    Intended Use

    The PG Pro microcatheter is intended for the peripheral vasculature for the infusion of diagnostic and therapeutic agents.

    Device Description

    The PG Pro Microcatheter is a single lumen catheter designed to be introduced over a steerable guidewire to access small, tortuous vasculature. The semi-rigid proximal section transitions to a flexible distal tip to facilitate advancement through vessels. Dual radiopaque markers at the distal end facilitate fluoroscopic visualization. The outer surface of the catheter is coated with a hydrophilic polymer to increase lubricity. A luer fitting on the catheter hub is used for the attachment of accessories.

    AI/ML Overview

    The provided text is for a 510(k) premarket notification for the "PG Pro Microcatheter". This document focuses on demonstrating substantial equivalence to a predicate device, rather than proving a device meets specific performance acceptance criteria for an AI/ML type of medical device through a clinical study.

    Therefore, the text does not contain the information requested regarding:

    1. A table of acceptance criteria and reported device performance for an AI/ML device.
    2. Sample size and data provenance for a test set.
    3. Number of experts and their qualifications for ground truth establishment.
    4. Adjudication method for a test set.
    5. MRMC comparative effectiveness study results.
    6. Standalone performance of an algorithm.
    7. Type of ground truth used (expert consensus, pathology, outcomes data).
    8. Sample size for a training set.
    9. How ground truth for the training set was established.

    Instead, the document describes:

    • Device Description: The PG Pro Microcatheter is a single lumen catheter designed for introducing over a guidewire to access small, tortuous vasculature for infusing diagnostic and therapeutic agents in the peripheral vasculature.
    • Predicate Device: Headway 27 Microcatheter (K142449). A reference device (Headway 17 Microcatheter, K101542) is also listed.
    • Comparisons: The document includes a detailed "TECHNOLOGICAL CHARACTERISTICS COMPARISON TABLE" (Table 1) comparing the PG Pro Microcatheter to its predicate and reference devices across properties like Indications For Use, Principle of Operation, Material(s), Proximal ID/OD, Distal ID/OD, Number of Marker bands, Coating, Effective Length, Tip Configuration, Guidewire Compatibility, Method of Supply, and Accessories. The table highlights differences such as the PG Pro being intended only for peripheral vasculature, using a Tungsten coil vs. stainless steel, having a different range of effective lengths, being straight configuration only, being compatible with larger guidewires, and not including a shaping mandrel.
    • Performance Data (Non-Clinical):
      • Biocompatibility testing: Conducted according to ISO10993-1, including cytotoxicity, sensitization, irritation, acute systemic toxicity, and hemocompatibility, as it is considered a limited (<24 hour) circulating blood contacting device.
      • Bench testing: A comprehensive list of tests was performed, including surface contamination, physical attributes, force at break, leakage (liquid, air), static and dynamic burst, durability and lubricity of hydrophilic coating, simulated use in flow model, catheter flow rate, kink resistance, stiffness, flexural fatigue, particulate testing, catheter dead space, torque strength, hub testing, corrosion testing, and radio detectability.
      • Animal Study: An acute large animal study was conducted with n=3 animals. The study assessed the device's safety profile, compatibility with therapeutic agents, and tissue response, concluding that the PG Pro Microcatheter performed similarly to the predicate device (Headway 27 Microcatheter).
    • Conclusion: The PG Pro Microcatheter is considered substantially equivalent to the identified predicate based on performance, intended use, design, materials, principle of operation, and overall technological characteristics. The nonclinical data, verification, and validation testing support this conclusion and demonstrate the device should perform as intended.

    In summary, this document is a 510(k) submission for a physical medical device (microcatheter) and relies on benchmarking against a predicate device, along with non-clinical (bench and animal) testing to demonstrate substantial equivalence, not the evaluation of an AI algorithm against specific performance criteria.