Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K210981

    Validate with FDA (Live)

    Device Name
    Pure Vu System
    Date Cleared
    2021-04-29

    (28 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Pure-Vu System is intended to connect to standard or slim colonoscopes and gastroscopes to facilitate intra-procedural cleansing of the GI tract by irrigating and evacuating the irrigation fluids (water), bodily fluids and other debris, e.g. blood.
    It is for use only by trained medical personnel located in hospitals, clinics and doctors' offices.

    Device Description

    The Pure-Vu System enables cleansing of the digestive tract during endoscopy using a standard or slim colonoscope with a length of 1630mm - 1710mm and an outer diameter range of 11.7mm - 13.3mm (slim) and 12.8mm - 13.7mm (standard), or gastroscope with length of 950-1030mm and an outer diameter of 10mm - 10.9mm. The Oversleeve, which fits over the endoscope and is connected to an external Workstation, generates fluid and gas to break up debris. The debris & fluids are removed through the suction channels of the Oversleeve into an external waste container.
    The Pure-Vu System consists of the following main components:
    • Oversleeve (OS) and Umbilical Section (US) - The disposable Oversleeve is mounted on Standard or Slim commercially available colonoscopes or gastroscopes to allow a physician to cleanse the GI tract and is connected to the external Workstation via a disposable US.
    • Workstation (WS) – The Workstation [WS] is reusable and supplies an irrigation mixture of water/saline and gas, and evacuates bodily fluid and matter. The Workstation [WS] includes the following components: A monitoring & Control Unit that controls the delivery of irrigation fluids and gas into the colon, and suction of bodily fluid and matter from the GI tract. Inlet Module that includes pumps and regulators enabling fluid & gas flow into the cleansing device. Outlet Module that includes pumps to evacuate fluid and matter from the Gl tract. A foot pedal activates the cleansing, suction and purging function, and switch between cleansing modes used by the physician.
    • Loading fixture - The loading fixture is reusable and aids in assembling the Oversleeve onto the endoscope. Unloading aids - A disposable luer lock syringe/ duckbill check valve assembly used in unloading the Oversleeve from the endoscope.

    AI/ML Overview

    This document is a 510(k) summary for the Motus GI Pure-Vu System, which is a medical device used for intra-procedural cleansing of the GI tract during endoscopy. The document describes the device, its intended use, and the performance data that supports its substantial equivalence to a predicate device.

    Here's an analysis of the requested information based on the provided text, keeping in mind that this is a 510(k) submission and not a typical clinical study report for AI-powered devices. The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to the engineering and design verification tests conducted for a medical device seeking 510(k) clearance, particularly as it relates to demonstrating substantial equivalence for design modifications. This is not a study for an AI algorithm's performance.

    1. A table of acceptance criteria and the reported device performance

    TestAcceptance Criteria (same as for predicate device)Reported Device Performance
    Steering Test (Gastro Oversleeve)Maximum 25% angle difference between a naked gastroscope and a gastroscope with an Oversleeve.Design verification and validation testing concluded that the design changes have no impact on the Pure-Vu System performance. (Implies criteria met)
    Head Pull Test (Gastro Oversleeve)Gastro Oversleeve shall not move linearly in relation to the gastroscope with less than 10N of force applied.Design verification and validation testing concluded that the design changes have no impact on the Pure-Vu System performance. (Implies criteria met)
    Loading and maintenance of pressure during Gastro Oversleeve loading procedure1. 100% successful loading procedures. 2. No damage to the flexible head identified by 100% visual inspection. 3. Maintain pressure of up to 400 mBar. 4. No damage to the gastroscope in 100% of the loading procedures verified by examining images using the gastroscope's optics.Design verification and validation testing concluded that the design changes have no impact on the Pure-Vu System performance. (Implies criteria met)

    Note: The document states that "Design verification and validation testing concluded that the design changes have no impact on the Pure-Vu System performance." This statement implies that the device met all the specified acceptance criteria for these tests. Specific numerical results validating each criterion are not provided in this summary, which is common for 510(k) summaries where the focus is on demonstrating substantial equivalence rather than presenting detailed raw data.

    2. Sample sized used for the test set and the data provenance

    The document does not explicitly state the sample sizes used for each of the performance tests (Steering Test, Head Pull Test, Loading and Pressure Maintenance Test). These are engineering verification and validation tests, not clinical studies. Therefore, data provenance in terms of country of origin or retrospective/prospective is not applicable in the same way it would be for a clinical trial involving patient data. These tests would have been performed in a lab setting by the manufacturer (Motus GI Medical Technologies Ltd., Israel).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable. The "ground truth" for these performance tests is based on objective, measurable physical properties and engineering specifications, not expert interpretation or clinical judgment. For example, angle differences are measured, force applied is measured, and visual inspections determine damage, not subjective expert opinion.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. As noted above, these are objective engineering tests, not involving human interpretation or adjudication processes like those in clinical studies.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a mechanical system for cleaning the GI tract, not an AI algorithm for image analysis or diagnosis. Therefore, an MRMC study and AI assistance are irrelevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a mechanical system, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" in this context is based on engineering specifications and physical measurements. For example:

    • Steering test: Measurement of physical angles and direct comparison.
    • Head Pull Test: Measurement of force required for linear movement.
    • Loading and maintenance of pressure test: Measurement of pressure, visual inspection for damage, and direct observation of loading success.

    8. The sample size for the training set

    Not applicable. This device is a mechanical system, not an AI-powered device that requires a "training set."

    9. How the ground truth for the training set was established

    Not applicable. As there is no training set for an AI algorithm, this question is not relevant.

    In summary:

    This 510(k) submission focuses on demonstrating the substantial equivalence of modifications to an existing mechanical medical device (Pure-Vu System) as defined by its physical performance characteristics. The "acceptance criteria" and "study" described are engineering and design verification tests to ensure the modified device performs as intended and introduces no new safety or effectiveness concerns, thereby maintaining substantial equivalence to the predicate device. It is crucial to understand that this document does not pertain to the validation of an AI/ML algorithm or its clinical performance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191220

    Validate with FDA (Live)

    Device Name
    Pure Vu System
    Date Cleared
    2019-06-06

    (30 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pure Vu System is intended to connect to standard colonoscopes to help facilitate intra-procedural cleaning of a poorly prepared colon by irrigating or cleaning the colon and evacuating the irrigation fluid (water), feces and other bodly fluids and matter, e.g. blood.

    It is for use only by trained medical personnel located in hospitals, clinics and doctor offices.

    Device Description

    The Pure-Vu System enables colon cleansing during colonoscopy using a standard or slim colonoscope with a length of 1630mm – 1710mm and a slim colonoscope outer diameter range of 11.7mm – 13.3mm and a standard colonoscope outer diameter range of 12.8mm – 13.7mm. The Oversleeve, which fits over the colonoscope and is connected to an external Workstation, generates fluid to break up feces. The fecal matter & fluids are removed through the suction channels of the Oversleeve into an external waste receptacle.
    The Pure-Vu System consists of the following main components:
    Oversleeve (OS) and Umbilical Section (US) - The disposable Oversleeve is mounted on Standard or Slim commercially available colonoscopes to allow a physician to cleanse the colon and is connected to the external Workstation via a disposable US. Workstation (WS) – The Workstation [WS] is reusable and supplies an irrigation mixture of water/saline and gas, and evacuates fecal material & fluids. The Workstation [WS] includes the following components: A monitoring & Control Unit that controls the delivery of irrigation fluids and gas into the colon, and suction of fluids and feces from the colon. Inlet Module that includes pumps and regulators enabling fluid & gas flow into the cleansing device. Outlet Module that includes pumps to evacuate fecal matter and fluids from the colon. A foot pedal activates the cleansing, suction and purging function, and switch between cleansing modes used by the physician. Loading fixture- The loading fixture is reusable and aids in assembling the Oversleeve onto the colonoscope. Unloading aids - A disposable 60cc lure lock syringe and 3-way stopcock used in unloading the Oversleeve from the colonoscope are provided in the disposable packaging.

    AI/ML Overview

    The provided document describes the Pure-Vu System, a device intended to facilitate intra-procedural cleaning of a poorly prepared colon during colonoscopy. This submission is a special 510(k) to improve the overall ease of use of the system by reducing the size and weight of the Workstation and making other minor modifications to the Oversleeve and loading fixture. The document asserts that these changes do not raise new issues of safety or effectiveness.

    Here's an analysis of the acceptance criteria and study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly present a table of acceptance criteria with specific numerical targets. Instead, it describes various tests performed to demonstrate that the modifications did not negatively impact the device's performance and that it remains substantially equivalent to the predicate device. The performance is reported as meeting "requirements" or demonstrating "efficiency and integrity."

    Acceptance Criteria (Implied)Reported Device Performance
    Functionality after simulated aging (1 year)"Pure Vu Disposables functionality meets the requirements following a simulated aging of one year."
    Sub-system and system functionality as intended"Simulation (bench test) was conducted and to support the sub system and system's use as intended and its substantial equivalence." (Specific tests like Dimensions, Steering (Angulation), Pressure, Bond Strength were conducted).
    Packaging efficiency and integrity after environmental conditioning and transport simulation"Disposable package undergone environmental conditioning and transport simulation (by DDL labs) in order to demonstrate efficiency and integrity of the packaging and their accompanying labels following the simulation."
    Absence of toxicological risk (Biocompatibility)"biocompatibility testing was done to determine if there was any toxicological risk to the patient as per ISO 10993-5." (Implied acceptance: no toxicological risk).
    Workstation functionality after environmental conditions and distribution cycle simulation"After the preconditioning simulations, the Pure Vu WS undergone verification testing including visual inspection and functionality tests." (Implied acceptance: device remains functional).
    Software validation for updated embedded software"The embedded updated software to comply with the WS design modifications was tested via software validation." (Implied acceptance: software functions correctly and complies with design).
    Safety and EMC compliance"Safety and EMC tests per IEC 60601." (Implied acceptance: device complies with relevant safety and electromagnetic compatibility standards).
    Usability and validation for WS and disposable device"Usability and validation testing for the WS and disposable device has been performed in two separate studies with 3 physicians and 15 nurses." (Implied acceptance: usable and validated by intended users).
    Loading and unloading validation"Validation testing for the new sealing plug used during loading and unloading was performed." (Implied acceptance: the new component functions as intended for loading and unloading). Also "Environmental conditions and distribution cycle simulation performed by DDL (USA) as per ASTM D4169-16 along with shipping integrity. After the preconditioning simulations, the Pure Vu WS undergone verification testing including visual inspection and functionality tests." (Implied acceptance: robust against environmental stresses and functional).
    Overall design changes not impacting performance"Design verification and validation testing concluded that the design changes have no impact on the Pure-Vu System performance."
    Substantial Equivalence to Predicate Device"substantially equivalent"

    2. Sample size used for the test set and the data provenance:

    • Sample Size:
      • Usability Testing: 3 physicians and 15 nurses for two separate studies for the Workstation and disposable device.
      • Other tests (e.g., disposable units, disposable package, workstation environmental, loading/unloading) do not specify a "sample size" in terms of clinical cases or patients. These are primarily engineering verification and validation tests performed on manufactured units.
    • Data Provenance: The document does not specify the country of origin for the participants in the usability studies or whether these studies were retrospective or prospective. Bench testing was performed by various laboratories (Carmel Environmental Laboratory, DDL labs, Hermon Laboratory, DDL (USA)).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • For the usability studies, the "experts" were the users themselves: 3 physicians and 15 nurses. Their qualifications are not further detailed beyond their professional roles.
    • For other engineering tests, the acceptance criteria are likely against established engineering specifications and regulations (e.g., ISO, ASTM, IEC standards), so no explicit "ground truth experts" in the medical sense are mentioned for these tests.

    4. Adjudication method for the test set:

    The document does not describe any formal adjudication method (e.g., 2+1, 3+1 consensus) for establishing ground truth, particularly in a medical diagnostic context. The usability studies would likely involve observation and feedback collection, but not a consensus determination of a medical finding.

    5. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done, and the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is an intra-procedural cleaning system for colonoscopy, not an AI-assisted diagnostic imaging device requiring human reader interpretation studies.

    6. If a standalone (i.e. algorithm only, without human-in-the-loop performance) was done:

    Not applicable. This is a medical device for physical intervention (cleaning), not a standalone diagnostic algorithm. The device works with a human operator (trained medical personnel).

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    For the usability studies, the "ground truth" would be user feedback and observed performance regarding ease of use and functionality, rather than a medical diagnostic ground truth. For the engineering tests, the ground truth is against established technical specifications, regulatory standards (e.g., ISO 10993 for biocompatibility, IEC 60601 for safety/EMC), and internal functional requirements.

    8. The sample size for the training set:

    Not applicable. This device is not an AI/ML algorithm that requires a training set.

    9. How the ground truth for the training set was established:

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181437

    Validate with FDA (Live)

    Device Name
    Pure Vu System
    Date Cleared
    2018-06-21

    (20 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pure Vu System is intended to connect to standard colonoscopes to help facilitate intra-procedural cleaning of a poorly prepared colon by irrigating or cleaning the colon and evacuating the irrigation fluid (water), feces and other bodily fluids and matter, e.g. blood.
    It is for use only by trained medical personnel located in hospitals, clinics and doctor offices.

    Device Description

    The Pure Vu system comprises the following components:
    Oversleeve - The Oversleeve is mounted on Standard or Slim commercially available colonoscopes to allow a physician to cleanse the colon.
    Workstation (WS) - The Workstation operates by using simultaneous irrigation and evacuation of colon content. Irrigation is based on a mixture of liquid (water or saline) and gas (air). The workstation includes:
    Monitoring & Control Unit that continuously monitors and controls irrigation and evacuation. Inlet Module that includes pumps and regulators enabling fluid & gas flow into the cleansing device. Outlet Module that includes pumps to evacuate fecal matter and fluids from the colon. External foot pedals that control the cleansing process to be operated by a physician.
    WS Connector (WSC) connects the Oversleeve to the WS and saline or water bag.
    Loading fixture to aid the nurse in mounting the Oversleeve onto a colonoscope.

    AI/ML Overview

    This 510(k) submission is for the Pure Vu System, which is intended to connect to standard colonoscopes to facilitate intra-procedural cleaning of a poorly prepared colon. The submission is for a modification of the device to comply with Slim colonoscopes and extends the hydrophilic coating on the Oversleeve.

    Here's a breakdown of the requested information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in a quantitative manner for specific performance metrics of the device itself (e.g., a specific percentage improvement in cleaning, or a defined force of suction).

    However, the "Performance Data" section states: "Performance tests were conducted for all modifications to the Pure Vu System. Specifically, the company performed the following bench tests: Steering test, Head pull test, Loading and maintenance of pressure during the loading procedure."

    The "Substantial Equivalence Discussion" then concludes: "Performance data demonstrate that the Pure-Vu System is substantially equivalent." This implies that the results of these bench tests met internal acceptance criteria for demonstrating that the modified device performs comparably to the predicate device. Without further details on the specific values or thresholds used in these tests, a quantitative table of acceptance criteria and reported performance cannot be generated from this document.

    2. Sample Size Used for the Test Set and Data Provenance

    The document only mentions "bench tests" for performance data. It does not provide details on specific sample sizes for these tests, nor does it indicate the provenance of any data (e.g., country of origin, retrospective/prospective). As an accessory to a medical device (colonoscope), it's possible that clinical studies involving patient data were not part of this specific 510(k) submission focused on modifications.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. The performance data mentioned are "bench tests," which typically do not involve establishing ground truth by human experts in the way clinical studies do.

    4. Adjudication Method for the Test Set

    This information is not provided in the document. As the performance data are "bench tests," an adjudication method by human experts is not applicable in this context.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study is not mentioned in this document. The submission focuses on modifications to an accessory device and bench testing for substantial equivalence, not a comparative effectiveness study involving human readers and AI.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No, the Pure Vu System is a mechanical device designed to assist a human operator (trained medical personnel) during a colonoscopy. It is not an AI algorithm with standalone performance.

    7. The Type of Ground Truth Used

    For the mentioned "bench tests" (Steering test, Head pull test, Loading and maintenance of pressure during the loading procedure), the "ground truth" would be the engineering specifications and expected physical properties or performance thresholds of the device components. This is not "expert consensus, pathology, or outcomes data" in the typical sense for medical image analysis or diagnostic devices.

    8. The Sample Size for the Training Set

    No training set is mentioned as this device is not an AI or machine learning algorithm requiring training data.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set mentioned.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1