Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K032541

    Validate with FDA (Live)

    Device Name
    POLESTAR N-20
    Date Cleared
    2003-09-12

    (25 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PoleStar N-20 is a Magnetic Resonance Diagnostic Device intended to produce transverse, sagittal, coronal, and oblique 2D and 3D images of the extremities and selected sections of the head. The images produced by the PoleStar N-20 reflect the spatial distribution of protons (Hydrogen Nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and T2*.

    Anatomical regions: extremities and selected sections of the head.

    Nuclei excited: H-1

    Diagnostic uses: T1, T2, T2* and density weighted imaging.

    The PoleStar N-20 is intended to be used intraoperatively in a standard operating room. When interpreted by trained physicians, the MR images provide information that can be useful in determining a diagnosis.

    Device Description

    The PoleStar N-20 utilizes a permanent magnet to acquire 2D single-slice, multi slice, and 3d volume images. A wide variety of pulse sequences are provided to the operator, including spin echo, gradient echo, fast spin echo, and steady state free precession acquisitions. The PoleStar N-20 is a widely open and compact Intraoperative MRI unit intended to be used in a typical pre-existing operating room. The PoleStar N-20 can be moved within the room between procedures, from the operating table to its Magnet Storage Cabinet, thus allowing the operating room to be used for any type of surgery.

    AI/ML Overview

    This document is a 510(k) summary for the Odin Medical Technologies PoleStar N-20, an intraoperative MRI device. The primary claim for substantial equivalence is based on a comparison to the predicate device, Odin's PoleStar N-10. This filing does not describe a study to prove acceptance criteria for device performance. Instead, it focuses on demonstrating that the PoleStar N-20 is substantially equivalent to an existing device (the PoleStar N-10) based on similar technical specifications and identical intended use.

    Therefore, most of the requested information regarding acceptance criteria, study design, sample sizes, expert involvement, and statistical analysis is not present in the provided text.

    Here's a breakdown of what can be extracted and what is missing:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't provide explicit "acceptance criteria" for performance that would typically be seen in a clinical validation study for a new diagnostic device. Instead, it offers a "Substantial Equivalence Comparison Table" that lists technical specifications of the new device (PoleStar N-20) against its predicate (PoleStar N-10). The "performance" in this context refers to these specifications matching or being comparable to the predicate.

    ParameterPoleStar N-10 (Predicate, K010850)PoleStar N-20 (New Device)Performance/Comparison
    Clinical applicationExtremities and selected sections of the headExtremities and selected sections of the headIdentical
    Magnet typePermanentPermanentIdentical
    Field strength0.12T0.13TSlight increase, deemed comparable
    5 gauss fringe field (radial/axial, m)1.52.2Minor difference, likely considered acceptable for similar use
    ShimmingPassive, activePassive, activeIdentical
    Gradient subsystem Strength mT/m2522Minor difference, likely considered acceptable
    Gradient subsystem Rise time to 10mT/m msec<1<1Identical
    Computer system - CPUPentium 586Pentium 586Identical
    Computer system - Memory size [MB]6464Identical
    Array processor4xDSP C44 TI4xDSP C44 TIIdentical
    Array processor - Memory size [MB]40004000Identical
    Storage mediaMagnetic disk, floppy diskMagnetic disk, floppy diskIdentical
    Number of images stored50005000Identical
    Imaging modesSingle, multislice, volume studySingle, multislice, volume studyIdentical
    Reconstruction time - single slice, sec<3/slice<3/sliceIdentical
    Reconstruction time - multislice, sec<3/slice<3/sliceIdentical
    Reconstruction time - volume sec<20/volume<20/volumeIdentical
    Cardiac gating (ECG/peripheral)NoNoIdentical
    Respiratory gatingNoNoIdentical
    AngiographyOptionalOptionalIdentical
    SpectroscopyNoNoIdentical
    Imaging - pulse sequenceSpin Echo, Fast Spin Echo, Gradient Echo, 2D 3DSpin Echo, Fast Spin Echo, Gradient Echo, 2D 3DIdentical
    Imaging - repetition time, msec10-5000 increments of 110-5000 increments of 1Identical
    Imaging - echo time, msec3-1503-150Identical
    Imaging - inversion time, msecN/AN/AIdentical
    Imaging - slice thickness, mm2-102-10Identical
    Imaging - scan orientationTransverse, coronal, sagittal, obliqueTransverse, coronal, sagittal, obliqueIdentical
    Imaging - measuring matrix64x64 to 256x256 steps of 1 in phase encoding64x64 to 256x256 steps of 1 in phase encodingIdentical
    Imaging - display matrix1024x7681024x768Identical
    Imaging - pixel intensity0-40950-4095Identical
    Surface coilsKnee, extremity, headKnee, extremity, headIdentical
    Bore diameter or WxH, cm24.5x3925.2x42Minor difference, likely considered acceptable
    Bore featuresOpen access to patientOpen access to patientIdentical
    Cooling system typeClosed loop water cooling (Gradients only)Closed loop water cooling (Gradients only)Identical
    Cryogen useNoNoIdentical
    Magnet weight, kg330400Minor difference, likely considered acceptable
    HxWxD, cm145x96x120153x97x120Minor difference, likely considered acceptable
    Field Of View (FOV), cm5-185-20Minor difference, likely considered acceptable
    Dicom 3.0 interfaceYesYesIdentical
    Power requirements3x208V, 16 Kva, <10000 BTU/hr A/C3x208V, 15 Kva, <10000 BTU/hr A/CMinor difference, likely considered acceptable

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Not Applicable. This document does not describe a clinical study with a test set of data. The substantial equivalence is based on technical specifications and intended use comparison.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Not Applicable. No ground truth establishment is described as there is no clinical test set in this filing. The filing mentions that "When interpreted by trained physicians, the MR images provide information that can be useful in determining a diagnosis," which refers to the eventual use of the device, not a study performed for this 510(k).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not Applicable. No test set or adjudication method is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This document describes an MRI diagnostic device, not an AI or CAD system, and therefore, an MRMC study related to AI assistance would not be relevant or expected for this device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No. This is a hardware device (MRI scanner) which requires human operation and interpretation. The concept of standalone algorithm performance does not apply.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not Applicable. As no clinical study or test set is described, no ground truth was established for this 510(k filing for the PoleStar N-20.**

    8. The sample size for the training set:

    • Not Applicable. This document does not describe a machine learning or AI-based device that would require a training set.

    9. How the ground truth for the training set was established:

    • Not Applicable. No training set or ground truth for a training set is described.

    In summary: The provided 510(k) summary focuses on demonstrating substantial equivalence of the PoleStar N-20 to its predicate, the PoleStar N-10, by comparing technical specifications and intended use. It does not include information about performance acceptance criteria validated through a clinical study with a test set, expert ground truth establishment, or AI-related metrics.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1