Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K223560

    Validate with FDA (Live)

    Date Cleared
    2023-08-21

    (266 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Plato 17 Microcatheter is intended for the introduction of therapeutic devices, infusion of diagnostic agents, such as contrast media, and delivery of embolization materials to the peripheral and neuro vasculature systems. The Plato 17 Microcatheter is not intended for use in the coronary vasculature.

    Device Description

    Scientia Vascular's Plato® 17 Microcatheter is a single lumen, open-ended catheter designed to be flexible and axially stable, to aid the physician in the accessing of the distal neuro and peripheral vasculature. The Plato® 17 Microcatheter is supplied sterile and is for single use only. The Plato® 17 Microcatheter is supplied in various tip configurations, including straight and pre-shaped (45° and 90°). The microcatheter shaft design includes nitinol, polymers of varying durometer, and an internal liner providing lubricity for therapeutic device, diagnostic agent, and embolization material delivery. The distal exterior section of the catheter shaft is hydrophilic coated to reduce friction during manipulation in vessels. The microcatheter includes two radiopaque tip markers to facilitate fluoroscopic visualization and a clear hub with luer lock and a strain relief.

    The microcatheter is provided with a steam shaping mandrel, peel-away introducer, and a ruler. The shaping mandrel, peel-away introducer, and ruler are accessories included to facilitate use of the microcatheter and are not intended to contact the patient's body.

    AI/ML Overview

    The provided document is a 510(k) summary for the Scientia Vascular, Inc. Plato® 17 Microcatheter. It does not contain information about studies related to AI/ML device performance, human reader improvement with AI assistance, or standalone algorithm performance. The device described is a microcatheter, a physical medical device, not a diagnostic imaging or AI-powered solution.

    Therefore, I cannot populate the table or provide information for points 2 through 9 as they are specific to AI/ML device evaluations.

    Here is the acceptance criteria and the (non-AI/ML related) study details as provided in the document:

    1. A table of acceptance criteria and the reported device performance

    Test NameAttribute and Acceptance CriteriaReported Device Performance
    Biocompatibility
    Cytotoxicity: MEM ElutionCell culture was observed for cytotoxic reactivity.Non-cytotoxic.
    Direct Contact and Extract method Hemolysis TestThe difference between the hemolytic indexes of the subject device and the negative control was evaluated.Non-hemolytic.
    Partial Thromboplastin Time (PTT) TestThe clotting time was observed for both the subject device and the predicate.Non-activator.
    Platelet Leukocyte CountsThe concentration of platelets and leukocytes were determined on an automated hematology analyzer.Non-activator.
    Functional Testing
    Accessory VerificationThe Peel-Away Introducer must facilitate introduction of catheter into a guide catheter.PASS
    Therapeutic Device Compatibility and Simulated UseThe Plato® 17 Microcatheter can successfully deliver representative therapeutic devices, including stent, in a tortuous neurovascular model. The Plato® 17 Microcatheter was evaluated for delivery to target location and functionality post therapeutic device delivery.PASS
    Embolization Material CompatibilityVisual inspection, lumen check, liquid leakage, static burst, tensile and force to withdraw testing was conducted for the Plato® 17 Microcatheter using representative embolization materials in a tortuous neurovascular model.PASS

    Since the provided document is for a physical medical device (microcatheter) and not an AI/ML-powered device, the following points are not applicable and cannot be answered from the given text:

    1. Sample size used for the test set and the data provenance: Not applicable for a physical medical device.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable for a physical medical device.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable for a physical medical device.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable as this is not an AI-assisted device.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable as this is not an AI-powered algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable for a physical medical device; ground truth refers to labels/diagnoses in AI/ML performance evaluation. For this device, "ground truth" would relate to confirming physical properties and functional performance against standards.
    7. The sample size for the training set: Not applicable as this is not an AI/ML device that requires a training set.
    8. How the ground truth for the training set was established: Not applicable as this is not an AI/ML device that requires a training set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K210601

    Validate with FDA (Live)

    Date Cleared
    2021-04-28

    (58 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Plato 17 Microcatheter is intended for the introduction of embolic coils and infusion of diagnostic agents, such as contrast media, to the peripheral and neuro vasculature systems. The Plato 17 Microcatheter is not intended for use in the coronary vasculature.

    Device Description

    Scientia Vascular's Plato® 17 Microcatheter is a single lumen, open-ended catheter designed to be flexible and axially stable, to aid the physician in the accessing of distal vasculature. The Plato 17 Microcatheter is supplied sterile and is for single use only. The Plato 17 Microcatheter is packaged with various tip configurations, including straight and pre-shaped (45° and 90°). The microcatheter shaft design includes nitinol, polymers of varying durometer and an internal liner providing lubricity for intraluminal delivered embolic coils and contrast media: the distal exterior section of the catheter shaft is hydrophilic coated to reduce friction during manipulation in vessels. The working length of the catheter is 160 cm. The microcatheter includes two radiopague tip markers to facilitate fluoroscopic visualization and a clear hub with luer lock and a strain relief. The Plato 17 Microcatheter is compatible with 0.014 inch guidewires or embolic coils.

    The microcatheter is provided with a steam shaping mandrel, peel-away introducer and a ruler. The shaping mandrel, peel-away introducer and ruler are accessories included to facilitate use of the microcatheter and are not intended to contact the patient's body.

    AI/ML Overview

    The Scientia Vascular, LLC's Plato® 17 Microcatheter was evaluated through non-clinical performance tests, including biocompatibility and functional testing, to demonstrate substantial equivalence to the predicate device, the Excelsior™ SL-10 Pre-Shaped Microcatheter (K042568) and reference device Excelsior XT-17 Microcatheter (K142565).

    1. Table of Acceptance Criteria and Reported Device Performance:

    TestAttribute and Acceptance CriteriaReported Device Performance
    Biocompatibility Testing (ISO 10993-1:2018)
    Cytotoxicity: MEM ElutionCell culture was observed for cytotoxic reactivity.Non-cytotoxic.
    SensitizationStudy animals with subject device were observed for dermal sensitization.No sensitization reaction.
    Intracutaneous Reactivity- by extractStudy animals with subject device were observed for dermal reaction.No significant dermal reactions at the injected site.
    Acute Systemic Toxicity - by extractStudy animals with the subject device were observed for abnormal clinical signs indicative of toxicity during the 72-hour test period.No signs of toxicity.
    Material-Mediated PyrogenicityStudy animals were observed for individual temperature rise.Non-pyrogenic.
    Direct Contact and Extract method Hemolysis TestThe difference between the hemolytic indexes of the subject device and the negative control was evaluated.Non-hemolytic.
    Partial Thromboplastin Time (PTT) TestThe clotting time was observed for both the subject device and the predicate.The difference in average clot time between the subject device and predicate was 7 seconds. The two samples are considered similar.
    Complement activation of SC5b-9Comparison of the subject device SC5b-9 value to the predicate device for all exposure times was performed.The subject device had similar or lower potential to activate the complement system when compared to the predicate.
    Hemocompatibility In-vitro Blood LoopThe catheters are placed in an in-vitro blood loop for three runs. The thrombus score for the subject device and predicate device is observed.The lab testing concluded that the subject device showed minimal thrombus formation and is considered thromboresistant.
    Functional Testing
    Visual and Dimensional InspectionVisual inspection of the catheter effective length to be free from surface defects, surface extraneous matter, and coating droplets. Catheter to meet the dimensional specifications per the design specifications.PASS
    Tensile StrengthThe minimum tensile force for all microcatheters must meet the tensile requirements of ISO 10555-1.PASS
    Tip FlexibilityComparable to the tip flexibility of the predicate catheter.PASS
    Flexural FatigueCatheter shall withstand cyclic fatigue without catheter damage.PASS
    Flow Characterization and Power InjectionCharacterize fluid (saline and contrast media) flow rate through the catheter. No leak or break in the catheter when subject to power injection.PASS
    Catheter Liquid Leak Under PressureNo liquid leakage on the hub or catheter body.PASS
    Catheter Static BurstMaximum catheter burst pressure > 300psi.PASS
    Torque Turns to FailureMaintain catheter integrity after applied hub rotations with the distal end held stationary.PASS
    Particulate Generation During Simulated UseThe amount of particulate matter that comes off the hydrophilic coated shaft during simulated use testing shall be determined and compared to the competitive products.PASS
    Coating Lubricity, Durability and IntegrityAfter 15 cycles the coating must maintain a minimum lubricity and maintain a minimum surface coating coverage.PASS
    Negative Collapse Pressure and Air IngressNo air ingress while under vacuum with a syringe held for 10 seconds. No catheter lumen collapse after applied vacuum with a syringe.PASS
    Kink RadiusThe catheter must maintain lumen integrity while wrapped around an anatomically relevant bend radius.PASS
    Tip Shape and Shape RetentionCatheter tip meets defined tip dimensional tolerances and must maintain tip shape.PASS
    Catheter ElongationCatheter shaft elongation at breakage should be comparable to the predicate.PASS
    Catheter Hub Luer ValidationPlato 17 Microcatheter hub will be compliant to requirements found in ISO 80369-7.PASS
    Simulated Use Model and Product Compatibility EvaluationSimulated use testing of the microcatheter was performed with accessory devices in a neurovascular model. The microcatheter and interfacing devices were delivered through the model and evaluated for effectiveness of delivery and functionality.PASS
    Accessory VerificationThe Peel-Away Introducer must facilitate introduction of catheter into a guide catheter. The tip shaping mandrel must facilitate steam shaping of the catheter tip and show no signs of corrosion. The ruler on the accessories card must meet design specifications.PASS
    RadiopacityCatheter tip markers to be visible under X-Ray fluoroscopy within the neuro vasculature.PASS
    Catheter Usability StudyPhysician evaluation of the microcatheter in a cadaver study demonstrates acceptable use when compared to current marketed devices. The microcatheter and interfacing devices were delivered through a representative procedure and evaluated for effectiveness of delivery and functionality.PASS
    Corrosion ResistanceNo corrosion is visible on the catheter samples in compliance with ISO 10555-1.PASS

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the specific sample sizes for each functional test performed. However, it mentions "study animals" for some biocompatibility tests (e.g., sensitization, intracutaneous reactivity, acute systemic toxicity, material-mediated pyrogenicity) and "catheter samples" for corrosion resistance. For the Catheter Usability Study, it mentions a "cadaver study." The provenance of the data (country of origin, retrospective/prospective) is not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    For the "Catheter Usability Study," it states "Physician evaluation of the microcatheter in a cadaver study." The number of physicians and their specific qualifications are not provided. For other functional and biocompatibility tests, "ground truth" is established by adherence to recognized standards and objective measurements, not necessarily expert consensus in the clinical sense.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    The document does not describe any specific adjudication method for establishing ground truth, as most tests are bench-top or animal studies with objective pass/fail criteria or comparisons to predicate devices/standards. For the "Catheter Usability Study" involving physician evaluation, the method of aggregation or adjudication of physician opinions is not detailed.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done as this device is a microcatheter, not an AI-assisted diagnostic tool. Therefore, the concept of human readers improving with AI assistance is not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This refers to a physical medical device (microcatheter), not an algorithm or AI system. Therefore, standalone algorithm performance is not applicable.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For biocompatibility tests, the ground truth is based on the biological responses observed in study animals or in vitro tests, evaluated against established scientific principles and standards (ISO 10993-1:2018). For functional tests, the ground truth is based on physical measurements, mechanical performance, and adherence to engineering standards (ISO 80369-7, ISO 80369-20, ISO 10555-1), as well as comparison to the predicate device where applicable. For the usability study, "acceptable use" by physicians in a cadaver study serves as a form of expert assessment.

    8. The sample size for the training set

    This device is a physical microcatheter, not a machine learning or AI model. Therefore, the concept of a "training set" is not applicable.

    9. How the ground truth for the training set was established

    As described in point 8, the concept of a "training set" and its associated ground truth is not applicable to this device submission.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1