Found 3 results

510(k) Data Aggregation

    K Number
    K242375

    Validate with FDA (Live)

    Date Cleared
    2024-10-07

    (59 days)

    Product Code
    Regulation Number
    888.3358
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OR30 Dual Mobility System is intended for use in primary and revision total hip arthroplasty in skeletally mature patients.

    Indications

    · Advanced degeneration of the hip joint as a result of degenerative, post-traumatic, or rheumatoid arthritis.

    · Fracture or avascular necrosis of the femoral head.

    · Failure of previous hip surgery; joint reconstruction, arthrodesis, hemiathroplasty, surface replacement athroplasty, or total hip replacement.

    · All forms of osteoarthritis.
    · Patients with hips at risk of dislocation.

    · Femoral neck fracture or proximal hip joint fracture.

    The OR3O Dual Mobility System is intended for single use only. The modular OR3O Liners and Inserts are to be implanted without bone cement.

    Device Description

    The OR3O Hip System was originally cleared via K220959. The OR3O Hip System is a modular dual mobility implant system containing two distinct articulating surfaces. The hip system consists of a metal acetabular shell, diffusion hardened oxdized zirconium (OXINIUM DH) alloy liner, a cross-linked polyethylene (XLPE) Insert and a femoral head.

    The purpose of this Taditional 510(k) is to add the MR safety information to the product labels and update the MR information within the package insert for the Snith & Nephew OR3O Dual Mobility System included within the scope of this 510(k). The subject OR30 Dual Mobility System is identical in design, technological characteristics, function, packaging, and sterilization to the commercially available predicate devices.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device (OR30 Dual Mobility System), seeking clearance to market based on substantial equivalence to predicate devices. It primarily focuses on the device's design, materials, indications for use, and a specific change related to MR safety information.

    This document does NOT contain information about acceptance criteria or a study proving that an AI/software device meets acceptance criteria. It discusses the testing done for MR compatibility for a physical implant, not a software or AI device. Therefore, it's impossible to extract the requested information (acceptance criteria for a software device, test set details, expert ground truth, MRMC study, standalone performance, training set details) from the provided text.

    The closest relevant section is:

    Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
    "Magnetic Resonance Imaging (MRI) compatibility testing was performed and the results of the testing performed were evaluated and used to generate recommended MRI safety labeling for the subject devices. Testing and evaluation were performed in accordance with the standards listed below:

    1. Magnetically induced displacement force (ASTM F2052)
    2. Magnetically induced torque (ASTM F2213)
    3. Radiofrequency (RF) induced heating (ASTM F2182)
    4. MR image artifact (ASTM F2119)
      Based on the MR assessment performed for the subject devices are MR Conditional and substantially equivalent to the predicate MRI 510(k) K211176."

    This describes physical testing of the implant's interaction with MRI, not the performance of an AI/software.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220959

    Validate with FDA (Live)

    Date Cleared
    2022-08-18

    (139 days)

    Product Code
    Regulation Number
    888.3358
    Age Range
    All
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended Use The OR3O Dual Mobility System is intended for use in primary and revision total hip arthroplasty in skeletally mature patients.

    Indications

    • · Advanced degeneration of the hip joint as a result of degenerative, post-traumatic, or rheumatoid arthritis.
    • · Fracture or avascular necrosis of the femoral head.
    • · Failure of previous hip surgery: joint reconstruction, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement.
    • · All forms of osteoarthritis.
    • · Patients with hips at risk of dislocation.
    • · Femoral neck fracture or proximal hip joint fracture.

    The OR30 Dual Mobility System is intended for single use only. The modular OR30 Liners and Inserts are to be implanted without bone cement.

    Mating components may be indicated for use without bone cement.

    Device Description

    The OR30 Hip System is a modular dual mobility implant system consists of diffusion hardened oxidized zirconium (OXINIUM DH) alloy liners with a highly polished zirconium oxide inner surface and a machined locking taper and backside of Zr2.5Nb alloy (F2384-10R16). The locking taper and outside profile is designed to mate with a dedicated Ti6Al4V R3 acetabular shell with OD sizes 44mm to 74mm or REDAPT Modular Press-fit acetabular shell with OD sizes 48mm to 74mm. For each assembled OR30 liner and R3 shell size from 44mm to 74mm and REDAPT Modular shell size from 48mm to 74mm, a dedicated insert is available. Inserts are made of highly crosslinked ultra-high-molecularweight polyethylene (ISO 5834-2/ASTM F648) that is irradiated with a 10 Mrad dose and remelted . These can be combined with oxidized zirconium or CoCr alloy femoral heads of sizes 22 mm (for Size 44mm-52mm) and 28 mm (for size 50mm-74mm). The final OR30 Dual Mobility construct will include an acetabular shell (with optional screws and optional screw hole covers), an OXINIUM DH Liner, a cross linked polyethylene (XLPE) Insert and a femoral head.

    The purpose of this Traditional 510(k) submission is to add additional components, namely additional sizes of the OR3O Dual Mobility OXINIUM DH Liners and OR3O Dual Mobility XLPE Inserts to the existing OR30 Dual Mobility System.

    This line extension will consist of two smaller size OXINIUM DH Liners and four additional XLPE inserts to allow for a wider range of implant options for physicians to choose from. The materials and manufacturing processes will be the same, with the exception of accounting for different sized implants.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from the FDA regarding the "OR30 Dual Mobility System," a hip joint prosthesis. The provided text primarily focuses on establishing substantial equivalence to existing predicate devices based on design, materials, indications for use, and performance data, including biocompatibility and mechanical testing.

    However, the document does NOT contain information about any study involving AI, human readers, or the establishment of ground truth for evaluating such a system. The "acceptance criteria" and "device performance" described in the document relate to the physical and biological characteristics of the orthopedic implant and its components, not to the performance of an AI or diagnostic system.

    Therefore, I cannot extract the information required by your prompt, as it pertains to AI/diagnostic system evaluation, which is absent from this regulatory submission for a hip joint prosthesis.

    The document covers:

    • Device Description: What the OR30 Dual Mobility System is, its components (liners, inserts, shells, femoral heads), and the materials used (OXINIUM DH, XLPE). It highlights that the current submission is for a line extension – adding smaller sizes of existing components.
    • Indications for Use: The medical conditions for which the hip system is intended (e.g., hip degeneration, fracture, failed previous hip surgery, osteoarthritis, risk of dislocation).
    • Comparison to Predicate Devices: A detailed comparison of the subject device (OR30 Dual Mobility System line extension) against previously cleared predicate devices, emphasizing similarities in intended use, fundamental scientific technology, materials, and indications for use.
    • Performance Data:
      • Biocompatibility: Evaluation according to ISO-10993 for permanent implants (>30 days body contact), confirming compliance with material standards (ASTM F2384-10R16 for Zr-2.5Nb alloy and ASTM F648-14 / ISO 5834-2 for XLPE). A Declaration of Conformity was provided.
      • Mechanical Testing: Listing of various ASTM and ISO standards for tests like push-out, torque to failure, lever-out, environmental fatigue, corrosion assessment, deformation, neck impingement, range of motion, and wear testing. It states that biomechanical testing and/or rationales were completed and "met the pre-determined acceptance criteria for each intended output." The submission concludes that the device does not represent a new worst-case scenario compared to predicates.
      • Non-Pyrogenicity Endotoxin Testing: Bacterial endotoxin testing was completed and met acceptable limits per FDA guidance and ANSI/AAMI ST72.

    In summary, this document is a regulatory submission for a physical medical device (orthopedic implant) and does not involve AI or diagnostic imaging. Therefore, it does not contain the specific information requested in your prompt regarding AI system acceptance criteria, study design for AI evaluation, expert ground truth, or MRMC studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191002

    Validate with FDA (Live)

    Date Cleared
    2019-10-31

    (198 days)

    Product Code
    Regulation Number
    888.3358
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended Use The OR3O Dual Mobility System is intended for use in primary and revision total hip arthroplasty in skeletally mature patients.

    Indications

    • · Advanced degeneration of the hip joint as a result of degenerative, post-traumatic, or rheumatoid arthritis.
    • · Fracture or avascular necrosis of the femoral head.

    · Failure of previous hip surgery: joint reconstruction, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement.

    • · All forms of osteoarthritis.
    • · Patients with hips at risk of dislocation.
    • · Femoral neck fracture or proximal hip joint fracture.

    The OR30 Dual Mobility System is intended for single use only. The modular OR30 Liners and Inserts are to be implanted without bone cement.

    Mating components may be indicated for use without bone cement.

    Device Description

    The OR30 Hip System is a modular dual mobility implant system. The system consists of diffusion hardened, oxidized zirconium alloy liners with a highly polished inner surface of zirconia and a machined locking taper and backside of Zr-2.5Nb alloy. The locking taper and machined outside profile is designed to mate with a dedicated Ti 6Al 4V R3 or REDAPT Modular Press-fit acetabular shell with OD sizes 48mm to 74mm. For each assembled OR30 Liner and R3 and REDAPT Modular Shell size from 48mm to 74mm, a dedicated plastic insert made of highly crosslinked ultrahigh-molecular-weight polyethylene (10Mrad irradiated and remelt-annealed UHMWPE (XLPE) according to ISO 5834-2/ASTM F648) is available. These can be combined with oxidized zirconium or CoCr alloy femoral heads of sizes 22mm (for size 48mm-52mm) and 28mm (for size 54mm-74mm). The final OR30 Dual Mobility construct will include an acetabular shell, an OXINIUM DH Liner, an XLPE Insert and a femoral head.

    The purpose of this 510(k) submission is to add additional options of dual mobility implants to the Smith & Nephew collection of hip implants.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device, the OR30 Dual Mobility System, and focuses on demonstrating substantial equivalence to previously cleared devices. It does not contain information about acceptance criteria and study data for a diagnostic device in the usual sense (e.g., performance metrics like sensitivity, specificity, or AUC). Instead, the performance data provided is for biocompatibility and mechanical testing of the implantable device.

    Therefore, many of the requested categories for a diagnostic device's acceptance criteria and study data (such as sample size for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, training set details) are not applicable to the information provided in this 510(k) summary.

    I will interpret the request within the context of the provided document, focusing on the "acceptance criteria" for the mechanical testing of the OR30 Dual Mobility System, as this is the closest equivalent to performance evaluation in this type of submission.

    Here's the information based on the document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance Metric (Mechanical Testing)Acceptance CriteriaReported Device Performance
    Push Out TestingPre-determined acceptance criteria (not explicitly defined in document)Met pre-determined acceptance criteria
    Torque to Failure TestingPre-determined acceptance criteria (not explicitly defined in document)Met pre-determined acceptance criteria
    Lever-Out TestingPre-determined acceptance criteria (not explicitly defined in document)Met pre-determined acceptance criteria
    Environmental Fatigue TestingPre-determined acceptance criteria (not explicitly defined in document)Met pre-determined acceptance criteria
    Deformation TestingPre-determined acceptance criteria (not explicitly defined in document)Met pre-determined acceptance criteria
    Wear TestingPre-determined acceptance criteria (not explicitly defined in document)Met pre-determined acceptance criteria
    Pull-Out TestingPre-determined acceptance criteria (not explicitly defined in document)Met pre-determined acceptance criteria
    Neck Impingement TestingPre-determined acceptance criteria (not explicitly defined in document)Met pre-determined acceptance criteria
    Range of MotionPre-determined acceptance criteria (not explicitly defined in document)Met pre-determined acceptance criteria
    Jump DistancePre-determined acceptance criteria (not explicitly defined in document)Met pre-determined acceptance criteria
    BiocompatibilityIn accordance with FDA's Draft Guidance for ISO-10993Report evaluation completed and rationales provided
    Non-Pyrogenicity Endotoxin TestingMet acceptable endotoxin limits per FDA Guidance and ANSI/AAMI ST72Met acceptable endotoxin limits

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • The document does not specify sample sizes for any of the mechanical or biocompatibility tests.
    • Data provenance is not mentioned. It can be inferred that testing was conducted as part of the device development and submission process, likely by the manufacturer or a contracted lab. This is industrial testing, not clinical data collection.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. This is a physical device intended to be implanted. "Ground truth" for mechanical testing is based on engineering specifications and adherence to standards, not expert clinical interpretation of data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable, as this refers to clinical evaluation and consensus building for diagnostic accuracy, which is not part of this submission.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is a physical implantable device, not a diagnostic AI system or a system requiring human interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a physical implantable device, not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • For mechanical testing: The "ground truth" or reference for performance is established by pre-determined acceptance criteria based on engineering standards (e.g., ASTM, ISO), regulatory guidelines, and comparison to predicate devices, ensuring the device meets safety and performance requirements for its intended use. The document states the device "met the pre-determined acceptance criteria for each intended output."
    • For biocompatibility: Ground truth is adherence to international standards like ISO-10993.
    • For non-pyrogenicity: Ground truth is adherence to established endotoxin limits in FDA Guidance and ANSI/AAMI ST72.

    8. The sample size for the training set

    • Not applicable. This is a physical implantable device; there is no "training set" in the context of machine learning or diagnostic device development.

    9. How the ground truth for the training set was established

    • Not applicable, as there is no training set mentioned in this context.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1