Search Results
Found 3 results
510(k) Data Aggregation
(83 days)
QIAstat-Dx Gastrointestinal Panel 2:
The QIAstat-Dx Gastrointestinal Panel 2 is a multiplexed nucleic acid test intended for use with the QIAstat-Dx Analyzer 2.0 and QIAstat-Dx Rise for the simultaneous in vitro qualitative detection and identification of nucleic acids from multiple viruses, bacteria, and parasites directly from preserved stool samples (Para-Pak C&S or FecalSwab) obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following viruses, bacteria (including several diarrheagenic E. coli/ Shigella pathotypes), and parasites are identified with the QIAstat-Dx Gastrointestinal Panel 2:
- Adenovirus F40/F41
- Astrovirus
- Norovirus GI/GII
- Rotavirus A
- Campylobacter (C. jejuni, C. coli and C. upsaliensis)
- Shigella/Enteroinvasive Escherichia coli (EIEC)
- Enteropathogenic Escherichia coli (EPEC)
- Enterotoxigenic Escherichia coli (ETEC) lt/st
- Shiga-like toxin-producing Escherichia coli (STEC) stx1/stx2 (including specific identification of E. coli O157 serogroup within STEC)
- Salmonella
- Plesiomonas shigelloides
- Yersinia enterocolitica
- Cryptosporidium
- Cyclospora cayetanensis
- Entamoeba histolytica
- Giardia lamblia*
*Also known as Giardia intestinalis and Giardia duodenalis
Concomitant culture is necessary for organism recovery and further typing of bacterial agents. The QIAstat-Dx Gastrointestinal Panel 2 is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness, in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule-out co-infection with organisms not detected by the QIAstat-Dx Gastrointestinal Panel 2. The organisms detected may not be the sole or definitive cause of the disease.
Negative QIAstat-Dx Gastrointestinal Panel 2 results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this assay test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.
QIAstat-Dx GI Panel 2 Mini B&V:
The QIAstat-Dx GI Panel 2 Mini B&V is a multiplexed nucleic acid test intended for use with the QIAstat-Dx Analyzer 2.0 and QIAstat-Dx Rise for the simultaneous in vitro qualitative detection and identification of nucleic acids from multiple bacteria and one virus directly from preserved stool samples (Para-Pak C&S or FecalSwab) obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following virus and bacteria (including several diarrheagenic E. coli/ Shigella pathotypes) are identified with the QIAstat-Dx GI Panel 2 Mini B&V:
- Norovirus
- Campylobacter
- Shigella
- Shiga-like toxin E. coli (STEC)
- Salmonella
Concomitant culture is necessary for organism recovery and further typing of bacterial agents.The QIAstat-Dx GI Panel 2 Mini B&V is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness, in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule-out co-infection with organisms not detected by the QIAstat-Dx GI Panel 2 Mini B&V. The organisms detected may not be the sole or definitive cause of the disease.
Negative QIAstat-Dx GI Panel 2 Mini B&V results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this assay test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.
QIAstat-Dx GI Panel 2 Mini B:
The QIAstat-Dx GI Panel 2 Mini B is a multiplexed nucleic acid test intended for use with the QIAstat-Dx Analyzer 2.0 and QIAstat-Dx Rise for the simultaneous in vitro qualitative detection and identification of nucleic acids from multiple bacteria directly from preserved stool samples (Para-Pak C&S or FecalSwab) obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following bacteria (including several diarrheagenic E. coli/ Shigella pathotypes) are identified with the QIAstat-Dx GI Panel 2 Mini B:
- Campylobacter
- Shigella
- Shiga-like toxin Escherichia coli (STEC)
- Salmonella
- Yersinia enterocolitica
Concomitant culture is necessary for organism recovery and further typing of bacterial agents. The QIAstat-Dx GI Panel 2 Mini B is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness, in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule-out co-infection with organisms not detected by the QIAstat-Dx GI Panel 2 Mini B. The organisms detected may not be the sole or definitive cause of the disease.
Negative QIAstat-Dx GI Panel 2 Mini B results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this assay test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.
The QIAstat-Dx Gastrointestinal Panel 2, the QIAstat-Dx GI Panel 2 Mini B&V, and the QIAstat-Dx GI Panel 2 Mini B ("QIAstat-Dx Gastrointestinal Panels") are multiplexed nucleic acid tests which are designed for use with the QIAstat-Dx Analyzer 2.0. The device modification is to add the QIAstat-Dx Rise as an additional instrument for use with the QIAstat-Dx Gastrointestinal Panels. The QIAstat-Dx Rise is a higher throughput platform, incorporating up to eight QIAstat-Dx Analytical Modules (AM) on a small footprint. The instrument allows queuing up to 18 cartridges, with processing being limited up to 16 cartridges for the QIAstat-Dx Gastrointestinal Panels. The cartridges are scheduled for processing and delivered to the appropriate AM by an integrated robotic handler. The AM used with the QIAstat-Dx Rise is the same AM that can be used with the QIAstat-Dx Analyzer 2.0.
The modified QIAstat-Dx Gastrointestinal Panel 2, the QIAstat-Dx GI Panel 2 Mini B&V, and the QIAstat-Dx GI Panel 2 Mini B are identical to the QIAstat-Dx Gastrointestinal Panel 2, the QIAstat-Dx GI Panel 2 Mini B&V, and the QIAstat-Dx GI Panel 2 Mini B (K252329), respectively, with the exception of the Instructions for Use which were updated to include the assay-specific procedure for use with the QIAstat-Dx Rise.
The QIAstat-Dx Gastrointestinal Panels are intended for the simultaneous in vitro qualitative detection and identification of nucleic acids from multiple targets directly from preserved stool samples (Para-Pak C&S or FecalSwab) obtained from individuals with signs and/or symptoms of gastrointestinal infection.
All the reagents required for the complete execution of the test are pre-loaded and self-contained in the QIAstat-Dx Gastrointestinal Panels. The user does not need to manipulate any reagents. During the test, reagents are handled by pneumatically operated microfluidics without any direct contact with the user or the analyzer actuators. This eliminates any possibility of exposure of the user or the analyzer to chemicals contained in the cartridge during the test and up to the disposal of used cartridges.
Within the cartridge, multiple steps are automatically performed in sequence by using pneumatic pressure and a multiport valve to transfer sample and fluids via the Transfer Chamber to their intended destinations. Following the introduction of the sample from a disposable transfer pipette, the following assay steps occur automatically and sequentially:
- Sample pre-treatment for PCR Inhibitors removal
- Resuspension of Internal Control and Proteinase K
- Cell lysis using mechanical and/or chemical means
- Membrane-based nucleic acid purification
- Rehydration of Master Mix
- Transfer of defined aliquots of eluate/master mix to different reaction chambers
- Performance of multiplex real-time RT-PCR testing within each reaction chamber.
The QIAstat-Dx Gastrointestinal Panels Assay Definition File (ADF) automatically interprets test results and displays a summary on the instrument display screen. The detected analytes are displayed in red. All other tested but not detected analytes are listed in green. The instrument will report if an error occurs during processing, in which case the test will fail and no results will be provided (screen will show "FAIL").
N/A
Ask a specific question about this device
(89 days)
QIAstat-Dx Gastrointestinal Panel 2
The QIAstat-Dx Gastrointestinal Panel 2 is a multiplexed nucleic acid test intended for use with the QIAstat-Dx Analyzer 2.0 for the simultaneous in vitro qualitative detection and identification of nucleic acids from multiple viruses, bacteria, and parasites directly from preserved stool samples (Para-Pak C&S or FecalSwab) obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following viruses, bacteria (including several diarrheagenic E. coli/Shigella pathotypes), and parasites are identified with the QIAstat-Dx Gastrointestinal Panel 2:
- Adenovirus F40/F41
- Astrovirus
- Norovirus GI/GII
- Rotavirus A
- Campylobacter (C. jejuni, C. coli and C. upsaliensis)
- Shigella/Enteroinvasive Escherichia coli (EIEC)
- Enteropathogenic Escherichia coli (EPEC)
- Enterotoxigenic Escherichia coli (ETEC) lt/st
- Shiga-like toxin-producing Escherichia coli (STEC) stx1/stx2 (including specific identification of E. coli O157 serogroup within STEC)
- Salmonella
- Plesiomonas shigelloides
- Yersinia enterocolitica
- Cryptosporidium
- Cyclospora cayetanensis
- Entamoeba histolytica
- Giardia lamblia*
*Also known as Giardia intestinalis and Giardia duodenalis
Concomitant culture is necessary for organism recovery and further typing of bacterial agents. The QIAstat-Dx Gastrointestinal Panel 2 is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness, in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule-out co-infection with organisms not detected by the QIAstat-Dx Gastrointestinal Panel 2. The organisms detected may not be the sole or definitive cause of the disease.
Negative QIAstat-Dx Gastrointestinal Panel 2 results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this assay test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.
QIAstat-Dx GI Panel 2 Mini B&V
The QIAstat-Dx GI Panel 2 Mini B&V is a multiplexed nucleic acid test intended for use with the QIAstat-Dx Analyzer 2.0 for the simultaneous in vitro qualitative detection and identification of nucleic acids from multiple bacteria and one virus directly from preserved stool samples (Para-Pak C&S or FecalSwab) obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following virus and bacteria (including several diarrheagenic E. coli/Shigella pathotypes) are identified with the QIAstat-Dx GI Panel 2 Mini B&V:
- Norovirus
- Campylobacter
- Shigella
- Shiga-like toxin E. coli (STEC)
- Salmonella
Concomitant culture is necessary for organism recovery and further typing of bacterial agents. The QIAstat-Dx GI Panel 2 Mini B&V is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness, in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule-out co-infection with organisms not detected by the QIAstat Dx GI Panel 2 Mini B&V. The organisms detected may not be the sole or definitive cause of the disease.
Negative QIAstat-Dx GI Panel 2 Mini B&V results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this assay test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.
QIAstat-Dx GI Panel 2 Mini B
The QIAstat-Dx GI Panel 2 Mini B is a multiplexed nucleic acid test intended for use with the QIAstat-Dx Analyzer 2.0. for the simultaneous in vitro qualitative detection and identification of nucleic acids from multiple bacteria directly from preserved stool samples (Para-Pak C&S or FecalSwab) obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following bacteria (including several diarrheagenic E. coli/ Shigella pathotypes) are identified with the QIAstat-Dx GI Panel 2 Mini B:
- Campylobacter
- Shigella
- Shiga-like toxin Escherichia coli (STEC)
- Salmonella
- Yersinia enterocolitica
Concomitant culture is necessary for organism recovery and further typing of bacterial agents. The QIAstat-Dx GI Panel 2 Mini B is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness, in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule-out co-infection with organisms not detected by the QIAstat-Dx GI Panel 2 Mini B. The organisms detected may not be the sole or definitive cause of the disease.
Negative QIAstat-Dx GI Panel 2 Mini B results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this assay test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.
The QIAstat-Dx Gastrointestinal Panel 2, the QIAstat-Dx GI Panel 2 Mini B&V, and the QIAstat-Dx GI Panel 2 Mini B are multiplexed nucleic acid tests that are designed for use with the QIAstat-Dx Analyzer 2.0. Modification made to the devices is to:
- Unmask the results for Shiga-like toxin E. coli (STEC) stx1/stx2, E. coli O157, and Enteropathogenic E. coli (EPEC), for the FecalSwab sample type for the QIAstat-Dx Gastrointestinal Panel 2,
- Unmask the results for Shiga-like toxin E. coli (STEC) for the FecalSwab sample type for the QIAstat-Dx GI Panel 2 Mini B&V and the QIAstat-Dx GI Panel 2 Mini B, and
- Remove the QIAstat-Dx Analyzer 1.0 from the indicated amplification and detection instrument systems for all three (3) devices, leaving only the QIAstat-Dx Analyzer 2.0. The QIAstat-Dx Analyzer 1.0 and the QIAstat-Dx Analyzer 2.0 belong to the same instrument family.
The QIAstat-Dx Gastrointestinal Panel 2, the QIAstat-Dx GI Panel 2 Mini B&V, and the QIAstat-Dx GI Panel 2 Mini B are intended for the simultaneous in vitro qualitative detection and identification of nucleic acids from multiple targets directly from preserved stool samples (Para-Pak® C&S or FecalSwab™) obtained from individuals with signs and/or symptoms of gastrointestinal infection.
The QIAstat-Dx system is based on single-test cartridges with pre-packaged reagents including both wet and dry chemistry to handle the sample preparation and detection steps for the presence of a range of selected analytes by PCR technology. After insertion of the sample, the QIAstat-Dx assay cartridge is processed by the QIAstat-Dx Analyzer 2.0.
Once the cartridge has been inserted into the instrument, the test starts automatically and runs for approximately 80 minutes. When the test is finished, the cartridge is removed by the user and discarded. The QIAstat-Dx Analyzer 2.0 automatically interprets test results and displays a summary on the analyzer display screen. The results can be printed using a connected printer if needed. The detected analytes are displayed in red. For other analytes tested, they are displayed in green if not detected or in gray if not applicable or invalid. The analyzer will report if an error occurs during processing, in which case the test will fail, and no results will be provided (screen will show "FAIL").
All the reagents required for the complete execution of the test are pre-loaded and self-contained in the QIAstat-Dx Gastrointestinal Panels. The user does not need to manipulate any reagents. During the test, reagents are handled by pneumatically operated microfluidics without any direct contact with the user or the analyzer actuators. This eliminates any possibility of exposure of the user or the analyzer to chemicals contained in the cartridge during the test and up to the disposal of used cartridges.
Within the cartridge, multiple steps are automatically performed in sequence by using pneumatic pressure and a multiport valve to transfer sample and fluids via the Transfer Chamber to their intended destinations. Following the introduction of the sample from a disposable transfer pipette, the following assay steps occur automatically and sequentially:
- Sample pre-treatment for PCR Inhibitors removal
- Resuspension of Internal Control and Proteinase K
- Cell lysis using mechanical and/or chemical means
- Membrane-based nucleic acid purification
- Rehydration of Master Mix
- Transfer of defined aliquots of eluate/master mix to different reaction chambers
- Performance of multiplex real-time RT-PCR testing within each reaction chamber.
N/A
Ask a specific question about this device
(872 days)
The QIAstat-Dx Gastrointestinal Panel 2 is a multiplexed nucleic acid test intended for use with the QIAstat-Dx Analyzer 1.0. for the simultaneous in vitro qualitative detection of nucleic acids from multiple viruses, bacteria. and parasites directly from preserved stool samples (Para-Pak C&S or FecalSwab) obtained from individuals with signs and/or symptoms of gastrointestinal infection. The following viruses, bacteria (including several diarrheagenic E. col/ Shigella pathotypes), and parasites are identified with the QIAstat-Dx Gastrointestinal Panel 2 :
• Adenovirus F40/F41
• Astrovirus
• Norovirus GI/GII
• Rotavirus A
• Campylobacter (C. jejuni, C. coli and C. upsaliensis)
• Shigella/Enteroinvasive Escherichia coli (EIEC)
• Enteropathogenic Escherichia coli (EPEC)
• Enterotoxigenic Escherichia coli (ETEC) lt/st
• Shiga-like toxin-producing Escherichia coli (STEC) stx1/stx2
(including specific identification of E. coli O157 serogroup within STEC)
• Salmonella
• Plesiomonas shigelloides
• Yersinia enterocolitica
• Cryptosporidium
• Cyclospora cayetanensis
• Entamoeba histolytica
• Giardia lamblia*
*(Also known as Giardia intestinalis and Giardia duodenalis)
Concomitant culture is necessary for organism recovery and further typing of bacterial agents.
The QIAstat-Dx Gastrointestinal Panel 2 is indicated as an aid in the diagnosis of specific agents of gastrointestinal illness, in conjunction with other clinical, laboratory, and epidemiological data. Positive results do not rule-out coinfection with organisms not detected by the QIAstat-Dx Gastrointestinal Panel 2. The organisms detected may not be the sole or definitive cause of the disease.
Negative QIAstat-Dx Gastrointestinal Panel 2 results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this assay test or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease.
QIAstat-Dx is based on single-test cartridges with pre-packaged reagents including both wet and dry chemistry to handle the sample preparation and detection steps for the presence of a range of selected analytes by PCR technology. After insertion of the sample, the OIAstat-Dx assay cartridge is processed by the OIAstat-Dx Analyzer 1.0.
Here's a summary of the acceptance criteria and study details for the QIAstat-Dx Gastrointestinal Panel 2, extracted from the provided text:
Acceptance Criteria and Device Performance for QIAstat-Dx Gastrointestinal Panel 2
The acceptance criteria for the QIAstat-Dx Gastrointestinal Panel 2 can be inferred from the performance metrics (Positive Percentage Agreement - PPA and Negative Percentage Agreement - NPA) reported for both prospective and retrospective clinical studies. While explicit 'acceptance criteria' values are not provided as a separate table, the reported performance demonstrates the device's ability to meet the necessary accuracy for clinical utility. The FDA's substantial equivalence determination implies these performance characteristics were found acceptable.
Implied Acceptance Criteria (based on reported performance):
- High PPA: The device should accurately detect the target pathogens when they are present. Most reported PPA values are above 90%, with many at 100%. Even lower values like 75% for Giardia lamblia in FecalSwab are within a statistically acceptable range given the confidence intervals.
- High NPA: The device should accurately report the absence of target pathogens when they are not present, minimizing false positives. Most reported NPA values are very high, often 99% or 100%.
Table of Acceptance Criteria (Implied) and Reported Device Performance
Given that specific numerical acceptance criteria (e.g., "PPA > X%") are not explicitly stated in the document, the table below showcases the reported clinical performance which serves as evidence of meeting the implicit acceptance criteria for reliable detection and non-detection of pathogens.
| Analyte (Sample Type) | Implied Acceptance Criterion (High PPA/NPA) | Reported Performance (PPA) | 95% Confidence Interval (PPA) | Reported Performance (NPA) | 95% Confidence Interval (NPA) |
|---|---|---|---|---|---|
| Viruses | |||||
| Adenovirus F40/F41 (FecalSwab) | High PPA, High NPA | 83.3% (5/6) | 43.7-97.0% | 100.0% (1214/1214) | 99.7-100.0% |
| Adenovirus F40/F41 (Para-Pak C&S) | High PPA, High NPA | 50.0% (1/2) | 9.5-90.6% | 99.9% (703/704) | 99.2-100.0% |
| Astrovirus (FecalSwab) | High PPA, High NPA | 100.0% (3/3) | 43.9-100.0% | 100.0% (1219/1219) | 99.7-100.0% |
| Astrovirus (Para-Pak C&S) | High PPA, High NPA | 100.0% (6/6) | 61.0-100.0% | 100.0% (700/700) | 99.5-100.0% |
| Norovirus GI/GII (FecalSwab) | High PPA, High NPA | 93.9% (31/33) | 80.4-98.3% | 99.6% (493/495) | 98.6-100.0% |
| Norovirus GI/GII (Para-Pak C&S) | High PPA, High NPA | 77.8% (14/18) | 54.8-91.0% | 100.0% (399/399) | 99.1-100.0% |
| Rotavirus A (FecalSwab) | High PPA, High NPA | 91.3% (21/23) | 73.2-97.6% | 99.8% (1197/1199) | 99.4-100.0% |
| Rotavirus A (Para-Pak C&S) | High PPA, High NPA | 100.0% (3/3) | 43.9-100.0% | 99.9% (702/703) | 99.2-100.0% |
| Bacteria | |||||
| Campylobacter (FecalSwab) | High PPA, High NPA | 97.0% (65/67) | 89.8-99.2% | 99.7% (1151/1155) | 99.1-99.9% |
| Campylobacter (Para-Pak C&S) | High PPA, High NPA | 96.8% (30/31) | 83.8-99.4% | 99.7% (675/677) | 98.9-99.9% |
| Plesiomonas shigelloides (FecalSwab) | High PPA, High NPA | N/A (0/0) | N/A | 99.8% (1220/1222) | 99.4-100.0% |
| Plesiomonas shigelloides (Para-Pak C&S) | High PPA, High NPA | 83.3% (5/6) | 43.7-97.0% | 99.7% (698/700) | 99.0-99.9% |
| Salmonella (FecalSwab) | High PPA, High NPA | 87.5% (14/16) | 64.0-96.5% | 100.0% (1206/1206) | 99.7-100.0% |
| Salmonella (Para-Pak C&S) | High PPA, High NPA | 95.0% (19/20) | 76.4-99.1% | 100.0% (688/688) | 99.4-100.0% |
| Yersinia enterocolitica (FecalSwab) | High PPA, High NPA | 93.8% (15/16) | 71.7-99.0% | 99.4% (1199/1206) | 98.8-99.7% |
| Yersinia enterocolitica (Para-Pak C&S) | High PPA, High NPA | 100.0% (3/3) | 43.9-100.0% | 99.3% (698/703) | 98.4-99.7% |
| Diarrheagenic E. coli/Shigella | |||||
| Enteropathogenic E. coli (EPEC) (Para-Pak C&S) | High PPA, High NPA | 87.7% (57/65) | 77.6-93.6% | 100.0% (632/632) | 99.4-100.0% |
| Enterotoxigenic E. coli (ETEC) lt/st (FecalSwab) | High PPA, High NPA | 90.0% (9/10) | 59.6-99.2% | 99.3% (427/430) | 98.0-99.8% |
| Enterotoxigenic E. coli (ETEC) lt/st (Para-Pak C&S) | High PPA, High NPA | 90.0% (9/10) | 59.6-99.2% | 98.7% (390/395) | 97.1-99.5% |
| Shiga-like toxin E. coli (STEC) stx1/stx2 (Para-Pak C&S) | High PPA, High NPA | 83.3% (5/6) | 43.6-97.0% | 99.3% (397/400) | 97.8-99.7% |
| E. coli O157 (Para-Pak C&S) | High PPA, High NPA | N/A (0/0) | N/A | 100.0% (5/5) | 56.6-100.0% |
| Shigella/Enteroinvasive E. coli (EIEC) (FecalSwab) | High PPA, High NPA | 100.0% (10/10) | 72.3-100.0% | 100.0% (1212/1212) | 99.7-100.0% |
| Shigella/Enteroinvasive E. coli (EIEC) (Para-Pak C&S) | High PPA, High NPA | 100.0% (2/2) | 34.2-100.0% | 99.9% (703/704) | 99.2-100.0% |
| Parasites | |||||
| Cryptosporidium (FecalSwab) | High PPA, High NPA | 50.0% (2/4) | 15.0-85.0% | 100.0% (1218/1218) | 99.7-100.0% |
| Cryptosporidium (Para-Pak C&S) | High PPA, High NPA | 100.0% (6/6) | 61.0-100.0% | 99.9% (699/700) | 99.2-100.0% |
| Cyclospora cayetanensis (FecalSwab) | High PPA, High NPA | 100.0% (3/3) | 43.9-100.0% | 100.0% (1219/1219) | 99.7-100.0% |
| Cyclospora cayetanensis (Para-Pak C&S) | High PPA, High NPA | 94.7% (18/19) | 75.4-99.1% | 100.0% (687/687) | 99.4-100.0% |
| Entamoeba histolytica (FecalSwab) | High PPA, High NPA | N/A (0/0) | N/A | 100.0% (1222/1222) | 99.7-100.0% |
| Entamoeba histolytica (Para-Pak C&S) | High PPA, High NPA | N/A (0/0) | N/A | 100.0% (706/706) | 99.5-100.0% |
| Giardia lamblia (FecalSwab) | High PPA, High NPA | 75.0% (6/8) | 40.9-92.9% | 98.4% (434/441) | 96.8-99.2% |
| Giardia lamblia (Para-Pak C&S) | High PPA, High NPA | 100.0% (1/1) | 20.7-100.0% | 100.0% (406/406) | 99.1-100.0% |
Study Details:
1. Sample sizes used for the test set and the data provenance:
- Test Set (Clinical Study):
- Total Specimens: 2808
- 1939 Prospective (1222 FecalSwab, 717 Para-Pak C&S)
- 119 Prospective Archived (Norovirus GI/GII: 81, STEC: 18 plus 20 negative specimens where relevant)
- 750 Retrospective Frozen Specimens
- Data Provenance: Multi-center international study conducted at thirteen clinical sites across 5 countries (4 sites in Europe and 9 sites in USA). Specimens were collected between May and July 2021.
- Total Specimens: 2808
2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
- The document does not specify the number or qualifications of experts used to establish the ground truth. Instead, it refers to the use of "one FDA-cleared test as comparator for most analytes" and a "composite comparator consisting of either three independent FDA-cleared test methods or two independent FDA-cleared tests methods and two validated PCR assays followed by bidirectional sequencing" for others. This implies that the ground truth was established through validated diagnostic methods rather than direct expert consensus on primary samples in many cases.
3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- For analytes where a composite comparator was used (Norovirus GI/GII, ETEC, STEC, and Giardia lamblia), the ground truth was determined by the majority of the three results.
- A positive composite comparator result: based on positive results for at least two comparator tests.
- A negative composite comparator result: based on negative results for at least two comparator tests.
- For other analytes, where "one FDA-cleared test method" was used, the comparator's result directly served as the ground truth.
- For cases with insufficient sample volume for complete composite comparator testing, a "worst-case model" was applied for PPA calculation.
4. If a muti-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- This information is not applicable as the device is an in vitro diagnostic (IVD) nucleic acid test for pathogen detection, not an AI-assisted diagnostic device interpreted by human readers for medical imaging or similar tasks. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not relevant to this device.
5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Yes, the performance presented (PPA and NPA) for the QIAstat-Dx Gastrointestinal Panel 2 is a standalone performance of the algorithm/device. The device automatically interprets test results and displays a summary, without a human interpretation loop for its core function.
6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- The ground truth was established using FDA-cleared comparator methods and validated PCR assays followed by bidirectional sequencing. This acts as a robust diagnostic ground truth based on established molecular and clinical laboratory standards. Pathology or outcomes data were not explicitly mentioned as primary ground truth sources for individual pathogen detection in this context.
7. The sample size for the training set:
- The document does not explicitly state a sample size for the training set. The clinical studies describe the evaluation of the device's performance using prospective, prospective archived, and retrospective samples, implying these are test sets rather than training sets. Analytical studies such as LoD and inclusivity also use specific strains and dilutions, but these are for analytical validation, not for training a machine learning model. For IVD devices like this, the "training set" concept (as understood in AI/ML) might be less direct, relying more on extensive analytical verification of probe/primer specificity and reactivity across a wide range of strains and concentrations, and then clinical validation.
8. How the ground truth for the training set was established:
- As the training set size is not provided, the method for establishing its ground truth is also not detailed. However, for the analytical studies (which could be considered a form of "training/validation data generation" in a broader sense for IVDs), the ground truth for LoD and inclusivity was established by using culture isolates from commercial suppliers (ZeptoMetrix® and ATCC®) or clinical samples positive for target analytes. These were "prepared in human stool matrix" and tested at known concentrations validated by in-house developed and validated qPCR assays for molecular unit titers.
Ask a specific question about this device
Page 1 of 1