    K152665

    Validate with FDA (Live)

    Date Cleared
    2015-10-26

    (39 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PolarCath Peripheral Dilatation System's intended use is for the dilatation of stenoses in the peripheral vasculature (iliac, femoral, popliteal, infrapopliteal, renal and subclavian arteries) and for the treatment of obstructive lesions of polytetrafluoroethylene (PTFE) access grafts or native arteriovenous dialysis fistulae. The PolarCath Peripheral Dilatation System is also indicated for post-deployed stent expansion of self-expanding peripheral vascular stents.

    Device Description

    The PolarCath™ Peripheral Dilatation System (PolarCath System) consists of a disposable catheter, a reusable Cryoplasty inflation unit, a disposable nitrous oxide cartridge, and a disposable catheter extension.

    The PolarCath System is designed for dilatation of stenotic lesions in peripheral arteries. The procedure consists of inserting a catheter over a quidewire to the target lesion, attaching the catheter to the catheter extension, attaching the catheter extension to the Inflation Unit, inserting a nitrous oxide cartridge into the Inflation Unit, and inflating the balloon catheter for a set time at a set pressure, using the controls on the Inflation Unit.

    The Cryoplasty inflation unit functions to: a) deliver the inflation media (liquid nitrous oxide) to the balloon; b) control the pressure inside the inner balloon; c) provide a vacuum between the inner and outer balloons; and d) control the length of time for delivery of the inflation media which is manually extracted at the end of the inflation time. The inflation unit monitors the temperature inside the balloon. Excess pressure is prevented by the presence of a relief valve in the inflation unit.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device called the PolarCath Peripheral Dilatation System. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study proving device performance against specific acceptance criteria in a clinical context.

    Therefore, many of the requested sections regarding clinical study design (sample size, data provenance, expert ground truth, adjudication, MRMC studies) cannot be fulfilled from this document. The document describes non-clinical testing.

    Here's an analysis of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance & The Study that Proves the Device Meets Acceptance Criteria:

    Testing TypeAcceptance Criteria (Implied)Reported Device Performance
    Performance Bench TestingPass all functional testing; meet all product specification requirements; demonstrate equivalent performance to the predicate device.The PolarCath System passed all functional testing and met all product specification requirements and demonstrated equivalent performance to the predicate device.
    Software Verification Validation TestingMeet all requirements of the Software Requirements Specification (SRS).PolarCath System Software met all requirements of the SRS.
    Electromagnetic Compatibility & Electrical SafetyMeet all acceptance criteria in accordance with IEC 60601-1:2005 and IEC 60601-1-2:2007.The PolarCath System met all acceptance criteria in accordance with: IEC 60601-1:2005 and IEC 60601-1-2:2007.

    The Study that Proves the Device Meets Acceptance Criteria:

    The document describes "non-clinical testing" as the study. This involved "Performance Bench Testing," "Software Verification Validation Testing," and "Electromagnetic Compatibility and Electrical Safety" testing. These tests were conducted to demonstrate that the materials chosen, manufacturing processes, and design meet established specifications and do not raise new questions of safety or effectiveness compared to the predicate device. No animal or clinical testing was required or performed for this submission.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified for any of the non-clinical tests. For bench testing, it refers to multiple aspects (Catheter Extension, Inflation Unit, Simulated Use, Transit Testing), implying a series of tests on various components/systems, but specific quantities are not given.
    • Data Provenance: Not applicable as these are non-clinical, bench-level tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable as this was non-clinical and did not involve expert-established ground truth in a clinical diagnostic sense. Performance was measured against engineering specifications and industry standards.

    4. Adjudication method for the test set:

    • Not applicable as this was non-clinical and did not involve expert adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is a medical device for physical intervention (dilatation system), not an AI-powered diagnostic or assistive tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable as this is a physical medical device, not an algorithm.

    7. The type of ground truth used:

    • For the non-clinical tests, the "ground truth" was established by engineering specifications, product requirements, and recognized industry standards (e.g., IEC 60601-1, IEC 60601-1-2). For the software, it was the Software Requirements Specification (SRS). For performance, it was also equivalent performance to the predicate device.

    8. The sample size for the training set:

    • Not applicable. This is not a machine learning or AI device that would have a training set.

    9. How the ground truth for the training set was established:

    • Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K071042

    Validate with FDA (Live)

    Date Cleared
    2007-06-13

    (62 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PolarCath™ Peripheral Dilatation System is indicated to dilate stenoses in the peripheral vasculature (iliac, femoral, popliteal, infrapopliteal, renal and subclavian arteries) and for the treatment of obstructive lesions of polytetrafluoroethylene (PTFE) access grafts or arteriovenous dialysis fistulae. The PolarCath™ Peripheral Dilatation System is also indicated for post-deployed stent expansion of self-expanding peripheral vascular stents.

    Device Description

    The PolarCath™ Peripheral Dilatation System consists of a Catheter, Inflation Unit, connecting cable and a rechargeable battery pack with recharging unit and battery receptacle. The inflation medium (liquid nitrous oxide) is provided in a disposable 14 gram cartridge.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a modification to the PolarCath™ Peripheral Dilatation System, specifically focusing on changes to the Inflation Unit's circuit board. This filing is for a device modification and does not involve a study to establish the diagnostic or predictive performance of an AI/ML algorithm. Therefore, many of the requested elements pertaining to such studies are not applicable.

    Here's an analysis of the provided information in the context of the requested questions:

    1. A table of acceptance criteria and the reported device performance

    Since this is a submission for a modification to an existing device, the "acceptance criteria" are related to ensuring the modified device performs equivalently to the predicate device. The performance reported here relates to the safety and functionality of the changed component.

    Acceptance Criteria (Implied)Reported Device Performance
    Maintain equivalent function of inflation units and circuitrySuccessful completion of verification testing which assessed the function of inflation units and their circuitry through 12 sequential runs, or inflations, of the balloon. The changes "have been proven not to adversely affect the function of the device."
    No adverse effect on overall design, function, material, intended use, labeling, or directions for use compared to the predicate deviceNo adverse effects reported. The changes only affect the Inflation Unit circuit board and not other aspects of the device.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The information states "12 sequential runs, or inflations, of the balloon" were performed during verification testing. This implies a sample size of 12 tests on the modified inflation units. There is no information provided about the country of origin or whether the data was retrospective or prospective. This type of testing is typically prospective, performed in a controlled lab setting, rather than on patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable. The "ground truth" for this type of engineering verification testing would be the pre-defined functional specifications of the device, assessed by engineers or technicians, not clinical experts.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable as there's no clinical "ground truth" being established by experts through adjudication for this type of engineering verification. The assessment would likely involve comparing device output/functionality against engineering specifications.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This submission is for a physical device modification, not an AI/ML algorithm that assists human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No, a standalone algorithm performance study was not done. This submission is for a physical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for this verification testing would be the functional specifications and performance parameters of the device's inflation unit and circuitry. It would be measured against engineering benchmarks, not clinical outcomes or expert consensus.

    8. The sample size for the training set

    This is not applicable. There is no "training set" as this is a physical device modification, not a machine learning model.

    9. How the ground truth for the training set was established

    This is not applicable for the same reason as above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K033622

    Validate with FDA (Live)

    Date Cleared
    2004-02-10

    (84 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PolarCath Peripheral Dilatation System is indicated to dilate stenosis in the peripheral vasculature (iliac, femoral, popliteal, infrapopliteal and renal arteries) and for the treatment of obstructive lesions of PTFE access grafts or artcriovenous dialysis fistulae.

    Device Description

    The PolarCath Peripheral Dilatation System consists of a Catheter, Inflation Unit, connecting cable and a rechargeable battery pack with recharging unit and battery receptacle. The inflation medium (liquid nitrous oxide) is provided in a disposable 14 gram cartridge.

    AI/ML Overview

    The provided text describes the 510(k) summary for the PolarCath™ Peripheral Dilatation System. However, it does not include detailed acceptance criteria or a study proving that the device meets those criteria.

    The "Performance Data" section explicitly states: "The safety and effectiveness of the modified PolarCath Peripheral Dilatation System is demonstrated with design control activities and bench testing on file at CryoVascular Systems, Inc." This indicates that the supporting data is held internally by the company and is not detailed in the provided 510(k) summary.

    Therefore, I cannot provide a table of acceptance criteria and reported device performance, nor can I answer the specific questions about sample size, data provenance, expert qualifications, adjudication methods, MRMC studies, standalone performance, or ground truth establishment for a study that is not described in this document.

    The document focuses on:

    • Device Description: What the PolarCath Peripheral Dilatation System is.
    • Indications for Use: What the device is intended for.
    • Substantial Equivalence: How it is similar to predicate devices.
    • Regulatory Information: Sponsor, contact person, product codes, classification, and the FDA's letter of clearance.

    In summary, based on the provided text, the specific study details and acceptance criteria you've requested are not available.