510(k) Data Aggregation

    K Number
    K183390

    Validate with FDA (Live)

    Device Name
    QNext and DG-PT
    Date Cleared
    2019-08-21

    (258 days)

    Product Code
    Regulation Number
    864.5425
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The QNext is a fully-automated, random-access instrument, intended for in vitro diagnostic use in clinical laboratories to perform hemostasis testing by detecting the changes in optical density.

    DG-PT is a thromboplastin reagent for the quantitative determination of Prothrombin Time on human plasma samples collected in 3.2% sodium citrate.

    The product is used for the evaluation of the extrinsic and common coagulation pathways in seconds and for the monitoring Oral Anticoagulant Therapy with warfarin in International Normalized Ratio (INR).

    For use with ONext.

    For clinical professional laboratory and prescription use only.

    For in vitro diagnostic use.

    The performance of this device has not been established in neonate and pediatric patient populations.

    Device Description

    QNext is designed to automatically perform all stages of the procedures associated to hemostasis tests allowing the operator to:

    • Absorb the workload involved in running hemostasis laboratory tests profiles, optimizing the execution of these profiles in the shortest time possible, and ensuring the maximum possible precision and accuracy in the results.
    • Increase the reliability of the analytical process, eliminating any possible errors in the identification ● and treatment of samples and products and in the revision and transcription of the results.
    • Reduce the risk of Operator contamination by minimizing interaction between the Operator and ● samples and products during the analytical process.

    To perform the operations for which it has been designed, QNext automatically follows the steps listed below:

    • Sample management: loading, positive identification, dilution (if required) and dispensation into cuvettes.
    • Reagent management: loading, positive identification, cooling, stirring, aspiration and dispensation into cuvettes.
    • Cuvette management: loading, transport, incubation during the reactions and management of used cuvettes.
    • Management of test requests. ●
    • Execution of test procedures.
    • Result management: optical measurement of the reactions, algorithm calculation of analytical parameters from reaction curves, validation of results, traceability, bi-directional transmission of requests and results to the LIS.
    • Management of disposable components. ●

    The data analyzed can be stored, displayed and printed. Additionally, the analyzer allows conducting integrated functions, such as the analysis of urgent samples or the Quality Control module.

    DG-PT consists of a glass vial containing lyophilized thromboplastin (tissue factor and phospholipids) from rabbit brain tissue, buffer, calcium ions and preservative. The closure system includes a stopper and a screw сар.

    DG-PT reagent is used to perform PT tests for:

    • the evaluation of the extrinsic and common coagulation pathways.
    • The monitoring Oral Anticoagulant Therapy with warfarin.

    The assay is based on the activation of the extrinsic coagulation pathway by the addition of the reagent to the plasma sample. The thromboplastin interacts with FVII and calcium ions activating a series of specific enzymes that comprise the extrinsic and common pathways of the coagulation cascade ultimately leading to the formation of a fibrin clot. The QNext reader measures the light change produced during the reaction.

    AI/ML Overview

    The provided text describes the performance testing of the ONext instrument and DG-PT reagent for Prothrombin Time (PT) determination, but it does not contain information related to an AI/ML powered device. Therefore, many of the requested criteria regarding AI/ML device performance (e.g., multi-reader multi-case studies, human-in-the-loop performance, training set details, expert ground truth adjudication) are not applicable to this document.

    However, I can extract and present the acceptance criteria and performance data for the described in-vitro diagnostic device:

    Device: ONext (fully-automated hemostasis instrument) and DG-PT (thromboplastin reagent for PT determination).

    Study Proving Device Meets Acceptance Criteria:
    The submission details several performance studies conducted according to CLSI guidelines to demonstrate the substantial equivalence of the ONext and DG-PT to predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined quantitative acceptance criteria thresholds for each study (e.g., "Precision CV must be less than X%"). Instead, it presents the results of various performance studies (Precision, Reference Interval, Sensitivity, Specificity, Method Comparison) which are implicitly understood to meet the necessary performance characteristics for substantial equivalence. The "performance" column therefore reflects the observed results from these studies.

    Performance MetricAcceptance Criteria (Implied)Reported Device Performance (ONext and DG-PT)
    PrecisionRepeatability and Within-laboratory precision should demonstrate acceptable variability.Repeatability (PT seconds / INR): Varies by sample level. For PT (seconds), 1.4-2.1% CV; for INR, 1.6-2.4% CV. Upper One-sided 95% Limits slightly higher. Within-laboratory Precision (PT seconds / INR): Varies by sample level. For PT (seconds), 2.4-5.7% CV; for INR, 2.2-8.2% CV. Upper One-sided 95% Limits higher. Reproducibility (PT seconds / INR): Varies by sample level. For PT (seconds), Repeatability 1.6-3.1% CV, Reproducibility 3.1-6.0% CV; for INR, Repeatability 1.9-3.7% CV, Reproducibility 5.5-9.2% CV. Upper One-sided 95% Limits higher. (See Tables 1 & 2 in source for full details)
    Reference IntervalEstablish a clear reference interval for healthy individuals.PT (seconds): 12.1 to 16.1 seconds PT (INR): 0.83 to 1.17 INR
    Sensitivity (Factor)Demonstrate ability to detect deficiencies in key coagulation factors.FII: 29 IU/dL FV: 45 IU/dL FVII: 44 IU/dL FX: 43 IU/dL
    Specificity (Interference)Demonstrate resistance to interference from common endogenous and exogenous substances.Maximum Concentration without Significant Interference (Normal/Abnormal samples): Hemoglobin: 10.0 g/L Conjugated bilirubin: 43.3 mg/dL Triglycerides: 3000 mg/dL Citrate: 0.6% UFH: 1.1 IU/mL (Normal), 0.7 IU/mL (Abnormal) LMWH: 2.3 IU/mL (Normal), 1.8 IU/mL (Abnormal) Observed Interference at All Concentrations Tested for: Rivaroxaban (e.g., 313, 625, 938, 1250 ng/mL for normal sample) Argatroban (e.g., 375, 750, 1125, 1500 ng/mL for normal sample) Dabigatran (e.g., 150, 300, 450, 600 ng/mL for normal sample)
    Method ComparisonResults should correlate well with a legally marketed predicate device.Results from two US sites (n=360): PT (s): Slope 0.913 (95% CI: 0.893-0.938), Intercept 1.118 (95% CI: 0.706-1.477), r=0.983 PT (INR): Slope 1.034 (95% CI: 1.010-1.060), Intercept -0.024 (95% CI: -0.054-0.009), r=0.980 Results from ex-US site (n=271): PT (s): Slope 0.985 (95% CI: 0.962-1.003), Intercept 0.272 (95% CI: -0.078-0.716), r=0.988 PT (INR): Slope 1.000 (95% CI: 0.978-1.022), Intercept -0.070 (95% CI: -0.099--0.042), r=0.990

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Precision Studies:
      • Repeatability and Within-laboratory: 7 patient-derived samples. Data provenance not explicitly stated (e.g., country of origin), assumed to be clinical samples. Retrospective/prospective not specified, but typically for precision studies, these would be collected prospectively or obtained from a biobank.
      • Reproducibility: 7 patient-derived samples. Data provenance not explicitly stated (e.g., country of origin), assumed to be clinical samples. Retrospective/prospective not specified.
    • Reference Interval Study: 243 reference samples. Data provenance not explicitly stated.
    • Sensitivity Study: Not specified, but involved a set of samples spanning from 100% to <1% of specific coagulation factors.
    • Specificity Study: Two sample levels (pooled normal and pooled abnormal samples) with various interfering substances added. Specific total sample count not given.
    • Method Comparison Study:
      • US Sites: 360 patient-derived samples covering the claimed analytical measurement range (AMR). Data provenance: from two US clinical sites.
      • Ex-US Site: 271 patient-derived samples covering the claimed AMR. Data provenance: from one ex-US clinical site.
      • Retrospective/Prospective: Not explicitly stated, but clinical validation studies often use prospectively collected samples or samples from a well-characterized biobank.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Not Applicable. This is an in-vitro diagnostic device for measuring Prothrombin Time. The "ground truth" is established by direct measurement of the blood coagulation time or the INR, not by expert interpretation of images or clinical data. Therefore, expert consensus for ground truth is not relevant to this type of device.

    4. Adjudication Method for the Test Set

    Not Applicable. As per point 3, the ground truth for this device is a direct quantitative measurement, not a subjective interpretation requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If So, What was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

    Not Applicable. This document describes a medical device for in-vitro diagnostics (measuring blood coagulation), not an AI/ML powered device for image interpretation or diagnosis. There are no "human readers" or "AI assistance" components mentioned in the context of this device's intended use or validation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    Not Applicable. This is not an AI/ML algorithm. The device, ONext, is a fully-automated instrument, meaning its performance is inherently "standalone" in terms of its operation and measurement of PT/INR. The "human-in-the-loop" would be the laboratory professional using the instrument, but their role is in operating the device and interpreting its quantitative output, not in assisting or being assisted by an algorithm in a diagnostic task.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance studies of the ONext and DG-PT device is:

    • Direct Measurement: The actual PT/INR values obtained from patient samples using the predicate device or a reference method.
    • Clinical Samples with Known Characteristics: For sensitivity studies, samples with known factor deficiencies were used. For specificity studies, samples with known concentrations of interfering substances were used.
    • Reference Intervals: Established from healthy donor samples.

    8. The Sample Size for the Training Set

    Not Applicable. As this is not an AI/ML powered device, there is no "training set" in the context of machine learning model development. The device's operational parameters and algorithms are determined through traditional engineering and chemistry principles, not data-driven machine learning training.

    9. How the Ground Truth for the Training Set was Established

    Not Applicable. See point 8.