510(k) Data Aggregation

    K Number
    K162086

    Validate with FDA (Live)

    Device Name
    MultiStim ECO
    Date Cleared
    2017-01-06

    (163 days)

    Product Code
    Regulation Number
    868.2775
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Pajunk's MultiStim ECO is intended for nerve stimulation during anaesthesia delivery; first for identification of peripheral nerves and second for localizations. This device is indicated for adults only.

    Device Description

    PAJUNK® GmbH Medizintechnologie is submitting this 510(k) for the MultiStim ECO handheld, battery powered peripheral nerve stimulator.
    The MultiStim ECO is a nerve stimulator that is suitable for stimulation technique. It is characterized in particular by the following properties:

    • small, compact device complies with the requirements of combined procedures -
    • -Simple operation - manual settings are limited to the parameters proven in practice
    • -No additional patient cable - cannula is directly connected to the device
      The MultiStim ECO settings are adapted to standard applications as well as the intuitively operated keyboard enable one-hand operation. A keystroke is all that is necessary for activation after switching on - an additional function selection is not required.

    Basic functions/ features:
    Fixed defined settings:
    The frequency is predefined at 1 Hz and the stimulation pulse width at 0.1 ms.
    Due to safety reasons, stimulation is first activated by a touch of the stimulation button.
    Variation of the current strength
    Depending on the special requirements of an application, the current strength can be gradually set on a six-level amplitude scale from 0.2 to 2.0 mA. Arrow buttons are used for the plus/minus key.
    Fixation by an adhesive electrode
    The MultiStim ECO has a pushbutton connection on the lower side of the device and is clicked directly on the adhesive electrode. The stimulateable cannula is connected directly to the MultiStim ECO.

    AI/ML Overview

    The provided text is a 510(k) summary for the MultiStim ECO, a peripheral nerve stimulator, seeking substantial equivalence to the MultiStim SENSOR. The document focuses on demonstrating that the new device performs as effectively as the predicate device.

    Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

    Acceptance Criteria and Reported Device Performance

    The core of the performance acceptance criteria appears to be the congruence of output graphs between the MultiStim ECO (subject device) and the MultiStim SENSOR (predicate device) across various current (mA) settings. The device's performance is tested against specific current outputs and the corresponding voltage (mV).

    Table of Acceptance Criteria and Reported Device Performance:

    Current (mA)Expected Voltage (mV) for CongruenceReported Device Performance (MultiStim ECO Output Voltage)Outcome of Comparison
    0.2Approx. 202202Congruent
    0.5Approx. 500500Congruent
    0.7Approx. 712712Congruent
    1.0Approx. 10001000Congruent
    1.5Approx. 15201520Congruent
    2.0Approx. 20402040Congruent

    Additional Implicit Acceptance Criteria:

    • Identical Intended Use: The MultiStim ECO must have the same intended use as the MultiStim SENSOR. This is explicitly stated and claimed to be identical, with a minor change in wording ("percutaneous" removed as obsolete).
    • Identical Basic Technical Description Relevant to Clinical Use: The fundamental method of impulse generation (quartz crystal) and the non-software hardware solution are presented as core to this.
    • Compliance with Recognized Standards: The device must meet a comprehensive list of IEC/EN standards for medical electrical equipment, including safety, electromagnetic compatibility, usability, and specific requirements for nerve and muscle stimulators. (See the list of standards provided on pages 10-11).
    • Physical and Functional Equivalence (where applicable): Various characteristics like dimensions, FDA classification, product code, biocompatibility, power supply, pulse form, and protection type are compared, and deviations are rationalized as either equivalent or improvements/simplifications that do not negatively impact safety or effectiveness.

    Study Details Proving Acceptance Criteria

    1. Sample Size and Data Provenance:

      • Test Set Sample Size: The document does not specify a "sample size" in terms of number of devices tested. It describes a bench test where "the output signals of values which are adjustable at the MultiStim ECO have been compared to those with the same settings at the MultiStim SENSOR." This implies testing of at least one MultiStim ECO device against a MultiStim SENSOR for the specified current settings. There is no indication of a large-scale clinical trial or patient data.
      • Data Provenance: The data is generated from bench testing ("the desired signal combination was adjusted and visualized with an oscilloscope"). The location of the test is not specified, but the manufacturer is based in Germany. The data is prospective in the sense that it was generated for this submission to verify performance.
    2. Number of Experts and Qualifications for Ground Truth:

      • This submission does not involve experts establishing ground truth in the typical sense of clinical image review or diagnostic performance. Instead, the "ground truth" is established by the engineering specifications and the measured output of the predicate device (MultiStim SENSOR), which is itself a legally marketed device. The comparison is objective, based on instrument readings (oscilloscope).
    3. Adjudication Method for the Test Set:

      • Not applicable. This was a bench test with objective measurements and comparison to a known standard (predicate device output).
    4. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study:

      • No MRMC study was done. This device is a nerve stimulator, not an imaging AI or diagnostic aid that would typically involve human readers interpreting results. The comparison is at the engineering/performance level, not human-in-the-loop effectiveness.
    5. Standalone Performance:

      • Yes, a standalone (algorithm only without human-in-the-loop performance) was done. The core performance testing involved measuring the output signals of the MultiStim ECO device in isolation and comparing them to those of the predicate device, using an oscilloscope. This is purely a device performance assessment independent of human interpretation or use in a clinical setting.
    6. Type of Ground Truth Used:

      • The "ground truth" is the measured output characteristics (voltage at specific current settings and pulse shape) of the predicate device (MultiStim SENSOR), along with compliance to relevant medical device standards (e.g., consistency of square pulse form, 1Hz frequency, 0.1ms pulse width). It's an engineering/device specification ground truth, not a clinical outcome or pathology report.
    7. Training Set Sample Size:

      • Not applicable. This is a hardware device without a "training set" in the context of machine learning or AI algorithms. The device's design is based on established electrical engineering principles and the performance of its predicate.
    8. How Ground Truth for Training Set was Established:

      • Not applicable, as there is no "training set." The design and "training" (if one could use the term loosely) of the device come from established medical device engineering practices and the existing predicate device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K132137

    Validate with FDA (Live)

    Device Name
    MULTISTIM
    Date Cleared
    2013-09-12

    (63 days)

    Product Code
    Regulation Number
    890.5850
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    • I. Relief of chronic intractable pain, acute post traumatic pain or acute post surgical pain (Interferential, TNS and Microcurrent waveforms);
      1. Relaxation of muscle spasm (all waveforms except Microcurrent);
      1. Prevention or retardation of disuse atrophy (all waveforms except Microcurrent);
      1. Increasing local blood flow (all waveforms except Microcurrent);
      1. Muscle re-education (all waveforms except Microcurrent);
      1. Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis (all waveforms except Microcurrent); and
      1. Maintaining or increasing range of motion (all waveforms except Microcurrent).
    Device Description

    Not Found

    AI/ML Overview

    This FDA 510(k) clearance letter for the Metron Multi-Stim Model MS-100A does not contain any information regarding acceptance criteria or a study proving the device meets those criteria.

    The document is a clearance letter, which means the FDA has determined the device is substantially equivalent to legally marketed predicate devices. This determination is based on a comparison to existing devices, and typically involves demonstrating similar technological characteristics and performance specifications. However, the letter itself does not provide the details of those performance specifications or the studies conducted to demonstrate them.

    Therefore, I cannot fulfill your request for the specific information regarding acceptance criteria and the study proving the device meets them because that information is not present in the provided text.

    The document states: "We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices..." This indicates a comparison to predicate devices, but the specifics of how that substantial equivalence was demonstrated (i.e., performance criteria and study results) are not elaborated upon in this letter.

    Ask a Question

    Ask a specific question about this device

    K Number
    K011308

    Validate with FDA (Live)

    Device Name
    MULTISTIM
    Manufacturer
    Date Cleared
    2001-11-29

    (213 days)

    Product Code
    Regulation Number
    868.2775
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1