510(k) Data Aggregation

    K Number
    K130585

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-07-15

    (132 days)

    Product Code
    Regulation Number
    872.1800
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PCH-2500 is a digital extra oral source x-ray system intended to take panoramic and cephalometric images of the oral and maxillofacial anatomy to provide diagnostic information for adult and pediatric patients. The device should be operated and used by dentists, x-ray technicians and other professionals licensed by the law of the state in which the device is used.

    Device Description

    PCH-2500 is a dental digital radiographic imaging system which is available in two different image acquisition modes. Specifically designed for dental radiography of the teeth or jaws. PCH-2500 can be equipped with four dedicated sensors for two different X-ray modalities : one panoramic (Xmaru1501CF), two cephalometric scan type (Xmaru2301CF and Xmaru3001CF) and three flat panel one shot ceph sensors (1210SGA, 910SGA and 1417PGA). The proposed device is available with two X-ray generator options. PCH-2500 offers the digital panoramic and cephalometric X-ray modality for dental radiographs. The multi platforms of PCH-2500 imaging mode provides a wide range of imaging options based on the patient diagnostic needs.

    AI/ML Overview

    Here's a summary of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Acceptance Criteria and Device Performance

    The provided document describes a 510(k) submission for the PCH-2500 dental digital radiographic imaging system. However, it does not explicitly list specific acceptance criteria with numerical targets that the device had to meet. Instead, the submission focuses on demonstrating substantial equivalence to a predicate device (also named PCH-2500, K122155). The core argument is that the new device is an "upgraded version of the predicate device with the same model name, the same indications for use and technical characteristics" (Section 2).

    The closest equivalent to "acceptance criteria" can be inferred from the technological characteristics comparison with the predicate device and the performance tests conducted to ensure safety and effectiveness.

    Table 1. Inferred Acceptance Criteria and Reported Device Performance

    CharacteristicInferred Acceptance Criteria (from predicate device characteristics or regulatory standards)Reported Device Performance (for new PCH-2500)Notes
    Indications for UseSame as predicate (take panoramic and cephalometric images of oral/maxillofacial anatomy for adult and pediatric diagnostic information).Identical to predicate device indications (Section 2, 5)This is a primary criterion for substantial equivalence.
    Performance Spec.Panoramic and cephalometric imaging capability.Panoramic and cephalometric imaging capability (Section 2)
    Input VoltageAC 100-120 / 200-240 V (predicate range)AC 100-240 V (includes predicate range) (Section 2)The new device has a wider, "Free Input Voltage" capability, which is an improvement but still covers the predicate's range. This change was explicitly addressed (Section 5).
    Tube Voltage50-90 kV (predicate range)50-99 kV (includes predicate range) (Section 2)The new device has an extended tube voltage range, indicating a potential performance enhancement while still meeting or exceeding the predicate's capability. This change was explicitly addressed with safety and image evaluation (Section 5).
    Tube Current4-10 mA (predicate range)4-16 mA (includes predicate range) (Section 2)The new device has an extended tube current range, indicating a potential performance enhancement while still meeting or exceeding the predicate's capability. This change was explicitly addressed with safety and image evaluation (Section 5).
    Exposure TimeMax. 20.2 s (predicate)Max. 20.2 s (Section 2)Identical.
    Focal Spot Size0.5 mm (predicate)0.5 mm (Section 2)Identical.
    Total FiltrationMin. 2.8 mmAl (predicate)Min. 2.8 mmAl (Section 2)Identical.
    Pixel Resolution5 lp/mm (for Xmaru series), 3.9 lp/mm (for SGA series) - (predicate)5 lp/mm (for Xmaru series), 3.9 lp/mm (for SGA series) (Section 2)Maintained identical pixel resolutions, demonstrating that new detectors (Xmaru3001CF, 1417PGA) perform at comparable levels. Non-clinical performance (MTF, DQE, NPS) was specifically evaluated (Section 6).
    Pixel Size100 x 100 µm (Xmaru), 127 x 127 µm (SGA) - (predicate)100 x 100 µm (Xmaru), 127 x 127 µm (SGA) (Section 2)Maintained identical pixel sizes.
    SoftwareDICOM 3.0 Format compatible, EasyDent for 2D Image Viewing (predicate)DICOM 3.0 Format compatible, EasyDent (Section 2, 7)Found to be the same as the predicate (Section 5, 7).
    Safety ConformityConformity to relevant IEC/EMC (e.g., IEC 60601-1, IEC 60601-1-2) and EPRC standards.Conformed to IEC 60601-1, IEC 60601-1-1, IEC 60601-1-3, IEC 60601-2-7, IEC 60601-2-28, IEC 60601-2-32, IEC 60601-1-2. Also met 21 CFR 1020.30, 31. (Section 6)Electrical, mechanical, environmental safety, and EMC testing were explicitly performed and found satisfactory (Section 6, 7). Separate safety tests for each generator were conducted (Section 5, 6).
    Imaging PerformanceNon-clinical & clinical considerations per FDA Guidance & acceptance test per IEC 61223-3-4. Not explicitly quantified as a specific numerical benchmark but implies maintenance of image quality.Satisfactory non-clinical performance (MTF, DQE, NPS) and clinical consideration reports for new image receptors. Separate clinical image evaluation performed for each X-ray generator. Acceptance test per IEC 61223-3-4 performed. (Section 6)The non-clinical performance (MTF, DQE, NPS) and clinical consideration reports for the new image receptors (Xmaru3001CF and 1417PGA) were provided. The outcome of an expert review of image comparisons was also relied upon for substantial equivalence (Section 6). "All test results were satisfactory" (Section 7).

    Study that Proves the Device Meets the Acceptance Criteria:

    The submission describes a set of tests to demonstrate substantial equivalence to the predicate device, rather than a single, large clinical study with specific acceptance criteria. The primary "study" is a compilation of various engineering tests, performance validations, and comparisons.

    1. Sample size used for the test set and the data provenance:

      • Test Set: Not explicitly stated as a "test set" in the traditional sense of a clinical trial. The evaluation primarily involved device characteristics and performance metrics.
      • Data Provenance: The studies are laboratory and engineering tests conducted by the sponsor's qualified individuals, likely at the manufacturing facility in Gyeonggi-Do, Republic of Korea, as indicated by the submitter's address. The data is retrospective in the sense that it relies on comparisons to an already cleared predicate device and established standards.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The document mentions "an expert review of image comparisons for both devices" (Section 6).
      • The number of experts and their specific qualifications (e.g., "radiologist with 10 years of experience") are not specified in the provided text. It only states "qualified individuals employed by the sponsor" conducted laboratory and clinical performance testing (Section 5).
    3. Adjudication method for the test set:

      • The document mentions "outcome of an expert review of image comparisons" (Section 6), but the specific adjudication method (e.g., 2+1, 3+1, none) is not described.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC comparative effectiveness study was done or reported. This device is a digital X-ray imaging system, not an AI-powered diagnostic tool. The focus is on the image acquisition system itself.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Not applicable. The device is an image acquisition system, not an algorithm. Its performance is evaluated based on image quality metrics and compliance with technical specifications and safety standards.
    6. The type of ground truth used:

      • For technical performance: The "ground truth" was based on established industry standards (e.g., IEC standards for safety, ISO/NEMA standards for image quality metrics like MTF, DQE, NPS), and comparison to the performance characteristics of the legally marketed predicate device.
      • For clinical image evaluation: The "ground truth" was based on expert review of images produced by the device, likely assessing diagnostic utility and quality against established clinical imaging benchmarks, but the exact nature of this "ground truth" is not detailed (e.g., if it involved pathology confirmation or outcomes data).
    7. The sample size for the training set:

      • Not applicable. This device is an X-ray imaging system, not a machine learning algorithm that requires a "training set" in the AI sense.
    8. How the ground truth for the training set was established:

      • Not applicable. As the device is not an AI algorithm, no training set or its associated ground truth establishment mechanism is relevant to this submission.

    In essence, the "study" demonstrating the device meets acceptance criteria is a comprehensive set of engineering, electrical, mechanical, and image quality tests, along with a direct comparison of all technical specifications to a pre-existing cleared predicate device (K122155). The conclusion of substantial equivalence is drawn from the satisfactory results of these tests and the identified similarities and acceptable changes from the predicate.

    Ask a Question

    Ask a specific question about this device

    K Number
    K122155

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-09-04

    (46 days)

    Product Code
    Regulation Number
    872.1800
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PCH-2500 (PaX-i) is a digital extra oral source x-ray system intended to take panoramic and cephalometric images of the oral and maxillofacial anatomy to provide diagnostic information for adult and pediatric patients. The device should be operated and used by dentists, x-ray technicians and other professionals licensed by the law of the state in which the device is used.

    Device Description

    PCH-2500 (PaX-i) is a dental digital radiographic imaging system which is available in two different image acquisition modes. Specifically designed for dental radiography of the teeth or jaws, PCH-(PaX-i) can be equipped with four dedicated sensors for two different X-ray modalities : one panoramic (Xmaru1501CF), one cephalometric scan type (Xmaru2301CF) and two flat panel one shot ceph sensors (1210SGA and 910SGA). PCH-2500 (PaX-i) offers the digital panoramic and cephalometric X-ray modality for dental radiographs. The multi platforms of PCH-2500 (PaX-i) imaging mode provides a wide range of imaging options based on the patient diagnostic needs.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided text, focusing on the new component (910SGA sensor) as this is a Special 510(k) submission primarily addressing changes to an existing device:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission is a Special 510(k) for PCH-2500 (PaX-i), primarily introducing a new cephalometric sensor (910SGA). The acceptance criteria for this sensor are implicitly set by its equivalence to the predicate device and the performance of other sensors within the PCH-2500 (PaX-i) system, as well as general performance standards for dental X-ray imaging.

    Acceptance Criteria CategorySpecific Criteria (Implicitly based on Predicate/Standards)Reported Device Performance (PCH-2500 (PaX-i) with 910SGA)
    Technical Specifications
    Pixel Resolution≥ 3.9 lp/mm (matching 1210SGA predicate sensor)3.9 lp/mm
    Pixel Size127 x 127 μm (matching 1210SGA predicate sensor)127 x 127 μm
    Image Receptor TypeAmorphous silicon TFT with scintillator (matching 1210SGA)Amorphous silicon TFT with scintillator
    SafetyCompliance with IEC 60601-1 and related standardsAll test results were satisfactory
    EMCCompliance with IEC 60601-1-2All test results were satisfactory
    PerformanceCompliance with IEC 61223-3-4 and IEC 61223-3-5All test results were satisfactory
    Image QualityImage quality comparable to predicate device (K113672) for diagnostic information of oral and maxillofacial anatomyExpert review of image comparisons showed substantial equivalence to predicate device.
    DICOM ConformanceCompliance with NEMA PS 3.1-3.18 (DICOM Set)Meets the provisions of NEMA PS 3.1-3.18
    Non-clinical PerformanceSatisfactory performance according to FDA Guidance for Solid State X-ray Imaging DevicesPerformed and satisfactory
    Clinical ConsiderationsSatisfactory considerations according to FDA Guidance for Solid State X-ray Imaging DevicesPerformed and satisfactory

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the numerical sample size for the "test set" (i.e., the images used for the expert review). It only mentions "image comparisons for both devices." The data provenance is not explicitly stated as retrospective or prospective, nor is the country of origin specified for the images used in the expert review.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document states an "expert review of image comparisons" was conducted, but it does not specify the number of experts used or their qualifications.

    4. Adjudication Method for the Test Set

    The document mentions an "expert review," but it does not describe an adjudication method (e.g., 2+1, 3+1). It implies a consensus or comparison against a standard, but no specific method is outlined.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done

    No, the document does not indicate that a Multi Reader Multi Case (MRMC) comparative effectiveness study was done. The evaluation primarily focused on demonstrating substantial equivalence to a predicate device through technical specifications, safety/EMC testing, and an expert review of image comparisons. There is no mention of human readers improving with AI assistance, as this is an imaging device, not an AI-assisted diagnostic tool.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was Done

    This question is not directly applicable in the context of this submission. The device is an X-ray imaging system, not an algorithm. Its performance is evaluated as an imaging system, which inherently produces images for human interpretation. The "non-clinical and clinical consideration report for the new SSXI detector, 910SGA" would assess the detector's capability to produce diagnostic images, effectively acting as its standalone performance.

    7. The Type of Ground Truth Used

    The ground truth for the "image comparisons" was established through expert review. This implies a comparison of diagnostic quality and features between images produced by the new 910SGA sensor and the predicate device's sensors. The ultimate "ground truth" for dental X-ray images is typically the clinical diagnosis or assessment derived by a qualified professional from those images.

    8. The Sample Size for the Training Set

    The document does not mention a training set or its sample size. This is a hardware submission, not a machine learning algorithm submission, so a traditional "training set" as understood in AI/ML is not applicable.

    9. How the Ground Truth for the Training Set Was Established

    Since there is no training set mentioned, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K113672

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-03-27

    (105 days)

    Product Code
    Regulation Number
    872.1800
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PCH-2500 is digital extra oral source x-ray system intended to take panoramic and cephalometric images of the oral and maxillofacial anatomy to provide diagnostic information for adult and pediatric patients. The device should be operated and used by dentists, x-ray technicians and other professionals licensed by the law of the state in which the device is used.

    Device Description

    PCH-2500 is a dental digital radiographic imaging system which is available in two different image acquisition modes. Specifically designed for dental radiography of the teeth or jaws, PCH-2500 can be equipped with three dedicated sensors, each for two X-ray modalities : panoramic (Xmarul 501CF), cephalometric scan type (Xmaru2301CF) and oneshot ceph sensor type (1210SGA). PCH-2500 offers the digital panoramic and cephalometric X-ray modality for dental radiographs. The multi platforms of PCH-2500 imaging mode provides a wide range of imaging option based on the customer's diagnostic needs.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the PCH-2500 device, based only on the provided text:

    Important Note: The provided text is a 510(k) summary for a medical device. This type of document is designed to demonstrate substantial equivalence to an already legally marketed device, not necessarily to provide a comprehensive clinical study report proving efficacy against a specific set of clinical acceptance criteria. Therefore, much of the requested information (like specific performance metrics for acceptance criteria, sample sizes for test/training sets, expert qualifications, and MRMC study details) is not present in the given text.


    1. Table of Acceptance Criteria and Reported Device Performance

    The 510(k) summary for PCH-2500 primarily focuses on demonstrating substantial equivalence to the predicate device, PaX-Flex3D, by comparing technical characteristics and performance specifications. There are no explicit "acceptance criteria" for diagnostic accuracy or clinical outcomes stated. Instead, the performance is presented as meeting certain technical specifications.

    CharacteristicAcceptance Criteria (from predicate/technical standards)Reported Device Performance (PCH-2500)
    Indications for UsePCH-2500 is a digital extra-oral source x-ray system intended to take panoramic and cephalometric images of the oral and maxillofacial anatomy to provide diagnostic information for adult and pediatric patients. Operated by licensed professionals.Same as Acceptance Criteria
    Performance SpecificationPanoramic and cephalometric imaging capability (similar to predicate, excluding CT)Panoramic and cephalometric imaging capability
    Tube Voltage50-90 kV (predicate)50-90 kV
    Tube Current4-10 mA (predicate)4-10 mA
    Focal Spot Size0.5 mm (predicate)0.5 mm
    Exposure Time9-24 s (predicate)Max 20.2 s
    Total Filtration2.8 mmAl (predicate)2.8 mmAl
    Pixel Resolution (Panoramic)5 lp/mm (predicate)5 lp/mm
    Pixel Resolution (Cephalometric - scan type)5 lp/mm (predicate)5 lp/mm
    Pixel Resolution (Cephalometric - one shot type)N/A (predicate lacked this specific mode/sensor)3.9 lp/mm
    Pixel Size (Panoramic)100 x 100 µm (predicate)100 x 100 µm
    Pixel Size (Cephalometric - scan type)100 x 100 µm (predicate)100 x 100 µm
    Pixel Size (Cephalometric - one shot type)N/A (predicate lacked this specific mode/sensor)127 x 127 µm
    Safety, EMC, Performance StandardsCompliance with IEC 60601-1, IEC 60601-1-1, IEC 60601-1-3, IEC 60601-2-7, IEC 60601-2-28, IEC 60601-2-32, IEC 60601-1-2, NEMA PS 3.1-3.18, DICOM Set, IEC 61223-3-4, IEC 61223-3-5.All test results were satisfactory.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided text does not specify a sample size for a test set in the context of clinical performance evaluation (e.g., number of patients/images used to assess diagnostic accuracy). The non-clinical performance and clinical consideration report for the SSXI detector (1210SGA) is "provided separately in this submission," but details are not given here.

    The data provenance for the device performance testing mentioned (e.g., electrical, mechanical, EMC, acceptance tests according to IEC standards) would be from in-house engineering and testing as per standard regulatory requirements, likely conducted in Korea Republic (manufacturer's country). No clinical trial data is detailed. The expert review of image comparisons is mentioned, but without specifics of the dataset.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The text mentions an "expert review of image comparisons" for both the PCH-2500 and the predicate device. However, it does not specify the number of experts, their qualifications, or how they established a "ground truth" for a test set. This review seems to be part of the substantial equivalence argument, likely comparing image quality or diagnostic utility rather than rigorously establishing a ground truth for a clinical condition.

    4. Adjudication Method for the Test Set

    Since the text does not detail a clinical test set with specific ground truth establishment, no adjudication method (e.g., 2+1, 3+1) is described.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    No MRMC comparative effectiveness study is mentioned or implied. The PCH-2500 is a digital X-ray imaging system, not an AI-powered diagnostic tool, and the submission predates widespread clinical AI in radiology.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

    Not applicable. The PCH-2500 is a hardware imaging device, not a standalone algorithm.

    7. The Type of Ground Truth Used

    For the technical performance aspects, the "ground truth" is adherence to established technical standards (e.g., IEC, NEMA) and the performance of the predicate device. For potential image quality assessments by experts (implied by "expert review of image comparisons"), the ground truth would inherently be expert opinion/consensus on image quality and diagnostic features. Pathology or outcomes data are not mentioned as being used for ground truth.

    8. The Sample Size for the Training Set

    Not applicable/Not provided. As this is an X-ray imaging system, there isn't a "training set" in the machine learning sense. Any calibration or factory testing would involve proprietary methods not detailed here.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable/Not provided. See point 8.