Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K221887

    Validate with FDA (Live)

    Date Cleared
    2022-09-24

    (87 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The nitrile examination glove is intended to be worn on the hands of examiner's to prevent contamination between patient and examiner. This is a single-use, powder-free, non-sterile device.

    Device Description

    The propose devices is powder free nitrile patient examination gloves, provided as non-sterile and disposable device. The proposed devices are provided with white, black and pink color. There are six sizes, extra-small, medium, large and extra-large, extra extra-large for optional.

    AI/ML Overview

    This is a 510(k) summary for a medical device (Nitrile Examination Gloves), not for an AI/ML powered device, therefore, the requested information elements related to AI/ML device studies (such as sample size for test set, data provenance, number of experts for ground truth, adjudication method, MRMC study, standalone performance, training set sample size, and ground truth establishment for training set) are not applicable.

    The document describes non-clinical testing performed to demonstrate substantial equivalence to a predicate device.

    Here's the table of acceptance criteria and reported device performance based on the provided document:

    Acceptance Criteria and Reported Device Performance

    Test MethodPurposeAcceptance CriteriaReported Device PerformanceResult
    ASTM D6319 - Physical Dimensions Test
    Length (XS)$\ge$ 220 mm226 mmPass
    Width (XS)70 $\pm$ 10 mm73 mmPass
    Length (S)$\ge$ 220 mm225 mmPass
    Width (S)80 $\pm$ 10 mm85 mmPass
    Length (M)$\ge$ 230 mm235 mmPass
    Width (M)95 $\pm$ 10 mm98 mmPass
    Length (L)$\ge$ 230 mm234 mmPass
    Width (L)110 $\pm$ 10 mm115 mmPass
    Length (XL)$\ge$ 230 mm235 mmPass
    Width (XL)120 $\pm$ 10 mm123 mmPass
    Length (XXL)$\ge$ 230 mm235 mmPass
    Width (XXL)$\ge$ 120 mm124 mmPass
    Thickness
    Finger Thickness$\ge$ 0.08 mmX-Small: 0.104 mm; Small: 0.106 mm; Medium: 0.104 mm; Large: 0.105 mm; X-Large: 0.106 mm; XX-Large: 0.105 mmPass
    Palm Thickness$\ge$ 0.05 mmX-Small: 0.059 mm; Small: 0.060 mm; Medium: 0.060 mm; Large: 0.062 mm; X-Large: 0.061 mm; XX-Large: 0.058 mmPass
    Physical Properties (Before Aging)
    Tensile Strength$\ge$ 14 MPaLot1: 17.6 MPa; Lot2: 18.5 MPa; Lot3: 18.2 MPaPass
    Ultimate Elongation$\ge$ 500%Lot1: 552%; Lot2: 573%; Lot3: 583%Pass
    Physical Properties (After Aging)
    Tensile Strength$\ge$ 14 MPaLot1: 19.8 MPa; Lot2: 20.3 MPa; Lot3: 20.1 MPaPass
    Ultimate Elongation$\ge$ 500%Lot1: 521%; Lot2: 518%; Lot3: 526%Pass
    ASTM D5151 - Freedom from PinholesFreedom from pinholesMeet the requirements of ASTM D5151 Test for AQL 2.5Lot1: $\le$ AQL 2.5; Lot2: $\le$ AQL 2.5; Lot3: $\le$ AQL 2.5Pass
    ASTM D6124 - Powder ResiduePowder Residue$\le$ 2.0 mgLot1: 1.2 mg; Lot2: 1.5 mg; Lot3: 1.2 mgPass
    ISO 10993-10 - IrritationTo determine if the finished device material is an irritantNon-irritatingUnder the conditions of the study, not an irritantPass
    ISO 10993-10 - SensitizationTo determine if the finished device material is a sensitizerNon-sensitizingUnder conditions of the study, not a sensitizerPass
    ISO 10993-11 - Systemic ToxicityTo determine if the finished device material extracts pose a systemic toxicity concernNon-acute systemic toxicityUnder conditions of the study, did not show acute systemic toxicity in vivoPass

    Study Details

    1. Sample size used for the test set and the data provenance: Not applicable. These are non-clinical bench tests on physical products, not an AI/ML test set. The tests were performed on "Lots" of the gloves (Lot1, Lot2, Lot3). The specific sample size for each test within these lots is not specified in this document but implied to be sufficient for meeting the ASTM/ISO standards. The data provenance is from non-clinical testing of the manufacturing output.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as these are objective physical and chemical tests, not requiring expert ground truth in the context of AI/ML evaluation.
    3. Adjudication method for the test set: Not applicable. The results are based on objective measurements against defined standards.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This concerns a physical medical device (gloves), not an AI/ML powered device.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This concerns a physical medical device (gloves), not an AI/ML powered device.
    6. The type of ground truth used: For physical/chemical properties (e.g., length, thickness, tensile strength, freedom from pinholes, powder residue), the "ground truth" is the quantitative measurement obtained through standardized test methods (ASTM and ISO standards) and compared against pre-defined numerical or qualitative acceptance criteria. For biocompatibility tests (irritation, sensitization, systemic toxicity), the ground truth is established by the outcome of the biological assays as per ISO 10993 standards.
    7. The sample size for the training set: Not applicable. This is not an AI/ML device.
    8. How the ground truth for the training set was established: Not applicable. This is not an AI/ML device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1