Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K230011

    Validate with FDA (Live)

    Device Name
    Mpact Extension
    Date Cleared
    2023-09-20

    (260 days)

    Product Code
    Regulation Number
    888.3358
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Mpact implants are designed for cementless use in total hip arthroplasty in primary or revision surgery. The patient should be skeletally mature.

    The patient's condition should be due to one or more of:

    · Severely painful and/or disabled joint: as a result of osteoarthritis, rheumatic arthritis, rheumatoid arthritis or psoriatic arthritis, congenital hip dysplasia, ankylosing spondylitis.

    • · Avascular necrosis of the femoral head.
    • · Acute traumatic fracture of the femoral head or neck.

    · Failure of previous hip surgery: joint reconstruction, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement where sufficient bone stock is present

    Device Description

    The Mpact Extension implants are a line extension to Medacta's Mpact (K103721) and Mpact Extensions (K122641, K132879, K183582 and K200391) Systems and are designed to be used with the Medacta Total Hip Prosthesis System. Specifically, the devices subject of this submission are:

    • Liners sizes 22/A and 28/A of each design type (flat, hooded, offset +4mm and face changing . +10°);
    • Acetabular Shells size Ø44 of each design type (No-hole, Two-holes and Multi-holes); ●
    • Acetabular Shells sizes from Ø42T to Ø58T of each design type (No-hole, Two-holes and Multi-● holes) allowing to be coupled with an increase size of liner.

    The Mpact Extension implants are sterile implantable devices intended to be used during Total Hip Arthroplasty: the acetabular shells together with their respective liners are used to replace the acetabulum.

    Analogously to the predicate devices, the subject acetabular shells are made of Ti6Al4V according to ASTM F136-13 and Ti coated according to ASTM F1580-18, while the subject liners are made of High-Cross ultra-high molecular weight polyethylene (UHMWPE).

    AI/ML Overview

    The provided text does not describe an AI medical device, but rather an orthopedic implant (hip joint prosthesis). Therefore, the questions regarding acceptance criteria, study details, expert qualifications, adjudication methods, AI assistance, standalone performance, and training set information are not applicable to this submission.

    The FDA 510(k) summary for the Medacta International S.A. Mpact Extension (K230011) outlines the device, its indications for use, comparison to predicate devices, and performance data for substantial equivalence.

    Here's a summary of the information provided in relation to the device's performance, but not in the context of AI:

    1. Acceptance Criteria and Reported Device Performance:

      The document does not explicitly present a table of acceptance criteria with corresponding device performance for an AI model. Instead, it lists the following non-clinical performance tests conducted to support substantial equivalence for the orthopedic implant:

      Acceptance Criteria (Implied by Test Type)Reported Device Performance
      Acetabular cup fatigue assessmentPerformed (Implicitly met industry standards for fatigue strength)
      Acetabular cup deformation assessmentPerformed (Implicitly met industry standards for resistance to deformation)
      PE liners stability (Axial Disassembly Force) per ASTM F1820Performed (Implicitly met ASTM F1820 standard)
      Evaluation of impingement riskPerformed (Rationale provided, implicitly acceptable risk)
      Evaluation of Range of Motion (ROM) per EN ISO 21535Performed (Implicitly met EN ISO 21535 requirements)
      PE HC liner wear testRationale for wear test provided (This suggests either the wear characteristics were considered equivalent to predicates, or specific wear tests were deemed not necessary based on similarity to previously cleared devices, or a rationale was provided to explain why testing demonstrated acceptable wear. The document states "Rationale for wear test" which implies a justification was given, rather than a direct report of a wear rate.)
      Bacterial Endotoxin Test (LAL test) per European Pharmacopoeia §2.6.14Performed (Implicitly met pyrogenicity standards)
      Pyrogen test per USP chapter <151>Performed (Implicitly met pyrogenicity standards; note that the devices are not labeled as non-pyrogenic or pyrogen-free)
      BiocompatibilityAssessed (Implicitly met biocompatibility standards)
      Shelf-Life evaluationPerformed (Implicitly demonstrated acceptable shelf-life)
    2. Sample Size for Test Set and Data Provenance: Not applicable as this is not an AI device. The performance tests for the orthopedic implant are laboratory-based mechanical and biological tests, not statistical analyses using a test set of medical data.

    3. Number of Experts and Qualifications: Not applicable. Ground truth for an AI device involves expert annotations; for an orthopedic implant, it's about meeting engineering and biological standards.

    4. Adjudication Method: Not applicable.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study: No clinical studies were conducted, and this is not a device that would typically involve human-in-the-loop AI assistance.

    6. Standalone Performance Study: Not applicable. This refers to AI algorithm performance without human interaction. The device is a physical implant.

    7. Type of Ground Truth Used: Not applicable in the AI sense. For the orthopedic implant, "ground truth" would be defined by established engineering standards (e.g., ASTM, ISO) for material properties, mechanical performance, and biocompatibility.

    8. Training Set Sample Size: Not applicable. This device is not an AI algorithm requiring a training set.

    9. Ground Truth for Training Set Establishment: Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200391

    Validate with FDA (Live)

    Device Name
    MPACT Extension
    Date Cleared
    2021-04-28

    (435 days)

    Product Code
    Regulation Number
    888.3358
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Mpact implants are designed for cementless use in total hip arthroplasty in primary or revision surgery. The patient should be skeletally mature.

    The patient's condition should be due to one or more of:

    · Severely painful and/or disabled joint: as a result of osteoarthritis, post-traumatic arthritis, rheumatoid arthritis or psoriatic arthritis, congenital hip dysplasia, ankylosing spondylitis.

    • · Avascular necrosis of the femoral head.
    • · Acute traumatic fracture of the femoral head or neck.

    · Failure of previous hip surgery: joint reconstruction, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement where sufficient bone stock is present.

    The Mpact 3D Metal Implants are designed for cementless use in total hip arthroplasty in primary or revision surgery. The patient should be skeletally mature.

    The patient's condition should be due to one or more of:

    · Severely painful and/or disabled joint: as a result of osteoarthritis, rheumatic arthritis, rheumatoid arthritis or psoriatic arthritis, congenital hip dysplasia, ankylosing spondylitis

    • Avascular necrosis of the femoral head

    · Acute traumatic fracture of the femoral head or neck

    · Failure of previous hip surgery: joint reconstruction, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement where sufficient bone stock is present.

    The Augments 3D Metal are intended to be used in combination with the Mpact 3D Metal Multi-hole acetabular cup in hip replacement surgeries.

    The Augments 3D Metal are indicated in cases of:

    • Congenital dysplasia.
    • · Acetabular fractures.

    · Revision of previous implants in presence of insufficient bone quality or seriously altered bone structures.

    The Versafitcup CC Trio and the Versafitcup CC Trio No-Hole are designed for cementless use in total hip arthroplasty in primary or revision surgery. The patient should be skeletally mature.

    The patient's condition should be due to one or more of:

    • · Severely painful and/or disabled joint: as a result of osteoarthritis, post-traumatic arthritis or
      psoriatic arthritis, congenital hip dysplasia, ankylosing spondylitis

    • Avascular necrosis of the femoral head

    • · Acute traumatic fracture of the femoral head or neck

    · Failure of previous hip surgery: joint reconstruction, arthrodesis, hemiarthroplasty, surface replacement arthroplastv, or total hip replacement where sufficient bone stock is present.

    Device Description

    The MPACT Extension implants subject of this submission are twelve (12) sizes of Cancellous Bone Screws.

    The subject devices are sterile implantable devices intended to be used during Total Hip Arthroplasty.

    Analogously to the predicate Cancellous Bone Screws flat head (K103721 and K132879), the subject Cancellous Bone Screws, are made of titanium alloy (Ti6-A14-V) according to ISO 5832-3:2016 Implants for Surgery - Metallic materials - Part 3: Wrought titanium 6-aluminum 4vanadium allov. As the predicate devices, they have been designed with a thread diameter of 6.5 mm to provide fixation of implants to the cancellous bone. They can be coupled with implants that have dedicated fixation holes with a spherical seat shape of Ø8mm.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from the FDA regarding a medical device called "MPACT Extension," which consists of Cancellous Bone Screws for total hip arthroplasty.

    Based on the provided text, a conventional "acceptance criteria" and "study proving the device meets the acceptance criteria" in the context of an AI/Software as a Medical Device (SaMD) study (as implied by the detailed breakdown requested in the prompt) were not performed. The document describes the acceptance criteria and study for a physical implantable medical device, specifically cancellous bone screws, not a software or AI product.

    Therefore, many of the requested points, such as AI-specific acceptance criteria, multi-reader multi-case studies, ground truth establishment for training sets, etc., are not applicable to this document.

    However, I can extract the relevant information provided for the device described:


    1. A table of acceptance criteria and the reported device performance (for the hardware device):

    The document does not explicitly present a table of numerical "acceptance criteria" for the performance of the cancellous bone screws in the way one might for a software or AI algorithm (e.g., target accuracy, sensitivity). Instead, acceptance is based on demonstrating substantial equivalence to predicate devices through various tests and material properties. The performance is assessed through meeting standards and showing compatibility.

    CategoryAcceptance Criteria (Implied by Regulatory Standards)Reported Device Performance
    MaterialMade of titanium alloy (Ti6-A14-V) conforming to ISO 5832-3:2016."Made of titanium alloy (Ti6-A14-V) according to ISO 5832-3:2016 Implants for Surgery - Metallic materials - Part 3: Wrought titanium 6-aluminum 4vanadium alloy."
    Fixation CapabilityDesigned for fixation of implants to the cancellous bone."Designed with a thread diameter of 6.5 mm to provide fixation of implants to the cancellous bone."
    CompatibilityCompatible with implants having dedicated fixation holes with a spherical seat shape of Ø8mm."They can be coupled with implants that have dedicated fixation holes with a spherical seat shape of Ø8mm."
    Sterility & SafetyMust be sterile, biocompatible, and non-pyrogenic. Shelf-life established.BIOCOMPATIBILITY evaluation: according to ISO 10993. SHELF-LIFE testing: according to ISO 11607-1 and ISO 11607-2. PYROGENICITY: Bacterial Endotoxin Test (LAL test) was conducted according to European Pharmacopoeia §2.6.14 (equivalent to USP chapter <85>). Pyrogen test according to USP chapter <151>. Sterility: "The subject devices are sterile implantable devices." Biocompatibility, sterility, and shelf life are listed as shared characteristics with predicate devices.
    Design ValidationDesign must be validated for functionality and compatibility.DESIGN VALIDATION: "Cancellous Bone Screw Ø6.5 design Validation rationale A1 item 3" and "Cancellous Bone Screw Ø6.5 acetabular shell compatibility Validation rationale A2 item 3."

    2. Sample size used for the test set and the data provenance:

    • Sample Size for Test Set: Not specified in the document in terms of traditional "test sets" for a software algorithm. The "test set" for this physical device would refer to the number of screws tested during design validation, biocompatibility, and sterility testing. This number is not provided.
    • Data Provenance: Not applicable in the context of a software data set (e.g., country of origin, retrospective/prospective). The document describes non-clinical studies performed on the device itself.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This pertains to a physical medical device, not a software/AI product requiring expert ground truth for image interpretation or similar. The "ground truth" for this device would be its physical and chemical properties and mechanical performance as evaluated through standard laboratory testing.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. This is a method for reconciling differences in expert annotations for AI/software ground truth.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not done. The device is a physical implant, not an AI or diagnostic software tool.
    • Effect size: Not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This refers to AI algorithm performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For this physical device, the "ground truth" is established through objective laboratory testing against recognized standards (e.g., ISO for materials, biocompatibility tests, pyrogenicity tests). It's about demonstrating the device's physical properties and safety characteristics meet established benchmarks, rather than interpretation of data by experts.

    8. The sample size for the training set:

    • Not applicable. This refers to an AI training set.

    9. How the ground truth for the training set was established:

    • Not applicable. This refers to an AI training set.

    Summary of the Study that Proves the Device Meets Acceptance Criteria (as described in the document):

    The acceptance of the MPACT Extension device (Cancellous Bone Screws) is based on demonstrating its substantial equivalence to established predicate devices (Medacta International SA, MPACT, K103721 and MPACT Extension, K132879). This substantial equivalence is supported by a series of non-clinical studies:

    • Design Validation: Specific rationales were provided for the design validation of the Ø6.5 Cancellous Bone Screw and its compatibility with acetabular shells. This ensures the device's design is adequate for its intended use.
    • Biocompatibility Evaluation: Performed according to ISO 10993, demonstrating the device is suitable for implantation without adverse biological reactions.
    • Shelf-Life Testing: Conducted according to ISO 11607-1 and ISO 11607-2, ensuring the device maintains its properties and sterility over its intended shelf life.
    • Pyrogenicity Testing: Bacterial Endotoxin Test (LAL test) per European Pharmacopoeia §2.6.14 (equivalent to USP chapter <85>) and Pyrogen test per USP chapter <151> were performed to ensure the device is not pyrogenic.

    No clinical studies were conducted for this 510(k) submission, as stated under "Clinical Studies" in Section VII. The substantial equivalence was determined based on the comparison of intended use, design, technological characteristics, and the non-clinical design validations.

    Ask a Question

    Ask a specific question about this device

    K Number
    K183582

    Validate with FDA (Live)

    Device Name
    MPACT Extension
    Date Cleared
    2019-03-21

    (90 days)

    Product Code
    Regulation Number
    888.3358
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MPACT Extension implants are designed for cementless use in total hip arthroplasty in primary or revision surgery. The patient should be skeletally mature.

    The patient's condition should be due to one or more of:

    · Severely painful and/or disabled joint: as a result of osteoarthritis, post-traumatic arthritis, rheumatoid arthritis or psoriatic arthritis, congenital hip dysplasia, ankylosing spondylitis

    · Avascular necrosis of the femoral head

    · Acute traumatic fracture of the femoral head or neck

    · Failure of previous hip surgery: joint reconstruction, arthrodesis, hemiarthrodesis, surface replacement arthroplasty, or total hip replacement where sufficient bone stock is present

    Device Description

    The MPACT implants subject of this submission are comprised of the following products:

    • 33 sizes of Offset +4mm PE liners; and
    • 32 sizes of Face-changing +10° PE liners.

    These liners are a component of a total hip joint prosthesis that is used to replace the acetabulum. The component is an inner liner made of High-Cross ultra-high molecular weight polyethylene (UHMWPE); which is inserted in the Acetabular shell.

    They are all a line extension to Medacta's MPACT (K103721), and MPACT Extensions (K122641 and K132879) Systems and are designed to be used with the Medacta Total Hip Prosthesis System. In detail, the Liners subject of the current submission are compatible with the MPACT No-Hole and Two-Hole (K132879), MPACT Multi-Hole and Rim-Hole (K132879); and the Mpact 3D Metal Acetabular Shells (K171966).

    As regards to the femoral head components, the MPACT Extension Liners can be combined with the CoCr Ball Heads (K072857 and K080885), MectaCer Biolox Option Heads (K131518), or with the MectaCer BIOLOX® Forte (K073337) or MectaCer BIOLOX® Delta Femoral Heads (K112115).

    AI/ML Overview

    The provided text describes the regulatory clearance for a medical device called "MPACT Extension" and details the non-clinical performance data used to demonstrate its substantial equivalence to predicate devices. It explicitly states that no clinical studies were conducted. Therefore, the acceptance criteria and study information related to human performance (e.g., sample sizes for test/training sets, expert ground truth, MRMC studies, standalone algorithm performance) are not applicable to this submission.

    Here's a breakdown of the relevant information from the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestStandard/GuidelineAcceptance Criteria (Implicit)Reported Device Performance (Implicit)
    Performance Tests
    Static/Dynamic Compressive Load TestingASTM F1820-13Meets requirements of ASTM F1820-13 for axial disassembly force of a modular acetabular device.Device passed the tests, supporting substantial equivalence.
    Static Lever-out TestingASTM F1820-13Meets requirements of ASTM F1820-13 for axial disassembly force of a modular acetabular device.Device passed the tests, supporting substantial equivalence.
    Range of MotionEN ISO 21535 (2007)Meets requirements of EN ISO 21535 for hip-joint replacement implants.Device passed the tests, supporting substantial equivalence.
    Biocompatibility
    Pyrogenicity (Bacterial Endotoxin Test)European Pharmacopoeia §2.6.14 (equivalent to USP chapter <85>)Meets requirements for bacterial endotoxin levels.Device passed the tests.
    Pyrogenicity (Pyrogen Test)USP chapter <151>Meets requirements for pyrogenicity determination.Device passed the tests.

    Note: The document states that testing was conducted according to written protocols with acceptance criteria based on standards. The specific numerical acceptance values or individual performance results are not explicitly listed in this summary, but the general statement "device passed the tests, supporting substantial equivalence" implies that the acceptance criteria were met for all listed non-clinical tests.

    2. Sample size used for the test set and the data provenance
    Not applicable. No human test set was used as this was a non-clinical evaluation.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    Not applicable. No human test set was used.

    4. Adjudication method for the test set
    Not applicable. No human test set was used.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done
    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. The document explicitly states: "No clinical studies were conducted."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    Not applicable. This device is a physical medical implant (hip joint component), not an algorithm or AI software for which standalone performance would be relevant.

    7. The type of ground truth used
    For the non-clinical performance tests, the "ground truth" was established by the specified industry standards (ASTM F1820-13, EN ISO 21535, European Pharmacopoeia §2.6.14, USP chapter <85>, USP chapter <151>). These standards define the methodologies and acceptable performance limits for mechanical properties and biocompatibility.

    8. The sample size for the training set
    Not applicable. No human training set was used. "Training set" typically refers to data used to train AI models; this device is a physical implant.

    9. How the ground truth for the training set was established
    Not applicable. No human training set was used.

    Ask a Question

    Ask a specific question about this device

    K Number
    K132879

    Validate with FDA (Live)

    Device Name
    MPACT EXTENSION
    Date Cleared
    2014-01-22

    (131 days)

    Product Code
    Regulation Number
    888.3358
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The hip prosthesis is designed for cementless use in total hip arthroplasty in primary or revision surgery.

    The patient should be skeletally mature.

    The patient's condition should be due to one or more of:

    .

    • Severely painful and/or disabled joint: as a result of osteoarthritis, posttraumatic . arthritis, rheumatoid arthritis or psoriatic arthritis, congenital hip dysplasia, ankylosing spondylitis.
    • Avascular necrosis of the femoral head.
    • Acute traumatic fracture of the femoral head or neck. ●
    • Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, ● hemiarthroplasty, surface replacement arthroplasty, or total hip replacement where sufficient bone stock is present.
    Device Description

    The Mpact Extension components are designed to be used with the Medacta Total Hip Prosthesis System. The Mpact Extension consists of the following components:

    Revised dimensions and labeling of the acetabular shells (Ti-6AI-4V, ASTM F136 and Ti, ASTM F1580 porous coating) cleared under K103721 and K122641 to correlate to the actual external diameter including the porous coating.

    Modified design of the hooded acetabular liners (UHMWPE ISO 5834-2 Type1 and HighCross® HXUHMWPE) cleared under K103721 and K122641. The modification of the hooded liner is specific to the external side of the anti-luxation shoulder, creating a chamfer in order to reduce the risk of psoas irritation.

    Additional sizes of the cancellous bone screws (flat head, 6.5mm diameter, Ti-6Al-4V ISO 5832-3) cleared under K103721 with lengths of 50, 55, 60, 65, and 70mm.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the "Mpact Extension" hip prosthesis. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than providing extensive clinical study data a new device might require. Therefore, the information typically requested in your prompt (e.g., sample sizes for test/training sets, number of experts for ground truth, MRMC studies, standalone performance) is largely not applicable or not provided in this specific document.

    Here's an analysis based on the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document states that performance testing was conducted "in accordance with various international standards and FDA guidance documents" and "as part of design verification to written protocols with pre-defined acceptance criteria." However, it does not explicitly list the specific numerical acceptance criteria for each test or the detailed, quantitative results against those criteria. It only provides a high-level summary that the results "met the acceptance criteria."

    Acceptance Criteria (General)Reported Device Performance (General)
    Pre-defined acceptance criteria for design verificationAll tested components and designs met the acceptance criteria.
    ASTM F543 for bone screwsTesting per ASTM F543 was completed on the bone screws, with the results meeting the acceptance criteria.
    Coating validationConducted (implied met criteria).
    Metal shell deformation resistance during impactionConducted (implied met criteria).
    Range of motionConducted (implied met criteria).
    WearConducted (implied met criteria).
    Pull-off stability of modular connection (fixed liner to shell)Conducted (implied met criteria).
    Lever-out stability of modular connection (fixed liner to shell)Conducted (implied met criteria).
    Torsional stability of modular connection (fixed liner to shell)Conducted (implied met criteria).

    2. Sample size used for the test set and the data provenance

    The document mentions that testing was conducted on the "worst case component size and option/design" for the Mpact Extension. It does not specify a numerical sample size for any of the tests, nor does it refer to a "test set" in the context of clinical data for a device performance study. The testing described is design verification, typically conducted on physical prototypes or simulations. Data provenance (country, retrospective/prospective) is not applicable here as it refers to a clinical study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable. The device underwent design verification testing against established engineering standards and internal protocols, not a clinical study requiring expert ground truth for interpretation of patient data.

    4. Adjudication method for the test set

    Not applicable, as this refers to a clinical study with image or data interpretation, which is not described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a hip prosthesis, not an AI-powered diagnostic tool, so MRMC studies are not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This refers to AI algorithms, and this device is a physical implant.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the design verification tests, the "ground truth" would be the engineering specifications, material properties, and performance requirements defined by relevant international standards (e.g., ASTM, ISO) and the manufacturer's internal design controls. The device was tested to ensure it met these pre-defined physical and mechanical properties.

    8. The sample size for the training set

    Not applicable. This refers to the training of an AI algorithm, which is not relevant to this device.

    9. How the ground truth for the training set was established

    Not applicable. This refers to the training of an AI algorithm, which is not relevant to this device.

    In summary:

    This 510(k) submission for the Mpact Extension focused on demonstrating substantial equivalence to predicate devices through design verification testing based on engineering standards and internal protocols. It does not include clinical trial data, AI performance metrics, or information typically associated with studies involving human data interpretation or AI algorithms. The "study" referenced is the series of design verification tests, which confirmed the device's physical and mechanical properties met established criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K122641

    Validate with FDA (Live)

    Device Name
    MPACT EXTENSION
    Date Cleared
    2012-09-28

    (30 days)

    Product Code
    Regulation Number
    888.3353
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The hip prosthesis is designed for cementless use in total hip arthroplasty in primary or revision surgery.

    The patient should be skeletally mature.

    The patient's condition should be due to one or more of:

    • Severely painful and/or disabled joint: as a result of osteoarthritis, posttraumatic arthritis, rheumatoid arthritis or psoriatic arthritis, congenital hip dysplasia, ankylosing spondylitis.
    • Avascular necrosis of the femoral head.
    • Acute traumatic fracture of the femoral head or neck.
    • Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement where sufficient bone stock is present.
    Device Description

    The Mpact Extension components are designed to be used with the Medacta Total Hip Prosthesis System. The Mpact Extension acetabular components that are the subject of this 510(k) consist of a multi-hole or rim-hole shell (Ti-6Al-4V, ASTM F136 and Ti, ASTM F1580 porous coating), a fixed liner in sizes "J" and "K" that is made of ultra-high molecular weight polyethylene (UHMWPE ISO 5834-2 Type1) or HighCross® highly crosslinked ultra-high molecular weight polyethylene (HXUHMWPE), and a cortical bone screw, flat head (Ti-6AJ-4V, ISO 5832-3).

    AI/ML Overview

    The provided document describes the 510(k) submission for the Mpact Extension, a hip prosthesis system. It outlines the performance testing conducted to demonstrate substantial equivalence to predicate devices, but it does not include the detailed information requested regarding sample sizes, data provenance, expert qualifications, or comparative effectiveness studies in the context of an AI/algorithm-based device.

    This document is for a traditional medical device (hip prosthesis), not an AI/ML-based medical device. Therefore, many of the requested categories (like "Number of experts used to establish ground truth," "Adjudication method," "MRMC comparative effectiveness study," "Training set sample size," etc.) are not applicable to this type of submission.

    However, I can extract the available information related to acceptance criteria and device performance based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test TypeAcceptance Criteria (Implied)Reported Device Performance
    Pull-out stability (modular connection of fixed liner to metal shell)(Implied: Meet established standards/predicate device performance for pull-out strength/resistance)Testing was conducted "as part of design verification to written protocols with pre-defined acceptance criteria." The testing "met all acceptance criteria and verifies that the performance of the Mpact Extension is substantially equivalent to the predicate devices." (Specific numerical values for acceptance criteria or performance are not provided in this summary.)
    Lever-out stability (modular connection of fixed liner to metal shell)(Implied: Meet established standards/predicate device performance for lever-out resistance)Testing was conducted "as part of design verification to written protocols with pre-defined acceptance criteria." The testing "met all acceptance criteria and verifies that the performance of the Mpact Extension is substantially equivalent to the predicate devices." (Specific numerical values for acceptance criteria or performance are not provided in this summary.)
    Torsional stability (modular connection of fixed liner to metal shell)(Implied: Meet established standards/predicate device performance for torsional resistance)Testing was conducted "as part of design verification to written protocols with pre-defined acceptance criteria." The testing "met all acceptance criteria and verifies that the performance of the Mpact Extension is substantially equivalent to the predicate devices." (Specific numerical values for acceptance criteria or performance are not provided in this summary.)
    Coating validation(Implied: Meet established standards for coating adhesion, integrity, and biocompatibility)Testing was conducted "as part of design verification to written protocols with pre-defined acceptance criteria." The testing "met all acceptance criteria and verifies that the performance of the Mpact Extension is substantially equivalent to the predicate devices." (Specific numerical values for acceptance criteria or performance are not provided in this summary.)
    Metal shell deformation resistance during impaction(Implied: Resist deformation during surgical impaction within acceptable limits as per standards/predicate device performance)Testing was conducted "as part of design verification to written protocols with pre-defined acceptance criteria." The testing "met all acceptance criteria and verifies that the performance of the Mpact Extension is substantially equivalent to the predicate devices." (Specific numerical values for acceptance criteria or performance are not provided in this summary.)
    Range of motion(Implied: Demonstrate an acceptable range of motion, comparable to predicate devices or physiological norms)Testing was conducted "as part of design verification to written protocols with pre-defined acceptance criteria." The testing "met all acceptance criteria and verifies that the performance of the Mpact Extension is substantially equivalent to the predicate devices." (Specific numerical values for acceptance criteria or performance are not provided in this summary.)
    Wear(Implied: Exhibit wear rates within acceptable limits as per standards/predicate device performance)Testing was conducted "as part of design verification to written protocols with pre-defined acceptance criteria." The testing "met all acceptance criteria and verifies that the performance of the Mpact Extension is substantially equivalent to the predicate devices." (Specific numerical values for acceptance criteria or performance are not provided in this summary.)
    Bone screw testing(Implied: Meet established standards for bone screw performance, e.g., thread pull-out strength, torsional strength, fatigue)Testing was conducted "as part of design verification to written protocols with pre-defined acceptance criteria." The testing "met all acceptance criteria and verifies that the performance of the Mpact Extension is substantially equivalent to the predicate devices." (Specific numerical values for acceptance criteria or performance are not provided in this summary.)

    Study Proving Acceptance Criteria Met:

    The study was described as "Performance testing of the Mpact Extension," conducted "in accordance with various international standards and FDA guidance documents." The testing was part of "design verification to written protocols with pre-defined acceptance criteria."

    Details on the Study (as much as available in the text):

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

      • Sample Size: The document states that testing was "conducted on the worst case component size and option/design." However, it does not specify the exact number of samples or units tested for each type of test (e.g., number of shells, liners, or screws).
      • Data Provenance: Not specified. This type of physical device testing typically wouldn't have "country of origin of the data" in the same sense as patient data.
      • Retrospective or Prospective: Not applicable in the context of physical product testing. It's prospective testing.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

      • Not applicable as this is performance testing of a physical medical device, not an AI/ML algorithm requiring expert interpretation for ground truth.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This is not an AI/ML device.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This is not an AI/ML device.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • The "ground truth" or reference standard for this type of testing is established by the relevant "international standards and FDA guidance documents" specific to orthopedic implants. These standards define the benchmark performance requirements for pull-out strength, wear, deformation, etc.
    7. The sample size for the training set:

      • Not applicable. This is not an AI/ML device.
    8. How the ground truth for the training set was established:

      • Not applicable. This is not an AI/ML device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1