    Intended Use

    The OmniTom Elite with PCD computed tomography (CT) system is intended to be used for x-ray computed tomography applications for anatomy that can be imaged in the 40 cm aperture, primarily head and neck.

    The CT system is intended to be used for both pediatric and adult imaging and as such has preset dose settings based upon weight and age. The CT images can be obtained either with or without contrast. The PCD CT system has multi-energy CT functionality with spectral capability for material decomposition and virtual monoenergetic images (VMI).

    Device Description

    The subject OmniTom Elite with PCD Computed Tomography (CT) system provides the same functionality as the previous version of the device, OmniTom Elite (K233767). Both computed tomography systems are identical in terms of the high resolution, 16 row, 40 cm bore, and 30 cm field of view. The lightweight translating gantry consists of a rotating disk with a solid-state x-ray generator, collimator, control computer, communications link, power slip-ring, data acquisition system, reconstruction computer, power system, brushless DC servo drive system (disk rotation) and an internal drive system (translation). The power system consists of batteries which provide system power while unplugged from the charging outlet. The system has the necessary safety features such as the emergency stop switch, x-ray indicators, interlocks, patient alignment laser and 110% x-ray timer. The gantry has omni-directional wheels that allow for robust diagonal, lateral, and rotational 360-degree movement and electrical drive system so the system can be moved easily to different locations.

    OmniTom Elite system with photon counting detector (PCD) provides the ability to capture CT data in multiple energy bands that can provide information on material composition of different tissues and contrast media. The multiple sets of CT data are acquired at the same time with configurable energy thresholds without any cross talk between images.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K233767

    Validate with FDA (Live)

    Device Name
    OmniTom Elite
    Date Cleared
    2024-06-10

    (199 days)

    Product Code
    Regulation Number
    892.1750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OmniTom Elite computed tomography (CT) system is intended to be used for X-ray computed tomography applications for anatomy that can be imaged in the 40cm aperture, primarily the head and neck.

    The CT system is intended to be used for both pediatric and adult imaging and as such has preset dose settings based upon weight and age The CT images can be obtained either with or without contrast.

    OmniTom Elite with photon counting detectors (PCD) configuration has multi-energy CT functionality with spectral capability for material decomposition and virtual monoenergetic images (VMI). OmniTom Elite with PCD is supported for adult imaging of the head and neck.

    Device Description

    The subject OmniTom Elite Computed Tomography (CT) system provides the same functionality as the previous version of the device, OmniTom Elite (K223447). Both computed tomography systems are identical in terms of the high resolution, 16 row, 40 cm bore, and 30 cm field of view. The lightweight translating gantry consists of a rotating disk with a solid-state x-ray generator, collimator, control computer, communications link, power slip-ring, data acquisition system, reconstruction computer, power system, brushless DC servo drive system (disk rotation) and an internal drive system (translation). The power system consists of batteries which provide system power while unplugged from the charging outlet. The system has the necessary safety features such as the emergency stop switch, x-ray indicators, interlocks, patient alignment laser and 110% x-ray timer. The gantry has omni-directional wheels that allow for robust diagonal, lateral, and rotational 360-degree movement and electrical drive system so the system can be moved easily to different locations.

    The ability to upgrade the OmniTom Elite system with photon counting detector (PCD) remains the same. PCD provides the ability to capture CT data in multiple energy bands that can provide information on material composition of different tissues and contrast media. The multiple sets of CT data are acquired at the same time with configurable energy thresholds without any cross talk between images.

    AI/ML Overview

    The provided text describes the OmniTom Elite Computed Tomography (CT) system and its updates, focusing on the Photon Counting Detector (PCD) configuration. However, it does not contain specific numerical acceptance criteria for performance metrics or detailed results from a study that quantifies these criteria. The document states that the device successfully meets "all the image quality criteria that are used for testing" and that images were "confirmed diagnostic quality," but it doesn't provide the actual criteria or the numerical outcomes.

    Therefore, many parts of the request cannot be fully answered with the given information.

    Here's an attempt to answer based on the provided text, highlighting where information is missing:

    Acceptance Criteria and Study Details for OmniTom Elite with PCD

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state numerical acceptance criteria for image quality metrics. It broadly claims that the device "meets all the image quality criteria that are used for testing" and that images were of "diagnostic quality."

    Metric / FeatureAcceptance Criteria (Not Explicitly Stated as Numerical Value)Reported Device Performance (Not Explicitly Stated as Numerical Value)
    NoiseSatisfactory for diagnostic qualitySuccessfully demonstrated comparable image quality to predicate
    Slice ThicknessSatisfactory for diagnostic qualitySuccessfully demonstrated comparable image quality to predicate
    Low Contrast ResolutionSatisfactory for diagnostic qualitySuccessfully demonstrated comparable image quality to predicate
    High Contrast ResolutionSatisfactory for diagnostic qualitySuccessfully demonstrated comparable image quality to predicate
    Radiation MetricsConformance with equipment performance standards & regulationsSuccessfully demonstrated comparable image quality to predicate
    Modulation Transfer FunctionSatisfactory for diagnostic qualitySuccessfully demonstrated comparable image quality to predicate
    Diagnostic Quality of Images (Clinical)Images confirmed to be of diagnostic quality by radiologistsConfirmed diagnostic quality of images

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: "Volunteers underwent CT scanning for a full head scan, the top of the skull, the midbrain, and through the sinuses." The number of volunteers is not specified.
    • Data Provenance: The study involved prospective scanning of volunteers. The country of origin is not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Number of Experts: "independent board-certified practicing radiologists." The number of radiologists is not specified.
    • Qualifications of Experts: "board-certified practicing radiologists."

    4. Adjudication method for the test set

    The document does not specify an adjudication method (e.g., 2+1, 3+1). It only states that images were "reviewed by independent board-certified practicing radiologists."

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study Done: No. The study described is a clinical evaluation of the device's image quality, not an MRMC study comparing human reader performance with and without AI assistance. The device itself is a CT system, and while it introduces new image processing features (like MD Plus for bone removal, VNC, ultra-high resolution), it doesn't appear to be an AI-assisted diagnostic tool for humans, but rather an imaging system that produces images that radiologists interpret.
    • Effect Size: Not applicable as an MRMC comparative effectiveness study with AI assistance was not performed or described.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, elements of standalone performance were evaluated.

    • Bench Testing: Software development and testing in accordance with IEC 62304 and FDA guidance for software functions, including risk management analysis and verification/validation testing, indicate standalone algorithm assessment for software functions.
    • Image Quality Evaluation: "Image quality metrics such as noise, slice thickness, low and high contrast resolution, radiation metrics, and modulation transfer function were measured utilizing phantom image quality tests." This constitutes standalone testing of the imaging system's output characteristics without human interpretation.

    7. The type of ground truth used

    • For Image Quality Evaluation: Physical phantom measurements were used to assess image quality metrics.
    • For Clinical Evaluation: The "ground truth" for diagnostic quality was established by the consensus or interpretation of "independent board-certified practicing radiologists." This is expert consensus on the diagnostic acceptability of the images. No mention of pathology or outcomes data as ground truth is made.

    8. The sample size for the training set

    The document does not provide information regarding a training set size or methodology, as the study described is an evaluation of the CT system's performance, not an AI algorithm that requires a distinct training phase. The MD Plus algorithm is mentioned, but its training details are not.

    9. How the ground truth for the training set was established

    Not applicable/no information provided, as details about a training set for an AI algorithm are not given.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223447

    Validate with FDA (Live)

    Device Name
    OmniTom Elite
    Date Cleared
    2023-06-01

    (198 days)

    Product Code
    Regulation Number
    892.1750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OmniTom Elite CT system is intended to be used for x-ray computed tomography applications for anatomy that can be imaged in the 40 cm aperture, primarily head and neck. The CT system is intended to be used for both pediatric and adult imaging and as such has preset dose settings based upon weight and age. The CT images can be obtained either with or without contrast.

    OmniTom Elite with photon counting detectors (PCD) configuration has multi-energy CT functionality with spectral capability for material decomposition and virtual monoenergetic images (VMI). OmniTom Elite with PCD is only supported for adult axial imaging for head and neck.

    Device Description

    The subject OmniTom Elite Computed Tomography (CT) system provides the same functionality as the previous version of the device, OmniTom Elite (K211711). Both computed tomography systems are identical in terms of the high resolution, 16 row, 40 cm bore, and 30 cm field of view. The lightweight translating gantry consists of a rotating disk with a solid-state x-ray generator, collimator, control computer, communications link, power slip-ring, data acquisition system, reconstruction computer, power system, brushless DC servo drive system (disk rotation) and an internal drive system (translation). The power system consists of batteries which provide system power while unplugged from the charging outlet. The system has the necessary safety features such as the emergency stop switch, x-ray indicators, interlocks, patient alignment laser and 110% x-ray timer. The gantry has omni-directional wheels that allow for robust diagonal, lateral, and rotational 360-degree movement and electrical drive system so the system can be moved easily to different locations.

    The ability to upgrade the OmniTom Elite system with photon counting detector (PCD) remains the same. PCD provides the ability to capture CT data in multiple energy bands that can provide information on material composition of different tissues and contrast media. The multiple sets of CT data are acquired at the same time with configurable energy thresholds without any cross talk between images.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details based on the provided text, formatted to answer your specific questions:


    Acceptance Criteria and Device Performance Study

    The NeuroLogica OmniTom Elite CT system, particularly with its Photon Counting Detector (PCD) configuration, underwent various tests to demonstrate its safety, effectiveness, and substantial equivalence to its predicate device. The primary focus of the new features described is the multi-energy CT functionality with spectral capability for material decomposition and virtual monoenergetic images (VMI).

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not provide a specific, quantified table of "acceptance criteria" for the new multi-energy CT functionality in terms of specific performance metrics tied to a pass/fail threshold. Instead, it describes various tests and their successful outcomes, demonstrating that the device performs as intended and is comparable to the predicate. The overall acceptance criterion is inferred to be that the new features do not raise new questions of safety or effectiveness and provide diagnostically acceptable images, similar to the predicate device.

    CategoryAcceptance Criteria (Inferred from text)Reported Device Performance
    Material DecompositionAbility to accurately decompose basic materials (water, iodine, calcium) and represent them in maps."For material decomposition, the multi-energy CT phantom was scanned, where three basic materials, water, iodine and calcium were decomposed. The materials were well represented in the respective decomposition maps, for example; iodine of varying densities highlighted on the iodine map."
    Virtual Monoenergetic Images (VMI)Ability to generate VMI. (Implicitly accepted if material decomposition works and images are diagnostic)VMI capability is listed as a feature and implied by the success of material decomposition and diagnostic image quality.
    Image Quality (PCD)Comparable image quality to the predicate device, meeting existing criteria for noise, slice thickness, low/high contrast resolution, radiation metrics, and MTF."Imaging metrics successfully demonstrated that the proposed device with PCD has comparable image quality with its previous version, predicate device (K211711) and meets all the image quality criteria that are used for testing." "PCD capabilities were tested to demonstrate properties of PCD in count mode and multi-energy mode... Imaging metrics demonstrated stability of the detector, analysis of the count rate and limited pileup loss at low dose scanning."
    Diagnostic AcceptabilityClinical scans produced by the PCD configuration must be diagnostically acceptable and equivalent to images from the predicate device (EID)."The clinical scans indicate the ability of the PCD configuration of the OmniTom Elite device to generate diagnostically acceptable images, which are equivalent to those produced by the OmniTom Elite with EID." "Multi-energy studies demonstrate the PCD's ability not only to produce diagnostically acceptable images but also to ability to decompose images using different basis materials such as iodine and calcium."
    Safety and ComplianceAdherence to recognized safety standards (e.g., IEC 60601 series, ISO 14971) and FDA regulations.Device is "in conformance with all applicable parts of the following FDA recognized consensus standards" and has passed "Electrical Safety and Electromagnetic Compatibility testing and proved to be in compliance with IEC 60601-1-2, and IEC 60601-2-44." Risks were controlled and mitigated.
    Software FunctionalityAll software requirements are met, and the new features do not negative affect existing functionality."Software verification and software validation testing was executed to confirm all software requirements were met. The proposed OmniTom Elite device was shown to meet all requirements and to not have any impact on imaging."
    Overall PerformanceDevice operates in accordance with specifications, meets user needs, and performs as well as similar CT devices."The results of the performance testing and conformance to the harmonized standards demonstrate that the subject device operates in accordance with specifications and meets user needs and intended use. The proposed OmniTom Elite CT system performs as well in its intended use as similar CT devices currently on the market."

    2. Sample Size for the Test Set and Data Provenance

    The document does not specify a numerical sample size (e.g., number of patients/scans) for the clinical evaluation (test set). It states, "Volunteers underwent CT scanning for a full head scan, from the upper jaw to the top of the skull to aid in generation of comparative images." The provenance is described as "Comparative study was conducted under an IRB" using "Volunteers." It implies a prospective study, but the geographic origin (country) is not specified.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    The clinical images "were reviewed by three independent board-certified practicing radiologists." No specific number of years of experience is provided, but 'board-certified' is a standard qualification.

    4. Adjudication Method for the Test Set

    The adjudication method is not explicitly stated beyond "reviewed by three independent board-certified practicing radiologists." It's unclear if consensus was required, if a majority rule was applied for disagreement, or if all three had to independently confirm diagnostic quality without formal adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    A multi-reader study was conducted to compare OmniTom Elite with PCD and OmniTom Elite with EID (existing detector technology). However, this was not a comparative effectiveness study designed to quantify how much human readers improve with AI vs. without AI assistance. The study's purpose was to confirm the diagnostic quality and equivalence of the images from the new PCD system compared to the existing EID system, verifying that the new technology does not compromise diagnostic utility. Therefore, an effect size of human improvement with AI assistance is not provided, as that was not the study's objective.

    6. Standalone (Algorithm Only) Performance

    The document does not describe a standalone (algorithm only without human-in-the-loop performance) study for the diagnostic interpretation of images. The performance data is primarily focused on the system's ability to generate diagnostically acceptable images, perform material decomposition, and adhere to image quality metrics. The "clinical evaluation" always refers to images being reviewed by radiologists. The material decomposition phantom study could be considered a form of standalone performance for that specific function.

    7. Type of Ground Truth Used for the Test Set

    The ground truth for the clinical evaluation test set was established by expert consensus/review of the images. "Three independent board-certified practicing radiologists" confirmed the "diagnostic quality of the images." For the material decomposition functionality, the ground truth was based on a manufactured "multi-energy CT phantom, where three basic materials, water, iodine and calcium were decomposed." This phantom represents a known, measurable composition.

    8. Sample Size for the Training Set

    The document does not provide any information regarding the sample size used for the training set (if any machine learning components were involved in the new features). Given that the submission focuses on hardware and software feature upgrades rather than a new AI diagnostic algorithm, a distinct "training set" may not be applicable in the traditional sense of machine learning model development. This is more of a system validation rather than an AI model clearance.

    9. How the Ground Truth for the Training Set Was Established

    As no training set is described in the document, there is no information on how its ground truth was established.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211711

    Validate with FDA (Live)

    Device Name
    OmniTom Elite
    Date Cleared
    2022-02-25

    (267 days)

    Product Code
    Regulation Number
    892.1750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OmniTom Elite computed tomography (CT) system is intended to be used for x-ray computed tomography applications for anatomy that can be imaged in the 40cm aperture, primarily head and neck.

    The CT system is intended to be used for both pediatric and adult imaging and as such has preset dose settings based upon weight and age. The CT images can be obtained either with or without contrast.

    OmniTom Elite with photon counting detectors (PCD) can generate spectral CT images at multiple energy levels. OmniTom Elite with PCD is only supported for adult imaging for anatomy that can be imaged in the 40cm aperture, primarily head and neck.

    Device Description

    The subject OmniTom Elite Computed Tomography (CT) system provides the same functionality as the previous version of the device, OmniTom Elite (K202526). We have included some other design changes to make the system more robust and reliable. Both computed tomography systems are identical in terms of the high resolution, 16 row, 40 cm bore, and 30 cm field of view. The lightweight translating gantry consists of a rotating disk with a solid-state x-ray generator, collimator, control computer, communications link, power slip-ring, data acquisition system, reconstruction computer, power system, brushless DC servo drive system (disk rotation) and an internal drive system (translation). The power system consists of batteries which provide system power while unplugged from the charging outlet. The system has the necessary safety features such as the emergency stop switch, x-ray indicators, interlocks, patient alignment laser and 110% x-ray timer. The gantry has omni-directional wheels that allow for robust diagonal, lateral, and rotational 360-degree movement and electrical drive system so the system can be moved easily to different locations.

    We have added the ability to upgrade the OmniTom Elite system with photon counting detector (PCD). OmniTom Elite with PCD is the same system as the predicate device OmniTom Elite energy integrating detector (EID) with the only difference being the detector array system, instead of the current gadolinium oxysulfide EID, it has a cadmium telluride based PCD.

    PCD provides the ability to capture CT data in multiple energy bands that can provide information on material composition of different tissues and contrast media. The multiple sets of CT data are acquired at the same time with configurable energy thresholds without any cross talk between images. OmniTom Elite with PCD has the capacity to more than double the EID detector spatial resolution.

    AI/ML Overview

    The provided text describes the OmniTom Elite CT system, which offers an optional Photon Counting Detector (PCD) upgrade. The submission aims to demonstrate substantial equivalence to its predicate device, OmniTom Elite with Energy Integrating Detector (EID) (K202526).

    Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally framed around demonstrating the safety, effectiveness, and comparable image quality of the OmniTom Elite with PCD to its EID counterpart, without raising new questions of safety or effectiveness. Specific quantitative acceptance criteria are not explicitly listed in easily consumable metrics such as sensitivity/specificity/accuracy, but rather implied through compliance with standards, successful performance tests, and expert review.

    Acceptance Criteria CategoryDescriptionReported Device Performance
    SafetyCompliance with relevant electrical safety, radiation safety, and electromagnetic compatibility standards. Risk analysis and mitigation.- "All components...are certified to meet those requirements [Federal Diagnostic Equipment Performance Standard and 21 CFR §1020.30 and §1020.33]." - "NeuroLogica adheres to recognized and established industry practices." - "OmniTom Elite CT system is designed and manufactured to comply with the FDA Quality System Regulations (21 CFR part 820) and ISO 13485:2016 requirements." - Conformance with numerous harmonized standards including ISO 14971 (Risk Management), IEC 60601-1 (Basic Safety), IEC 60601-1-2 (EMC), IEC 60601-1-3 (Radiation Protection), IEC 60601-2-44 (CT Equipment Specific), NEMA XR 25, 26, 28, 29, 21 CFR §1020.30, §1020.33. - "The OmniTom Elite with PCD underwent Electrical Safety and Electromagnetic Compatibility testing and proved to be in compliance with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-44." - "The risk management analysis identified potential hazards which were controlled and mitigated during development."
    Effectiveness/FunctionalityDevice performs as intended for CT imaging, including new PCD capabilities (multi-energy, spectral CT), without negative impact from design changes.- "The proposed OmniTom Elite device demonstrated that the new features did not exhibit any negative effects on the requirements in place, as well as they did not exhibit any concerns that may impact safety and effectiveness." - "Software verification and software validation testing was executed to confirm all software requirements were met. The proposed OmniTom Elite device was shown to meet all requirements and to not have any impact on imaging." - PCD capabilities were tested to demonstrate properties of PCD in count and multi-energy mode. Imaging metrics demonstrated stability of the detector, analysis of the count rate and limited pileup loss at low dose scanning. - "Comparative study was conducted... The comparative study involved scanning the same subject using both the EID and the PCD CT systems at the same anatomical sections... These images were reviewed by two independent board-certified radiologists and confirmed diagnostic quality of the images demonstrating the effectiveness of the subject device for clinical scanning."
    Image QualityImage quality metrics (noise, slice thickness, resolution) are comparable to the predicate device and meet established criteria.- "Imaging metrics such as noise, slice thickness, low and high contrast resolution, radiation metrics, and modulation transfer function were measured utilizing phantom image quality tests in accordance with the equipment performance standards for diagnostic x-ray systems administered by the FDA." - "Imaging metrics successfully demonstrated that the proposed device with PCD has comparable image quality with its previous version, predicate device (K202526) that had EID and meets all the image quality criteria that are used for testing."
    PCD Specific FunctionalityAbility to generate spectral CT images at multiple energy levels with configurable thresholds, minimal crosstalk, and enhanced spatial resolution for adults in head/neck. Stable detector and acceptable count rate.- "PCD provides the ability to capture CT data in multiple energy bands that can provide information on material composition... The multiple sets of CT data are acquired at the same time with configurable energy thresholds without any cross talk between images." - "OmniTom Elite with PCD has the capacity to more than double the EID detector spatial resolution." - "No crosstalk noticed between images." - "Imaging metrics demonstrated stability of the detector, analysis of the count rate and limited pileup loss at low dose scanning."
    Usability/Movement ControlEnhanced movement controls for simpler/easier navigation, improved durability, and ease of operator transport.- "The addition of a drive camera (SmartDrive) and new transport wheels (OmniWheels) for simpler navigation; the drive bar mechanism was enhanced for durability, easier manufacturability and ease of operator overall transportation of device." - "The improvements in movement performance were implemented to solely allow the operator to transport the device with ease."
    Software IntegritySoftware developed and tested in accordance with relevant standards and guidance; no negative impact on imaging.- "The software contained in the proposed device has been developed & tested in accordance with IEC 62304, and the FDA guidance for Content of Premarket Submissions for Software Contained in Medical Devices." - "Software verification and software validation testing was executed to confirm all software requirements were met." - "The proposed OmniTom Elite device was shown to meet all requirements and to not have any impact on imaging."

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: The text states, "Volunteers underwent scanning to aid in generation of comparative images." It does not specify the exact number of volunteers. It implies a small number, as it talks about "the same subject" being scanned.
    • Data Provenance: Not explicitly stated, but the mention of "an IRB" (Institutional Review Board) implies a prospective study involving human subjects. The country of origin is not specified but is likely the US given the FDA submission. It was a prospective study involving volunteers.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Number of Experts: "two independent board-certified radiologists."
    • Qualifications: "board-certified radiologists." No further details on years of experience are provided, but board certification implies a high level of expertise in medical imaging interpretation.

    4. Adjudication Method for the Test Set

    The text states: "These images were reviewed by two independent board-certified radiologists and confirmed diagnostic quality of the images demonstrating the effectiveness of the subject device for clinical scanning."

    This indicates a consensus or agreement-based approach for diagnostic quality, where both radiologists had to confirm it. There is no mention of a formal adjudication process (e.g., 2+1 or 3+1 where a third or fourth reader resolves discrepancies), beyond the agreement on diagnostic quality. It seems the goal was to confirm diagnostic quality, not necessarily to create a "ground truth" for specific disease detection where adjudication for disagreements would be critical.

    5. If a Multi-Reader, Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a formal MRMC comparative effectiveness study to assess improvement in human readers with AI vs. without AI assistance was not conducted or described. The study described was a comparative imaging study to show that images from the new PCD system were of diagnostic quality and comparable to the EID system. The focus was on the device's image quality and diagnostic utility as a standalone imaging system, not its impact on human reader performance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done

    Yes, a standalone assessment was a primary component of the submission. The "Image quality evaluation" section describes rigorous phantom testing and measurements of "noise, slice thickness, low and high contrast resolution, radiation metrics, and modulation transfer function." This demonstrates the algorithm's (and hardware's) performance in generating images. The clinical evaluation with radiologists also served to confirm the diagnostic quality of the images generated by the device itself, implicitly assessing its standalone performance.

    7. The Type of Ground Truth Used

    The ground truth for the device's image quality and diagnostic utility was established through:

    • Phantom Tests: For objective image quality metrics (noise, resolution, MTF, etc.).
    • Expert Consensus/Opinion: For confirming the "diagnostic quality" of the clinical images by two independent board-certified radiologists. This isn't a "pathology" or "outcomes data" ground truth, but rather a subjective assessment by experts of the image's clinical utility.

    8. The Sample Size for the Training Set

    The document does not mention a training set sample size. This is because the device is a CT scanner, not an AI/ML diagnostic algorithm that requires a separate training set of patient data for model development. The "training" here refers to the engineering and validation processes (design verification, software testing, bench testing, etc.) that ensure the system functions correctly and produces high-quality images.

    9. How the Ground Truth for the Training Set Was Established

    As explained above, there isn't a "training set" in the sense of patient data used to train an AI model. The system's "ground truth" (i.e., its correct operational parameters and expected image output) was established through:

    • Engineering Specifications and Design Requirements: What the system was designed to achieve.
    • Reference Standards: Adherence to IEC 60601-x standards, NEMA, and FDA performance standards.
    • Phantom Studies: Objective measurements against known targets.
    • Bench Testing: Verification and validation of hardware and software components against their specified functions.
    Ask a Question

    Ask a specific question about this device

    K Number
    K202526

    Validate with FDA (Live)

    Device Name
    OmniTom Elite
    Date Cleared
    2020-12-23

    (113 days)

    Product Code
    Regulation Number
    892.1750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OmniTom Elite CT system is intended to be used for x-ray computed tomography applications for anatomy that can be imaged in the 40cm field of view, primarily head and neck.

    The CT system is intended to be used for both pediatric and adult imaging and as such has preset dose settings based upon weight and age. The CT images can be obtained either with or without contrast.

    Device Description

    The OmniTom Elite is an improved version of OmniTom computed tomography system, providing enhanced functionality. It still has the same high resolution, multi row, 40 cm bore, and 30 cm field of view x-ray computed tomography system. The lightweight translating gantry consists of a rotating disk with a solid state x-ray generator, Gd2O2S detector array, collimator, control computer, communications link, power slip-ring, data acquisition system, reconstruction computer, power system, brushless DC servo drive system (disk rotation) and an internal drive system (translation). The power system consists of batteries which provide system power while unplugged from the charging outlet. The system has the necessary safety features such as the emergency stop switch, x-ray indicators, interlocks, patient alignment laser and 110% x-ray timer. The gantry has omnidirectional wheels that allow for robust diagonal, lateral, and rotational 360-degree movement and electrical drive system so the system can be moved easily to different locations.

    AI/ML Overview

    The provided text is a 510(k) summary for the OmniTom Elite CT system, which is a medical device. This type of regulatory document focuses on establishing substantial equivalence to a predicate device rather than providing detailed acceptance criteria and a complete study report for a novel AI/software component, particularly for diagnostic performance metrics like those specified in the prompt.

    The document primarily discusses comparative technical characteristics and conformance to established standards for computed tomography X-ray systems, rather than an AI or software algorithm's diagnostic performance. Therefore, many of the requested points, especially those related to diagnostic performance study design (test set size, ground truth, experts, MRMC, standalone performance), are not applicable or not present in this document.

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't present a formal table of "acceptance criteria" in terms of diagnostic performance metrics (e.g., sensitivity, specificity, AUC) for an AI algorithm. Instead, it compares the technical specifications of the subject device (OmniTom Elite) to its predicate device (OmniTom). The "performance" assessment focuses on image quality metrics measured using phantoms and overall system safety and effectiveness.

    AttributeAcceptance Criteria (Implicit - at least comparable to predicate)Reported Device Performance (OmniTom Elite)
    Indications for UseSame as predicate (x-ray CT for 40cm FOV, primarily head and neck, pediatric and adult imaging, with/without contrast)Same as predicate
    Aperture (cm)4040
    Image Field of View (cm)3030
    Detector MaterialGadolinium OxysulfideGadolinium Oxysulfide
    Detector Coverage (mm)16 x 0.625 mm16 x 0.625 mm
    Slice Configurations1616
    Spatial Resolution (Soft tissue kernel) (lp/cm)Comparable to predicate (6.0 lp/cm)6.7 lp/cm (Improved)
    Spatial Resolution (High resolution kernel) (lp/cm)Comparable to predicate (14.0 lp/cm)14.8 lp/cm (Improved)
    X-ray Tube TypeGrounded/Fixed Anode/MonoblockGrounded/Fixed Anode/Monoblock
    Heat Storage (MHU)0.720.72
    CoolingWater-GlycolWater-Glycol
    Tube Current (Ma)5 – 455 – 45
    Tube Voltage Range (Kv)80, 100, 12080, 100, 120
    Rotation time (s)1, 21, 2
    Gantry Weight (lbs)17001700
    Mobile / StationaryMobileMobile
    Battery / Wall powerLithium Iron PhosphateLithium Iron Phosphate
    Input VoltageSingle Phase 90 – 264 VAC/1300 watts peakSingle Phase 90 – 264 VAC/1300 watts peak
    Max Input Power (Kva)5.4 Kw5.4 Kw
    PACS DICOM 3.1YesYes
    2D Scout ScanYesYes
    Bolus TrackingNo (predicate)Yes (New feature)
    Axial ScanYesYes
    Helical ScanYesYes
    Dynamic ScanYesYes
    2D ViewingYesYes
    MPR ViewingYesYes
    3D ViewingYesYes
    Max Scan Range (cm)4040
    Scan LocalizerLaserLaser
    Cardiac/Respiratory GatingNoNo
    Patient TableNoNo
    Image recon speed max (images per second)16 image/s16 image/s
    X-ray Warning LightsYesYes
    110% X-ray TimerYesYes
    Emergency StopYesYes
    Operator X-ray On SwitchYesYes
    Quality Test PhantomYesYes
    Quality Test ReportYesYes
    X-ray FilterBowtieBowtie
    Administrator PrivilegesYesYes
    Dose DisplayYesYes
    Dose Report / AuditYesYes
    Protocol Override ProtectionYesYes
    Dose CheckYesYes
    Pediatric ProtocolsYes (By Age/weight)Yes (By Age/weight)
    Automatic Exposure ControlYesYes
    BiocompatibilityN/A (predicate)Yes (Applicable for subject device)
    SterilityN/AN/A
    Chemical SafetyN/AN/A
    EM EmissionsETL TestingETL Testing
    Electrical Safety (IEC 60601-1)ETL TestingETL Testing
    Mechanical Safety (IEC 60601-1)ETL TestingETL Testing
    Where Used(Mobile) Radiology, ICU, ED, OR, Clinic, Office(Mobile) Radiology, ICU, ED, OR, Clinic, Office, Ambulance (New location)
    Metal Artifact Reduction (MAR)Not explicitly stated for predicate but stated as a new feature for the subject device. Performance assessed by "successfully mitigating the streak artifacts... without raising any new questions of safety or effectiveness."Performs as indicated, mitigating streak artifacts without new safety/effectiveness concerns.
    Image Quality (Noise, Slice thickness, Low and High contrast resolution, Radiation metrics, MTF)Must demonstrate comparable image quality to predicate.Successfully demonstrated comparable image quality as the predicate and meets all image quality criteria used for testing the OmniTom.
    Diagnostic Quality of ImagesImages reviewed by a board-certified radiologist must be of diagnostic quality.Images reviewed by a board-certified radiologist were of diagnostic quality.

    Study Proving Acceptance Criteria:

    The document states that "internal verification and validation activities and external testing of product safety and EMC / EMI was completed successfully." The "study" mentioned for image quality and diagnostic relevance primarily relies on phantom testing and expert review of those phantom-generated images.

    • "Image quality metrics such as noise, slice thickness, low and high contrast resolution, radiation metrics, and modulation transfer function were measured utilizing phantom image quality tests in accordance with the equipment performance standards for diagnostic x-ray systems administered by the FDA. The OmniTom Elite system successfully demonstrated that it has comparable image quality as the predicate device OmniTom (K171183) and meets all the image quality criteria that are used for testing the OmniTom."
    • "Clinical performance of the CT system was evaluated using an ACR Phantom was used to measure image metrics, such as CT number linearity, image slice thickness, image noise, low contrast resolution and high contrast resolution. The data provided clinical demonstration of the operation of the device and the images reviewed by a board-certified radiologist were of diagnostic quality."

    2. Sample size used for the test set and the data provenance

    • Test Set Sample Size: Not specified in terms of patient cases. The evaluation involved "phantom image quality tests" and an "ACR Phantom."
    • Data Provenance: The data appears to be from internal lab testing using phantoms, not human patient data. Therefore, country of origin and retrospective/prospective designations are not applicable.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Number of Experts: "a board-certified radiologist" (singular, implied one) reviewed the images from the ACR Phantom.
    • Qualifications of Experts: "board-certified radiologist." No further details on years of experience or subspecialty are provided.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Adjudication Method: Not applicable/not specified. The review was by a single "board-certified radiologist" for diagnostic quality of phantom images. There's no indication of multiple readers or an adjudication process for a "ground truth" derived from patient cases.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No. This document describes a CT scanner (hardware and basic software functions), not an AI diagnostic algorithm. The "Metal Artifact Reduction (MAR)" is a post-processing algorithm, but its evaluation does not appear to be a comparative effectiveness study involving human readers with and without AI assistance. The focus for MAR was simply demonstrating artifact mitigation without new safety/effectiveness concerns.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone Performance: Not applicable in the context of diagnostic AI. The document primarily concerns the performance of the CT scanner hardware and integrated software features. While MAR is an algorithm, its "standalone" performance is described in terms of its effect on artifacts in phantom images, not as a diagnostic accuracy measure on patient data.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Ground Truth: For image quality, the ground truth is against expected values and characteristics of reference phantom images. For the diagnostic quality assessment, the "ground truth" for ensuring images were "of diagnostic quality" was the subjective assessment by the board-certified radiologist of the phantom images. This is not a ground truth for a disease state based on patient data.

    8. The sample size for the training set

    • Training Set Sample Size: Not applicable. This document does not describe the development or validation of a machine learning model that would require a training set. The software updates mentioned are general enhancements and workflow improvements, not an AI algorithm for diagnostic inference.