Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K062053

    Validate with FDA (Live)

    Date Cleared
    2006-08-07

    (18 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    RadiForce GS320 is intended to be used in displaying and viewing digital images for diagnosis of X-ray or MRI, etc. by trained medical practitioners. The device is not specified for digital mammography system.

    Device Description

    RadiForce GS320 is a 54cm (21.3") Monochrome LCD display for medical image viewing. GS310 displays high-definition medical imaging.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical display monitor (EIZO NANAO CORPORATION Monochrome LCD Monitor, RadiForce GS320), not a device that utilizes AI or provides diagnostic interpretations. Therefore, many of the requested fields regarding acceptance criteria, study details, expert involvement, and ground truth are not applicable to this type of device and are not present in the submission.

    The 510(k) process for a monitor like this focuses on demonstrating substantial equivalence to a predicate device through technical specifications and intended use, rather than clinical performance studies that directly assess AI accuracy or human reader improvement.

    Here's the information that can be extracted or noted as not applicable based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    For medical display monitors, "acceptance criteria" and "reported device performance" are typically defined by technical specifications and adherence to standards, rather than clinical accuracy metrics. The comparison is primarily against a predicate device's technical specifications.

    CharacteristicAcceptance Criteria (Predicate Device K060845 - RadiForce G310)Reported Device Performance (RadiForce GS320)
    Panel Size and Type53 cm (20.8") TFT Monochrome LCD panel54 cm (21.3") TFT Monochrome LCD panel
    Pixel Pitch0.207 x 0.207mm0.2115 x 0.2115mm
    Ratio of Sub-pixel Opening Areas4: 4: 4Same as predicate
    Grayscale Tones1,024 from a pallet of 8,161Same as predicate
    Viewing AnglesH: 170°, V: 170°Same as predicate
    Scanning Frequency (H, V)31-100kHz, 48-71.5Hz (VGA Text: 69-71Hz) Frame synchronous mode: 59-61HzSame as predicate
    Native Resolutions1536 x 2048Same as predicate
    Brightness700 cd/m² (Typical)Same as predicate
    Contrast Ratio900: 1 (typical)850: 1 (typical)
    DOT Clock165 MHzSame as predicate
    Response Time50 ms (typical)Same as predicate
    Active Display Size (H x V)318 x 424 mm324 x 433 mm
    Viewable Image Size529 mm (20.8") (diagonal)541mm (21.3") (diagonal)
    Luminance CalibrationBuilt-in swing calibration sensor providedSame as predicate
    Input SignalsDVI Standard 1.0Same as predicate
    Input TerminalsDVU-D 24 pinSame as predicate
    USB Ports / Standard1 upstream, 2 downstream1 upstream, 2 downstream/Standard Rev.2.0
    PowerAC100-120V, 200-240V, 50/60HzSame as predicate
    Power ManagementDVI-DMPMSame as predicate
    Dimensions (W x H x D)With Stand: 368 x 515.5 mm – 597.5 x 209 mm Without Stand: 368 x 486 x 90 mmWith Stand: 376 x 522.5 mm – 604.5x 208.5 mm Without Stand: 376 x 500 x 92 mm
    Certifications & StandardsTUV/GM, CE Medical Device Directive, CB (EN60601-1), cTUVus (UL2601-1, CSA C22.2 No. 601-1), VCCI-B, FCC-B, Canadian ICES-003-A, CCCSame as predicate

    The key acceptance criterion here is demonstrating "substantial equivalence" to the predicate device (RadiForce G310, K060845), meaning the new device has similar technological characteristics and intended use, and does not raise new questions of safety or effectiveness. The reported performance of the GS320 is compared directly against the specifications of the GS310. The submission states: "Comparison table of the principal characteristics of 2 devices in Attachment 1 shows that new and predicate devices are substantially equivalent in the areas of technical characteristics, general functions." The modifications noted are a slightly larger panel size and a lower contrast ratio, with the submission implicitly arguing these do not negate substantial equivalence for the intended use.

    2. Sample size used for the test set and the data provenance

    N/A. This is a medical display monitor, not an AI or diagnostic device that would typically undergo testing with a "test set" of patient data. The evaluation is based on engineering specifications and comparison to a predicate device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    N/A. No ground truth or expert evaluation of diagnostic performance is mentioned or required for this type of device submission.

    4. Adjudication method for the test set

    N/A. No test set for diagnostic performance is mentioned.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    N/A. This is not an AI-assisted device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    N/A. This is a medical display monitor, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    N/A. Ground truth is not relevant for a medical display monitor's 510(k) submission, which focuses on technical specifications and substantial equivalence.

    8. The sample size for the training set

    N/A. This device does not involve a training set as it's a hardware monitor, not a machine learning algorithm.

    9. How the ground truth for the training set was established

    N/A. Not applicable for a hardware device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1