Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Passeo-35 Xeo peripheral dilatation catheter is indicated to dilate stenosis in the iliac, femoral, popliteal and infrapopliteal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Passeo-35 Xeo is also recommended for post-dilatation of balloon expandable and self-expanding stents in the peripheral vasculature.

    The Passeo-18 peripheral dilatation catheter is indicated to dilate stenosis in the femoral, popliteal and infrapopliteal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.

    Passeo-14 is indicated for balloon dilatation of the stenotic portion of a lower limb artery for the purpose of improving perfusion.

    The Oscar Peripheral Multifunctional Catheter system is indicated for percutaneous transluminal interventions in the peripheral vasculature to provide support during access into and to dilate stenoses in femoral, popliteal and infrapopliteal arteries. The product is also intended for injection of radiopaque contrast media for the purpose of angiography.

    The Pantera LEO is indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion and for post dilatation of coronary stents.

    The Pantera Pro is indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion. The Pantera Pro (balloon diameter 2.0 – 4.0 mm) is also indicated for post-delivery expansion of balloon expandable stents.

    Device Description

    BIOTRONIK's Passeo 35 Xeo Catheter is an over-the-wire (OTW) balloon dilatation catheter, indicated for dilatation of stenotic segments in peripheral vessels. The Passeo 35 Xeo Catheter is a dual lumen design with both lumens contained within one tube. The smaller lumen is the balloon inflation/deflation lumen. The larger lumen permits the use of guide wires with a maximum diameter of 0.035" to facilitate advancement of the Passeo 35 Xeo Catheter towards and through the lesion(s) to be dilated.

    The Passeo-18 peripheral dilatation catheter is intended for dilatation of stenotic segments in peripheral vessels and arteriovenous dialysis fistulae. The dilatation balloon is designed to inflate to a known diameter at a specific inflation pressure consistent with the compliance chart on the label. One radiopaque marker is located at each end of the balloon to facilitate fluoroscopic visualization and positioning of the balloon catheter towards and across the lesion. The dilatation catheter includes a soft tapered tip to facilitate advancement of the catheter. The dilatation catheter has two Luer-ports at the proximal end. One port (inflation port) serves for connecting an inflation device to inflate/deflate the balloon. The other port enables flushing of the guide wire lumen. The dilatation catheter has a hydrophobic silicone coating on the shaft outer surface and a hydrophobic patchwork coating on the balloon.

    The Passeo-14 peripheral dilatation catheter is intended for the dilatation of stenotic segments in lower limb arteries. The dilatation balloon is designed to inflate to a known diameter at a specific inflation pressure consistent with the compliance chart on the label. One radiopaque marker is located at each end of the balloon to facilitate fluoroscopic visualization and positioning of the balloon catheter towards and across the lesion. The dilatation catheter includes a soft tapered tip to facilitate advancement of the catheter. The dilatation catheter has two Luer-ports at the proximal end. One port (inflation port) serves for connecting an inflation device to inflate/deflate the balloon. The other port enables flushing of the guide wire lumen.

    The Oscar Peripheral Multifunctional Catheter system is an intravascular balloon catheter system, supplied with a retractable sheath (Oscar Support Catheter), a flexible catheter (Oscar Dilator) and a PTA balloon (Oscar PTA balloon), allowing a variable guide wire support and injection of fluids, and adjustable length inflatable balloon up to 180 mm. The over the wire (OTW) catheter has a retractable sheath allowing the balloon to be inflated at various lengths as determined by the physician. The balloon lengths are graduated with evenly spaced radiopaque markers. The Oscar Peripheral Multifunctional Catheter system is a 4F and 6F catheter system with a shaft working length of 60 cm or 120 cm, compatible with 0.014" (Oscar 4F) and 0.018" (Oscar 6F) guide wires. The device uses a semi-compliant balloon with a size dependent rated burst pressure and an indicated clinical use range of 6 atm to 16 atm. The balloon expands to a set nominal diameter (2.0, 2.5, 3.0, 4.0, 5.0, 6.0 and 7.0 mm). If required, separate PTA balloon catheters in same size range are also available separately.

    Pantera LEO is a PTCA rapid exchange system with a balloon at the distal end of the catheter. A Luer port at the proximal end enables the attachment of an inflation device for the inflation of the balloon. The catheter provides a lumen, which enables the use of a guide wire to position the catheter. Radiopaque balloon markers aid in the placement of the catheter's balloon segment under fluoroscopy.

    Pantera Pro is a PTCA rapid exchange system with a balloon at the distal end of the catheter. A Luer port at the proximal end enables the attachment of an inflation device for the inflation of the balloon. The catheter provides a lumen, which enables the use of a guide wire to position the catheter. Radiopaque balloon markers aid in the placement of the catheter's balloon segment under fluoroscopy.

    AI/ML Overview

    The provided FDA 510(k) clearance letter (K250706) is for BIOTRONIK's peripheral and coronary dilatation catheters. This document describes several devices: Passeo-35 Xeo, Passeo-18, Passeo-14, Oscar Peripheral Multifunctional Catheter System, Pantera LEO PTCA Catheter, and Pantera Pro PTCA Catheter.

    It's important to note that this 510(k) Summary does not describe an AI/ML powered device. Instead, it focuses on the physical and performance characteristics of medical devices (catheters). Therefore, many of the requested criteria, such as "number of experts used to establish ground truth," "adjudication method," "MRMC comparative effectiveness study," "standalone algorithm performance," and details about training data, are not applicable to this type of submission.

    The "performance testing" mentioned throughout the document refers to bench testing and engineering assessments to ensure the physical device meets its design specifications and performs safely and effectively. It does not involve AI model performance evaluation with clinical data.

    Here's an attempt to answer the applicable parts of your request based on the provided text, with clear indications where the information is not present or not relevant to an AI/ML device:


    Acceptance Criteria and Study to Prove Device Meets Criteria

    The acceptance criteria for these devices are implicitly derived from the design specifications and are demonstrated through performance (bench) testing. The study proving these devices meet the acceptance criteria is a series of performance (bench) tests.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document describes technological characteristics and states that "All necessary performance testing was conducted... to ensure that the devices conform to the design specification." The "Subject Devices" column in the tables explicitly states "Identical" for most characteristics, meaning they meet the same specifications as the predicate devices. The only explicit change mentioned is that "Components will be brought into compliance with ISO 80369-7:2021" for Luer connectors and manifolds. The full set of specific performance acceptance criteria values (e.g., specific burst pressure thresholds, flexibility requirements, etc.) are not explicitly detailed in this summary but are indirectly stated as being met for "consistent performance during its intended use."

    Here's a summary derived from the comparison tables, focusing on the acceptance of "identical" characteristics:

    Acceptance Criteria CategorySpecific Criteria (from Predicate Device)Reported Device Performance (Subject Device)
    Intended UserPhysicians competent in PTA proceduresIdentical
    Method of PlacementStandard percutaneous access to site over a guide wire, with fluoroscopic visualizationIdentical
    SterilizationEO gas, SAL 10⁻⁶Identical
    Shelf Life3 yearsIdentical
    Radiopaque MarkersPresence, material, number, length, and spacing as specified for each device typeIdentical
    Usable LengthSpecified lengths (e.g., 90, 130, 170 cm for Passeo-35 Xeo)Identical
    Introducer Sheath CompatibilitySpecified F sizes and balloon diameter/length compatibilityIdentical
    Crossing ProfileMaximum inches/mm as specifiedIdentical
    Guide Wire CompatibilitySpecified guide wire diameter (e.g., 0.035" for Passeo-35 Xeo)Identical
    Shaft Outer DiameterSpecified F sizeIdentical
    Balloon DiameterSpecified range of diameters (e.g., 3.0-12.0 mm for Passeo-35 Xeo)Identical
    Balloon LengthSpecified range of lengths (e.g., 20-250 mm for Passeo-35 Xeo)Identical
    Balloon WrappingSpecified folds (e.g., 3-5 folds for Passeo-35 Xeo)Identical
    Balloon Nominal PressureSpecified atmIdentical
    Balloon RBP (Rated Burst Pressure)Specified atm, often varying by balloon diameter/lengthIdentical
    Guiding Catheter Compatibility (PTCA catheters)Minimum F size/IDIdentical
    Distal Outer Shaft/Balloon Coating (PTCA catheters)Hydrophilic/Hydrophobic coating as specifiedIdentical
    Luer Connectors and ManifoldsPreviously Luer lock connectors (L2)Will be brought into compliance with ISO 80369-7:2021 (This is the only explicitly noted difference/update)

    Overall Performance Conclusion: "The collective results of the performed testing demonstrated that the materials chosen, the manufacturing processes, and design of the components meet the established specifications necessary for consistent performance during its intended use. In addition, the collective bench testing demonstrates that the proposed device does not introduce new issues of safety or effectiveness when compared to the predicate device."

    2. Sample size used for the test set and the data provenance

    The document specifies "performance testing" or "bench testing," which refers to laboratory-based evaluations of the physical device. It does not mention a test set with patient data or any sample size related to clinical data. The provenance is not applicable as this concerns physical device characteristics, not data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable as the evaluation is not based on expert-labeled ground truth for an AI/ML algorithm but on physical measurements and engineering tests of the device.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable as the evaluation is not based on expert consensus or adjudication of clinical cases for an AI/ML algorithm.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable as the device is a physical medical instrument (catheter), not an AI/ML diagnostic or assistive tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable as the device is a physical medical instrument (catheter), not an AI/ML algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for these devices is the established engineering specifications, material properties, and performance benchmarks that the predicate devices have successfully met. These are verified through various physical and mechanical bench tests (e.g., burst pressure, flexibility, lubricity, dimensional accuracy, etc.). The document indicates these tests confirm the new devices' conformity to these specifications.

    8. The sample size for the training set

    This information is not applicable as the device is a physical medical instrument (catheter) and does not involve AI/ML training.

    9. How the ground truth for the training set was established

    This information is not applicable as the device is a physical medical instrument (catheter) and does not involve AI/ML training or a training set with ground truth.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222065

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-02-16

    (218 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Passeo-35 Xeo peripheral dilatation catheter is indicated to dilate stenosis in the iliac, femoral, popliteal and infrapopliteal arteries and for the treatment of obstructive or synthetic arteriovenous dialysis fistulae. Passeo-35 is also recommended for post-dilatation of balloon expanding stents in the peripheral vasculature.

    Device Description

    BIOTRONIK's Passeo 35 Xeo Catheter is an over-the-wire (OTW) balloon dilatation catheter, indicated for dilatation of stenotic segments in peripheral vessels. The Passeo 35 Xeo Catheter is a dual lumen design with both lumens contained within one tube. The smaller lumen is the balloon inflation/deflation lumen. The larger lumen permits the use of guide wires with a maximum diameter of 0.035″ to facilitate advancement of the Passeo 35 Xeo Catheter towards and through the lesion(s) to be dilated.

    AI/ML Overview

    This is information on a medical device, not AI/ML. The provided text is a 510(k) premarket notification for the "Passeo-35 Xeo Peripheral Dilatation Catheter." It describes the device, its indications for use, and a comparison to a predicate device. The information requested in the prompt, such as acceptance criteria, study details, sample sizes, expert qualifications, and ground truth, is typically relevant for studies evaluating the performance of AI/ML-driven medical devices, especially for diagnostic or prognostic purposes.

    Since the document provided is for a physical medical device (a catheter), these AI/ML-centric questions are not applicable. The performance data section focuses on bench testing of physical characteristics and safety, not on evaluating an algorithm's performance.

    Therefore, I cannot provide the requested information as the document does not pertain to an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K142379

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-12-05

    (101 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Passeo-35 Peripheral Dilatation Catheter is indicated to dilate stenosis in the renal, iliac, femoral, popliteal and infrapopliteal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.

    Device Description

    The Passec-35 Peripheral Dilatation Catheter (Passeo-35) is intended for dilatation of stenotic segments in peripheral vessels. One radiopaque marker is located at either end of the balloon to facilitate fluoroscopic visualization and positioning of the balloon catheter towards and across the lesion. The dilatation balloon is designed to inflate to a known diameter at a specific inflation pressure consistent with the compliance chart on the label. The balloon catheter includes a tapered soft tip to facilitate advancement of the catheter.

    The balloon catheter shaft has two Luer ports at the proximal end. One port) serves for connecting an inflation device to inflate/deflate the balloon. The other port) enables insertion of the guide wire. The balloon catheter is a dual lumens contained within one tube. The smaller lumen is the balloon inflation/deflation lumen permits the use of guide wires with a maximum diameter of 0.035" to facilitate advancement of the Passeo-35 catheter towards and through the lesion(s) to be dilated. The balloon catheter is compatible with introducer) sizes according to the recommendations on the label. The balloon catheter has a silicone coating (hydrophobic) to improve the trackability characteristics.

    AI/ML Overview

    This document describes the Biotronik Passeo-35 Peripheral Dilatation Catheter (K142379) and its substantial equivalence to a predicate device (K082933). The submission primarily focuses on non-clinical performance testing for additional device size configurations and material changes.

    Here's an analysis of the provided information concerning acceptance criteria and supporting studies:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document provides a comprehensive list of non-clinical tests performed on the Passeo-35 Peripheral Dilatation Catheter. The "Test Conditions / Specifications" column often implicitly or explicitly states the acceptance criteria, while the "Summary Results" column indicates whether these criteria were met.

    Due to the length, a complete table is not feasible here. Instead, a representative sample of acceptance criteria and results is provided.

    Test NameAcceptance Criteria (or implicit in description)Reported Device Performance
    Biocompatibility Testing
    CytotoxicityNo cytotoxic effects of the test article on L929 cells."Growth analyses of cells cultured with test article extract showed no cytotoxic effects of the test article."
    SensitizationNo reactions identified as sensitization."No reactions were identified as sensitization."
    Irritation / Intracutaneous reactivityNo signs of irritation for polar and non-polar extracts."There were no signs of irritation of the polar and non-polar test article extracts injected intracutaneously into rabbits."
    Acute systemic toxicityNo compound-related mortalities, no significant body weight loss, clinically normal animals, no gross pathology."There were no compound related mortalities, no significant body weight loss was recorded and all animals appeared clinically normal. At necropsy no evidence of gross pathology of organs was found. The test article was considered to have no acute toxic characteristics."
    PyrogenicityNo single animal showed a temperature increase higher or equal to 0.5°C above its initial temperature."No single animal showed a temperature increase higher or equal to 0.5°C above its initial temperature. The test article is considered to be non-pyrogenic."
    HemocompatibilityNo significant changes in cell counts, TAT complex generation identical to reference, no evidence of hemolysis."There were no significant changes in the cell counts and the coagulation system activation value measured as TAT complex generation of the test article was identical in comparison with the reference. There was no evidence of hemolysis."
    GC-MS / FT-IRNo differences in detected substances / >99% correlation to predicate materials."GC/MS fingerprint analysis... yielded no differences in the type of detected substances... Materials had greater than 99% correlation according to FT-IR analysis."
    Non-Clinical Performance Testing
    Visual and Dimensional InspectionAll inspectional acceptance criteria (e.g., no defects, correct printing, correct marker positioning, specified dimensions) met."Inspectional acceptance criteria were met."
    Crossing ProfileAcceptance criteria for crossing profile met; within specs of predicate."Acceptance criteria for crossing profile were met. Crossing profile is within specs of predicate."
    Simulated Use / Balloon Prep, Deployment, RetractionAcceptance criteria met; performs similar to predicate in simulated use."Acceptance criteria for balloon prep, deployment and retraction were met. Test shows device performs similar to predicate in a simulated use environment."
    Deflated Balloon ProfileWithdrawal force less than minimal tensile; deflated balloon diameter less than max crossing profile."Acceptance criteria were met: withdrawal force is less than minimal tensile and deflated balloon diameter is less than max crossing profile."
    TrackabilityAcceptance criteria for comparison to the predicate met."Acceptance criteria for comparison to the predicate were met."
    PushabilityComparable or better (greater) than the existing Passeo-35 range."Acceptance criteria for comparison to predicate devices were met."
    Torqueability / Torque StrengthQualitative assessment showed rotation transfers to distal tip; all balloons passed inflation test."Qualitative assessment showed rotation transfers to distal tip on average between 10 and 16 rotations and all balloons passed inflation test. Device meets torqueability and torque strength acceptance criteria."
    Pullback and Reintroduction testPullback and reintroduction comparable or better than comparator product; product specifications met for 1st and 2nd pullback force; safe retrieval after burst."Pullback and reintroduction was comparable or better than comparator product. Product specifications were met for 1st and 2nd pullback force values. Balloon can be safely retrieved after burst."
    Balloon Inflation / Deflation TimeDeflation time according to specifications within instructions for use."Inflation time was characterized and deflation time was determined to be according to specifications within the instructions for use."
    Flexibility and kink testNo kink, function not compromised at 20mm radius, guidewire movable, successful inflation/deflation."No kink occurred nor was the function of the device compromised at the clinically relevant radius of 20 mm. The guide wire was movable during testing and it was possible to inflate and to deflate the balloon to and from NP."
    Particulate EvaluationNumber of particulates released either the same as (p-value >0.05) or less than (p-value <0.05) comparator device."The number of particulates released by the Passeo-35 LE were either the same as (p-value >0.05) or were less than the comparator device (p-value <0.05)."
    X-ray VisibilityMarkers visible with contrast values approximately equivalent to a 10mm thick aluminum plate (based on predicate)."The markers are unchanged from the predicate and the predicate (K082933) results show that markers are visible with contrast values approximately equivalent to a 10mm thick aluminum plate."
    Corrosion ResistanceNo signs of corrosion visible on tested samples."No signs of corrosion were visible on the samples tested."
    Compatibility with Contrast mediaNo visible damage or deformation after 1-hour in contrast solution; balloon inflation to RBP possible."No visible damage or deformation after 1-hour in contrast solution. Balloon inflation to RBP was possible after contrast exposure."
    Balloon Compliance RadialAll balloon diameters within tolerance and radial compliance meets acceptance criteria."All balloon diameters are within tolerance and radial compliance meets acceptance criteria."
    Balloon Compliance AxialDifference between balloon length at RBP and NP within specification."The difference between balloon length at RBP and NP is within specification."
    Balloon RBPBalloons met acceptance criteria for lower 99.9% quantile at 95% confidence interval for all sizes."Balloons met acceptance criteria for lower 99.9% quantile at 95% confidence interval for all sizes."
    Balloon Rated Fatigue Testing90% of balloons survive the test with at least 95% confidence."Results demonstrate that 90% of the balloons will survive the test with at least 95% confidence."
    Tensile Strength of the Entire CatheterTensile strength performance (Fmax) for distal and proximal balloon sections met performance specifications."Tensile strength performance (Fmax) for distal and proximal balloon sections met performance specifications."
    Tip pull testAll tested samples met acceptance criteria for mean tensile strength."All tested samples met acceptance criteria for mean tensile strength."
    Resistance to KinkMeasured mean catheter kink radius well within acceptance criteria."Measured mean catheter kink radius was well within the acceptance criteria."
    Adhesive Strength of Catheter hydrophobic CoatingNo delamination or residual particles observed on tape or catheter; all samples passed."No delamination or residual particles were observed on the tape or the catheter. All samples tested passed."
    Connector TestConnector meets acceptance criteria and complies with ISO 594-2."Connector meets acceptance criteria and complies with ISO 594-2."
    Indelibility and adhesive strength of printingAll samples showed readable printing following exposure to solvent."All samples tested showed readable printing following exposure to solvent."
    Post-Dilatation TestAll samples withstood 10 inflation/deflation cycles within the stent without bursting."All samples withstood 10 inflation/deflation cycles within the stent without bursting."
    Visual Inspection of packaging after exposure to adverse environmental conditionsAll devices passed inspection of shipping container, box integrity, pouch integrity, labeling integrity, product fixation and device integrity."All devices passed inspection of shipping container, box integrity, pouch integrity, labeling integrity, product fixation and device integrity."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes (number of devices or tests) for each specific non-clinical test. It generally refers to "test article," "samples," or "all sizes." For example, for "Balloon RBP," it mentions "all sizes," implying multiple devices were tested across the range of diameters and lengths.

    • Test Set: No specific number for a single "test set" is provided across all tests. Each test likely utilized a relevant number of samples appropriate for the specific test's statistical requirements, though these numbers are not detailed.
    • Data Provenance: The studies are described as "non-clinical performance testing," indicating they were conducted in a laboratory or simulated environment (e.g., "anatomical model," "Crossover model," "polymer tube"). There is no mention of country of origin, but given it's a submission to the FDA by BIOTRONIK, Inc. (Lake Oswego, Oregon), it's reasonable to infer the testing was conducted either internally by the manufacturer or by a contracted lab, adhering to international standards (e.g., ISO, ASTM, DIN EN ISO, USP). The studies are prospective in the sense that they were conducted specifically for this submission to demonstrate the performance of the modified device.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    This information is not applicable to the provided document. The studies are non-clinical, mechanical, and biocompatibility tests. They do not involve expert interpretation or human assessment of image data or clinical outcomes that would require a "ground truth" established by experts in the sense of a medical device intended for diagnostic or prognostic purposes.

    4. Adjudication Method for the Test Set

    This information is not applicable as there is no expert human reading involved in these non-clinical tests. The results are typically quantitative measurements or qualitative observations by trained technicians or automated systems according to defined protocols.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was done. The document explicitly states: "The determination of substantial equivalency on this subject device does not rely upon the clinical data. There is no clinical data submitted in this application." Therefore, there is no effect size reported for human readers with or without AI assistance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop) Performance

    This information is not applicable. The device is a physical medical device (catheter), not a software algorithm or AI that would have "standalone performance."

    7. Type of Ground Truth Used

    The concept of "ground truth" as typically used for AI/software in medicine (e.g., pathology, clinical outcomes) is not directly applicable here. For these non-clinical tests, "ground truth" would correspond to:

    • Engineering specifications and standards: e.g., ISO, ASTM, DIN EN ISO, USP standards that define acceptable performance ranges and test methodologies.
    • Predicate device performance: The performance of the previously cleared Passeo-35 (K082933) serves as a benchmark or "ground truth" for demonstrating substantial equivalence for many parameters.
    • Material properties: Established properties of materials (e.g., for biocompatibility tests).
    • Defined physical measurements: e.g., specified balloon diameters, lengths, pressures.

    8. Sample Size for the Training Set

    This information is not applicable. There is no "training set" as this is a physical device undergoing non-clinical validation studies, not an AI or machine learning model.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable for the reasons stated above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K130161

    Validate with FDA (Live)

    Device Name
    PASSEO-35
    Manufacturer
    Date Cleared
    2013-04-23

    (90 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Passeo-35 peripheral dilatation catheter is indicated to dilate stenosis in the renal, iliac, femoral, popliteal and infrapopliteal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.

    Device Description

    The Passeo-35 peripheral transluminal angioplasty (PTA) balloon catheter is indicated for dilatation of stenotic segments in peripheral vessels. One radiopague marker is located at either end of the balloon to facilitate fluoroscopic visualization and positioning of the balloon catheter towards and across the lesion. The dilatation balloon is designed to inflate to a known diameter at a specific inflation pressure consistent with the compliance chart on the label. The balloon catheter includes a tapered soft tip to facilitate advancement of the catheter. The balloon catheter shaft has two Luer ports at the proximal end. One port (inflation port) serves for connecting an inflation device to inflate/deflate the balloon. The other port (guidewire port) enables insertion of the guide wire. The balloon catheter is a dual lumen design with both lumens contained within one tube. The smaller lumen is the balloon inflation lumen. The larger lumen permits the use of guide wires with a maximum diameter of 0.035" to facilitate advancement of the Passeo-35 catheter towards and through the lesion(s) to be dilated. The balloon catheter is compatible with introducer sheath (introducer) sizes according to the recommendations on the label. The balloon catheter has a silicone coating to improve the trackability and pushability characteristics.

    AI/ML Overview

    This 510(k) K130161 is for a Special 510(k) notification, indicating that the device (Passeo-35 PTA Catheter) is a modification of a previously cleared device (Passeo-35 PTA Catheter, K082933).

    Crucially, the document states:

    • "The determination of substantial equivalency on this subject device does not rely upon the nonclinical test data. There is no nonclinical test data submitted in this application."
    • "The determination of substantial equivalency on this subject device does not rely upon the clinical data. There is no clinical data submitted in this application."

    This means that no new studies (clinical or non-clinical) were conducted or submitted to demonstrate that the modified device meets any acceptance criteria. The substantial equivalence is based solely on the fact that the changes are minor (only labeling changes) and the device's technological characteristics, materials, and manufacturing remain unchanged from the previously cleared predicate device.

    Therefore, the requested information about acceptance criteria and studies cannot be provided from this document, as none were performed for this specific 510(k) submission. Any original studies or acceptance criteria would have been part of the original K082933 submission.

    For completeness, if this were a submission requiring new clinical or non-clinical data, the table and details would be structured as follows:


    1. Table of Acceptance Criteria and Reported Device Performance

    (This information is not available in the provided document as no new studies were conducted for this Special 510(k).)

    Acceptance Criteria (e.g., specific endpoints for safety/effectiveness)Reported Device Performance
    No new criteria or performance data for this submission.N/A

    2. Sample size used for the test set and the data provenance

    (This information is not available in the provided document as no new studies were conducted for this Special 510(k).)

    • Sample Size (Test Set): N/A
    • Data Provenance: N/A (e.g., country of origin, retrospective/prospective)

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    (This information is not available in the provided document as no new studies were conducted for this Special 510(k).)

    • Number of Experts: N/A
    • Qualifications: N/A

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    (This information is not available in the provided document as no new studies were conducted for this Special 510(k).)

    • Adjudication Method: N/A

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    (This information is not available in the provided document as no new studies were conducted for this Special 510(k). This is also not an AI device.)

    • MRMC Study Conducted: No
    • Effect Size: N/A

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    (This information is not available in the provided document as no new studies were conducted for this Special 510(k). This is not an AI device.)

    • Standalone Study Conducted: No

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    (This information is not available in the provided document as no new studies were conducted for this Special 510(k).)

    • Type of Ground Truth: N/A

    8. The sample size for the training set

    (This information is not available in the provided document as no new studies were conducted for this Special 510(k). This is not an AI device.)

    • Training Set Sample Size: N/A

    9. How the ground truth for the training set was established

    (This information is not available in the provided document as no new studies were conducted for this Special 510(k). This is not an AI device.)

    • Ground Truth Establishment (Training Set): N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K082933

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2008-11-03

    (33 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Passeo-35 peripheral dilatation catheter is indicated to dilate stenosis in the renal, iliac, femoral, popliteal and infrapopliteal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.

    Device Description

    The Passeo-35 peripheral transluminal angioplasty (PTA) balloon catheter is indicated for dilatation of stenotic segments in peripheral vessels. One radiopaque marker is located at either end of the balloon to facilitate fluoroscopic visualization and positioning of the balloon catheter towards and across the lesion. The dilatation balloon is designed to inflate to a known diameter at a specific inflation pressure consistent with the compliance chart on the label. The balloon catheter includes a tapered soft tip to facilitate advancement of the catheter. The balloon catheter shaft has two Luer ports at the proximal end. One port (inflation port) serves for connecting an inflation device to inflate/deflate the balloon. The other port) enables insertion of the guide wire. The balloon catheter is a dual lumen design with both lumens contained within one tube. The smaller lumen is the balloon inflation/deflation lumen. The larger lumen permits the use of guide wires with a maximum diameter of 0.035" to facilitate advancement of the Passeo-35 catheter towards and through the lesion(s) to be dilated. The balloon catheter is compatible with introducer sheath (introducer) sizes according to the recommendations on the label. The balloon catheter has a silicone coating to improve the trackability and pushability characteristics.

    AI/ML Overview

    The provided text is a 510(k) premarket notification for the BIOTRONIK Passeo-35 PTA Catheter. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than providing extensive clinical study data with acceptance criteria and detailed performance metrics as would be found in a Premarket Approval (PMA) application or a more comprehensive clinical trial report.

    Therefore, the input does not contain the specific information required to complete the detailed table and answer all questions related to acceptance criteria, study design, and performance. The document primarily describes the device, its intended use, and identifies predicate devices.

    Here's what can be extracted and what is missing:

    1. A table of acceptance criteria and the reported device performance:

    This information is not available in the provided document. 510(k) submissions typically rely on design verification and validation testing (bench and/or animal studies) to demonstrate that a new device performs as intended and is as safe and effective as a predicate device. Specific acceptance criteria and quantitative performance results for a clinical study with human subjects, in the format requested, are not included.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    This information is not available in the provided document. There is no mention of a clinical "test set" with human subjects for the Passeo-35, nor details on data provenance. The submission focuses on device characteristics and substantial equivalence to predicates.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    This information is not available in the provided document. "Ground truth" establishment by experts is a concept typically associated with studies involving diagnostic accuracy or image interpretation, which is not the focus of this device (a PTA catheter).

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    This information is not available in the provided document. Adjudication methods are relevant for studies where expert consensus is needed for outcome assessment, which, again, is not detailed here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not available in the provided document. An MRMC study is not relevant for a PTA catheter, nor is AI assistance mentioned.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    This information is not available in the provided document. A "standalone algorithm" is not applicable to an interventional medical device like a PTA catheter.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    This information is not available in the provided document. As a physical device for dilatation, the concept of "ground truth" for the device's performance often relates to physical and mechanical properties (e.g., balloon diameter at pressure, burst strength, trackability), verified through engineering and bench testing, rather than clinical "ground truth" derived from expert consensus, pathology, or outcomes data in a clinical trial.

    8. The sample size for the training set:

    This information is not available in the provided document. The concept of a "training set" is typically associated with machine learning or AI models, which are not described here.

    9. How the ground truth for the training set was established:

    This information is not available in the provided document for the reasons stated above.

    Summary based on available information:

    The document describes the BIOTRONIK Passeo-35 PTA Catheter as a device for dilating stenotic segments in peripheral vessels (renal, iliac, femoral, popliteal, infrapopliteal arteries) and for treating obstructive lesions in arteriovenous dialysis fistulae.

    The device's safety and effectiveness are established through demonstration of substantial equivalence to predicate devices:

    The submission implies that the device meets its intended performance through its design, materials, and manufacturing, consistent with the cleared predicate devices. Specific clinical acceptance criteria and results from human studies are not detailed in this 510(k) summary. The FDA's clearance (K082933) indicates that they reviewed the submission and determined the device is substantially equivalent to legally marketed predicate devices.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1