Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K100952

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-08-05

    (121 days)

    Product Code
    Regulation Number
    888.3070
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Synthes USS are non-cervical spinal fixation devices intended for posterior pedicle screw fixation (T1-S2/ilium), posterior hook fixation (T1-L5), or anterolateral fixation (T8-L5). Pedicle screw fixation is limited to skeletally mature patients with the exception of the Small Stature USS, which includes small stature and pediatric patients. These devices are indicated as an adjunct to fusion for all of the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis, Scheuermann's Disease), tumor, stenosis, and failed previous fusion (pseudoarthrosis).

    When treating patients with Degenerative Disc Disease (DDD), transverse bars are not cleared for use as part of the posterior pedicle screw construct.

    When used with the 3.5/6.0 mm parallel connectors, the Synthes USS 6.0 mm rod systems can be linked to the CerviFix 3.5 mm Systems. In addition, when used with 3.5/5.0 mm parallel connectors, the Synthes Small Stature USS can be linked to the CerviFix 3.5 mm Systems. When used with the 5.0/6.0 mm parallel connectors, the Synthes Small Stature USS can be linked to the Synthes USS 6.0 mm rod systems.

    When used with the 3.5/6.0 mm and 4.0/6.0 mm tapered rods, the Synthes USS 6.0 mm rod systems can be linked to the CerviFix 3.5 mm and 4.0 mm Systems, respectively. When used with the 3.5/5.5 mm and 4.0/5.5 mm tapered rods, Matrix can be linked to the CerviFix 3.5 mm and 4.0 mm Systems, respectively. When used with the 5.5/6.0 mm tapered rods, the Synthes USS 6.0 mm rod systems can be linked to the Matrix System.

    In addition, Synthes USS 6.0 mm rod systems can be interchanged with all USS 6.0 mm rods and transconnectors.

    Device Description

    The Synthes Matrix System is an addition to Synthes' existing posterior thoracolumbar spine systems. The Matrix System consist of a family non-cervical spinal fixation devices intended for posterior pedicle screw fixation (TI-S2), posterior hook fixation (TI-L5) or anterolateral fixation (T8-L5). The implants include pedicle bone screws, polyaxial pedicle screws, monoaxial pedicle screws, polyaxial heads, reduction screws, reduction heads, locking caps, transconnectors, transverse bars, rods and hooks. The implants are primarily manufactured from titanium (ASTM F67-06), titanium alloy (ASTM F1295-05), cobalt-chromium-molybdenum alloy (ASTM F1537-08) or nitinol (ASTM F2063-05), similar to the predicates.

    AI/ML Overview

    The Synthes Matrix System, a spinal fixation device, demonstrated substantial equivalence to its predicate devices through non-clinical performance testing.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state "acceptance criteria" with numerical targets in the conventional sense. Instead, the performance data focused on demonstrating that the Synthes Matrix System performs "as well as or better than" the predicate devices in various biomechanical tests. The implied acceptance criterion is that the device's performance in these tests is comparable to or exceeds that of the predicate devices.

    Test NameStandard AccordanceReported Device Performance
    Cantilever Bend in conjunction with Rod Push-throughASTM F1798 – 97Performs as well as or better than predicate devices.
    Static Compression BendASTM F1717 – 09Performs as well as or better than predicate devices.
    Static TorsionASTM F1717 – 09Performs as well as or better than predicate devices.
    Dynamic Compression BendASTM F1717 – 09Performs as well as or better than predicate devices.
    Compress and Return TestingASTM F1717 – 09Performs as well as or better than predicate devices.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not specify the exact sample sizes (e.g., number of implants tested) for each non-clinical test. The testing was conducted by Synthes, implying an internal or contracted laboratory setting, but the country of origin of the data is not explicitly stated. The nature of these tests (biomechanical/mechanical) means they are retrospective in the sense that they are conducted on manufactured devices in a controlled environment, not on live patients.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

    This question is not applicable. The study involved non-clinical biomechanical testing of medical devices, not evaluations by human experts to establish "ground truth" for clinical conditions or diagnoses. The ground truth for these tests is defined by the objective measurements obtained from the testing equipment according to the specified ASTM standards.

    4. Adjudication Method for the Test Set:

    This question is not applicable. Adjudication methods like "2+1" or "3+1" are relevant for expert consensus in clinical or imaging studies, not for non-clinical mechanical testing where performance is determined by objective measurements.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, an MRMC comparative effectiveness study was not done. The submission explicitly states, "Clinical data was not needed for this device." The evaluation was based solely on non-clinical performance data.

    6. Standalone Performance (Algorithm Only Without Human-in-the-Loop Performance):

    This question is not applicable. This is a physical medical device (spinal fixation system), not a software algorithm or AI-powered system where "standalone" performance would be a relevant metric.

    7. Type of Ground Truth Used:

    The "ground truth" for the non-clinical tests was established by objective physical measurements obtained from testing the devices according to specific ASTM standards (ASTM F1798-97 and ASTM F1717-09). These standards define the methodology for evaluating the mechanical properties of spinal implants.

    8. Sample Size for the Training Set:

    This question is not applicable. There is no information about a "training set" as this is a physical medical device, not a machine learning model or algorithm. The non-clinical tests evaluate the physical properties and performance of the manufactured device.

    9. How the Ground Truth for the Training Set Was Established:

    This question is not applicable. As stated above, there is no "training set" in the context of this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1