Found 1 results

510(k) Data Aggregation

    K Number
    K213842

    Validate with FDA (Live)

    Date Cleared
    2022-10-03

    (298 days)

    Product Code
    Regulation Number
    888.3358
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ONVOY™ Acetabular System is intended for use in reconstruction of the articulating surface of the acetabular portion of the hip that is severely disabled and/or very painful resulting from:

    1. Non-inflammatory degenerative joint disease including osteoarthritis, and avascular necrosis,
    2. Rheumatoid arthritis.
    3. Correction of functional deformity.
    4. Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques.
    5. Revision of previously failed total hip arthroplasty.
      The ONVOY™ Acetabular System is used in conjunction with Globus/Stelkast hip systems. The acetabular components of this system are intended for cementless fixation.
    Device Description

    The ONVOY™ Acetabular System consists of shells, liners, and bone screws that are used as part of a complete total hip system in conjunction with a femoral head and femoral stem in total hip arthroplasty. Implants are available in various configurations and sizes to fit a wide variety of patient anatomy. Shells are available in a cluster hole design, liners are available in hooded and non-hooded designs used in conjunction with the shells, and bone screws provide additional fixation when inserted through the shell.
    ONVOY™ acetabular shells are manufactured from titanium alloy, with a commercially pure titanium vacuum sintered coating. The liners are manufactured from ultra-high molecular weight polyethylene (UHMWPE) with Vitamin E. Bone screws are manufactured from titanium alloy.

    AI/ML Overview

    This document is a 510(k) premarket notification from the FDA for a medical device called the ONVOY™ Acetabular System. A 510(k) submission is for demonstrating that a medical device is substantially equivalent to a legally marketed predicate device, not for proving that a device meets specific performance criteria beyond substantial equivalence. Therefore, the information typically requested in your prompt regarding acceptance criteria, specific studies to prove criteria, sample sizes, expert ground truth, adjudication methods, MRMC studies, standalone performance, and training data is not part of this type of regulatory submission.

    This document confirms substantial equivalence through:

    • Indications for Use: The ONVOY™ Acetabular System has the same intended use as its predicate devices, which is for reconstructing the articulating surface of the acetabular portion of the hip in cases of non-inflammatory degenerative joint disease, rheumatoid arthritis, functional deformity, certain fractures, and revision of failed hip arthroplasty. It is used with Globus/StelKast hip systems and intended for cementless fixation.
    • Technological Characteristics: The subject implants possess the same technological characteristics as the predicate devices, including design, material composition (titanium alloy shells with commercially pure titanium vacuum sintered coating, UHMWPE liners with Vitamin E, and titanium alloy bone screws), and range of sizes.
    • Performance Data (Mechanical Testing): Mechanical testing was conducted in accordance with several ASTM and ISO standards (F1820, F2582, ISO 7206-12, F543, F3090, ISO 21535). These tests cover aspects like push-out, torque-out disassembly, impingement, deformation, screw testing, shell fatigue, and range of motion. The results of these tests "demonstrate substantial equivalence to the predicate devices."
    • Bacterial Endotoxin Testing (BET): BET was conducted in accordance with ANSI/AAMI ST-72:2011.

    Therefore, it's not possible to provide the specific details you requested because a 510(k) summary focuses on demonstrating substantial equivalence to existing devices, rather than establishing and meeting novel acceptance criteria through independent performance studies. The document indicates that the performance data shows substantial equivalence, which is the "acceptance criteria" for a 510(k) submission itself.

    Here's a breakdown based on the information provided and what is not provided in a 510(k) summary:

    Information RequestedProvided in Document?Details from Document
    1. Table of Acceptance Criteria & Reported Device PerformanceNo*The document states that "Performance data demonstrate substantial equivalence to the predicate devices." For a 510(k), demonstrating substantial equivalence is the acceptance criterion. Specific quantitative thresholds for various performance metrics would be assessed against the predicate device's established performance or relevant standards (e.g., ASTM/ISO). The document lists the types of mechanical tests performed (push-out, torque-out disassembly, impingement, deformation, screw testing, shell fatigue, range of motion) and standards (ASTM F1820, ASTM F2582, ISO 7206-12, ASTM F543, ASTM F3090, and ISO 21535), as well as Bacterial Endotoxin Testing (ANSI/AAMI ST-72:2011). However, the specific quantitative results of these tests and the precise acceptance criteria used to deem them "substantially equivalent" are not provided in this summary.
    2. Sample Size for Test Set & Data ProvenanceNoNot specified in a 510(k) summary. These details would be in the full submission, but generally pertain to engineering tests, not clinical data for an AI/diagnostic device.
    3. Number of Experts & Qualifications for Ground Truth for Test SetNot ApplicableThis is a mechanical orthopedic device, not an AI/diagnostic device requiring expert interpretation of images or signals for ground truth.
    4. Adjudication Method for Test SetNot ApplicableThis is a mechanical orthopedic device, not an AI/diagnostic device.
    5. MRMC Comparative Effectiveness Study & Effect Size (Human + AI vs. Human only)NoNot applicable. This is a medical device (hip prosthesis) and not an AI-assisted diagnostic or decision support tool.
    6. Standalone Performance Done (Algorithm Only)NoNot applicable. This is a medical device (hip prosthesis), not an algorithm.
    7. Type of Ground Truth UsedNot Applicable"Ground truth" in the context of this device would be established by physical measurements and test results according to specified standards, not typically by expert consensus or pathology in the way an AI diagnostic device would.
    8. Sample Size for Training SetNoNot applicable. This is a mechanical device, not an AI/ML algorithm that requires a training set.
    9. How Ground Truth for Training Set Was EstablishedNoNot applicable.

    In summary, this document is a regulatory approval notice based on demonstrating substantial equivalence to existing predicate devices through comprehensive mechanical and material testing, rather than presenting a study against novel, pre-defined acceptance criteria for a new type of performance claim (like an AI model's accuracy).