510(k) Data Aggregation

    K Number
    K161163

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-07-13

    (79 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PROBEAT-V with X-ray Limiting Accessory is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.

    Device Description

    The PROBEAT-V with X-ray Limiting Accessory is a proton beam irradiation system, which provides a therapeutic proton beam for clinical treatment. It is designed to deliver a proton beam with the prescribed dose, dose distribution and directed to the prescribed patient treatment site. The equipment to perform the above work is composed of two main components: (1) a beam delivery system whose primary responsibility is to ensure that the desired prescription parameters are properly delivered and (2) equipment necessary to generate the proton beam and direct it to the beam delivery system for patient treatment.

    The X-ray Beam Limiting accessory for use in conjunction with the cleared PROBEAT-V system is designed to limit the patient's X-ray exposure during for patient positioning with the Cone Beam CT (CBCT). The device has 4 motorized leaves which are controlled by the PROBEAT-V imaging software to position the leaves in order to limit the patient's radiation exposure during patient positioning.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the PROBEAT-V Proton Beam Therapy System with X-ray Limiting Accessory. The document focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed study report with specific acceptance criteria and detailed performance data on the X-ray Limiting Accessory itself.

    Based on the provided information, I can extract the following:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a specific table of acceptance criteria with numerical targets. Instead, it states that "In all cases, the device performed as intended and met all pre-specified success criteria." The performance evaluation focused on the integration of the X-ray Beam Limiting Device with the PROBEAT-V system.

    Acceptance Criteria CategoryReported Device Performance
    Slice Pitch AccuracyMet all pre-specified success criteria and performed as intended.
    Spatial AccuracyMet all pre-specified success criteria and performed as intended.
    Spatial ResolutionMet all pre-specified success criteria and performed as intended.
    Isocenter AccuracyMet all pre-specified success criteria and performed as intended.
    Performance with Proton Beam Therapy Positioning SystemMet all pre-specified success criteria and performed as intended.
    Electrical SafetyEvaluated and met all pre-specified success criteria and performed as intended.
    Electromagnetic CompatibilityPerformed and met all pre-specified success criteria and performed as intended.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not provide details on sample sizes for the test set or data provenance (country of origin, retrospective/prospective). The testing appears to be a validation of the integrated system in an engineering and performance testing context rather than a clinical study with patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. The testing described appears to be technical performance testing of the device itself (e.g., accuracy, resolution, safety) and not a diagnostic or therapeutic output requiring interpretation by medical experts to establish ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. The testing described appears to be technical performance testing of the device itself and not a diagnostic or therapeutic output requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document describes a proton beam therapy system with an X-ray limiting accessory, not an AI-based diagnostic or therapeutic assistance system.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This refers to the performance of the device's X-ray limiting accessory and its integration with the PROBEAT-V system. The "Performance Data" section describes this standalone performance testing for the integrated system, including slice pitch accuracy, spatial accuracy, spatial resolution, isocenter accuracy, and performance with the proton beam therapy positioning system, as well as electrical safety and electromagnetic compatibility. The device performed as intended and met all pre-specified success criteria in these tests.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the technical performance tests mentioned (slice pitch accuracy, spatial accuracy, etc.), the "ground truth" would be established by validated measurement standards and reference devices used in engineering and quality control, rather than clinical ground truth like pathology or expert consensus. For electrical safety and electromagnetic compatibility, the ground truth would be compliance with relevant international and national standards.

    8. The sample size for the training set

    Not applicable for this type of device and submission. This is not an AI/machine learning device requiring a training set.

    9. How the ground truth for the training set was established

    Not applicable for this type of device and submission. This is not an AI/machine learning device requiring a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K153614

    Validate with FDA (Live)

    Date Cleared
    2016-04-21

    (126 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PROBEAT-V is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.

    The PROBEAT-V includes optional accessories to assist the radiation oncologist in the delivery of proton radiation to defined target volumes while sparing surrounding normal tissue and critical organs from excess radiation.

    Device Description

    The PROBEAT-V is a proton beam irradiation system, which provides a therapeutic proton beam for clinical treatment. It is designed to deliver a proton beam with the prescribed dose, dose distribution and directed to the prescribed patient treatment site. The equipment to perform the above work is composed of two main components: (1) a beam delivery system whose primary responsibility is to ensure that the desired prescription parameters are properly delivered and (2) equipment necessary to generate the proton beam and direct it to the beam delivery system for patient treatment.

    AI/ML Overview

    This document describes the PROBEAT-V and its accessories, focusing on modifications to the minimum MU and the introduction of a high-accuracy camera system for patient positioning. The document states that testing was conducted to confirm the device meets pre-defined acceptance criteria. However, it does not provide specific details about the acceptance criteria or the study that proves the device meets them.

    Therefore, I cannot answer your request based on the provided text. The document states that "All the tests were successfully conducted and confirmed that the device meets the pre-defined acceptance criteria," but it does not elaborate on:

    1. A table of acceptance criteria and the reported device performance: No table or specific performance metrics are provided.
    2. Sample size used for the test set and the data provenance: No information on sample size or data origin is given.
    3. Number of experts and their qualifications: No details about experts are mentioned.
    4. Adjudication method: Not described.
    5. MRMC comparative effectiveness study: No indication of such a study.
    6. Standalone performance: Not explicitly addressed in terms of a separate study.
    7. Type of ground truth: Not specified.
    8. Sample size for the training set: Not applicable as this is not a description of an AI/ML algorithm development.
    9. How ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K152592

    Validate with FDA (Live)

    Device Name
    PROBEAT-V
    Date Cleared
    2015-11-02

    (53 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PROBEAT-V is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.

    Device Description

    The PROBEAT-V is a proton beam irradiation system, which provides a therapeutic proton beam for clinical treatment. It is designed to deliver a proton beam with the prescribed dose, dose distribution and directed to the prescribed patient site. The equipment to perform the above work is composed of two main components: (1) a beam delivery system whose primary responsibility is to ensure that the desired prescription parameters are properly delivered and (2) equipment necessary to generate the proton beam and direct it to the beam delivery system for patient treatment.

    The beam delivery system is composed of the following components:

    • Gantry Room
      • Rotating Gantry
      • Scanning Nozzle
      • Patient Positioning System
      • Cone Beam CT / X-ray Imaging System
    • Fixed Beam Room
      • Scanning Nozzle
      • Patient Positioning System
      • Cone Beam CT

    The beam production system is composed of the following components:

    • Accelerator system (LINAC, Synchrotron)
    • Beam transport system (Low/High Energy Beam Transport systems)
    AI/ML Overview

    The provided text is a 510(k) summary for the PROBEAT-V proton beam therapy system. It describes the device, its intended use, and performance data. However, it does not contain the detailed information required to answer all the questions about specific acceptance criteria and a study proving the device meets those criteria, particularly regarding AI performance, expert ground truth, sample sizes for training/test sets, and adjudication methods, as this is neither an AI device nor does it have relevant information about AI device testing.

    Here's a breakdown of what can be extracted from the document:

    1. A table of acceptance criteria and the reported device performance

    The document provides a general statement about performance testing without specific quantitative acceptance criteria. It broadly states that the device "functioned as intended and met its specifications."

    Acceptance Criteria (Stated broadly)Reported Device Performance
    Mechanical performance of rotating gantry and patient couchFunctioned as intended and met specifications
    Beam dose shape and beam doseFunctioned as intended and met specifications
    Beam stop control, dose monitor, area safety, and mechanical interlocksFunctioned as intended and met specifications
    Electrical safety (IEC 60601-1)Met specifications
    Electromagnetic compatibility (IEC 60601-1-2)Met specifications

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The document describes general performance testing but does not specify sample sizes or data provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This information is not applicable/provided. The PROBEAT-V is a therapeutic device (proton beam therapy system), not a diagnostic or AI-driven decision support system that typically requires expert-established ground truth for image interpretation or diagnosis. The performance testing focuses on the physical and functional aspects of the hardware and beam delivery.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This information is not applicable/provided for the same reasons as the previous point. Adjudication methods are typically used in studies involving subjective assessments or classifications by human readers, which is not the nature of the described performance testing for a proton beam therapy system.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no indication that an MRMC comparative effectiveness study was done, nor is AI assistance mentioned. The device is a proton beam therapy system, not an AI-driven image analysis or decision support tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable/provided. The PROBEAT-V is a physical medical device for radiation therapy, not an algorithm. Its performance testing relates to its mechanical, electrical, and beam delivery functionalities.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The concept of "ground truth" as typically applied to diagnostic studies (e.g., pathology for a tumor diagnosis) is not directly applicable here. The performance testing described (mechanical, beam, safety interfaces, electrical safety, EMC) likely uses objective, measurable physical parameters and engineering standards as its "ground truth" or reference for meeting specifications. For example, a "beam dose shape" would be compared against a pre-defined expected profile, or electrical safety against a standard like IEC 60601-1.

    8. The sample size for the training set

    This information is not applicable/provided. The PROBEAT-V is a hardware medical device; it does not involve a "training set" in the context of machine learning. Its design and manufacturing are based on engineering principles and established medical device standards.

    9. How the ground truth for the training set was established

    This information is not applicable/provided for the same reasons as the previous point.

    Ask a Question

    Ask a specific question about this device

    K Number
    K151132

    Validate with FDA (Live)

    Device Name
    PROBEAT-V
    Date Cleared
    2015-06-01

    (34 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PROBEAT-V is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.

    Device Description

    The PROBEAT-V is a proton beam irradiation system, which provides a therapeutic proton beam for clinical treatment. It is designed to deliver a proton beam with the prescribed dose, dose distribution and directed to the prescribed patient site. The equipment to perform the above work is composed of two main components: (1) a beam delivery system properly delivered and (2) equipment necessary to generate the proton beam and direct it to the beam delivery system for patient treatment.

    The beam delivery system is composed of the following components"

    • o Gantry Room
      • · Rotating Gantry
      • . Scanning Nozzle
      • . Patient Couch (or Patient Positioning System)
      • . X-ray Imaging System
    • o Fixed Beam Room
      • Scanning Nozzle .
      • . Patient Couch (or Patient Positioning System)
      • . X-ray Imaging System

    The beam production system is composed of the following components"

    • Accelerator system (LINAC, Synchrotron) O
    • Beam transport system (Low/High Energy Beam Transport systems) O
    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the Hitachi PROBEAT-V, a medical charged-particle radiation therapy system. It outlines performance data but does not contain information about specific acceptance criteria, study methodologies, or AI performance metrics as requested.

    Therefore, I cannot provide a table of acceptance criteria and reported device performance or information about sample sizes, expert ground truth, adjudication methods, MRMC studies, standalone performance, or training set details as these are not present in the provided document.

    The document indicates that Hitachi performed validation testing to evaluate:

    • Mechanical performance: Specifically for the rotating gantry and patient couch.
    • Beam performance: To assess beam dose shape and beam dose.
    • Safety interface testing: To evaluate beam stop control, dose monitor, area safety, and mechanical interlocks.
    • Electrical safety and electromagnetic compatibility: In accordance with IEC 60601-1 and IEC 60601-1-2.

    The conclusion of these tests was that "In all instances, the PROBEAT-V functioned as intended and met its specifications." This statement broadly implies that the device met its internal acceptance criteria for these functional and safety aspects, but the specific numerical or qualitative criteria for "met its specifications" are not detailed.