Found 2 results

510(k) Data Aggregation

    K Number
    K103594

    Validate with FDA (Live)

    Device Name
    PONTO PRO POWER
    Manufacturer
    Date Cleared
    2011-06-15

    (189 days)

    Product Code
    Regulation Number
    874.3302
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ponto Pro Power is intended for the following patients and indications:

    • Patient with conductive or mixed hearing losses, who can still benefit from amplification of the sound. The pure tone average (PTA) bone conduction (BC) threshold of the indicated ear should be better than or equal to 55 dB HL (measured at 0.5, 1, 2 and 3 kHz)
    • Bilateral fitting is applicable for most patients having a symmetrically conductive or mixed hearing loss. The difference between the left and right sides' BC thresholds should be less than 10 dB on average measured at 0:5, 1, 2 and 4 kHz, or less than 15 dB at individual frequencies.
    • Patients who have a profound sensorineural hearing loss in one ear and normal hearing in the opposite ear (i.e. single sided deafness or "SSD"). The pure tone average (PTA) air conduction (AC) threshold of the hearing ear should then be better than or equal to 20 dB HL (measured at 0.5, 1, 2 and 3 kHz).
    • Also indicated for any patient who is indicated for an air-conduction contralateral routing of signals (AC CROS) hearing aid, but who for some reason cannot or will not use an AC CROS.

    The placement of a bone anchored implant is contraindicated for patient below the age of 5.

    The Ponto Pro Power sound processor is intended to be connected to the Oticon Medical bone anchored implant system or to the Baha@ Abutment snap coupling from Cochlear Bone Anchored Solutions AB (ref no. 90410, 90305, 90434, 90480).

    Device Description

    Ponto Pro Power is a bone conduction hearing aid which is connected to an implant with a skin penetrating abutment which has been surgically anchored in the bone behind the ear. Vibrations generated by the sound processor are transmitted directly through the skull bone to the cochlea as bone conduction sound. Ponto Pro Power has a coupling so that it can easily be connected and disconnected from the abutment by the user or alternatively from a head band accessory, to function as a conventional bone conductor. Using a computer based fitting system, Genie Medical; Ponto Pro Power can be adjusted to the patient's individual hearing requirements. Ponto Pro Power contains a number of advanced features: Wind Noise Reduction, Feedback Management System (Feedback Manager and Dynamic Feedback Cancellation), Automatic Gain Control, Speech Guard, BC In-situ Audiometry, Automatic Multiband Adaptive Directionality, Tri-state Noise Reduction, Data Logging, and Learning Volume Control.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Oticon Medical Ponto Pro Power bone-anchored sound processor, based on the provided text:

    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria are primarily established through substantial equivalence to predicate devices, particularly the Baha Intenso (K081606). The device performance is verified through bench testing and clinical evaluation comparing it to the predicate.

    CharacteristicAcceptance Criteria (Predicate: Baha Intenso K081606)Reported Device Performance (Ponto Pro Power)Device Meets Criteria?
    Indications for UseImprovement of hearing for patients with conductive and mixed hearing losses, bilateral fitting, and single-sided deafness.Equivalent indications for use.Yes
    Pure tone average bone conduction threshold (conductive/mixed hearing loss)Better than or equal to 55 dB HL (measured at 0.5, 1, 2, 3 kHz)Verified to successfully fit the same patient population, i.e., those with bone-conduction thresholds of < 55 dB HL.Yes
    Max gain at 1600 Hz45 dB47 dBYes (within acceptable equivalence, and potentially improved)
    Wind Noise ReductionN/A (new feature)Bench performance data verify that device responds to airflow by reducing overall gain and switching mode.Yes
    Dynamic Feedback Cancellation (DFC)N/A (new feature)Bench performance data show feedback is identified and cancelled without artifacts.Yes
    Speech GuardN/A (new feature)Bench testing shows ability to maintain linear processing while responding to rapid environmental sounds, less distortion than traditional compression.Yes
    BC In-situ AudiometryN/A (new feature, but equivalent to Baha BP 100)Difference compared to conventional unmasked BC hearing thresholds (audiometer with RadioEar B-71 transducer) is on average less than 5 dB.Yes
    Aided ThresholdsEssentially similar to predicate device (Baha Intenso)Essentially similar for both devices; no significant differences observed.Yes
    Speech Perception in QuietEquivalent to predicate device (Baha Intenso)Verified to be substantially equivalent to predicate device.Yes
    Speech Perception in NoiseEquivalent to predicate device (Baha Intenso)Verified to be substantially equivalent to predicate device.Yes
    Subjective Evaluation (Wind Noise & Feedback)Equivalent to predicate device (Baha Intenso)Verified to be substantially equivalent to predicate device.Yes

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size:
      • Clinical Evaluation comparing Ponto Pro Power to Baha Intenso: 20 experienced users of Intenso; 18 subjects with mixed hearing losses and 2 with single-sided deafness. Only the 18 subjects with conductive and mixed hearing losses were analyzed for the equivalence study.
      • BC In-situ Audiometry verification: 25 hearing-impaired patients.
    • Data Provenance: The document does not explicitly state the country of origin for the clinical study data but mentions the submitter is Oticon Medical AB in Sweden. The study is described as "unpublished" and "funded by the submitter," indicating it was a prospective study conducted specifically for this submission.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    • For the clinical evaluation comparing Ponto Pro Power to Baha Intenso: The "ground truth" here is the performance of the predicate device (Baha Intenso), against which the Ponto Pro Power is compared. The study involved "experienced users of Intenso," implying their established hearing experience with the predicate device served as a baseline.
    • For BC In-situ Audiometry verification: The ground truth for audiometric BC thresholds was established using "conventional unmasked BC hearing thresholds...with an audiometer using a RadioEar B-71 bone conduction transducer." This is a standard clinical method for establishing audiometric ground truth.
    • Number of Experts: Not explicitly stated how many audiologists or other experts conducted measurements or assessments, but it implies standard clinical practice by qualified professionals. Their specific qualifications (e.g., years of experience) are not provided.

    4. Adjudication Method for the Test Set

    The document does not describe a formal adjudication method (like 2+1 or 3+1 consensus) for the clinical data. The clinical evaluation involved direct comparisons of objective measures (aided thresholds, speech perception) and subjective measures (APHAB, SSQ, questionnaires on user experience, satisfaction, and preference). For the BC In-situ Audiometry, it was a direct comparison of measurement methods.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No Multi-Reader Multi-Case (MRMC) comparative effectiveness study is mentioned. The study involved individual patients being tested with both the predicate device and the new device. The "before-and-after" or "cross-over" nature of the clinical study (users tested with Intenso prior to testing with Ponto Pro Power) allowed for direct comparison within individuals.

    • Effect size of human readers improvement with AI vs. without AI assistance: Not applicable, as this is a hearing aid device, not an AI diagnostic tool primarily interpreted by human "readers." The "AI" features (Wind Noise Reduction, DFC, Speech Guard, Automatic Multiband Adaptive Directionality, Tri-state Noise Reduction, Learning Volume Control) are integrated into the device's signal processing to directly improve the listener's experience, not to assist human interpretation of data.

    6. Standalone (Algorithm Only) Performance Study

    Yes, standalone (algorithm only) performance was done through bench performance testing for the new sound processing features:

    • Wind Noise Reduction: Tested when exposed to a "preset air flow" to verify response (reducing gain, switching mode).
    • Dynamic Feedback Cancellation (DFC): Tested to show feedback is identified and cancelled without artifacts.
    • Speech Guard: Tested "at different gain settings, input levels, and different types of noise input" to confirm linear processing and instantaneous response to environmental sounds.
    • BC In-situ Audiometry: Compared the device's measurement (via the sound processor) directly against a conventional audiometer with a specified transducer to evaluate its accuracy as a measurement tool.

    7. Type of Ground Truth Used

    • For demonstrating substantial equivalence in overall hearing improvement: The ground truth was the performance of the legally marketed predicate device (Baha Intenso), as measured by aided thresholds, speech perception in quiet, speech perception in noise, and subjective user feedback.
    • For BC In-situ Audiometry: The ground truth was conventional audiometric measurements obtained using a standard audiometer and RadioEar B-71 bone conduction transducer.
    • For new sound processing features (Wind Noise Reduction, DFC, Speech Guard): The ground truth was established by engineering design criteria and expected functional behavior verified through controlled bench tests.

    8. Sample Size for the Training Set

    The document does not explicitly mention a "training set" in the context of device development or any machine learning algorithms. The device's features (like Learning Volume Control or Automatic Multiband Adaptive Directionality) likely use internal algorithms that may have been developed and refined using R&D data, but no formal "training set" with a specified sample size is detailed in this regulatory summary.

    9. How the Ground Truth for the Training Set Was Established

    Since no formal "training set" is described for algorithm development, the method for establishing its ground truth is not provided. The Ponto Pro Power's advanced features rely on signal processing and environmental analysis, which would typically be developed based on audiology principles, acoustic modeling, and potentially large datasets of sound environments and user preferences, but these details are not part of this 510(k) summary.

    Ask a Question

    Ask a specific question about this device

    K Number
    K090996

    Validate with FDA (Live)

    Device Name
    PONTO PRO
    Manufacturer
    Date Cleared
    2009-07-29

    (112 days)

    Product Code
    Regulation Number
    874.3302
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ponto Pro is intended for the following patients and indications;

    • Patients with conductive or mixed hearing losses, who can still benefit from amplification of the sound. The pure tone average (PTA) bone conduction (BC) threshold for the indicated ear should be better than 45 dB HL (measured at 0.5, 1, 2 and 3 kHz).
    • Bilateral fitting is applicable for most patients having a symmetrically conductive or mixed hearing loss. The difference between the left and right sides' BC thresholds should be less than 10 dB on average measured at 0.5, 1, 2 and 4 kHz, or less than 15 dB at individual frequencies.
    • Patients who have a profound sensorineural hearing loss in one ear and normal hearing in the opposite ear, who for some reason will not or cannot use an AC CROS. The pure tone average (PTA) air conduction (AC) threshold of the hearing ear should then be better than 20 dB HL (measured at 0.5. 1, 2 and 3 kHz).

    The placement of a bone anchored implant is contraindicated for patients below the age of 5.

    The Ponto Pro sound processor is intended to be connected to the Oticon Medical bone anchored implant system or to the Baha® Abutment snap coupling from Cochlear BAS. In addition, the Oticon Medical abutments can be used for connection of the Baha® sound processors with snap coupling from Cochlear BAS (Baha Divino®, Baha IntensoTM, Baha Cordelle II, Baha BP100),

    Device Description

    The Ponto Pro is a bone conduction-type hearing aid. The Ponto Processor is connected to an implant with a skin penetrating abutment which as been surgically anchored in the bone behind the ear. Vibrations generated by the sound processor are transmitted directly through the skull bone to the cochlea as bone conduction sound. The Ponto Pro sound processor has a coupling so that it can be easily connected to and disconnected from the abutment by the user. The Ponto Pro can alternatively be connected to head band accessories, to function as a conventional bone conductor. Using a computer based fitting system the Ponto Pro can be adjusted to the patient's individual hearing requirements. The sound processor is intended to work with the Oticon Medical bone anchored implant system or the Baha® Abutment snap coupling from Cochlear BAS (ref no. 90410, 90305, 90434, 90480). In addition, the Oticon Medical abutments can be used for connection of the Baha® sound processors with snap coupling from Cochlear BAS (Baha Divino®, Baha Intenso"" Baha Cordelle II, Baha BP100; ref no. 90500, 90510, 90511, 90502, 90512, 90503, 90513, 90730, 90731, 90732, 90733, HBC 400-0, HCB 401-0, HCB 402-0, 91301, 91302, 91303, 91304, 91305).

    AI/ML Overview

    The provided document describes the Ponto Pro bone anchored sound processor, which falls under the classification of a Class II medical device (Hearing Aid, Bone Conduction, Product Code: LXB). The device is intended for improving hearing in patients with conductive and mixed hearing losses, bilateral fitting, and single-sided deafness.

    The submission is a 510(k) summary, which typically demonstrates substantial equivalence to a predicate device rather than presenting full clinical study data for new acceptance criteria. Therefore, the information provided focuses on the comparison to predicate devices and functional testing, not a detailed study with specific acceptance criteria and detailed device performance outcomes in a traditional clinical trial sense.

    Here's an analysis based on the provided text, addressing your points where information is available:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in the formal sense found in a clinical trial report or a performance goal. Instead, the submission relies on demonstrating substantial equivalence to predicate devices. The "reported device performance" is essentially that it functions equivalently to these established devices.

    CharacteristicAcceptance Criteria (Implied by Predicate Equivalence)Reported Device Performance (Ponto Pro)
    DesignBone conduction sound processor connected to surgically placed implantBone conduction sound processor connected to surgically placed implant
    Intended UseImprove hearing for patients with conductive/mixed hearing loss, bilateral fitting, single-sided deafnessSame as predicate; specific audiometric thresholds are given for intended use
    MaterialSound processor coupling: PEEK; Implant: TitaniumSound processor coupling: PEEK; Implant: Titanium
    Power RequirementZinc-air batteryZinc-air battery
    Max Gain33dB33dB
    Frequency Response125 Hz - 8 kHz125 Hz - 8 kHz
    Sound ProcessingDigitalDigital
    Patient FittingIndividual adjustment via computer-based fitting systemIndividual adjustment via computer-based fitting system
    AccessoriesVarious accessories for direct audio input (DAI)Various accessories for direct audio input (DAI)
    Safety & EfficacyNo new concerns compared to predicate devicesFunctional testing and software validation confirm equivalence, raising no new safety/efficacy concerns.

    2. Sample Size Used for the Test Set and Data Provenance

    The document explicitly states: "Testing of the Ponto Pro include functional testing and software validation. These tests verify that the Ponto Pro is functionally equivalent to the OBC, why we have come to the conclusion that further testing will not raise new issues of safety and efficacy."

    This indicates that the "test set" for performance evaluation was likely functional testing and software validation of the device itself, rather than a clinical study on a patient population.

    • Sample Size: Not applicable in the context of a patient-based test set, as no patient study is detailed for proving functional equivalence in this section. The "testing" refers to internal device verification.
    • Data Provenance: Not applicable for a patient-based test set. The functional and software testing would have been conducted by the manufacturer (Oticon Medical AB in Sweden).

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not provided. Since the submission relies on functional and software testing to demonstrate equivalence, and not on a clinical ground truth evaluation by experts for a patient test set, this detail is not relevant to the provided text.

    4. Adjudication Method for the Test Set

    Not applicable. There's no mention of a human-reviewed "test set" for which adjudication would be required. The "testing" described is functional and software-based.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, a multi-reader multi-case (MRMC) comparative effectiveness study is not mentioned or indicated. The submission focuses on substantial equivalence based on technological characteristics and functional testing.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    The device is a hearing aid, which by its nature requires human interaction (the patient wearing it and an audiologist fitting it). The concept of "standalone algorithm only" performance doesn't directly apply in the same way it would for an AI diagnostic imaging algorithm. The "functional testing and software validation" could be considered a form of standalone testing of the device's inherent functionality.

    7. The Type of Ground Truth Used

    For the purpose of this 510(k) submission, the "ground truth" for demonstrating safety and effectiveness relied on:

    • Predicate Device Performance: The established safety and efficacy of the legally marketed predicate devices (OBC, Baha Divino, and air conduction hearing aids with digital sound processing). The Ponto Pro's performance is deemed acceptable because it is substantially equivalent to these devices.
    • Functional Testing and Software Validation: Internal verification that the Ponto Pro performs according to its design specifications, ensuring it functions as intended and safely, similar to its predicates.

    No external "expert consensus, pathology, or outcomes data" is described as being used to establish a ground truth for a novel performance claim beyond equivalence.

    8. The Sample Size for the Training Set

    Not applicable. As this is a medical device (a hearing aid), not an AI algorithm trained on a dataset of patient cases, there is no "training set" in the context of machine learning. The device's design and software are developed through engineering processes, and its functionality is validated through testing and comparison to predicate devices, not by training on a data set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" in the context of this device. The "ground truth" for the device's development and validation is based on established audiological principles, engineering standards, and the performance of predicate devices.