Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K111925

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-05-01

    (300 days)

    Product Code
    Regulation Number
    862.3250
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Psychemedics Microplate EIA for Cocaine is an enzyme immunoassay (EIA) for the preliminary qualitative detection of cocaine in human head and body hair samples using a cocaine calibrator at 5 ng /10 mg hair cutoff for the purpose of identifying cocaine use. This product is intended exclusively for in-house professional use only and not for sale to anyone.

    The Psychemedics EIA Cocaine Assay provides only a preliminary analytical test result. To obtain a quantitative analytical result or to confirm positive results, a more specific alternate chemical method (e.g. GC/MS) must be used. Clinical consideration and professional judgment should be applied to the interpretation of any drug-of-abuse test result.

    Device Description

    The test consists of two parts; a pre-analytical hair treatment procedure (to convert the solid matrix of hair to a measurable liquid matrix) and the screening assay, the Psychemedics Microplate EIA for Cocaine. The drug is recovered from the hair using a patented method (U.S. Patent #8,084,215). The screening portion of the test system consists of (1) microplate wells coated with multiple drugs including benzoylecgonine conjugated to bovine serum albumin (BSA) (patent pending), monoclonal mouse anti-cocaine, goat anti-mouse secondary antibody conjugated to HRP (horseradish peroxidase), substrate [3, 3', 5, 5' tetramethylbenzidine (TMB)], HCl to acidify the final reaction, and wash buffer for washing the plates. Absorbance in the wells is read with a microplate reader.

    AI/ML Overview

    Acceptance Criteria and Study for Psychemedics Microplate EIA for Cocaine in Hair

    The Psychemedics Microplate EIA for Cocaine in Hair is an enzyme immunoassay (EIA) for the preliminary qualitative detection of cocaine and its metabolites in human head and body hair. The device's performance was evaluated against a predicate device (Psychemedics Cocaine Assay, K010868) and confirmed by LC/MS/MS (liquid chromatography/mass spectrometry/mass spectrometry).

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided 510(k) summary does not explicitly state pre-defined acceptance criteria for the accuracy of the new device relative to the predicate or LC/MS/MS. However, the study aims to demonstrate "substantial equivalence" to the predicate and "substantially equivalent" results to LC/MS/MS. Based on the presented data, the implicit acceptance criteria appear to be:

    • High agreement with the predicate device.
    • High agreement with LC/MS/MS for confirmed samples.
    • No significant impact of cosmetic treatments on negative or positive samples.
    • Effective wash procedure to mitigate external contamination.
    • Stability of calibrator and control solutions.
    • Cross-reactivity within acceptable limits for related compounds, and minimal to no cross-reactivity or interference from other common compounds.
    • Equivalent recovery to the predicate device.

    Here's a summary of the reported device performance against these implicit criteria:

    Acceptance Criteria (Implicit)Reported Device Performance
    Agreement with Predicate DeviceFour hundred-twenty-five hair samples were compared. <1% discordance between the EIA and predicate RIA. Importantly, the 4 discordant samples (positive by predicate, negative by EIA) were also found negative by LC/MS/MS, suggesting the EIA's results were correct.
    Agreement with LC/MS/MS (Confirmation Study)Strong agreement for the 256 samples confirmed by LC/MS/MS.
    (From provided table)EIA Positive (when LC/MS/MS is) * ≥ Cutoff, And < +50% of Cutoff: 12/12 samples (100%) * ≥ +50% of Cutoff and < +100% of Cutoff: 5/5 samples (100%) * ≥ +100% of cutoff: 101/101 samples (100%)EIA Negative (when LC/MS/MS is) * Zero: 118/118 samples (100%) * ≥ 10% and <50% of Cutoff: 9/9 samples (100%) * ≥ -50% of Cutoff and < Cutoff: 3/3 samples (100%)The table shows that the EIA correctly identified all samples well above the cutoff as positive and all samples well below the cutoff as negative. There were no false positives when LC/MS/MS was zero or significantly below the cutoff, and no false negatives when LC/MS/MS was significantly above the cutoff. There were 8 samples that were EIA Positive where LC/MS/MS showed "≥ -50% of Cutoff and > Cutoff", and 12 samples for "≥ Cutoff, And < +50% of Cutoff", indicating good performance at or around the cutoff.
    Impact of Cosmetic Treatments- Negative Samples: No significant differences observed; all 20 negative samples for each treatment type (bleach, perm, dye, relaxer, shampoo) remained negative after treatment (total 100 samples).- Positive Samples: No positive samples became negative after any of the cosmetic treatments (bleach, perm, dye, relaxer, shampoo). Mean B/B₀ x 100 values for positive samples showed slight variations but remained positive after treatment.
    Effectiveness of Wash Procedure- High Contamination (1000 ng cocaine/mL water): Amounts on hair reduced from 38.6-98.7 ng/10 mg to 1.1-3.0 ng/10 mg after washing. None at or above cutoff before wash criterion applied.- Moderate Contamination (1000 ng cocaine/mL saline): Amounts on hair reduced from 8.5-29.3 ng/10 mg to 0.2-1.0 ng/10 mg after washing. All samples were negative (below cutoff) even without wash criterion.
    Stability of Calibrator and Control SolutionsStability of cocaine in methanol in the presence of other drugs of abuse was shown to exceed 1 year.
    Cross-reactivity and Interference- Cross-reactivity: Seven related compounds showed cross-reactivity ranging from 2.5% (Norbenzoylecgonine) to 76.9% (Cocaethylene).- No Cross-reactivity: Seventy-nine other compounds showed no cross-reactivity.- No Interference: One hundred-forty-one compounds tested at +/-50% of the cutoff showed no interference.
    Recovery StudyRecovery of cocaine and benzoylecgonine from hair of cocaine users was shown to be substantially equivalent to that of the predicate device.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size:

      • Agreement Testing: 425 hair samples were used for comparison between the predicate device and the new EIA.
      • LC/MS/MS Confirmation: 256 of the 425 samples were further confirmed by LC/MS/MS.
      • Cosmetic Treatment Study: 20 negative hair samples per treatment type (5 types = 100 samples), and 12 positive hair samples per treatment type (5 types = 60 samples).
      • Contamination Study: 8 hair samples per contamination method (water and saline = 16 samples).
    • Data Provenance: The document does not explicitly state the country of origin of the data or whether the studies were retrospective or prospective. Given the nature of a 510(k) submission for a diagnostic device, the studies are typically conducted in a controlled laboratory environment using collected human samples, implying real-world (prospective or retrospectively collected and stored) human hair samples. The samples included both head and body hair (15.8% were body hair samples).

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • The ground truth for the agreement and confirmation studies was established by LC/MS/MS, which is a highly specific and quantitative analytical chemical method, considered the "gold standard" for drug confirmation in forensic toxicology.
    • No human "experts" in the sense of clinicians or radiologists are mentioned as establishing the ground truth for the test set results. The ground truth refers to the analytical confirmation by LC/MS/MS, performed by trained laboratory personnel.

    4. Adjudication Method for the Test Set

    • For the agreement testing, discrepancies between the new EIA device and the predicate RIA device (<1% of samples) were adjudicated by LC/MS/MS. The document states, "The samples were negative by LC/MS/MS, demonstrating that the EIA negative results, although discordant with the predicate, are correct." This indicates that LC/MS/MS served as the ultimate adjudicator for discordant results.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • No. This product is an in-vitro diagnostic (IVD) assay, not an imaging device for human interpretation. Therefore, a multi-reader, multi-case comparative effectiveness study involving human readers is not applicable. The device's performance is determined by its analytical accuracy against a gold standard (LC/MS/MS), not by human interpretation improvement.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Yes. The primary studies presented (Agreement Testing, LC/MS/MS Confirmation, Cosmetic Treatment, Contamination, Cross-reactivity, and Recovery) all demonstrate the standalone performance of the Psychemedics Microplate EIA for Cocaine in Hair. The output is a preliminary qualitative result (positive/negative based on cutoff), without direct "human-in-the-loop" interpretation for the initial screening result itself. The "human-in-the-loop" aspect comes in the professional judgment for clinical interpretation of any drug-of-abuse test result and the requirement for confirmation by a more specific method for positive preliminary results.

    7. The Type of Ground Truth Used

    • The primary ground truth used for validating the device's accuracy is LC/MS/MS (Liquid Chromatography/Mass Spectrometry/Mass Spectrometry). This is a highly sensitive and specific analytical method for identifying and quantitating substances.
    • For the agreement study, the predicate device (Psychemedics Cocaine Assay, K010868) also served as an initial comparison point, though LC/MS/MS was used to resolve discrepancies.

    8. The Sample Size for the Training Set

    • The document does not explicitly state the sample size for a training set. For an IVD assay like this, development typically involves internal validation and optimization, but the 510(k) summary focuses on the performance studies used for regulatory submission. It is common for the public summary to detail testing data rather than development/training data for such devices.

    9. How the Ground Truth for the Training Set Was Established

    • As the document does not explicitly mention a "training set" in the context of machine learning or AI models, it also does not describe how the ground truth for a training set was established. For an immunoassay, the "training" involves optimizing reagents, protocols, and cutoff values, likely guided by samples with known cocaine concentrations confirmed by methods like GC/MS or LC/MS/MS during the assay development phase.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1