510(k) Data Aggregation

    K Number
    K221425

    Validate with FDA (Live)

    Device Name
    Primary Relief
    Manufacturer
    Date Cleared
    2022-09-13

    (120 days)

    Product Code
    Regulation Number
    882.5890
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Primary Relief is a percutaneous electrical nerve stimulator (PENS) system indicated for up to 3 days for symptomatic relief of post-operative pain following cesarean section delivery or up to 3 days adjunctive symptomatic relief of post-operative pain following cardiac surgery.

    Device Description

    The Primary Relief is designed to aid in the reduction of post-operative pain by the method of cranial electrical stimulation at the auricular stimulation points. The Primary Relief is a wearable, battery-operated device that is designed to administer periodical low level electrical pulses to the ear over three days / 72 hours from the time of activation of the device. The electrical pulse from the device will be delivered to the branches of cranial nerves on the ear through a set of wire assembly and stimulation needles. Three zinc air batteries with 1.4 V each provide the required stimulation energy for 72 hours. There stimulation electrode and one ground electrode which constitute of a needle and lead/ wire with the snap-fit ring. The stimulation needles are inserted at three specific points, which have the ability to stimulate the cranial nerves. The ground electrode is inserted at one specific point (constant in all treatments) which forms the functional earthing to the device. This constant current source guarantees equivalent stimulation energy regardless of the individual impedance of the stimulation pattern consists of rectangular pulses with differing inter-pulse intervals. A 3-pin connector is provided, which is used to check the output voltage of the device once it is activated and before applying to the patient with any one of the voltage measuring devices available in the market with the appropriate regulatory compliance.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Primary Relief device, based on the provided FDA 510(k) summary:

    Description of Acceptance Criteria and Proving Study

    The acceptance criteria for the Primary Relief device appear to be primarily focused on demonstrating its effectiveness in reducing post-operative pain, specifically after cardiac surgery, to support its expanded indications for use. The study aimed to show a statistically significant reduction in pain scores compared to a sham device.


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied)Reported Device Performance (Primary Endpoints)
    Statistically significant reduction in pain intensity at rest.Group A (Treatment) score was 1.19 units (SE=0.16; P-value < 0.001) less than the control group (Group B), for the group-wise change in pain score at rest. This indicates a statistically significant improvement in pain relief for the treatment group at rest.
    Statistically significant reduction in pain intensity during cough.Group A (Treatment) score was 1.25 units (SE=0.23; P-value < 0.001) less than the control group (Group B), for the group-wise change in the pain score during cough. This also indicates a statistically significant improvement in pain relief for the treatment group during cough.
    Safety: No serious adverse events.No complications or adverse events were observed in any of the subjects during the study period across both arms. Device met safety expectations.
    Effect maintained throughout the study period.The effect was maintained throughout the study period and during rest and cough situations. This suggests sustained efficacy over the 72-hour treatment period.

    2. Sample size used for the test set and the data provenance

    • Sample Size: 60 subjects (29 in the intervention group and 31 in the control/sham group).
    • Data Provenance: Single center, prospective study. The country of origin is not explicitly stated, but the submitter's address is in San Mateo, California, USA, and the FDA is a US regulatory body, suggesting the study likely took place in the US or was conducted under US regulatory standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The ground truth for pain was established through Numeric Rating Scale (NRS) scores reported by the patients themselves. This is a patient-reported outcome measure and does not involve expert adjudication in the traditional sense for image interpretation or diagnosis. The study notes that the primary and secondary endpoints were measured using "validated and widely used tools."


    4. Adjudication method for the test set

    There was no external adjudication method (like 2+1 or 3+1) described for the pain scores, as these are direct patient-reported outcomes via NRS. The study itself was a double-blinded prospective study, meaning neither the subjects nor the study staff administering the treatment (and presumably collecting pain scores) knew whether the patient received the active or sham device. This blinding helps maintain the integrity of the data collection for patient-reported outcomes.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC study was not done. This study is for a medical device (Transcutaneous Electrical Nerve Stimulator for Pain Relief), not an AI-powered diagnostic tool requiring human reader interpretation of images. Therefore, the concept of "human readers improve with AI vs without AI assistance" is not applicable here.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Yes, in spirit, a standalone performance was assessed. The device (Primary Relief) delivers electrical stimulation for pain relief. Its performance is directly measured by its effect on patient pain scores. While a human applies the device, the device itself functions autonomously for 72 hours, delivering stimulation. The clinical study evaluated its direct effect (standalone effect) on pain scores compared to a sham device.

    7. The type of ground truth used

    The ground truth for the primary efficacy endpoint (pain intensity) was patient-reported outcomes using the Numeric Rating Scale (NRS). For secondary efficacy endpoints, the summary mentions "five measures intended to assess the pulmonary function," which would also be objective physiological measurements or patient-reported outcomes, but the specific ground truth establishment method for these is not detailed in the summary.


    8. The sample size for the training set

    The document does not explicitly mention a separate "training set" in the context of the clinical study for the expanded indication. The 60 subjects were part of the test set or the clinical validation study for this specific 510(k) submission. For medical devices like this, there isn't typically a "training set" in the machine learning sense for the device's functionality. The "training" for the device's design and parameters would have occurred during its development and prior predicate clearance (K213188).


    9. How the ground truth for the training set was established

    As there's no explicitly defined "training set" for an AI model in this context, the question regarding its ground truth establishment is not directly applicable. If we consider the development of the device itself (including its electrical parameters), the ground truth would have been established through preclinical studies, engineering specifications, and previous clinical trials that led to the predicate device's clearance (K213188), demonstrating its general safety and effectiveness for a similar indication (post-operative pain following cesarean section delivery).

    Ask a Question

    Ask a specific question about this device

    K Number
    K213188

    Validate with FDA (Live)

    Device Name
    Primary Relief
    Manufacturer
    Date Cleared
    2022-01-31

    (124 days)

    Product Code
    Regulation Number
    882.5890
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Primary Relief is a percutaneous electrical nerve stimulator (PENS) system indicated for up to 3 days for symptomatic relief of post-operative pain following cesarean section delivery.

    Device Description

    The Primary Relief is a wearable, battery-operated device that is designed to administer periodical low level electrical pulses to the ear over 3 days / 72 hours from the time of activation of the device. The electrical pulse from the device will be delivered to the branches of cranial nerves on the ear through a set of wire assembly and stimulation needles. Three zinc air batteries with 1.4 V each provide the required stimulation energy for 72 hours. There are three stimulation electrode and one ground electrode which constitute of a needle and lead/ wire with the snap-fit ring. The stimulation needles are inserted at three specific points, which have the ability to stimulate the cranial nerves. The ground electrode is inserted at one specific point (constant in all treatments) which forms the functional earthing to the device. This constant current source guarantees equivalent stimulation energy regardless of the individual impedance of the skin. The stimulation pattern consists of rectangular pulses with differing inter-pulse intervals. A 3-pin connector is provided, which is used to check the output voltage of the device once it is activated and before applying to the patient with any one of the voltage measuring devices available in the market with the appropriate regulatory compliance.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the DyAnsys Primary Relief device, based on the provided FDA 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document doesn't explicitly list "acceptance criteria" in the traditional sense of specific quantitative thresholds the device needed to meet. Instead, it focuses on demonstrating substantial equivalence to a predicate device (SPRINT PNS System) and safety/effectiveness through clinical and non-clinical testing.

    Here’s a table summarizing the performance aspects evaluated and the reported results:

    Feature/CriterionAcceptance/Evaluation ApproachReported Device PerformanceComments
    Preamble (Substantial Equivalence)Comparison to predicate device (SPRINT PNS System K202660) and reference device (Drug Relief K173861) based on intended use, indications for use, and basic operating principle.Demonstrated substantial equivalence to identified predicate device.Formally stated as the conclusion of the 510(k) submission.
    Indications for UseShow that the indications for use do not alter the intended therapeutic use or affect safety/effectiveness relative to the predicate."The Primary Relief is a percutaneous electrical nerve stimulator (PENS) system indicated for up to 3 days for symptomatic relief of post-operative pain following cesarean section delivery."Differs from predicate but deemed not to raise new safety/effectiveness questions.
    Technological CharacteristicsComparison of design, power, pulse characteristics, battery, dimensions, materials, etc., with predicate(s).Minor technological differences (electrode placement, duration of patient contact) do not raise new safety/effectiveness questions.Detailed comparisons provided in Tables 6.1 and 6.2.
    Electrical Safety (EMC & Safety)Nonclinical testingPerformed and validated effectiveness.Standard regulatory requirement for electrical devices.
    BiocompatibilityNonclinical testingPerformed.Essential for devices with tissue contact.
    Performance Bench TestingValidate effectiveness of each unit, verify and compare effectual output with reference device, monitor continuous performance for 72 hours.Equivalent performance specifications when compared to predicate and reference devices. Pulse width, duration, amplitude, and current values captured for Drug Relief.Confirms functional integrity and consistency.
    Software Verification & ValidationNonclinical testingPerformed.Ensures software functions as intended and safely.
    Primary Efficacy Endpoint (Pain Intensity Reduction)Clinical study: Comparison of pain scores (NRS) between intervention (Primary Relief) and control (standard analgesics) groups.Minimally invasive nerve stimulation reduces pain score faster. High least-square means difference post six hours (p-value < 0.001) suggests intervention is more effective.Specific quantitative acceptance criteria (e.g., % reduction) for pain relief are not explicitly stated, but statistical significance is reported.
    Secondary Efficacy Endpoints (Quality of Life)Clinical study: Assessment of six quality of life measures using NRS.Significant positive response in the device intervention group (p-value < 0.04) compared to the control group.Shows improvement beyond just pain scores.
    Safety (Adverse Events)Clinical study: Monitoring for complications and adverse events during the study and follow-up.No complications or adverse events were observed in any participants during the study period across the two arms.Crucial for demonstrating safety.

    2. Sample Size for Test Set and Data Provenance

    • Sample Size for Test Set: 44 participants.
      • Intervention group: 22 participants (mean age = 24.7 ± 2.6 years)
      • Control group: 22 participants (mean age = 26.6 ± 3.9 years)
    • Data Provenance: Single center, prospective study. The country of origin is not explicitly stated in the provided text, but the company address is San Mateo, CA, USA, suggesting the study likely took place in the US or was conducted under US regulatory standards.

    3. Number of Experts and Qualifications for Ground Truth

    • The ground truth in this study is based on patient-reported outcomes using the Numerical Rating Scale (NRS) for pain intensity and quality of life measures.
    • No external experts were explicitly stated to establish ground truth for individual cases in the test set. The patients themselves provided the primary data. Medical professionals would have been involved in the administration of the study, data collection, and assessment of participants, but not in establishing an "expert ground truth" as might be seen in imaging studies.

    4. Adjudication Method

    • None as described. The primary and secondary endpoints were measured directly from patient self-reports using the Numerical Rating Scale (NRS). This method does not typically involve an adjudication process among experts for interpretation.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No. This was a clinical trial comparing a device to a control group for pain relief, not an MRMC study comparing human reader performance with and without AI assistance for interpretation tasks. The device itself is a therapeutic stimulus, not an AI diagnostic aid.

    6. Standalone (Algorithm Only) Performance

    • Yes, in the sense of the device operating independently. The Primary Relief device is a therapeutic device that delivers electrical stimulation. Its "performance" refers to its ability to generate specific electrical pulses and the clinical outcome from those pulses.
    • The "standalone performance" was evaluated through:
      • Performance bench testing: Validating effectiveness of each unit, comparing effectual output with a reference device, and monitoring continuous performance for 72 hours.
      • Clinical study: The device was administered to the intervention group without human-in-the-loop interpretation or modification of its core function during the treatment period. The human "in the loop" was the prescriber and the patient. While a healthcare professional inserts the needles, the device then operates autonomously to deliver stimulation.

    7. Type of Ground Truth Used

    • Patient-reported outcomes: Specifically, the Numerical Rating Scale (NRS) for:
      • Pain intensity (primary efficacy endpoint)
      • Six measures intended to assess quality of life (secondary efficacy endpoints)

    8. Sample Size for the Training Set

    • Not applicable / Not explicitly mentioned. The provided document describes a clinical study to evaluate the device, not a machine learning model that would have a "training set." The Primary Relief device is a hardware therapeutic device, not an AI algorithm that learns from data.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As there is no machine learning component described, there is no "training set" or "ground truth for the training set" in the context of AI development.