510(k) Data Aggregation

    K Number
    K073487

    Validate with FDA (Live)

    Date Cleared
    2008-02-11

    (61 days)

    Product Code
    Regulation Number
    866.5510
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    System reagent for the quantitative determination of IgM immunoglobulins in human serum and plasma on OLYMPUS analyzers. For in vitro diagnostic use.

    Device Description

    In this Olympus procedure:

    • When a sample is mixed with R1 buffer and R2 antiserum solution, human IgM reacts specifically with anti-human IgM antibodies to yield insoluble aggregates.
    • Immune complexes formed in solution scatter light in proportion to their size, shape and concentration.
    • Turbidimeters measure the reduction of incidence light due to reflection, absorption or scatter.
    • In the Olympus procedure, the decrease in intensity of light transmitted (increase in absorbance) through particles suspended in solution is as a result of complexes formed during the antigen-antibody reaction.
    AI/ML Overview

    The provided 510(k) summary describes a new Olympus IgM reagent (OSR6X173) and compares its performance to a predicate device. This document is a premarket notification for a diagnostic reagent, which falls under IVD (in vitro diagnostic) devices. For IVD devices, the "study" typically refers to analytical and clinical performance studies, not AI-specific studies like MRMC or standalone algorithm performance. Therefore, many of the requested items related to AI device performance (e.g., number of experts, adjudication methods, MRMC studies, standalone performance, training set details) are not directly applicable or reported in this type of submission.

    Here's an analysis based on the provided text, focusing on the available information:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state "acceptance criteria" in a separate section. Instead, it presents performance characteristics of the new device and the predicate device side-by-side, implying that the new device's performance is acceptable if it is comparable to or better than the predicate. For the purpose of this analysis, I will treat the performance values of the predicate device and the new device as demonstrating acceptable performance through substantial equivalence.

    Performance CharacteristicAcceptance Criteria (Predicate Device)Reported Device Performance (New Olympus IgM reagent OSR6X173)
    Precision (Total CV%)
    AU400/400eSample 1: 1.74%Sample 2: 1.18%Sample 3: 1.49%Sample 1: 4.03%Sample 2: 2.95%Sample 3: 2.60%
    AU600/640/640eSample 1: 2.16%Sample 2: 1.48%Sample 3: 1.35% (for AU600)Sample 1: 1.5%Sample 2: 1.3% (for AU640/640e)Sample 1: 3.44%Sample 2: 3.29%Sample 3: 4.08%
    AU2700/5400Sample 1: 2.49%Sample 2: 2.57%Sample 3: 3.07%Sample 1: 3.79%Sample 2: 3.31%Sample 3: 3.52%
    Assay RangeNot explicitly stated, inferred from Method Comparison20-500 mg/dL
    Hook Effect/ProzoneMay occur with highly elevated IgM samples > 3,500 mg/dL polyclonalMay occur with highly elevated IgM samples > 10,000 mg/dL polyclonal
    Method Comparison (Linear Regression)Slope: 0.968Intercept: 0.6R: 0.998N: Not clearly stated in predicate table (appears truncated)Slope: 1.006Intercept: 2.8R: 1.000N: 107
    Interfering Substances (Bilirubin)
    AU400/400eInterference less than 2% up to 40 mg/dLInterference less than 4% up to 40 mg/dL
    AU600/640/640eInterference less than 5% up to 40 mg/dLInterference less than 3% up to 40 mg/dL
    AU2700/5400Interference less than 10% up to 40 mg/dLInterference less than 8% up to 40 mg/dL
    Interfering Substances (Hemolysis)
    AU400/400eInterference less than 2% up to 500 mg/dLInterference less than 4% up to 500 mg/dL
    AU600/640/640eInterference less than 10% up to 500 mg/dLInterference less than 3% up to 500 mg/dL
    AU2700/5400Interference less than 5% up to 500 mg/dLInterference less than 3% up to 500 mg/dL
    Interfering Substances (Lipemia)
    AU400/400eInterference less than 10% up to 300 mg/dLInterference less than 10% up to 300 mg/dL
    AU600/640/640eInterference less than 10% up to 400 mg/dLInterference less than 10% up to 300 mg/dL
    AU2700/5400Interference less than 10% up to 400 mg/dLInterference less than 10% up to 200 mg/dL
    Interfering Substances (Ascorbic Acid)
    AU400/400eInterference less than 2% up to 20 mg/dLNot Tested
    AU600/640/640eInterference less than 1% up to 20 mg/dLNot Tested
    AU2700/5400Interference less than 3% up to 20 mg/dLNot Tested
    Functional SensitivityNot Specified< 20 mg/dL on AU400/400e, AU600/640/640e & AU2700/5400

    Study Proving Device Meets Acceptance Criteria:

    The study described is a comparative analysis against a legally marketed predicate device (Olympus (OSR6X46) IgM Reagent, K950900). The manufacturer carried out various analytical performance tests, including precision, assay range determination, prozone effect evaluation, method comparison (linear regression), and interference studies. The data presented demonstrates that the new device's analytical performance is substantially equivalent to the predicate device, or in some cases, improved (e.g., higher prozone limit).

    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size:
      • Method Comparison: N = 107 (for linear regression). This represents 107 human serum/plasma samples analyzed.
      • Precision: The table lists 3 samples for each instrument type (AU400/400e, AU600/640/640e, AU2700/5400). It's implied that these samples were run multiple times to calculate the Total CV%.
      • Interfering Substances: The exact number of samples for each interference study is not specified, but the data indicates the upper limit of interferent concentration at which interference remained below a specified threshold.
    • Data Provenance: The document states the intended use is for "human serum and plasma." The country of origin for the data is not specified in this summary. The studies appear to be prospective analytical performance evaluations performed by the manufacturer as part of the 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This is not applicable for this type of IVD device. The ground truth for IgM measurement is established through quantitative laboratory methods, not by expert interpretation in an AI context. The predicate device's measurements serve as the reference for method comparison.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    Not applicable. Adjudication methods are relevant for subjective interpretations (e.g., radiology reads) where discrepancies between experts need to be resolved. This device measures a quantitative analyte.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is an IVD reagent and not an AI-assisted diagnostic imaging or interpretation device.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    This is not an AI algorithm. The device is a chemical reagent used in a laboratory analyzer. Its performance is inherently "standalone" in the sense that it directly measures IgM concentration, without human interpretive input for the measurement itself, but operated by laboratory personnel.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" a.k.a. reference method for the quantitative determination of IgM in the method comparison study is the predicate device (Olympus (OSR6X46) IgM Reagent, K950900), which is a legally marketed and presumably validated method for IgM quantification. The calibrator traceability is to the International Reference Preparation CRM470, which serves as a metrological standard.

    8. The sample size for the training set:

    Not applicable. This is not an AI device that requires a training set. The "training" for such a system would involve validating the chemical formulation and assay parameters through analytical studies.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set in the context of an AI device. The development of the reagent involves chemical formulation and optimization, with performance verified against established analytical standards and reference methods.