Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K210981
    Device Name
    Pure Vu System
    Date Cleared
    2021-04-29

    (28 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Pure-Vu System is intended to connect to standard or slim colonoscopes and gastroscopes to facilitate intra-procedural cleansing of the GI tract by irrigating and evacuating the irrigation fluids (water), bodily fluids and other debris, e.g. blood.
    It is for use only by trained medical personnel located in hospitals, clinics and doctors' offices.

    Device Description

    The Pure-Vu System enables cleansing of the digestive tract during endoscopy using a standard or slim colonoscope with a length of 1630mm - 1710mm and an outer diameter range of 11.7mm - 13.3mm (slim) and 12.8mm - 13.7mm (standard), or gastroscope with length of 950-1030mm and an outer diameter of 10mm - 10.9mm. The Oversleeve, which fits over the endoscope and is connected to an external Workstation, generates fluid and gas to break up debris. The debris & fluids are removed through the suction channels of the Oversleeve into an external waste container.
    The Pure-Vu System consists of the following main components:
    • Oversleeve (OS) and Umbilical Section (US) - The disposable Oversleeve is mounted on Standard or Slim commercially available colonoscopes or gastroscopes to allow a physician to cleanse the GI tract and is connected to the external Workstation via a disposable US.
    • Workstation (WS) – The Workstation [WS] is reusable and supplies an irrigation mixture of water/saline and gas, and evacuates bodily fluid and matter. The Workstation [WS] includes the following components: A monitoring & Control Unit that controls the delivery of irrigation fluids and gas into the colon, and suction of bodily fluid and matter from the GI tract. Inlet Module that includes pumps and regulators enabling fluid & gas flow into the cleansing device. Outlet Module that includes pumps to evacuate fluid and matter from the Gl tract. A foot pedal activates the cleansing, suction and purging function, and switch between cleansing modes used by the physician.
    • Loading fixture - The loading fixture is reusable and aids in assembling the Oversleeve onto the endoscope. Unloading aids - A disposable luer lock syringe/ duckbill check valve assembly used in unloading the Oversleeve from the endoscope.

    AI/ML Overview

    This document is a 510(k) summary for the Motus GI Pure-Vu System, which is a medical device used for intra-procedural cleansing of the GI tract during endoscopy. The document describes the device, its intended use, and the performance data that supports its substantial equivalence to a predicate device.

    Here's an analysis of the requested information based on the provided text, keeping in mind that this is a 510(k) submission and not a typical clinical study report for AI-powered devices. The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to the engineering and design verification tests conducted for a medical device seeking 510(k) clearance, particularly as it relates to demonstrating substantial equivalence for design modifications. This is not a study for an AI algorithm's performance.

    1. A table of acceptance criteria and the reported device performance

    TestAcceptance Criteria (same as for predicate device)Reported Device Performance
    Steering Test (Gastro Oversleeve)Maximum 25% angle difference between a naked gastroscope and a gastroscope with an Oversleeve.Design verification and validation testing concluded that the design changes have no impact on the Pure-Vu System performance. (Implies criteria met)
    Head Pull Test (Gastro Oversleeve)Gastro Oversleeve shall not move linearly in relation to the gastroscope with less than 10N of force applied.Design verification and validation testing concluded that the design changes have no impact on the Pure-Vu System performance. (Implies criteria met)
    Loading and maintenance of pressure during Gastro Oversleeve loading procedure1. 100% successful loading procedures. 2. No damage to the flexible head identified by 100% visual inspection. 3. Maintain pressure of up to 400 mBar. 4. No damage to the gastroscope in 100% of the loading procedures verified by examining images using the gastroscope's optics.Design verification and validation testing concluded that the design changes have no impact on the Pure-Vu System performance. (Implies criteria met)

    Note: The document states that "Design verification and validation testing concluded that the design changes have no impact on the Pure-Vu System performance." This statement implies that the device met all the specified acceptance criteria for these tests. Specific numerical results validating each criterion are not provided in this summary, which is common for 510(k) summaries where the focus is on demonstrating substantial equivalence rather than presenting detailed raw data.

    2. Sample sized used for the test set and the data provenance

    The document does not explicitly state the sample sizes used for each of the performance tests (Steering Test, Head Pull Test, Loading and Pressure Maintenance Test). These are engineering verification and validation tests, not clinical studies. Therefore, data provenance in terms of country of origin or retrospective/prospective is not applicable in the same way it would be for a clinical trial involving patient data. These tests would have been performed in a lab setting by the manufacturer (Motus GI Medical Technologies Ltd., Israel).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable. The "ground truth" for these performance tests is based on objective, measurable physical properties and engineering specifications, not expert interpretation or clinical judgment. For example, angle differences are measured, force applied is measured, and visual inspections determine damage, not subjective expert opinion.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. As noted above, these are objective engineering tests, not involving human interpretation or adjudication processes like those in clinical studies.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a mechanical system for cleaning the GI tract, not an AI algorithm for image analysis or diagnosis. Therefore, an MRMC study and AI assistance are irrelevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a mechanical system, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" in this context is based on engineering specifications and physical measurements. For example:

    • Steering test: Measurement of physical angles and direct comparison.
    • Head Pull Test: Measurement of force required for linear movement.
    • Loading and maintenance of pressure test: Measurement of pressure, visual inspection for damage, and direct observation of loading success.

    8. The sample size for the training set

    Not applicable. This device is a mechanical system, not an AI-powered device that requires a "training set."

    9. How the ground truth for the training set was established

    Not applicable. As there is no training set for an AI algorithm, this question is not relevant.

    In summary:

    This 510(k) submission focuses on demonstrating the substantial equivalence of modifications to an existing mechanical medical device (Pure-Vu System) as defined by its physical performance characteristics. The "acceptance criteria" and "study" described are engineering and design verification tests to ensure the modified device performs as intended and introduces no new safety or effectiveness concerns, thereby maintaining substantial equivalence to the predicate device. It is crucial to understand that this document does not pertain to the validation of an AI/ML algorithm or its clinical performance.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1