    Intended Use

    Precision Flow® HVNI is intended for use to add warm moisture to breathing gases from an external source for administration to a neonate/infant, pediatric and adult patients in the hospital and subacute institutions settings. It adds heat and moisture to a blended medical air/ oxygen mixture and assures the integrity of the precise air/oxygen mixture via an integral oxygen analyzer. The flow rates may be from 1 to 40 liters per minute via nasal cannula.

    Precision Flow® HVNI provides high velocity nasal insufflation (HVNI) with simultaneous oxygen delivery to augment breathing of spontaneously breathing patients suffering from respiratory distress and/or hypoxemia in the hospital setting. Precision Flow® HVNI is not intended to provide total ventilatory requirements of the patient and not for use during field transport.

    Device Description

    The Precision Flow® HVNI system delivers high flow rates of heated and humidified blended breathing gas through high flow nasal cannulas. The Precision Flow® HVNI system can connect to air and oxygen source. The Precision Flow® HVNI may be operated with limited performance at gas inlet pressures as low as 4 psi (28 kPa). For the full specified range of gas flows and oxygen percentages, both gas inlet pressures must be at minimum 40 psi (276 kPa).The main unit contains an integrated blender that delivers the targeted gas mixture to the disposable patient circuit (DPC). The Disposable Patient Circuit contains:

    • Water path: tubing from sterile water supply to vapor transfer cartridge
    • . Vapor transfer (humidification) cartridge: 2 types; low flow (1-8 lpm) and high flow (5-40 lpm)
    • Delivery Tube: triple lumen tubing
    • . Nasal Cannula

    The device automatically senses cartridge type. The available set temperature range is 33 °C to 39 °C. The device also contains a backup battery to provide power only for 15 minutes.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the Precision Flow® HVNI, based on the provided text:

    Acceptance Criteria and Device Performance

    The acceptance criteria for the Precision Flow® HVNI are primarily non-clinical, focusing on safety, performance standards, software validation, and labeling requirements. The clinical studies establish safety and effectiveness by demonstrating non-inferiority or comparable outcomes to existing treatments.

    Acceptance Criteria CategoryAcceptance CriteriaReported Device Performance/Evidence
    Biocompatibility/MaterialsPatient-contacting components must be demonstrated to be biocompatible (Special Control 1). Tests include Cytotoxicity, Sensitization, Intracutaneous Reactivity, Extractables and Leachables, Volatile Organic Compounds (VOC), and Particulate Matter (PM2.5).All biocompatibility tests passed, demonstrating the biocompatibility of the device, in accordance with ISO 10993-1.
    Shelf Life/Reprocessing/SterilityCleaning instructions for reusable components must be validated (Special Control 3). Acceptance criteria for cleaning validation: no visible soil, protein level <6.4 micrograms/cm², TOC <12 micrograms/cm². Low-level disinfection validated by ≥ 6 log reduction of specified test organisms.Cleaning validation study met acceptance criteria: no visible soil, protein level <6.4 micrograms/cm², and TOC <12 micrograms/cm². Low-level disinfection demonstrated ≥ 6 log reduction of E. coli, S. aureus, K. pneumoniae, and P. aeruginosa.
    EMC/Electrical/Mechanical/Thermal SafetyElectrical safety, thermal safety, mechanical safety, electromagnetic compatibility, and radiofrequency identification (RFID) testing must be performed (Special Control 4). Standards: IEC 60601-1-2:2014, AAMI / ANSI ES60601-1:2005/(R) 2012 and A1:2012, AIM Standard 7351731 (RFID reference).Device passed tests in accordance with IEC 60601-1-2:2014, AAMI / ANSI ES60601-1:2005/(R) 2012 and A1:2012. Tested for RFID exposure using AIM Standard 7351731 as a reference.
    SoftwareSoftware verification, validation, and hazard analysis must be performed (Special Control 5). Adequate software documentation consistent with a "Major" level of software concern (FDA Guidance, May 11, 2005).Adequate software documentation provided. Software validation and verification testing demonstrated that the device met its design, implementation, and cybersecurity requirements.
    Non-clinical Performance TestingDevice performs as intended under anticipated conditions for use (Special Control 2). Specific tests: a. Alarm testing b. Continuous use thermal stability testing c. Humidity output testing d. Blender performance (FiO2 blending accuracy) Additional tests: Battery performance, Cannula pressure drop (blocked tube alarm not initiated). Standards: AAMI / ANSI / IEC 60601-1-8:2006 & A1:2012, ISO 8185:2007, ISO 80601-2-74.Blender performance: Verified set FiO2 at various flowrates (21%-100% FiO2) with high/low flow cartridges. Thermal Stability: Output temperature remained within specifications for 15 days across full range of settings, and another 15 days at max temp (43°C) and max flow (8/40 LPM). Battery performance: No change in set flow and FiO2 observed. Nasal cannula performance: Acceptable pressure drop at applicable flows, blocked tube alarm not initiated. Alarm testing, continuous use thermal stability, humidity output, and blender performance were implied to be successful as per "Additional testing".
    Clinical Effectiveness (Adults)Non-inferiority to NIPPV in avoidance of failure (including intubation/mechanical ventilation) within 72 hours of initiation (Primary Endpoints for Doshi et al. study).Doshi et al. study: Intubation within first 72 hours: 7% (HVNI) vs. 13% (NIPPV). Demonstrated non-inferiority (Wald p<0.001) for intubation rates. While arm failures (crossover) were higher in HVNI (26% vs 17%), the intubation rate was lower, suggesting clinical usefulness. Secondary findings: HVNI favored for pCO2 normalization at 240 min (p=0.02), physician perception (respiratory response, comfort/tolerance, simplicity of use), and patient perception of dyspnea was similar.
    Clinical Effectiveness (Neonates)Safety and effectiveness comparable to CPAP/nCPAP/NIPPV for respiratory distress/hypoxemia, without increasing adverse outcomes (pneumothorax, infection, ROP surgery, extubation failure, BPD, air leaks, IVH, NEC, sepsis). Non-inferiority for intubation rates within 72 hours of therapy initiation (Lavizzari et al. study).McQueen et al.: Compared to VON database (CPAP predominant), HFT did not increase pneumothorax (4.8% vs 4.4%) or nosocomial infection (14.8% vs 15.1%). Lower ROP surgery (2.2% vs 3.2%). Supported effectiveness and safety. Collins et al.: Non-inferior extubation failure rates for HFT (22%) vs nCPAP (34%). Similar clinical outcomes (BPD), trend for shorter supplemental oxygen in HFT. Improved nasal trauma score with HFT. Lavizzari et al.: Non-inferiority for primary endpoint (mechanical ventilation within 72 hrs). Failure rates: HFT (10.8%) vs nCPAP/BiPAP (9.5%). Lower bound of 95% CI (-6.0%) below zero met non-inferiority criteria. Similar secondary outcomes (duration of support, surfactant, air leaks, BPD). Kugelman et al.: Comparable failure to intubate rates for HFT (34.2%) vs NIPPV (31.6%). No clinically meaningful differences in secondary outcomes (air leak, nasal trauma, BPD, IVH, NEC, sepsis).
    LabelingLabeling must meet 21 CFR 801.109, include specific information (FiO2 ranges, flowrates, continuous monitoring warning, condensation warning, alarm description/function, not a CPAP device warning, aspiration risk warning) (Special Control 6).Labeling meets 21 CFR 801.109 and includes all specified warnings and descriptions as per Special Control 6.
    Adverse EventsNo device-related serious adverse events.In the Doshi et al. adult clinical study, there were no adverse events related to the use of the device. All adverse events were related to the underlying patient condition. The neonate studies also did not raise new safety concerns.

    Study Information

    Due to the nature of this particular device (high flow humidified oxygen delivery device) and the provided text, the "studies" are clinical trials and published literature, rather than an AI/ML-specific validation study with test sets, ground truth experts, or training sets in the typical sense. The regulatory submission leverages existing clinical evidence.

    For the Adult Clinical Study (Doshi et al.):

    1. Sample size used for the test set and the data provenance:

      • Sample Size: 204 adult patients (104 in HVNI arm, 100 in NIPPV arm).
      • Data Provenance: Multi-center, prospective, randomized clinical trial conducted at 5 sites. Country of origin not explicitly stated but implies US context for FDA submission.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This is a clinical trial where clinical outcomes (e.g., intubation, failure to tolerate device, physiological parameters like O2 saturation, PaO2, PCO2, relief of respiratory distress, deteriorating medical status) were the ground truth. These outcomes were assessed by treating clinicians (physicians, nurses, respiratory therapists) at the study sites. The "experts" are the medical professionals directly involved in patient care and decision-making within the clinical trial setting. Specific qualifications beyond "physician" are not detailed in the text.
    3. Adjudication method for the test set:

      • Not explicitly stated as a formal adjudication panel in the typical AI/ML context. Treatment failure was defined by clear clinical criteria (need for intubation, device intolerance, failure to oxygenate/ventilate, failure to relieve distress, deteriorating medical status). These assessments are inherent to clinical practice and were likely made by the treating clinicians in real-time, within the trial's protocol.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, this was not an MRMC study comparing human readers with and without AI assistance. It was a comparative effectiveness study between two medical devices/therapies (HVNI vs. NIPPV).
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • No, this device is a medical device for patient breathing support; it is not an AI algorithm performing a diagnostic task. Its performance is measured by its clinical impact on patients.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Outcomes Data: Primary ground truth was clinical outcomes data, specifically "treatment failure rate, defined as the need for intubation, and arm failure rate, defined as the decision for crossover to the alternate therapy, within 72 hours of initiation of assigned therapy." Other outcomes included vital signs, blood gas analysis (e.g., pCO2), physician perception, and adverse events.
    7. The sample size for the training set:

      • Not applicable in the context of this device being a medical hardware system, not an AI/ML algorithm requiring a specific training set of patient data.
    8. How the ground truth for the training set was established:

      • Not applicable for the same reason as above.

    For the Neonate Patient Population (McQueen et al., Collins et al., Lavizzari et al., Kugelman et al.):

    1. Sample size used for the test set and the data provenance:

      • McQueen et al.: Retrospective cohort study. HFT group: Data from five centers, no specific n given for test set, but stated "used Vapotherm HVNI as their primary and predominant mode." Compared to Vermont Oxford Network (VON) database: 176,599 very low birth weight baby admissions (control/comparison data).
      • Collins et al.: 132 low birth weight infants (< 2500g, <32 weeks gestation). Randomized (67 HFT, 65 nCPAP). Data provenance: Randomized Controlled Trial.
      • Lavizzari et al.: Randomized trial, 298 infants (158 HFT, 158 nCPAP/BiPAP). Data provenance: Prospective, randomized clinical non-inferiority trial.
      • Kugelman et al.: 76 infants (<35 week gestational age, >1000g). Randomized (38 HFT, 38 NIPPV). Data provenance: Randomized Pilot Study.
      • Data Provenance for all neonate studies: Published literature, likely international based on author names and journal types, but not explicitly stated for each study.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Similar to the adult study, ground truth involved clinical outcomes assessed by treating medical professionals (neonatologists, nurses, etc.) within the context of their respective clinical trials or retrospective data collection. Specific expert qualifications beyond clinical roles are not detailed within the provided text.
    3. Adjudication method for the test set:

      • Not explicitly stated as a formal adjudication panel. Outcomes were generally defined by protocol-driven clinical events (e.g., extubation failure, need for intubation, physiological parameters).
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. These are clinical studies comparing therapies, not AI assistance for human readers.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • No.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Outcomes Data: Primary ground truths included extubation failure rates, rates of intubation/mechanical ventilation, occurrences of pneumothorax, nosocomial infection, ROP surgery, bronchopulmonary dysplasia (BPD), air leaks, IVH, NEC, sepsis, duration of respiratory support, need for surfactant, time to full feeds, length of stay, mortality, and nasal trauma scores.
    7. The sample size for the training set:

      • Not applicable.
    8. How the ground truth for the training set was established:

      • Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K111640

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-10-07

    (116 days)

    Product Code
    Regulation Number
    868.5450
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Precision Flow® - Heliox is intended to warm and add moisture to breathing gases from an external source for administration to neonate/infant, pediatric and adult patients in the hospital and sub-acute institutions. It adds heat and moisture to a blended medical heliox (79% helium. 21% oxygen)/oxygen mixture and assures the integrity of the precise heliox (79% helium, 21% oxygen)/oxygen mixture via an integral oxygen analyzer. The flow rates may be from 1 to 40 liters per minute via nasal cannula.

    Device Description

    The Precision Flow® - Heliox consists of two parts: The main unit which contains all the electrical and electronic components including the electronic blender and flow controllers. All the sensors are located in the main unit. The main unit has no water pathways and the gas pathway contains only dry gas at room temperature, and consequently does not need internal cleaning or disinfection. The disposable components comprising the disposable water module, vapor transfer cartridge and heated delivery tube. Conditions in the circulating water and gas streams are sensed remotely via the interface between the main unit and the disposable module.

    AI/ML Overview

    The Vapotherm Precision Flow® - Heliox is a medical device intended to warm and humidify breathing gases, specifically a heliox/oxygen mixture, for neonate/infant, pediatric, and adult patients in hospital and sub-acute settings.

    Here's a breakdown of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document lists various standards that the Precision Flow® - Heliox met, with some exceptions. It also includes functional and design verification tests. The "Reported Device Performance" for these tests is generally stated as: "In all instances, the Precision Flow® - Heliox functioned as intended and the results observed were as expected." Specific quantitative performance data is limited within this summary but is generally implied by compliance with the referenced standards.

    Acceptance CriteriaReported Device Performance
    I. Compliance with Standards:"In all instances, the Precision Flow® - Heliox functioned as intended and the results observed were as expected." (Implied compliance with the intent of the standard, with specific exceptions noted)
    ISO 62304: Medical Device Software - Software Life Cycle ProcessMet
    ISO 8185:2007: Respiratory tract humidifiers for medical useMet
    ISO 10993-5:2009: Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicityMet
    ISO 10993-10:2010: Tests for Irritation and SensitizationMet
    ISO 11195:1995: Gas mixers for medical use - Stand-alone gas mixersMet
    ISO 14971:2007: Medical devices - Application of risk management to medical devicesMet
    ANSI/AAMI/ISO 15223-1:2007/A1:2008: Medical Devices - Symbols to be used with Medical Device Labels, Labeling and information to be supplied - Part 1: General RequirementsMet
    ANSI/AAMI ES60601-1:2005: Medical electrical equipment – Part 1: General requirements for basic safety and essential performanceMet
    ANSI/AAMI/IEC 60601-1-2:2007 3rd edition: Medical Electrical Equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic Compatibility - Requirements and testsMet (with exceptions: Section 5: Device cannot produce hazardous radiation; Section 6: Anesthetic mixtures not applicable)
    EN60601-1-4:1999: Medical Electrical Equipment: Part 1: General Requirements for Safety 4. Collateral Standard: Programmable Electrical Medical SystemsMet
    IEC 60601-1-8 2006-10: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systemsMet
    IEC 60529 IPX1: Drip ProofMet
    AAMI TIR 32:2004: Software Risk ManagementMet
    AAMI TIR 36:2007: Validation of SoftwareMet
    ISTA -1A:2001: Procedure 1A: Packaged-Products weighing 150 LB (68 KG) or LessMet
    EPA/625/R-96/010b (VOC) Method TO-15 (1999): Compendium Method TO-15 Determination of Volatile Organic Compounds (VOCs) in Air collected in specially-Prepared canisters and analyzed by chromatography/mass spectrometry (GC/MS)Met
    NIOSH Method 0500 (1994): NIOSH manual of Analytical methods (NMNM) fourth edition August 1994 (Particulate Matter Testing)Met
    II. Design Verification Plan (Required Tests):"In all instances, the Precision Flow® - Heliox functioned as intended and the results observed were as expected."
    Functional Performance TestMet
    Alarms & Fault Conditions TestMet
    Patient Connector Attachment TestMet
    Water Connector Attachment TestMet
    Extended Life TestMet
    Membrane Switch and Adjustment Dial TestMet
    Software Verification TestMet
    Fluid & Gas Pathways Leak TestMet
    Cartridge Insertion TestMet
    Door TestMet
    IPX TestMet
    Environmental Temperature/Humidity Extremes TestMet
    III. Performance Characteristics (Implied by Comparison for Substantial Equivalence):
    Oxygen % Range21 to 100% (Equivalent to predicates)
    FiO2 Accuracy2 to 40 lpm: +/-2%; 1 to 2 lpm: +/-5% (Equivalent to predicates)
    Max Flow Range40 lpm (Equivalent to predicates)
    Alarm Sound Level (Medium Priority)47 dB @ 1m (Equivalent to predicate)
    Alarm Sound Level (Low Priority)45 dB @ 1m (Equivalent to predicate)

    2. Sample Size Used for the Test Set and Data Provenance

    The 510(k) summary does not provide specific sample sizes for particular performance tests or explicitly mention a "test set" in the context of clinical or image-based studies. The performance data section broadly states: "In all instances, the Precision Flow® - Heliox functioned as intended and the results observed were as expected." This implies that the design verification tests listed were conducted on the device.

    • Sample Size for Test Set: Not explicitly stated for each test, but implied to be sufficient for demonstrating compliance with cited standards and internal design verification plans.
    • Data Provenance (Country of Origin, Retrospective/Prospective): Not specified. The tests are likely laboratory or bench testing, not clinical studies involving patient data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • This information is not applicable to this type of device and study. The testing for the Precision Flow® - Heliox involves engineering and performance validation against objective technical standards (e.g., flow rates, temperature, alarm function, electrical safety), not subjective interpretation requiring expert consensus.

    4. Adjudication Method for the Test Set

    • This information is not applicable as the evaluation relies on objective measurements against engineering specifications and regulatory standards, not subjective interpretations requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of AI vs. Without AI Assistance

    • No, an MRMC comparative effectiveness study was not done. This device is a medical gas humidifier/blender, not an AI-powered diagnostic or decision-support tool. It does not involve human readers interpreting cases with or without AI assistance.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • No, this is not applicable. The device is hardware with integrated software for control, not a standalone algorithm. Its performance is inherent to its physical operation and control systems.

    7. The Type of Ground Truth Used

    • The "ground truth" for this device's performance is established by objective engineering specifications, technical standards (e.g., ISO, ANSI/AAMI, IEC), and manufacturer-defined design requirements. This includes accuracy of gas blending, flow rates, temperature control, alarm thresholds, electrical safety, and biocompatibility.

    8. The Sample Size for the Training Set

    • This is not applicable in the context of a traditional "training set" for machine learning or AI models. The device's software is likely developed using standard software engineering practices and verified accordingly, rather than trained on a dataset.

    9. How the Ground Truth for the Training Set Was Established

    • This is not applicable for the reasons stated above. The "ground truth" for the device's operational parameters is derived from engineering design, relevant medical standards, and intended physiological effects of humidified gas delivery.
    Ask a Question

    Ask a specific question about this device

    K Number
    K072845

    Validate with FDA (Live)

    Device Name
    PRECISION FLOW
    Manufacturer
    Date Cleared
    2008-07-17

    (287 days)

    Product Code
    Regulation Number
    868.5450
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Precision Flow™ is intended for use to add warm moisture to breathing gases from an external source for administration to a neonate/infant, pediatric and adult patients in the hospital, sub-acute institutions, and home settings. It adds heat and moisture to a blended medical air/oxygen mixture and assures the integrity of the precise air/oxygen mixture via an integral oxygen analyzer. The flow rates may be from 1 to 40 liters per minute via nasal cannula.

    Device Description

    The Precision Flow™ consists of two parts: The main unit which contains all the electrical and electronic components including the electronic blender and flow controllers. All the sensors are located in the main unit. The main unit has no water pathways and the gas pathway contains only dry gas at room temperature, and consequently does not need internal cleaning or disinfection. The disposable components comprising the disposable water module, vapor transfer cartridge and heated delivery tube. Conditions in the circulating water and gas streams are sensed remotely via the interface between the main unit and the disposable module.

    AI/ML Overview

    This 510(k) summary describes a medical device, the Vapotherm Precision Flow™, which is a respiratory gas humidifier. It is not an AI/ML powered device, therefore, many of the requested categories related to AI performance, ground truth, and expert evaluation are not applicable.

    Here's an analysis of the provided text based on your request, highlighting the non-applicability of AI/ML specific criteria:

    Acceptance Criteria and Device Performance (Non-AI/ML Device)

    The Vapotherm Precision Flow™ is cleared based on demonstrating substantial equivalence to predicate devices, primarily through non-clinical performance data (testing) and a comparison of its technological characteristics to existing devices. The "acceptance criteria" here are implicitly that the device performs its intended functions (humidifying and blending medical gases) safely and effectively, and that these performance characteristics are comparable to legally marketed predicate devices without raising new safety or effectiveness concerns.

    Since this is not an AI/ML device, performance metrics like sensitivity, specificity, accuracy, or AUC are not applicable or reported in this 510(k) summary. Instead, the performance is demonstrated through various engineering, safety, and biocompatibility tests.

    Acceptance Criteria (Implied)Reported Device Performance
    Biocompatibility: No adverse biological reactions.Biocompatibility Tests (LLNA, Intracutaneous Reactivity, MEM Elution): Performed, implying acceptable results (otherwise it wouldn't be marketed).
    Emissions: No harmful volatile organic compounds or particulates.Volatile Organic Compounds (VOC), Particulate Matter tests: Performed to recognized standards, implying acceptable results.
    Electrical Safety: Meets international safety standards.EMC Test, Emissions Test, Safety Inspection Test, IEC 60601-1, IEC 60601-1-4, IEC 60601-1-8: Performed, ensuring compliance with general requirements for safety, programmable electrical medical systems, and alarm systems.
    Microbiological Safety: No detectable bacteria after extended use.Bio Burden, 30 Day Comparative Use Testing: "No detectible bacteria in any water condensation samples from the Precision Flow® devices either initially (3 days) or after 30 days of operation."
    Software Functionality: Verified and validated.Ximedica TRP 1097, TRP 1071 (Black Box), Unit Test Case Listing - 3VAP1004/TRP 1092 (White Box): Software verified and validated in accordance with applicable FDA guidance.
    Thermal Stability: Maintains temperature regulation.Thermal Stability: Performed, indicating proper temperature control.
    Blender Accuracy: Maintains precise air/oxygen mixture.Blender Comparison Performance: Performed, indicating comparable performance to predicate blenders, assuring integrity of the precise air/oxygen mixture via an integral oxygen analyzer.
    Substantial Equivalence: Comparable to predicate devices.The device has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate devices. Performance data demonstrate it is as safe and effective as the predicate devices, with minor differences raising no new issues.

    Non-Applicable (N/A) for AI/ML Specific Questions:

    Given that K072845 describes a humidifier, not an AI/ML-powered device, the following categories are not applicable. The submission process for such a device focuses on engineering specifications, safety testing, and performance validation against established standards and predicate devices, rather than statistical performance against a ground truth dataset.

    2. Sample size used for the test set and the data provenance: N/A (Not an AI/ML device, no "test set" in the context of an algorithm's classification/prediction performance). Testing involved physical hardware, software, and biological compatibility, not a data-driven test set.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: N/A (Not an AI/ML device, no ground truth established by experts for algorithmic performance).

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: N/A (Not an AI/ML device, no adjudication of algorithm outputs).

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: N/A (Not an AI/ML device. No human-in-the-loop AI assistance involved).

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: N/A (Not an AI/ML device, no standalone algorithm performance to evaluate).

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): N/A (Not an AI/ML device). The "ground truth" for this device relates to physical measurements meeting specifications (e.g., oxygen percentage, temperature, bacterial counts being zero).

    8. The sample size for the training set: N/A (Not an AI/ML device, no "training set"). Software verification and validation are described, but this refers to traditional software engineering processes, not machine learning model training.

    9. How the ground truth for the training set was established: N/A (Not an AI/ML device, no "training set" or ground truth in the ML context).

    Ask a Question

    Ask a specific question about this device

    K Number
    K971799

    Validate with FDA (Live)

    Date Cleared
    1998-02-06

    (267 days)

    Product Code
    Regulation Number
    882.5550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Precision Flow OSV II is an implantable system used in the treatment of patients with hydrocephalus, to shunt CSF from the ventricles to the peritoneal cavity or other appropriate drainage site such as the heart's right atrium.

    Device Description

    The Orbis Sigma Valve and the Precision Flow OSV IT are both variable resistance valves that manuain a drainage rate close to the rate of CSF secretion (~ 20 ml/h) within the physiological range of intracranial pressure. The mechanism incorporates a safety pressure relief mode to prevent accidental intracranial hypertension. The principle of operation consists of a flow restricting diaphragm, a seat, and a notched pln. The diaphragm, held between two polysulfone parts of the casing, reacts to variations in pressure. A synthetic ruby seat is inserted into the center of the diaphragm. The clearance between the seat aperture and the notched synthetic ruby nin varies depending on seat movements along the pin as pressure varies. Precision Flow OSV II and Orbis Sigma Valve are characterized by the same three stages of operation, that are defined below : Stage / - Low Differential Pressure (DP), Stage II - Flow Regulation, Stage III - Pressure Rellef Mode.

    AI/ML Overview

    This document, K971799, is a 510(k) premarket notification for the Precision Flow OSV II, a Central Nervous System (CNS) Fluid Shunt System. The primary method used to demonstrate safety and effectiveness, and thus meet acceptance criteria, is through substantial equivalence to a predicate device, the Cordis Orbis Sigma Valve.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The core "acceptance criteria" are the performance specifications of the predicate device, which the Precision Flow OSV II claims to replicate. The study described focuses on demonstrating that the new device's performance is identical to the predicate device.

    Performance AttributeAcceptance Criteria (from Predicate)Reported Device Performance (Precision Flow OSV II)
    Intended UseTreatment of hydrocephalus, shunting CSF from ventricles to peritoneal cavity or other appropriate drainage site.Identical to predicate device.
    Operating PrincipleVariable resistance valve maintaining drainage rate close to CSF secretion rate (~20 ml/h) within physiological ICP range, with safety pressure relief mode.Identical to predicate device.
    Hydrodynamic Specifications- Stage I (Low Differential Pressure): Flow rate 5-18 ml/hr at DP 30-120 mm H2O.Identical to predicate device.
    - Stage II (Flow Regulation): Flow rate 18-30 ml/hr at DP 120-300 mm H2O.Identical to predicate device.
    - Stage III (Pressure Relief Mode): Rapid flow rate for IVP normalization, then reverts to Stage II or I.Identical to predicate device.
    BiocompatibilityDocumented biocompatibility of materials in predicate devices.No new issues raised; materials are the same as predicate devices.
    Design, Materials, Manufacturing Methods, SpecificationsEquivalent to predicate device.Equivalent to predicate device.

    2. Sample Size Used for the Test Set and Data Provenance

    This submission does not describe a clinical study with a "test set" in the sense of a new patient cohort for the Precision Flow OSV II. Instead, it relies on the established performance of its predicate, the Orbis Sigma Valve.

    • Sample Size for Test Set: Not applicable. The "test" is a comparison to the established specifications of the predicate device, not an independent clinical trial on a new set of patients for the Precision Flow OSV II.
    • Data Provenance: The hydrodynamic specifications and operating principles are derived from the Orbis Sigma Valve. The document references a clinical study and use over more than seven years validating the concept for the Orbis Sigma Valve (Sainte Rose, Hooven MD, Hirsch JF: A new approach in the treatment of hydrocephalus. J Neurosurg. 86:213-226, 1987). This suggests retrospective data from clinical use of the predicate, likely from various international locations where the device was marketed.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Number of Experts: Not explicitly stated for establishing "ground truth" for the new device.
    • Qualifications of Experts: The "concept of the flow regulation linked to the OSV hydrodynamic specifications has been developed by Pr C. Sainte Rose M.D." This indicates that the initial specifications were developed by a medical professional. For the 510(k) submission, FDA reviewers (Celia M. Witten, Ph.D., M.D., Director, Division of General and Restorative Devices, Office of Device Evaluation) acted as experts in reviewing the claim of substantial equivalence.

    4. Adjudication Method for the Test Set

    Not applicable, as there was no "test set" requiring adjudication in the context of a new clinical study for the Precision Flow OSV II. The device's performance is asserted as identical to, and therefore meeting the established performance of, its predicate.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This device is a physical medical implant (shunt), not an imaging or diagnostic AI tool that would typically involve a multi-reader study. The evaluation focuses on physical and hydrodynamic equivalence.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done

    Not applicable. This device is a physical CNS shunt system and does not involve an algorithm or AI.

    7. The Type of Ground Truth Used

    • The "ground truth" for the predicate device's performance was established through:
      • Expert Consensus/Development: The concept developed by Pr C. Sainte Rose M.D.
      • Clinical Studies and Use: "Validated by clinical studies and use over more than seven years" (for the predicate Orbis Sigma Valve). This implies outcomes data from a real-world patient population.
    • For the Precision Flow OSV II, the ground truth is that its design, materials, manufacturing methods, and performance specifications are identical/equivalent to this established predicate.

    8. The Sample Size for the Training Set

    Not applicable. There is no training set in the context of machine learning or AI, as this is a physical medical device. The "training" for its design and performance would be the extensive research, development, and clinical validation conducted for its predicate device over several years.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" in the AI sense. The "ground truth" for the predicate device's efficacy and performance, which forms the basis for demonstrating equivalence for the Precision Flow OSV II, was established through a combination of:

    • Expert Design and Theoretical Principles: Pr C. Sainte Rose M.D.'s development of the flow regulation concept and hydrodynamic specifications.
    • Clinical Validation: "Clinical studies and use over more than seven years" for the predicate Orbis Sigma Valve, which would have involved patient outcomes data to confirm the device's intended therapeutic effect and performance in vivo.