510(k) Data Aggregation

    K Number
    K253323

    Validate with FDA (Live)

    Device Name
    Protrieve Sheath
    Manufacturer
    Date Cleared
    2025-10-29

    (29 days)

    Product Code
    Regulation Number
    870.5150
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ClotTriever Thrombectomy System is indicated for:

    • The non-surgical removal of thrombi and emboli from blood vessels.
    • Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel.

    The ClotTriever Thrombectomy System is intended for use in the peripheral vasculature including deep vein thrombosis (DVT).

    The Protrieve Sheath is indicated for use as a conduit for the insertion of endovascular devices into the vasculature while minimizing blood loss associated with such insertions.

    Device Description

    The Protrieve Sheath is a sterile, single-use over-the-wire introducer sheath with a hydrophilic coating, distal self-expanding funnel, aspiration port, and proximal hub. A dilator is provided to aid insertion and positioning of the sheath. Other provided accessories include a 60cc large bore syringe that provides a vacuum source and collects aspirated contents. Radiopaque markers aid sheath positioning under fluoroscopic visualization. The dilator tip is radiopaque, and there is a radiopaque marker band near the distal end of the sheath. Target vessels include, but are not limited to, the iliofemoral, upper and lower extremity, inferior vena cava (IVC), and superior vena cava (SVC).

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K230331

    Validate with FDA (Live)

    Device Name
    Protrieve Sheath
    Date Cleared
    2023-03-07

    (28 days)

    Product Code
    Regulation Number
    870.5150
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ClotTriever Thrombectomy System is indicated for:

    • · The non-surgical removal of thrombi and emboli from blood vessels.
    • · Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel.
      The ClotTriever Thrombectomy System is intended for use in the peripheral vasculature including deep vein thrombosis (DVT).
      The Protrieve Sheath is indicated for use as a conduit for the insertion of endovascular devices into the vasculature while minimizing blood loss associated with such insertions.
    Device Description

    The Protrieve Sheath is a sterile, single-use over-the-wire introducer sheath with a hydrophilic coating, distal self-expanding funnel, aspiration port, and proximal hub. A dilator is provided to aid insertion and positioning of the sheath. Other provided accessories include a 60 cc large bore syringe that provides a vacuum source and collects aspirated contents. Radiopaque markers aid sheath positioning under fluoroscopic visualization. The dilator tip is radiopaque, and there is a radiopaque marker band near the distal end of the sheath. Target vessels include, but are not limited to, the iliofemoral, upper and lower extremity, inferior vena cava (IVC), and superior vena cava (SVC).

    AI/ML Overview

    The FDA document for the Protrieve Sheath (K230331) details acceptance criteria and performance data primarily for non-clinical testing related to a design modification. The submission emphasizes substantial equivalence to a predicate device and an expansion of the Indications for Use.

    It's important to note that this document does not describe a study involving AI assistance for human readers, nor does it detail a standalone algorithm performance study. The device in question is a medical sheath, not an AI/ML powered diagnostic device. Therefore, many of the requested points, such as MRMC studies, expert ground truth for imaging, and training set details, are not applicable to this type of medical device submission.

    Here's the information gleaned from the provided text regarding acceptance criteria and performance, as applicable to this medical device:

    1. Table of Acceptance Criteria and Reported Device Performance

    Device: Protrieve Sheath (K230331)

    Acceptance Criteria CategorySpecific Test PerformedAcceptance CriteriaReported Device Performance
    Functional PerformanceSimulated Use, Flushing – Protrieve SheathNot explicitly stated in terms of specific numeric or qualitative thresholds for flushing."Test results demonstrated that all acceptance criteria were met."
    Functional PerformanceLeakage Testing – Protrieve SheathNot explicitly stated in terms of specific numeric or qualitative thresholds for leakage."Test results demonstrated that all acceptance criteria were met."
    Mechanical PerformanceTensile Testing – Protrieve SheathNot explicitly stated in terms of specific numeric thresholds for tensile strength."Test results demonstrated that all acceptance criteria were met."

    Note: The document states that "verification tests were identified and performed only for the attributes impacted by the slide actuator modification to support substantial equivalence." This implies that the acceptance criteria for these tests were established to ensure the modified device performs comparably to the previous version and the predicate device. However, the specific quantitative or detailed qualitative criteria are not explicitly listed in the provided summary. The general statement "Test results demonstrated that all acceptance criteria were met" confirms successful performance against these internal criteria.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: The document does not specify the exact number of units or test iterations conducted for the "Simulated Use, Flushing," "Leakage Testing," and "Tensile Testing." In non-clinical device testing, sample sizes are typically determined by statistical methods to ensure robustness and often align with recognized standards.
    • Data Provenance: This was a non-clinical bench testing study, not a clinical study involving patient data. Therefore, data provenance in terms of country of origin or retrospective/prospective collection is not applicable. The testing was performed by Inari Medical, Inc. to support regulatory submission.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Not Applicable. This submission pertains to a physical medical device (introducer sheath) and its performance in non-clinical bench testing. There is no "ground truth" established by experts in the context of diagnostic outcomes or image interpretation, as would be relevant for an AI/ML-based device. The "ground truth" for these tests are the pre-defined performance specifications of the device.

    4. Adjudication Method for the Test Set

    • Not Applicable. As this is non-clinical bench testing, there is no need for expert adjudication of results. The results are typically quantitative measurements against pre-defined engineering specifications.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • No. An MRMC study is relevant for diagnostic devices, particularly those involving image interpretation where human readers' performance with and without AI assistance is compared. The Protrieve Sheath is an introducer sheath; therefore, an MRMC study was not performed. The document explicitly states "Clinical testing was not required to support substantial equivalence."

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    • No. This is not an AI/ML algorithm. Therefore, a standalone algorithm performance study was not conducted.

    7. The Type of Ground Truth Used

    • Engineering Specifications and Industry Standards: The "ground truth" for the non-clinical testing (Simulated Use, Flushing, Leakage, Tensile Testing) would be the device's technical specifications and relevant performance requirements derived from engineering design, risk analysis (FMEA), and applicable medical device standards. The statement "all acceptance criteria were met" indicates performance aligned with these pre-defined standards.

    8. The Sample Size for the Training Set

    • Not Applicable. This is a medical device, not an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set was Established

    • Not Applicable. As per point 8, there is no training set for this type of medical device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K220415

    Validate with FDA (Live)

    Device Name
    Protrieve Sheath
    Date Cleared
    2022-07-12

    (148 days)

    Product Code
    Regulation Number
    870.5150
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ClotTriever Thrombectomy System is indicated for:

    • · The non-surgical removal of thrombi and emboli from blood vessels.
    • · Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel.

    The ClotTriever Thrombectomy System is intended for use in the peripheral vasculature including deep vein thrombosis (DVT).

    Device Description

    The ClotTriever Thrombectomy System is a single-use, sterile medical device designed to remove thrombi and emboli from the peripheral vasculature. The ClotTriever Thrombectomy System consists of ClotTriever 13 Fr and 16 Fr Sheaths and Protrieve Sheath ("Sheath"), and the ClotTriever/ClotTriever Bold Catheter (“Catheter"), each packaged separately.

    The sheath is an introducer sheath with a distal self-expanding funnel, aspiration port, and proximal hub. A dilator is provided to aid insertion and positioning of the sheath. Other provided accessories include a 60 cc large bore syringe that provides a vacuum source and collects aspirated contents. Radiopaque markers aid sheath positioning under fluoroscopic visualization. The dilator tip is radiopaque, and there is a radiopaque marker band near the distal end of the sheath. Target vessels include, but are not limited to, the iliofemoral, upper and lower extremity, inferior vena cava (IVC), and superior vena cava (SVC).

    AI/ML Overview

    The provided text is a 510(k) summary for the Protrieve Sheath, a medical device. It details the device's technical characteristics and the non-clinical testing performed to establish substantial equivalence to a predicate device.

    Crucially, this document states: "Clinical testing was not required to support substantial equivalence." This means there was no clinical study involving human subjects for this specific submission to prove the device meets performance criteria, especially as it relates to AI/machine learning performance.

    Therefore, I cannot fulfill your request to describe the acceptance criteria and the study proving the device meets those criteria in the context of an AI/ML device, as the provided document does not pertain to an AI/ML device or a clinical study for performance.

    The acceptance criteria mentioned in the document relate to the non-clinical (engineering and material) testing of the physical device to demonstrate mechanical and material properties are consistent with the predicate device, not performance benchmarks for an AI algorithm.

    If you would like to know about the non-clinical acceptance criteria and the tests performed for this specific device, I can provide that information based on the text. However, it will not involve AI/ML performance, ground truth, expert consensus, or clinical effectiveness studies in the way your prompt describes.