Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K052038

    Validate with FDA (Live)

    Date Cleared
    2005-08-16

    (19 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Peripheral Cutting Balloon™ catheters are indicated for Percutaneous Transluminal Angioplasty of obstructive lesions of synthetic or native arteriovenous dialysis fistulae.

    Device Description

    The PCB: small MR and OTW features a balloon with 3 or 4 atherotomes (microsurgical blades) mounted longitudinally on its outer surface. When the PCB device is inflated, the atherotomes score the plaque, creating initiation sites for crack propagation. Percutaneous Angioplasty (PTA) with the PCB device allows dilatation of the target lesion with less pressure, minimizing barotrauma.

    Radiopaque markers are placed on the guidewire tubing at the ends of the atherotomes to provide visual reference points for balloon positioning within the vessel. The Rated Burst Pressure (RBP) is 12 atm. The devices are available in nominal balloon diameters from 2.0 mm to 4.0 mm and 1.5 cm length.

    The OTW devices are attached to a Y-connector at one end; the MR is attached to a female luer. The catheter body for the PCB: small OTW has two lumens. The outer lumen is the balloon inflation lumen. The inner lumen is used to pass the catheter over a guidewire.

    The proximal shaft for the PCB:small MR is a hypotube. This hypotube contains the balloon inflation lumen. The distal shaft is made of Pebax material and has a lumen for balloon inflation as well as a guidewire lumen. The guidewire lumen is colored green for case of guidewire insertion. The guidewire exit port is 24 cm from the catheter tip. This port facilitates rapid exchange of the catheter. The PCB:small MR and OTW are coated with Bioslide BL at the distal section.

    AI/ML Overview

    This document is a 510(k) summary for a medical device called the "Peripheral Cutting Balloon (PCB): small Monorail and Over-the-Wire Delivery systems". It is a premarket notification to the FDA to demonstrate that the device is substantially equivalent to legally marketed predicate devices.

    The document does not describe the acceptance criteria and a study that proves the device meets specific acceptance criteria in the way you've outlined. Instead, it describes a submission seeking substantial equivalence to predicate devices based on bench testing and biocompatibility testing. This is a different regulatory pathway than proving performance against specific clinical efficacy or safety endpoints using a standalone or comparative effectiveness study.

    Therefore, many of the requested fields cannot be directly extracted from this document, as the information is not present. Here's what can be inferred or stated based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    This document does not provide a table of acceptance criteria with specific performance metrics (e.g., sensitivity, specificity, accuracy, etc.) for a diagnostic or AI device. It mentions "Performance Testing" which consists of "Bench testing and biocompatibility testing" to support substantial equivalence. The specific results or defined acceptance criteria for these tests are not detailed.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    Not applicable. The submission is based on bench testing and biocompatibility testing, not a clinical test set with patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    Not applicable. Ground truth as typically understood for diagnostic performance is not relevant to bench and biocompatibility testing of a physical medical device.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. There is no mention of a test set requiring adjudication in the context of diagnostic performance.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a physical device (a catheter) and not an AI or diagnostic imaging device involved in human interpretation or a MRMC study.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the "Performance Testing" mentioned (bench and biocompatibility), the "ground truth" would be established by standard engineering and biological testing protocols and criteria. Specific details are not provided in this summary.

    8. The sample size for the training set

    Not applicable. This device does not involve a "training set" in the context of an AI or machine learning model.

    9. How the ground truth for the training set was established

    Not applicable. See point 8.

    Summary based on the document:

    This document describes a 510(k) submission for a medical device (a peripheral cutting balloon catheter) seeking to establish substantial equivalence to legally marketed predicate devices.

    • Acceptance Criteria & Device Performance: The document states that "Bench testing and biocompatibility testing support a determination of substantial equivalence." No specific quantitative acceptance criteria or performance metrics are provided in this summary. The device's performance is, implicitly, "substantially equivalent" to the predicate devices in terms of design, fundamental technology, and intended use.
    • Study Type: The "study" mentioned for demonstrating equivalence involved "Bench testing and biocompatibility testing." These are laboratory-based evaluations rather than clinical trials with patient data.
    • Purpose: The goal was to show that the new device is as safe and effective as pre-existing devices, not to establish novel efficacy or superior performance through clinical trials.
    Ask a Question

    Ask a specific question about this device

    K Number
    K051254

    Validate with FDA (Live)

    Date Cleared
    2005-06-22

    (37 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Peripheral Cutting Balloon™ catheters are indicated for Percutaneous Transluminal Angioplasty of obstructive lesions of synthetic or native arteriovenous dialysis fistulae.

    Device Description

    The Peripheral Cutting Balloon™ (PCB) has the features of a conventional angioplasty catheter, with advanced microsurgical capabilities. The PCB features a non-compliant balloon with three or four atherotomes (microsurgical blades) mounted longitudinally on its outer surface. The device is inserted over a guidewire and delivered to the target lesion. Upon inflation, the atherotomes score the plaque as the balloon expands and pushes the plaque radially. The scores serve as notches for stress concentration and allow dilatation at lower pressures, minimizing barotrauma.

    Three of the PCB devices (1 cm, 2 cm and sPCB OTW) are offerred in Over-the-Wire delivery configuration, the sPCB is also available in Monorail configuration. The OTW devices are attached to a Y-connector at one end, the sPCB MR is attached to a female luer. The OTW catheters feature a co-axial shaft consisting of an inflation lumen and a guidewire lumen; the sPCB MR features a proximal shaft using a single lumen hypotube, which is attached to a coaxial distal section consisting of an inflation and guidewire lumen. The guidewire exit port for the sPCB MR is 24 cm from the catheter tip.

    Radiopaque markers are placed on the guidewire tubing at the ends of the atherotomes to provide visual reference points for balloon positioning within the vessel. The sPCB devices are coated with Bioslide at the distal section, the 1 cm and 2 cm PCB are coated with MDX 4-4159.

    The Rated Burst Pressure (RBP) for all devices is 10 atm. The devices are available in nominal balloon diameters from 2.0 mm to 8.0 mm and blade lengths of 1 cm to 2 cm (Tables 1-2).

    AI/ML Overview

    This document is a 510(k) summary for the Boston Scientific Peripheral Cutting Balloon™ (PCB) and primarily focuses on establishing "substantial equivalence" to predicate devices rather than presenting a performance study with detailed acceptance criteria and reported device performance.

    Therefore, many of the requested sections (Table of acceptance criteria, sample sizes for test/training sets, ground truth details, expert qualifications, MRMC study, standalone performance) are not directly addressed in the provided text.

    However, I can extract the information that is present:

    Device Name: Peripheral Cutting Balloon™ (PCB)

    1. A table of acceptance criteria and the reported device performance

    The document states: "Bench testing and biocompatibility testing support a determination of substantial equivalence. The results of these tests provide reasonable assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use."

    This implies that the acceptance criteria were met by the bench and biocompatibility tests, but the specific criteria and numerical performance results are not provided in this summary.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The summary refers to "bench testing" but does not detail the sample sizes or data provenance for these tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided as the device's substantial equivalence is based on bench testing and biocompatibility rather than a clinical study requiring expert ground truth establishment.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no indication of an MRMC comparative effectiveness study as this is a medical device for angioplasty, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This information is not applicable as this is a physical medical device, not an algorithm. The "performance testing" mentioned refers to bench and biocompatibility tests of the device itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the "performance testing" described (bench and biocompatibility), the ground truth would be established by engineering specifications, material standards, and biological safety standards. The document does not elaborate on these specifics.

    8. The sample size for the training set

    This information is not applicable as there is no mention of a "training set" in the context of this traditional medical device submission.

    9. How the ground truth for the training set was established

    This information is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K040155

    Validate with FDA (Live)

    Date Cleared
    2004-07-01

    (160 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Peripheral Cutting Balloon™ catheters are indicated for Percutaneous Transluminal Angioplasty of obstructive lesions of synthetic arteriovenous dialysis fistulae.

    Device Description

    The Peripheral Cutting Balloon (PCB) is available in nominal balloon diameters of 5.0 mm to 8.0 mm (Table 1). The device features a non-compliant balloon with four Atherotomes (microsurgical blades) mounted longitudinally on its outer surface. The proximal end of the balloon is equipped with a stainless folding spring to enhance balloon refold. The spring consists of a collar, proximal to the balloon cone; and four fingers that extend across the balloon cone and terminate before the pads that hold the atherolomes to the surface of the balloon. The catheter body has two lumens. The outer lumen is the balloon inflation lumen. The inner lumen is used to pass the catheter over a guidewire. Radiopaque markers are placed on the guidewire tubing at the ends of the atherotomes to provide visual reference points for balloon positioning within the vessel.

    AI/ML Overview

    The provided text describes the 510(k) summary for the Boston Scientific Peripheral Cutting Balloon. It includes information on performance testing (bench and biocompatibility) and a clinical study (Cutting EDGE trial). However, the information provided focuses on the substantial equivalence to predicate devices rather than specific acceptance criteria thresholds the device had to meet. The study's primary purpose was to compare the Peripheral Cutting Balloon to conventional angioplasty, not to prove the device met predetermined acceptance criteria against a specific numerical benchmark.

    Therefore, I cannot populate all the requested fields as the document does not explicitly state "acceptance criteria" for the device's performance in the way a traditional validation study might. Instead, it presents comparative efficacy and safety data against a control group (conventional angioplasty) and references general guidelines (KDOQI) for expected patency rates.

    However, I can extract information related to the clinical study and its findings.

    Here's the information based on the provided text, with "N/A" for criteria not explicitly stated:

    Acceptance Criteria and Device Performance

    Acceptance Criteria (Stated or Implied)Reported Device Performance (Peripheral Cutting Balloon - PCB)
    Primary Patency (6 months)47.9% PCB vs. 40.5% PTA (p=0.37)
    For stenosed grafts51.3% PCB vs. 46.9% PTA (p=0.84)
    For thrombosed grafts43.1% PCB vs. 32.0% PTA (p=0.15)
    Procedural SuccessNot significantly different from PTA (p=0.24)
    Hemodialysis Access Circuit Patency (6 months)Not significantly different from PTA (p=0.45)
    Mean Number of Target Lesion Reinterventions (6 months)Not significantly different from PTA (p=0.22)
    For thrombosed grafts0.7 PCB vs. 1.0 PTA (marginally significant, p=0.06)
    Device-Related Adverse Events (Dissections)1.7% (3/173) PCB vs. 0% PTA
    Device-Related Adverse Events (Perforations)2.9% (5/173) PCB vs. 0% PTA
    Overall Safety and EffectivenessNot significantly different from conventional angioplasty

    Notes on Acceptance Criteria:

    • The document implies that "not significantly different from conventional angioplasty in overall safety and effectiveness" was a key criterion for demonstrating substantial equivalence.
    • For patency rates, the document references KDOQI guidelines as benchmarks for "achievable" or "typical" rates, suggesting these were considered relevant for comparison. For example, for stenosed grafts, KDOQI suggests ~50% at 6 months, and the PCB achieved 51.3%. For thrombosed grafts, KDOQI suggests ~40% at 3 months and 20% at 6 months; the PCB's 43.1% at 6 months was noted as "more than double the expected rate."
    • For complication rates, the document compares the PCB's perforation rate (2.9%) to literature-based expected rates of 2% to 4%, indicating that falling within this range was acceptable.

    Study Details

    1. Sample Size used for the test set and the data provenance:

      • Sample Size: 340 patients (173 in Peripheral Cutting Balloon arm, 167 in PTA arm).
      • Data Provenance: The study was a "prospectively randomized clinical trial," and data was collected from "27 centers in the US."
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Number of Experts: Not explicitly stated for specific ground truth establishment.
      • Qualifications of Experts: An "Adverse events were adjudicated by a CEC" (Clinical Events Committee), and a "DSMB reviewed safety information throughout the enrollment portion of the study." Specific qualifications of CEC or DSMB members are not detailed.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • "Adverse events were adjudicated by a CEC" (Clinical Events Committee). The specific method (e.g., 2+1) is not provided.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • This was a clinical trial comparing a device (Peripheral Cutting Balloon) to another medical intervention (conventional angioplasty - PTA), not an MRMC study involving human readers or AI. So, N/A.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • This information pertains to an AI/algorithm. This device is a physical medical device. So, N/A.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The study used outcomes data, specifically "primary patency through 6 months post-procedure," "access circuit primary patency," "procedural success," "number of target lesion reinterventions," and "occurrence of adverse events."
      • Adverse events were "adjudicated by a CEC."
    7. The sample size for the training set:

      • This document describes a clinical trial for substantial equivalence, not a machine learning study. Therefore, there's no "training set" in the context of an algorithm. So, N/A.
    8. How the ground truth for the training set was established:

      • As there is no training set for an algorithm, this question is N/A.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1