Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K140571

    Validate with FDA (Live)

    Date Cleared
    2014-05-15

    (70 days)

    Product Code
    Regulation Number
    876.5820
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NxStage PureFlow SL module is an optional accessory to the NxStage System One™ that prepares dialysate for use during hemodialysis, as prescribed by the physician.

    Device Description

    The NxStage PureFlow SL module is an optional accessory to the NxStage System One used to prepare water for hemodialysis that meets ANSI/AAMI/ISO 13959:2009 and proportion it with dialysate concentrate to produce dialysate per ANSI/AAMI/ISO 11663:2009. The PureFlow SL consists of the Control Unit (CU), the water Pre-Treatment Unit, the optional OPTA Kit, the Purification Pack (PAK), and the Dialysate Sack (SAK) with Dialysate Concentrate.

    AI/ML Overview

    Here's an analysis of the provided text regarding acceptance criteria and the supporting study:

    The provided document is a 510(k) Summary for a device modification, not a comprehensive study report with detailed performance data. Therefore, many of the requested details about specific acceptance criteria, study sizes, ground truth establishment, and MRMC effectiveness are not explicitly stated. The document focuses on demonstrating substantial equivalence to a predicate device based on design control requirements and general performance verification.

    Here's a breakdown of the available information structured as requested:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not provide a table with specific quantitative acceptance criteria or detailed numerical performance results. Instead, it makes a general statement about meeting design specifications and being suitable for its intended use.

    However, based on the Technological Characteristics Comparison Table (Table 1) and the description, we can infer some qualitative "acceptance criteria" through comparison with the predicate device and relevant standards:

    Feature/Criteria (Inferred)Reported Device Performance
    Indication for UseSame as predicate device (NxStage PureFlow SL module is an optional accessory to the NxStage System One that prepares dialysate for use during hemodialysis, as prescribed by the physician).
    Water purification technologySame as predicate device (Deionization).
    UltrafiltrationSame as predicate device (Redundant ultrafiltration replaced within 12 weeks).
    Water QualityMeets or exceeds requirements per ANSI/AAMI/ISO 26722:2009 (Water treatment equipment for hemodialysis applications).
    Proportioning methodSame as predicate device (Mixed to use Batch via volumetric dosing of purified water with specified amount of electrolyte concentrate).
    Conductivity MeasurementSame as predicate device (System measures final batch conductivity prior to each treatment per ANSI/AAMI/ISO 26722:2009).
    Dialysate ConcentrateSame as predicate device (Standard hemodialysis concentrate per ANSI/AAMI/ISO 13958:2009).
    PackagingSame as predicate device (Packaged in flexible LLDPE bags with standard luer lock connector, ranging in size from 20L to 60L).
    BufferSame as predicate device (Lactate).
    Mix-to-Use TimeSame as predicate device (96-hours).
    BioburdenSame as predicate device (AAMI/ANSI/ISO 11663: <100 CFU/ml).
    EndotoxinSame as predicate device (AAMI/ANSI/ISO 11663: <0.5 EU/ml).
    Batch sizeSame as predicate device (Range – 20 to 60 L; 40, 50 and 60L currently available).
    Fluid warming MethodSame as predicate device (Integrated fluid warming pad).
    Validation of Design Control RequirementsCertification of compliance to 21 CFR §820.30 Design Control Requirements.
    General PerformancePredetermined acceptance criteria were met. Device is suitable for labeled indications for use.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not specified. The document states "Performance, verification and validation testing was conducted," but no specific sample sizes for this testing are provided.
    • Data Provenance: Not specified. The testing is referred to as "Non-Clinical Test/Performance Testing - Bench," implying internal lab testing, but no country of origin or whether it was retrospective or prospective is mentioned.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable/Not specified. This is a device modification for a hemodialysis system, not an imaging or diagnostic AI device where expert reads establish ground truth. The "ground truth" for this type of device relates to its adherence to engineering specifications and performance standards.

    4. Adjudication Method for the Test Set

    Not applicable/Not specified. Since expert consensus on ground truth is not relevant here, no adjudication method would be mentioned.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a component of a hemodialysis system and does not involve AI or human readers for diagnostic interpretation.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable in the context of an AI algorithm. The device's performance is inherently "standalone" in its function of preparing dialysate, but this is mechanical/chemical performance, not AI.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance relies on engineering specifications, design control requirements, and compliance with established industry standards (e.g., ANSI/AAMI/ISO 13959:2009, ANSI/AAMI/ISO 11663:2009, ANSI/AAMI/ISO 26722:2009, ANSI/AAMI/ISO 13958:2009). The device's performance against these standards constitutes its "ground truth."

    8. The Sample Size for the Training Set

    Not applicable. This device is not an AI/ML device that requires a training set.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set for this type of device.

    Study Proving Device Meets Acceptance Criteria:

    The study proving the device meets acceptance criteria is described as "Performance, verification and validation testing" (Section F). This testing was conducted to characterize the performance of the modified device. The submission states that "predetermined acceptance criteria were met. Results of this testing have documented that the proposed NxStage PureFlow SL is substantially equivalent to the predicate device and is suitable for the labeled indications for use."

    The entire submission is a "Special 510(k) Device Modification" which means the focus is on certifying compliance to 21 CFR §820.30 Design Control Requirements and demonstrating that the modifications do not alter the safely and effectiveness, making it substantially equivalent to the cleared predicate device (K111174). The study is essentially the internal design verification and validation activities conducted by NxStage Medical, Inc.

    Ask a Question

    Ask a specific question about this device

    K Number
    K111174

    Validate with FDA (Live)

    Date Cleared
    2011-09-19

    (145 days)

    Product Code
    Regulation Number
    876.5820
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NxStage PureFlow SL module is an optional accessory to the NxStage System One that prepares dialysate for use during hemodialysis, as prescribed by the physician.

    Device Description

    The NxStage PFSL module is an optional accessory to the NxStage System One that is used to treat water for hemodialysis per ANSI/AAMI/ISO 13959:2009 and ANSI/AAMI/ISO 26722:2009 and proportion it with dialysate concentrate to produce dialysate per ANSI/AAMI/ISO 11663:2009. The PFSL module consists of the Control Unit (CU), the water Pre-Treatment Unit, the optional OPTA Kit, the Purification Pack (PAK), and the Dialysate Sack (SAK) with Dialysate Concentrate.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the NxStage PureFlow SL Subsystem, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criterion (Standard)Reported Device Performance (Result for 96-hour MTU)
    Bioburden: < 100 CFU/ml (ANSI/AAMI/ISO 11663:2009)Met criterion (< 100 CFU/ml)
    Endotoxin: < 0.5 EU/ml (ANSI/AAMI/ISO 11663:2009)Met criterion (< 0.5 EU/ml)
    Mix-to-Use (MTU) DurationExtended to 96-hours (from 72-hours)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated. The document mentions "validation batches of dialysate" were used for testing.
    • Data Provenance: The study was conducted by NxStage Medical, Inc., with manufacturing occurring in Fresnillo, Zacatecas, Mexico. The context suggests this was an internal validation study to support a device modification. The study is retrospective in the sense that it evaluates the modified device's performance against established standards, but it's prospective for the testing itself to confirm the extended MTU time.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • This type of study does not involve human expert adjudication for image interpretation or diagnosis. The ground truth is established via objective laboratory measurements against defined standards.

    4. Adjudication Method

    • Not applicable. This is a technical performance study, not a diagnostic accuracy study relying on human interpretation.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Status

    • Not applicable. This is a performance validation study for a medical device's technical specifications, not a comparative effectiveness study involving human readers and AI. The device is a "proportioning hemodialysis system and accessories," not an AI diagnostic tool.

    6. Standalone Performance Study Status

    • Yes, a standalone performance study was done. The document states: "Performance and verification and validation testing was conducted to characterize performance of the proposed device. This included stability, bioburden and endotoxin testing on validation batches of dialysate to support the extension of the MTU duration to 96-hours." This refers to testing the device itself, without human intervention in the performance evaluation.

    7. Type of Ground Truth Used

    • Objective laboratory measurements and compliance with industry standards: The ground truth for bioburden and endotoxin levels was established by laboratory testing of dialysate samples and comparison against the specific limits set by ANSI/AAMI/ISO 11663:2009. The MTU duration extension was validated through stability testing.

    8. Sample Size for the Training Set

    • Not applicable. This device is not an AI/ML model that requires a "training set" in the traditional sense. It's a hardware device with defined performance parameters.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable, as there is no training set for an AI/ML model. The "ground truth" (i.e., the acceptable performance criteria) for the device's technical specifications was established by recognized industry standards (ANSI/AAMI/ISO 11663:2009, ANSI/AAMI/ISO 13959:2009, and ANSI/AAMI/ISO 26722:2009).
    Ask a Question

    Ask a specific question about this device

    K Number
    K080919

    Validate with FDA (Live)

    Date Cleared
    2008-10-15

    (197 days)

    Product Code
    Regulation Number
    876.5820
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NxStage PureFlow SL module is an optional accessory to the NxStage System One that prepares dialysate for use during hemodialysis, as prescribed by the physician.

    Device Description

    The NxStage PFSL module is an optional accessory to the NxStage System One that is used to treat water for hemodialysis and proportion it with dialysate concentrate to produce dialysate per ANSI/AAMI RD52:2004 and ANSI/AAMI RD62: 2006. The PFSL module consists of the Control Unit (CU), the water Pre-Treatment Unit, the Purification Pack (PAK), and the Dialysate Sack (SAK) with Dialysate Concentrate.

    AI/ML Overview

    The provided text describes a Special 510(k) submission for a device modification, which primarily focuses on demonstrating substantial equivalence to a predicate device. This type of submission relies on design control requirements and performance testing to ensure the modified device meets design specifications and remains as safe and effective as the original. As such, the information you've requested regarding detailed acceptance criteria, specific study designs (like MRMC or standalone performance), sample sizes for test and training sets, and expert qualifications for ground truth establishment is not typically present in a 510(k) summary for such a modification.

    The 510(k) summary focuses on demonstrating that the NxStage PureFlow SL (PFSL) module with modifications is substantially equivalent to the legally marketed predicate device (cleared through K060296).

    Here's an analysis based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a specific table of acceptance criteria with detailed reported performance metrics. Instead, it states:

    "Performance, verification and validation testing was conducted to characterize performance of the proposed PFSL to provide a basis of comparison to the predicate device as all features are not identical. Results of this testing have documented that the proposed PFSL module is substantially equivalent to the predicate device and is suitable for the labeled indications for use."

    This indicates that internal performance specifications were met to ensure substantial equivalence, but the specific numerical acceptance criteria and results are not disclosed in this summary. The general acceptance criterion is that the modified device performs comparably to the predicate device and meets its design specifications, preparing dialysate per ANSI/AAMI RD52:2004 and ANSI/AAMI RD62: 2006.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify the sample size used for any test set or the data provenance. The testing mentioned is "Performance, verification and validation testing," which typically refers to bench testing rather than clinical study data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable and not provided. The study described is a non-clinical, bench-testing comparison for substantial equivalence, not a clinical study requiring expert-established ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable and not provided. The study described is a non-clinical, bench-testing comparison for substantial equivalence.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. This device is an accessory for hemodialysis, and the evaluation is non-clinical, focusing on technological characteristics and performance rather than human reader interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    A "standalone" performance evaluation in the context of an algorithm or AI would not apply here. However, the non-clinical "Performance, verification and validation testing" can be considered a standalone evaluation of the device's functional performance against its design specifications and the predicate device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device modification is its design specifications and compliance with relevant industry standards (ANSI/AAMI RD52:2004 and ANSI/AAMI RD62: 2006) for preparing dialysate. The performance was characterized to demonstrate substantial equivalence to the predicate device.

    8. The sample size for the training set

    This information is not applicable and not provided. This is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established

    This information is not applicable and not provided. This is not an AI/ML device that requires a training set.

    In summary:

    This 510(k) focuses on a "Special 510(k) Device Modification" which primarily relies on demonstrating compliance with 21 CFR §820.30 Design Control Requirements and conducting performance, verification, and validation (bench) testing to confirm that the modified device remains substantially equivalent to its predicate device. This type of submission does not typically involve the detailed clinical study metrics, expert consensus, or AI-specific evaluation criteria often associated with novel medical devices or AI-powered diagnostics.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1