Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K143386

    Validate with FDA (Live)

    Date Cleared
    2015-03-13

    (107 days)

    Product Code
    Regulation Number
    878.3300
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ProGrip™ laparoscopic self-fixating mesh is indicated for the reinforcement of soft tissues during repair of inguinal hernia defects by laparoscopic approach.

    Device Description

    Both the predicate and the proposed mesh are made of knitted monofilament polyester with monofilament polylactic acid resorbable grips on one side and a resorbable film made of collagen from porcine origin and glycerol, on the other side. The grips allow positioning and fixation of the mesh to the surrounding tissue, while the collagen film facilitates mesh handling and deployment. The mesh incorporates a green band to facilitate orientation.

    The monofilament polylactic acid grips are bioresorbable and provide the fixation of the mesh to surrounding tissue for at least 8 weeks. The polylactic acid grips degrade and resorb in vivo by hydrolysis and are metabolized by the body into CO2 and H2O.

    AI/ML Overview

    The provided text describes a 510(k) submission for Sofradim Production's ProGrip™ Laparoscopic Self-Fixating Mesh, seeking to add new sizes to an existing product family. The document focuses on demonstrating substantial equivalence to a predicate device, rather than presenting a standalone study with detailed acceptance criteria and performance metrics for a medical device that produces quantitative or qualitative outputs (like an AI/ML algorithm).

    Therefore, based on the provided text, it's not possible to provide a table of acceptance criteria and reported device performance in the way typically expected for an AI/ML device or a device with measurable performance output. The "performance data" mentioned in the document refers to physical characteristics and compatibility of the mesh sizes with trocars, not clinical outcomes or diagnostic accuracy.

    Here's an analysis based on the information provided, recognizing the limitations:

    1. Table of Acceptance Criteria and Reported Device Performance

    As this submission is for a physical medical device (surgical mesh) and primarily concerns adding new sizes, the "acceptance criteria" are related to physical properties and compatibility, not a performance metric like sensitivity, specificity, or accuracy for an AI/ML device.

    Acceptance Criteria (Bench Testing)Reported Device Performance
    For 16x12cm and 16x14cm mesh sizes:
    - Passage through 10mm and/or 12mm trocarTests demonstrate compatibility and successful passage. (Specific numerical results or pass/fail statements are not provided, but the conclusion states equivalence).
    - Visual inspection after trocar passageTests demonstrate satisfactory visual appearance after passage. (Specific details are not provided, but the conclusion states equivalence).
    - Gripping point measurement and comparison with specification (≤27 gripping points for 16x12cm and 16x14cm) after trocar passageMeasurements were taken and compared with the specification. (Specific numerical results are not provided, but the conclusion states equivalence, implying compliance with the ≤27 gripping points specification).

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated. The text mentions "bench performance evaluations were completed to verify that the proposed additional sizes... are compatible with 10mm and/or 12mm trocars." It doesn't specify how many meshes of each size were tested.
    • Data Provenance: This is bench testing of physical devices (surgical mesh). The country of origin of the data is not specified, but the submitter is "Sofradim Production (subsidiary of Covidien LLC)" with an address in Trevoux, France. The data is prospective in the sense that it was generated for this submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    Not applicable. The "ground truth" for this type of physical device testing is the direct measurement and observation of physical properties and functionality (e.g., whether the mesh passes through the trocar, visual integrity, number of gripping points). It does not involve expert interpretation or diagnosis.

    4. Adjudication Method for the Test Set

    Not applicable. This is objective bench testing of physical properties, not a subjective assessment requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This submission is for a physical surgical mesh, not an AI/ML diagnostic or assistive device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    Not applicable. This is for a physical surgical mesh.

    7. The Type of Ground Truth Used

    The ground truth for the bench testing was based on:

    • Direct observation of physical events (passage through trocar).
    • Visual inspection criteria.
    • Quantitative measurement against a defined specification (e.g., number of gripping points).

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/ML device; therefore, there is no training set in that context. The "training" in manufacturing comes from established material compositions and manufacturing processes which are stated to be unchanged from the predicate device.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no AI/ML training set, this question is not relevant. The "ground truth" for the predicate device's design and manufacturing processes would have been established through extensive design validation, material testing, and historical performance data, none of which are detailed here beyond the statement of "ProGrip™ Laparoscopic Self-Fixating Mesh new sizes are composed of biocompatible materials that are in compliance with ISO 10993-1 standard." and that materials and manufacturing processes have not changed.

    Ask a Question

    Ask a specific question about this device

    K Number
    K123479

    Validate with FDA (Live)

    Date Cleared
    2012-12-11

    (28 days)

    Product Code
    Regulation Number
    878.3300
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PROGRIP™ Laparoscopic Self-Fixating Mesh is indicated for the reinforcement of soft tissues during repair of inguinal hernia defects by laparoscopic approach.

    Device Description

    The PROGRIP™ Laparoscopic Self-Fixating Mesh is available in anatomical and rectangular shapes. The mesh is made of knitted monofilament polyester with monofilament polylactic acid resorbable grips on one side and a resorbable collagen film on the other side. The film is made up of collagen from porcine origin and glycerol. The grips allow positioning and fixation of mesh to the surrounding tissue, while the collagen film facilitates mesh handling and deployment. The mesh presents a green band that facilitates mesh orientation No changes to product have been made in this submission.

    AI/ML Overview

    This 510(k) submission (K123479) for the PROGRIP™ Laparoscopic Self-Fixating Mesh is not a study that proves the device meets specific acceptance criteria through clinical trials or performance testing as would be described for an AI/ML device.

    Instead, this submission is a request to the FDA to remove a specific final product specification (lower specification of extractible pH) for an already legally marketed device (PROGRIP™ Laparoscopic Self-Fixating Mesh, K120897). The core assertion is that this change does not impact the device's intended use, does not alter its fundamental scientific technology, and does not require additional performance data.

    Therefore, many of the requested elements for an AI/ML device study (like sample sizes for test/training sets, expert ground truth establishment, MRMC studies, etc.) are not applicable and not present in this document.

    Here's an breakdown based on the provided text, focusing on what is relevant to a 510(k) submission for a modification to a predicate device:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't present a table with typical performance metrics (like sensitivity, specificity, AUC) for a new device. Instead, it discusses the impact of a specification change on the effectiveness for intended use.

    Acceptance Criteria (Conceptual, based on the submission)Reported Device Performance (Argument for Equivalence)
    Original Device Performance: Maintained despite specification change. The product must continue to be "effective for intended use" (reinforcement of tissues during surgical repair, specifically inguinal hernia).The submission argues that "The removal of the collagen film lower specification of extractible pH, as a final product specification, does not affect the product performance." It further states that "The combination of the upper extractible pH and the lot release final product specification (gripping point test) ensure the product effectiveness for intended use." No quantitative performance data is provided as it's a change to a predicate.
    Technological Characteristics: Remain substantially equivalent to the predicate device."The subject PROGRIP™ Laparoscopic Self-Fixating Mesh is identical to the predicate device PROGRIP™ Laparoscopic Self-Fixating Mesh (K120897) in terms of its physical, technological characteristics and performance characteristics."
    Safety: No new safety concerns introduced by the specification change.Implied by the lack of new performance data requirement and the assertion that "This change does not impact the intended use and does not alter the fundamental scientific technology of the device."

    2. Sample size used for the test set and the data provenance:

    • Not applicable. This submission is for a modification to a previously cleared medical device, not a new device requiring a clinical performance study with a test set. The justification relies on the existing predicate device's performance and the argument that the change is minor.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. No expert review panel or ground truth establishment for a test set is mentioned, as this is not a study evaluating diagnostic or predictive performance.

    4. Adjudication method for the test set:

    • Not applicable. No test set or adjudication method is described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a surgical mesh, not an AI/ML diagnostic or assistive tool for human readers. Therefore, an MRMC study is irrelevant to this submission.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is not an algorithm or AI device.

    7. The type of ground truth used:

    • Not applicable. As there is no "test set" in the context of an AI/ML study, there is no discussion of ground truth derived from pathology, outcomes data, or expert consensus. The "ground truth" for this submission is implicitly the established safety and efficacy profile of the predicate device (K120897), which this modified device claims to be substantially equivalent to, despite a minor change in a manufacturing specification.

    8. The sample size for the training set:

    • Not applicable. This is not an AI/ML device, so there is no training set mentioned.

    9. How the ground truth for the training set was established:

    • Not applicable. No training set exists for this device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1