable to this type of device submission.

    The document primarily focuses on demonstrating substantial equivalence to a predicate device through:

    • Bench testing: Verifying physical properties, mechanical integrity, and resistance to fluids.
    • Biocompatibility testing: Ensuring the materials are safe for patient contact.
    • Sterilization and packaging integrity testing: Confirming the device remains sterile and its packaging intact.
    • User-related risk analysis: Assessing potential human factors issues.

    Below is an attempt to address your request based only on the information provided, highlighting why certain sections are not applicable to this physical device.


    Acceptance Criteria and Device Performance for Qitexio® 4-Way Stopcock

    As the Qitexio® 4-Way Stopcock is a physical medical device, not a software or AI-based device, the concept of "acceptance criteria" and "device performance" in this context refers to meeting established engineering, material, and safety standards rather than diagnostic accuracy or algorithmic performance metrics. The information provided in the 510(k) summary focuses on demonstrating that the device meets these standards and is substantially equivalent to a legally marketed predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    The 510(k) summary does not present a single, consolidated table of quantitative acceptance criteria with specific performance values. Instead, it refers to compliance with recognized consensus standards (e.g., ISO, ASTM, USP) and internal criteria. The "Conclusions" column in Table 2 (Performance Testing) serves as the "reported device performance" by stating that the test results were "satisfactory for the requirements" or "compliant with Medex internal criteria."

    Here's an attempt to synthesize the closest equivalent based on the provided text, focusing on the key performance attributes mentioned:

    Type of Performance Test / Acceptance Criteria CategorySpecific Acceptance Criteria (Inferred from standards/statements)Reported Device Performance (as stated in document)
    Mechanical PerformanceBore connectors (Luer Lock standard): Compliance with ISO 80369-7, ISO 80369-20."The test results are satisfactory for the requirements."
    Stopcock Body Pressure Withstand: Compliance with ISO 80369-7, ISO 80369-20. Max pressure resistance of 550 psi."The test results are satisfactory for the requirements." Device withstands "medium pressure up to 500 psi" (in direct comparison to predicate) and "Up to 550 psi" (under its own specifications).
    Mechanical resistance to Lipiodol: Maintenance of physical performance (breakaway torque, mean torque, withstands pressure, stop resistance torque) after 24h exposure to Lipiodol."Lipiodol® analysis report is compliant with the Medex internal criteria." "maintains its physical performance characteristics after 24 hrs of exposure to Lipiodol®."
    Stopcock Lever Rotation & Tightness: Verification of breakaway torque and mean torque during rotation."Verification results met the Medex internal criteria."
    Sterilization & Package IntegritySterile Barrier Integrity: Compliance with ISO 11607-1, ASTM F1980-21, ASTM F1886/F1886M-16."The test results are satisfactory for the requirements."
    Microbiological Validation of Sterilization: Compliance with ISO 11137-1, ISO 11137-2, ISO 11737-1, ISO 11737-2."The test results are satisfactory for the requirements."
    BiocompatibilityCytotoxicity: Compliance with ISO 10993-5."The test results are satisfactory for the requirements."
    Irritation & Sensitization: Compliance with ISO 10993-10.(No specific conclusion stated, but implies compliance as part of overall biocompatibility)
    Acute Systemic Toxicity: Compliance with ISO 10993-11.(No specific conclusion stated, but implies compliance as part of overall biocompatibility)
    Hemocompatibility: Compliance with ISO 10993-4, ISO 10993-12, ASTM F756-17.(No specific conclusion stated, but implies compliance as part of overall biocompatibility)
    Chemical Characterization (Extractables): Compliance with ISO 10993-17, ISO 10993-18, ISO/TS 10993-19.(No specific conclusion stated, but implies compliance as part of overall biocompatibility)
    Material-mediated pyrogen: Compliance with USP<151>.(No specific conclusion stated, but implies compliance as part of overall biocompatibility)
    Particulate ContaminationCompliance with USP<788>."The test results are satisfactory for the requirements."
    PyrogenicityBacterial Endotoxins: Compliance with USP <85>, ANSI/AAMI ST72."The test results are satisfactory for the requirements."
    TransportationSimulated Transportation Study: Compliance with ISTA 3A-2018."The test results are satisfactory for the requirements."
    User Related Risk AnalysisNo use-related critical task identified per IEC 62366-1 and FDA Guidance "Human Factors and Usability Engineering to Medical Devices.""No Use-related critical task was identified due to differences between the predicate and the Qitexio® 4-Way Stopcock."
    Shelf LifeVerified by accelerated and real-time aging for 3 years."The shelf life of the subject device has been verified... 3 years."

    Information Not Applicable to This Device Type (Physical Medical Device)

    The following points are not applicable to this 510(k) submission as it pertains to a physical, non-AI/software medical device. If this were an AI/SaMD, this information would be critical.

    • 2. Sample size used for the test set and the data provenance: Not applicable. Performance testing for a physical device refers to bench testing on physical samples, not a data test set for an algorithm. There is no "data provenance" in the sense of clinical data from patients.
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for a physical device is established through engineering and laboratory standards and measurements, not expert consensus on clinical data.
    • 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. Adjudication is for resolving discrepancies in expert labeling of data, which is irrelevant for physical device testing.
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. MRMC studies are used for evaluating the impact of AI on human reader performance, which doesn't apply to a mechanical stopcock.
    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This refers to AI algorithm performance, not a physical device.
    • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable. Ground truth for this device is based on established engineering and material science standards (e.g., "Luer Lock standard," "ISO," "ASTM," "USP" requirements).
    • 8. The sample size for the training set: Not applicable. There is no "training set" for a physical medical device.
    • 9. How the ground truth for the training set was established: Not applicable. No training set for a physical medical device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K203738

    Validate with FDA (Live)

    Device Name
    Qitexio
    Manufacturer
    Date Cleared
    2022-03-14

    (447 days)

    Product Code
    Regulation Number
    880.5860
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Qitexio Luer Lock Syringes are manual devices used to inject fluids into the body. They are also used for delivery of Lipiodol (Ethiodized Oil) Injection.

    Device Description

    Medex, a subsidiary of Guerbet Group, designed and manufactures the Qitexio® Luer Lock Syringe, a disposable, handheld syringe available in 3 different volumes (1, 3 and 20 mL). Oitexio® Syringes are standard syringes which are verified to endure the potentially damaging chemical effects of Lipiodol® (Ethiodized Oil), an oil-based contrast media manufactured by Guerbet.

    Qitexio® Luer Lock Syringes are manual devices used to inject fluids into the body. Qitexio® Syringes are also used for delivery of Lipiodol® (Ethiodized Oil) Injection.

    They are single-use devices, sterilized by gamma radiation.

    Standard five-piece piston syringe constructed with a clear hollow barrel into which is inserted a closely fitting movable plunger, O-Ring, rotative finger flange, and backstop. Fitting offered with male luer lock.

    AI/ML Overview

    This document describes the premarket notification (510(k)) for the Qitexio Luer Lock Syringes, a medical device. The information provided heavily focuses on proving the substantial equivalence of the new device to a predicate device, the Merit Syringe (K173601).

    The acceptance criteria for this type of device (a piston syringe) are primarily defined by conformity to established FDA recognized consensus standards and a battery of performance, biocompatibility, and sterilization tests. The document does not describe an AI/ML-driven device or study. Therefore, some of the requested information regarding AI/ML study design, such as MRMC studies, ground truth establishment for training sets, and expert adjudication, are not applicable to this submission.

    Here's a breakdown of the acceptance criteria and proof, based on the provided text:


    1. Table of Acceptance Criteria and Reported Device Performance

    For the Qitexio Luer Lock Syringes, acceptance criteria are generally met by demonstrating conformity to recognized consensus standards and by passing specific performance tests. The document repeatedly states that the device "complies with" or "met the established acceptance criteria" for these standards and tests. Specific numerical performance results are not provided in this summary, as is common for 510(k) summaries which focus on demonstrating equivalence rather than presenting raw data.

    Acceptance Criteria CategorySpecific Criteria (Standard/Test)Reported Device Performance
    Performance TestingISO 7886-1 Sterile Hypodermic Syringes For Single Use - Part 1: Syringes For Manual UseComplies; specific for Lipiodol® (Ethiodized Oil) Injection delivery.
    ISO 80369-7 Small-bore connectors for liquids and gases in healthcare applications — Part 7: Connectors for intravascular or hypodermic applicationsComplies.
    Functional performance testing (with preconditioning: double sterilization, accelerated/real-time aging)Met established criteria; performs as intended.
    BiocompatibilityISO 10993 Biological Evaluation of Medical Devices (Cytotoxicity, Irritation, Sensitization, Acute Systemic Toxicity, Hemolysis, Material-Mediated Pyrogenicity, Chemical Characterization)Non-toxic, biocompatible for intended use; all tests passed.
    USP <788> Particulate Matter in Injections (Method 1 Light Obscuration Particle Count)Met USP <788> 1B acceptance criteria (for ≤ 100 mL volume).
    SterilizationISO 11137-1:2006 (Sterilization, Validation, and Routine Control of a Sterilization Process for Medical Devices)Validated sterilization cycle; SAL 10-6 achieved; dose 25-50 kGy.
    ISO 11137-2:2013 (Establishing the Sterilization Dose)Determined 25 kGy requirement.
    USP 39 -NF 34 and European Pharmacopoeia (9th Edition) requirements for Material-mediated PyrogenicityConfirmed non-pyrogenicity (Rabbit Pyrogen study).
    Limulus Amebocyte Lysate (LAL) test for Bacterial endotoxinsMet criteria: < 20 endotoxin units. Device non-pyrogenic.
    Package IntegrityASTM F 1886/F 1886 M (Visual Inspection)Passed (after max dose sterilization; and after max dose + ISTA 3A).
    ISO 11607-1 (Peel Characteristics, Peel Test)Passed (after max dose sterilization; and after max dose + ISTA 3A).
    ASTM F 88/F 88 M (Peel Test)Passed (after max dose sterilization; and after max dose + ISTA 3A).
    ASTM F 1929 (Dye Test)Passed (after max dose sterilization).
    ASTM F 2096 (Bubble test)Passed (after max dose + ISTA 3A Simulated Transportation).
    Shelf-LifeDevice functionality and package integrity after 3 years aging (real-time or accelerated)Acceptable results supporting a 3-year shelf-life.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the exact sample sizes (number of syringes) used for each individual test (e.g., performance, biocompatibility). It mentions "samples" or "representative device samples" were used.

    • Data Provenance: Not explicitly stated regarding country of origin for the test data. However, the manufacturer is Medex, a subsidiary of Guerbet Group, located in France. The testing was conducted to comply with international standards (ISO, ASTM, USP, European Pharmacopoeia) recognized by the FDA for market clearance in the US.
    • Retrospective or Prospective: The testing described is prospective, conducted specifically to gain regulatory clearance for the device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Not Applicable. This submission is for a physical medical device (Piston Syringe), not an AI/ML diagnostic tool requiring expert interpretation as ground truth for a test set. The "ground truth" for this device's performance is established by objective measurements against engineering and biological standards.

    4. Adjudication Method for the Test Set

    • Not Applicable. As above, this is not an AI/ML diagnostic tool requiring human adjudication of subjective interpretations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No. This type of study is typically performed for imaging or diagnostic AI/ML devices where the AI assists human readers. It is not relevant for a Piston Syringe.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Not Applicable. This is not an algorithm or software device. The "performance" is that of the physical syringe itself.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance and safety is established through:

    • Conformity to recognized consensus standards (e.g., ISO 7886-1 for syringe function, ISO 10993 for biocompatibility, ISO 11137 for sterilization).
    • Objective bench testing measurements (e.g., particulate counts, endotoxin levels, peel force, functional performance like fluid injection).
    • Biological assay results (e.g., cytotoxicity, irritation, pyrogenicity).
    • Material characterization to ensure compatibility with intended fluids (e.g. Lipiodol).

    8. The Sample Size for the Training Set

    • Not Applicable. This is not an AI/ML device that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable. As this is not an AI/ML device, there is no training set or associated ground truth establishment. Device design and manufacturing processes are validated through engineering principles and quality system regulations (21 CFR Part 820).