510(k) Data Aggregation

    K Number
    K141484

    Validate with FDA (Live)

    Device Name
    OPEN FMSEALER
    Date Cleared
    2014-07-30

    (55 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Open FMsealer is intended for use in general and gynecological surgery and other open surgical procedures where ligation of vessels, including lymph vessels, is desired. The system creates a vessel ligation (seal) by the application of heat (coagulation) to vessels interposed between the jaws of the device. The Open FMsealer can be used to seal and ligate vessels up to 7mm and tissue bundles as large as will fit in the jaws of the instrument.

    The Open FMsealer has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these purposes.

    Device Description

    An electrosurgical cutting and coagulation device (and accessories) is a device intended to remove tissue and control bleeding by use of high-frequency electrical current (see 21 CFR §878.4400). It is classified as a Class II (510(k)) device.

    The Open FMsealer is a device intended for sealing of vessels and for soft tissue cutting and coagulation. The device is designed to be used only with the FMwand Generator (K130606). The two-piece device consists of a sterile, single-use patient-contacting shears section (Open FMsealer) and an accessory cable (Power Module or PM) that connects the FMsealer to the generator. The dual footpedal connects to the rear of the generator and can be used to activate the FMsealer if desired by the user.

    The Open FMsealer is a hand-held surgical instrument intended for use in open surgical procedures where ligation of vessels or cutting and coagulation of soft tissue is desired. The Open FMsealer consists of a scissor-like handle with two actuation buttons and jaws to grasp the tissue to be affected.

    The Open FMsealer is connected to the FMwand Generator (K130606) by the Power Module. The connectors on both ends of the Power Module are designed to fit only the FMwand Generator and FMwand sterile, disposable accessories. The power module and FMwand Generator were previously cleared as part of K130606.

    This system creates sealing, cutting and coagulation by the application of heat and compression to tissue bundles and vessels interposed between the iaws of the instrument. Like the predicate device, the Open FMsealer includes actuation buttons that can be used to activate the device.

    The Open FMsealer has been shown (in animal studies) to cut and seal vessels up to and including 7mm in diameter, lymphatics and tissue bundles as large as will fit into the jaws of the device.

    AI/ML Overview

    This 510(k) summary describes the "Open FMsealer" intended for vessel sealing and soft tissue cutting and coagulation in surgical procedures. The device uses ferromagnetic induction to generate heat, similar to its predicate device, the FMwand Ferromagnetic Surgical System (K130606). It also claims substantial equivalence to the LigaSure Vessel Sealing System (K043273).

    Here's an analysis of the provided information according to your requested criteria:


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in the format of pass/fail thresholds for specific quantitative metrics. Instead, it describes performance characteristics and states that the device "passed all bench tests" and "is at least as effective as the predicate device." The performance testing focused on demonstrating substantial equivalence to the predicate devices.

    Performance Feature / Acceptance Criteria (Implied)Reported Device Performance
    Vessel Sealing CapabilityCuts and seals vessels up to and including 7mm in diameter.
    Tissue Sealing/Cutting CapabilitySeals and ligates tissue bundles as large as will fit into the jaws of the device. Cuts and coagulates soft tissue.
    Electrical Safety Standards ComplianceComplies with IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, and IEC 60601-2-2.
    BiocompatibilityMaterials in patient-contacting portions are known to be biocompatible or passed testing according to ISO 10993-5 (Cytotoxicity), 10993-10 (Acute Systemic Toxicity), and 10993-11 (Sensitization and Irritation).
    SterilizationProvided sterile, single-patient use. Sterilization method is Ethylene Oxide: SAL 10-6.
    Functional Equivalence"At least as effective as the predicate device" for vessel sealing strength, tissue histological analysis, and acute/chronic animal studies.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document mentions "animal studies" for performance testing, including "acute and chronic animal studies" and "testing on animals and animal tissues." However, specific sample sizes (number of animals, number of vessels/tissue bundles, etc.) are not provided in the summary.
    • Data Provenance: The studies were prospective animal studies. The country of origin for the data is not specified but implicitly US-based, given the FDA submission.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    The document does not mention the use of human experts to establish ground truth for the animal studies. It refers to "tissue histological analysis" which would be performed by pathologists, but no details are given regarding the number or qualifications of such experts.


    4. Adjudication Method for the Test Set

    Since human experts are not mentioned as establishing ground truth in a consensus-based manner, no adjudication method (e.g., 2+1, 3+1) is described or applicable in the context of this summary. The results of the animal studies and histological analyses would likely be interpreted by the researchers performing the studies and potentially reviewed by veterinary pathologists.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The clearance of this device was "not based on human clinical testing." The studies focused on animal models and bench testing to demonstrate technical and performance equivalence to predicate devices.


    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    This question is not applicable to the Open FMsealer. This is a physical surgical device, not an AI algorithm. Its performance is inherent in its design and operation, not through an algorithm operating independently.


    7. The Type of Ground Truth Used

    The ground truth for the performance evaluations appears to be based on:

    • Animal Outcomes Data: Specifically, successful vessel sealing and tissue coagulation, indicated by "vessel sealing strength" and outcomes from "acute and chronic animal studies."
    • Pathology/Histology: "Tissue histological analysis" was performed, which relies on microscopic examination of tissue by pathologists to confirm the effects of the device (e.g., cell necrosis, collagen denaturation, seal integrity).
    • Bench Test Measurements: Quantitative measurements of physical parameters (e.g., thermal performance, seal strength).

    8. The Sample Size for the Training Set

    Not applicable. This device is a physical surgical instrument and does not use a "training set" in the context of machine learning.


    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As above, there is no "training set" for this device.