510(k) Data Aggregation

    K Number
    K221245

    Validate with FDA (Live)

    Date Cleared
    2022-12-14

    (226 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PTA balloon dilatation catheter is indicated for the percutaneous transluminal angioplasty (PTA) of the peripheral vascular system, including the iliac, femoral, popliteal, infra-popliteal and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.

    Device Description

    The 0.0180" PTA balloon dilatation catheter is mainly composed of tip tube, inner tube, balloon, marker bands, outer tube, stress release tube and hub. Among them the balloon is the most important part of the catheter. In order to dilate different stenosis, the balloon should be dilated to difference dimension by inflating different pressure. The soft tip at the end of the balloon is to make the balloon catheter more easily to push to the stenosis position. The inner tubing which connects to the tip tubing is for guide wire passage and the pushing road. The two marker bands which wrapping on the inner tubing are for positioning the balloon location by cooperating in vitro monitoring equipment. The distal of the outer tube is connected with the balloon by a variable-diameter stretching structure, which can make the balloon get smaller winding diameter. The proximal of the outer tube is connected with the hub as the inflation passage and the pushing shaft. The hub is for connecting the outer pressure filling equipment. The stress release tubing is for protecting the outer tube from bending. The outer surface of the catheter is coated with a hydrophilic coating that ranges 400mm in length from the tip tube to the proximal end of the balloon, including part of the outer tube.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a PTA Balloon Dilatation Catheter. It outlines the device's characteristics, its comparison to a predicate device, and the performance data used to demonstrate substantial equivalence. However, it does not involve an AI/ML powered device, therefore no information about AI/ML specific aspects likes performance data, sample sizes for training/testing, ground truth establishment, or expert involvement for ground truth creation.

    Here's the information based on the provided text, with "N/A" for criteria not applicable to this type of device submission:

    Acceptance Criteria and Device Performance Study

    The submission demonstrates substantial equivalence for the PTA Balloon Dilatation Catheter to a legally marketed predicate device (SABER .018 Percutaneous Transluminal Angioplasty (PTA) Dilatation Catheter, K133843). The acceptance criteria are based on meeting performance standards primarily through bench testing and biocompatibility assessments, ensuring the device is suitable for its intended use and does not raise new questions of safety or effectiveness compared to the predicate.

    1. Table of Acceptance Criteria and Reported Device Performance

    The general acceptance criteria are that the device's design and construction are suitable for its intended use, and its performance is equivalent to the predicate device, as recommended by the draft Peripheral Percutaneous Transluminal Angioplasty (PTA) and Specialty Catheters - Premarket Notification (510(k)) Submissions (FDA; January 13, 2020). Specific acceptance criteria were applied to each bench test listed below.

    Acceptance Criterion (General)Reported Device Performance
    Device design and construction suitable for intended use, meeting relevant recommendations."The results met all acceptance criteria and ensure that the Balloon Dilatation Catheter design and construction are suitable for its intended use..."
    Biocompatibility of body-contacting materials in accordance with ISO 10993-1 and FDA draft guidance."The results of the testing show that the subject device included in this submission met all acceptance criteria and the subject device is biocompatible."
    Performance characteristics (e.g., burst pressure, flexibility, coating integrity) equivalent to or better than predicate, and safe for use.See specific bench tests below. "The results met all acceptance criteria..."

    Specific Bench Tests Conducted:

    • Visual inspection
    • Dimensional verification
    • Hydration test
    • Leakage test
    • Balloon inflation and deflation time
    • Balloon fatigue
    • Radiopacity
    • Simulated use test/guidewire and introducer sheath compatibility test
    • Tip pull strength
    • Catheter bond strength
    • Torque strength
    • Balloon rated burst pressure
    • Balloon compliance
    • Catheter body burst pressure
    • Flexibility and kinking
    • Coating integrity
    • Particulate evaluation
    • Infusion flow rate
    • Corrosion resistance test
    • Packaging performance Test (Visual inspection, Sealing strength, Sealing integrity, Sterility)
    • Shelf life testing

    Specific Biocompatibility Tests Conducted:

    • Cytotoxicity
    • Sensitization
    • Intracutaneous reactivity
    • Acute systemic toxicity
    • Hemocompatibility (hemolysis, partial thromboplastin time, complement activation, and in vivo thromboresistance)
    • Material-mediated pyrogenicity

    All these tests "met all acceptance criteria."

    2. Sample size used for the test set and the data provenance

    The document indicates that bench testing and biocompatibility testing were performed. However, it does not specify the exact sample sizes used for each of these tests.

    Data Provenance: Not explicitly stated, but typically, these tests are conducted under controlled laboratory conditions by the manufacturer or a contracted testing facility.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    N/A. This is a medical device (PTA Balloon Dilatation Catheter) submission, not an AI/ML device. Ground truth is established through physical and chemical testing against engineering specifications and industry standards, not expert annotation of data.

    4. Adjudication method for the test set

    N/A. This is a medical device (PTA Balloon Dilatation Catheter) submission, not an AI/ML device. The "acceptance criteria" are based on quantitative measurements from bench and biocompatibility tests.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    N/A. This is a medical device (PTA Balloon Dilatation Catheter) submission, not an AI/ML device. Therefore, no MRMC study or AI assistance evaluation was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    N/A. This is a medical device (PTA Balloon Dilatation Catheter) submission, not an AI/ML device.

    7. The type of ground truth used

    The "ground truth" for this device is based on established engineering specifications, industry standards (e.g., ISO 10993-1 for biocompatibility), and performance benchmarks derived from the predicate device and relevant FDA guidance documents. This is determined through physical measurements, chemical analyses, and simulated use tests.

    8. The sample size for the training set

    N/A. This is a medical device (PTA Balloon Dilatation Catheter) submission, not an AI/ML device. There is no concept of a "training set" in this context.

    9. How the ground truth for the training set was established

    N/A. This is a medical device (PTA Balloon Dilatation Catheter) submission, not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K141150

    Validate with FDA (Live)

    Date Cleared
    2014-09-25

    (143 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sterling Monorail PTA Balloon Dilatation Catheter is indicated for Percutaneous Transluminal Angioplasty in the peripheral vasculature, including iliac, femoral, popliteal, infra-popliteal, renal, and carotid arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for post-dilatation of balloon expandable and self-expanding stents in the peripheral vasculature.

    Device Description

    The Sterling Monorail Percutaneous Transluminal Angioplasty (PTA) Balloon Dilatation Catheter is a high performance balloon catheter for peripheral indications. The device features an ultra low profile, semi-compliant balloon combined with a low profile tip. The catheter is compatible with either 0.014 in (0.36 mm) or 0.018 in (0.46 mm) guidewires.

    The Sterling Monorail PTA Balloon Dilatation Catheter is a Monorail brand rapid exchange catheter with a semi-compliant balloon fixed at the distal tip. The balloon catheter has a coaxial shaft design. The outer lumen is used for inflation of the balloon, and the wire lumen permits the use of guidewires 0.014 in / 0.018 in (.36 mm / .46 mm) to facilitate advancement of the catheter to and through the stenosis to be dilated. The balloon is designed to provide an inflatable segment of known diameter and length at recommended pressures.

    The catheter includes a tapered tip to facilitate advancement of the catheter to and through the stenosis. Two radiopaque marker bands (one proximal and one distal), in conjunction with fluoroscopy, enable accurate positioning of the balloon. Markers on the 80 cm and 90 cm effective length catheters indicate the exit of the dilatation catheter tip out of the guiding catheter (one at 50 cm and two at 60 cm). Markers on the 135 cm and 150 cm effective length catheters indicate the exit of the dilatation catheter tip out of the guiding catheter (one at 90 cm and two at 100 cm). The effective lengths of the balloon catheter are 80 cm, 90 cm, 135 cm, and 150 cm. A needle with a luer port is included for flushing the distal inner lumen prior to the insertion of appropriate guidewires.

    AI/ML Overview

    The provided text is a Food and Drug Administration (FDA) 510(k) premarket notification for a medical device called the Sterling™ Monorail™ PTA Balloon Dilatation Catheter. This document primarily focuses on establishing "substantial equivalence" to predicate devices, rather than presenting a detailed clinical study for novel device performance claims. Therefore, much of the requested information regarding acceptance criteria and performance studies for a new AI/software-based device will not be found in this type of submission.

    However, I can extract the information available and explain why some of the requested points are not applicable to this document.

    1. A table of acceptance criteria and the reported device performance

    The document states that the "Sterling Monorail PTA Balloon Dilatation Catheter met all acceptance criteria for the bench and biocompatibility testing with results similar to the predicate." However, the specific quantitative acceptance criteria for each test and the corresponding reported performance values are not explicitly listed in a table format in this document. Instead, it provides a general statement of compliance.

    It lists the types of tests performed:

    Test TypeSpecific Tests Mentioned
    Biocompatibility- MEM Elution Cytotoxicity
    - Hemocompatibility (Direct Contact)
    - Chemical Characterization-USP Physicochemical
    - Natural Rubber Latex
    Bench Testing (in-vitro performance)- Bond Integrity
    - Working Length
    - Deflation Time
    - Balloon Rated Burst Pressure (RBP)
    - Balloon Multiple Inflation
    - Crossing Profile
    - Full Catheter Tensile Extension and Deflation
    - Balloon Burst Mode
    - Balloon Compliance
    - Balloon Nominal Diameter
    - Burst in a Stent
    - Balloon Body Length
    - Guidewire Movement
    - Sheath Withdrawal
    - Balloon Multiple Inflation in a Stent
    - Particulate Evaluation
    - Proximal Balloon Bond and Shaft Tensile Strength
    - Marker Band to Balloon Alignment
    - Torque After Conditioning

    2. Sample sizes used for the test set and the data provenance

    The document does not specify sample sizes for each of the listed tests. The provenance of the data is generally in-vitro bench testing and biocompatibility testing, likely conducted by the manufacturer (Boston Scientific) in controlled laboratory environments. This is typical for a 510(k) submission and not considered "country of origin" in the sense of clinical data. It is neither retrospective nor prospective clinical data; it is in-vitro (laboratory) data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable to a 510(k) submission for a mechanical medical device like a balloon catheter. "Ground truth" established by human experts is typically relevant for AI/software devices where human interpretation is involved. For this device, the "truth" is determined by physical measurements and chemical analyses against established engineering and biological standards.

    4. Adjudication method for the test set

    This question is also not applicable. Adjudication methods are used in studies where there is subjective interpretation of data, often by multiple human readers, to establish a consensus or "ground truth." For the bench and biocompatibility tests of this device, the results are typically objective measurements against predefined specifications.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable. The device described is a physical medical instrument (a balloon catheter), not an AI/software device. Therefore, no MRMC study involving human readers and AI assistance would be performed for this device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable. The device is a physical balloon catheter, not an algorithm or software. It does not perform any standalone algorithmic analysis.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the bench testing, the "ground truth" is defined by established engineering specifications and medical device standards (e.g., ISO, ASTM, internal company specifications for performance characteristics like burst pressure, deflation time, etc.). For biocompatibility, the ground truth is against established biological safety standards (e.g., ISO 10993-1). This is not expert consensus, pathology, or outcomes data in the clinical sense, but rather compliance with scientific and engineering benchmarks.

    8. The sample size for the training set

    This question is not applicable. This device is not an AI/software device that would use a "training set."

    9. How the ground truth for the training set was established

    This question is not applicable, as there is no training set for this type of medical device.