Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K110444

    Validate with FDA (Live)

    Device Name
    MIRAGE MICRO
    Manufacturer
    Date Cleared
    2011-03-16

    (28 days)

    Product Code
    Regulation Number
    868.5905
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Mirage Micro channels airflow non-invasively to a patient from a positive airway pressure device such as a continuous positive airway pressure (CPAP) or bilevel system.

    The Mirage Micro is:

    • to be used by adult patients (>66lb / >30kg) for whom positive airway pressure has been prescribed.
    • intended for single patient re-use in the home environment and multipatient re-use in the hospital/institutional environment.
    Device Description

    The headgear colour of the modified device has changed. The color of the outer layer of the fabric is being changed from dark blue to tan. All other aspects of the device remain unchanged.

    The modified Mirage Micro is safe when used under the conditions and purposes intended as indicated in the labelling provided with the product.

    The modified Mirage Micro is a prescription device supplied nonsterile.

    Both the masks incorporate vent holes to provide continuous air leak to flush out the dead space within the mask and minimize the amount of CO2 rebreathed by the patient. The design of the mask components is such that the incorporation of these vent-holes does not interfere with the intended performance of the masks. Both masks connect to a conventional air delivery hose between the mask and the positive airway-pressure source via standard conical connectors (ref: ISO 5356-1:2004)

    Both masks have provisions for connecting oxygen and pressure sensing tubing via luer ports.

    Both masks are constructed using molded plastic components and fabric headgear. All the components of both the masks are fabricated using materials deemed safe. (ref: ISO 10993-1).

    Both the modified device and the previously cleared device can be reused in the hospital / institution environment.

    AI/ML Overview

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance CriteriaReported Device Performance
    Same intended use as the predicate device (Mirage Micro Mask K072940).The modified Mirage Micro has the same intended use: channeling airflow non-invasively to adult patients from a positive airway pressure device (CPAP or bilevel system), for single-patient re-use at home and multi-patient re-use in hospital/institutional environments.
    Similar technological characteristics to the predicate device.The only change is the headgear color (from dark blue to tan). All other aspects (vent holes, connectors, provisions for oxygen/pressure sensing, construction materials, re-use capability) remain identical to the predicate device.
    Does not raise new questions of safety and effectiveness.The change in headgear color is asserted not to affect technological characteristics when used with a compatible ResMed flow generator, and the outer layer of the fabric does not come into contact with the patient, therefore not requiring ISO 10993-1. This implies no new safety or effectiveness concerns.
    At least as safe and effective as the previously cleared Mirage Micro (K072940).Due to the cosmetic nature of the change (headgear color) and the assertion that it does not affect the device's function or patient contact, the device is considered substantially equivalent in safety and effectiveness to the predicate.
    Bench testing not required as the change has no effect on technological characteristics when used with a compatible ResMed flow generator.No bench testing was performed due to the nature of the change. This implicitly serves as an acceptance criterion being met by the product's unchanged functional aspects.
    The headgear color change (from dark blue to tan) does not affect performance.The document states, "The headgear colour of the modified device has changed... All other aspects of the device remain unchanged." and "The headgear performance of the modified device and the previously cleared device are substantially equivalent." This indicates the color change did not impact performance.
    All components are fabricated using materials deemed safe (ref: ISO 10993-1)."All the components of both the masks are fabricated using materials deemed safe. (ref: ISO 10993-1)." This implies that while the outer layer of the headgear fabric doesn't come into patient contact, other components still adhere to this standard, ensuring overall safety.
    Design of vent-holes does not interfere with the intended performance of the masks."Both the masks incorporate vent holes to provide continuous air leak to flush out the dead space within the mask and minimize the amount of CO2 rebreathed by the patient. The design of the mask components is such that the incorporation of these vent-holes does not interfere with the intended performance of the masks." This characteristic is maintained from the predicate.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not applicable. This document is for a Special 510(k) for a minor device modification (headgear color change). No new clinical or performance test data was generated with a specific test set. The submission relies on a comparison to the predicate device and the assertion that the change is cosmetic and does not impact performance.
    • Data Provenance: Not applicable for a new test set. The basis for substantial equivalence is the prior clearance of the predicate device (Mirage Micro Mask K072940) and the established safety and effectiveness of vented nasal masks with CPAP/Bilevel therapy.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. No new ground truth was established for a test set as part of this Special 510(k) submission.

    4. Adjudication method for the test set:

    • Not applicable. No test set requiring adjudication was used.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This device is a passive medical device (a mask); it does not involve AI or human readers for interpretation. Therefore, an MRMC study is not relevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • No. This device is a passive medical device and does not involve any algorithms or software-driven performance that would require standalone testing.

    7. The type of ground truth used:

    • The ground truth for the safety and effectiveness of the device is based on the prior clearance and established performance of the predicate device (Mirage Micro Mask K072940). The current submission argues that the modified device is substantially equivalent because the change is purely cosmetic (headgear color) and does not alter the fundamental design, materials, or intended use that were proven safe and effective by the predicate.

    8. The sample size for the training set:

    • Not applicable. This device is a passive medical device, and this submission is for a minor modification. It does not involve machine learning or AI, and therefore no training set.

    9. How the ground truth for the training set was established:

    • Not applicable. As there is no training set involved, no ground truth needed to be established for it.
    Ask a Question

    Ask a specific question about this device

    K Number
    K100772

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-06-09

    (83 days)

    Product Code
    Regulation Number
    868.5905
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Mirage Micro channels airflow noninvasively to a patient from a positive airway pressure device such as a continuous positive airway pressure (CPAP) or bilevel system. The Mirage Micro is: - to be used by adult patients (>66lb / >30kg) for whom positive airway pressure has been prescribed. - to be used for single-patient re-use in the home environment and multi-patient reuse in the hospital/institutional environment.

    Device Description

    Not Found

    AI/ML Overview

    This appears to be a 510(k) clearance letter from the FDA for a medical device (Mirage Micro™). It does not contain the information required to answer your prompt.

    The document is a regulatory approval, confirming that the device is substantially equivalent to legally marketed predicate devices. It discusses regulatory compliance, labeling, and reporting requirements.

    Specifically, the document does NOT include any of the following:

    • A table of acceptance criteria or reported device performance.
    • Information on sample sizes for test sets, data provenance, or training sets.
    • Details about expert involvement, ground truth establishment, or adjudication methods.
    • Any mention of multi-reader multi-case (MRMC) comparative effectiveness studies or standalone algorithm performance.

    Therefore, I cannot provide the requested information based on the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K072940

    Validate with FDA (Live)

    Device Name
    MIRAGE MICRO
    Manufacturer
    Date Cleared
    2007-12-13

    (57 days)

    Product Code
    Regulation Number
    868.5905
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Mirage Micro channels airflow non-invasively to a patient from a positive airway pressure device such as a continuous positive airway pressure (CPAP) or bilevel system.

    The Mirage Micro is:

    • to be used by adult patients (>66lb / >30kg) for whom positive airway pressure has been prescribed.
    • intended for single patient re-use in the home environment and multi-patient re-use in the hospital/institutional environment.
    Device Description

    The Mirage Micro provides seal such that airflow from a positive pressure source is directed to the patient's nose. The mask is held in place with adjustable headgear that straps the mask to the face.

    Mirage Micro is safe when used under the conditions and purposes intended as indicated in the labeling provided with the product.

    Mirage Micro is a prescription device supplied nonsterile.

    AI/ML Overview

    The provided document is a 510(k) summary for the ResMed Mirage Micro Traditional, a vented nasal mask, seeking substantial equivalence to predicate devices. It is not a study that proves the device meets acceptance criteria in the way typically seen for AI/ML-based medical devices or diagnostics. Instead, it demonstrates substantial equivalence through comparisons of technological characteristics and performance data via bench testing.

    Therefore, many of the requested fields (e.g., sample size for test set, number of experts, adjudication method, MRMC study, standalone performance, training set size) are not applicable to this type of submission.

    Here's a breakdown of the available information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not explicitly state formal "acceptance criteria" with numerical thresholds in the sense of a diagnostic device. Instead, it relies on demonstrating substantial equivalence to predicate devices through comparative performance and technological characteristics. The "reported device performance" is essentially a statement of equivalence.

    Acceptance Criteria (Implied)Reported Device Performance
    CO2 performance substantially equivalent to predicate device."The CO2 performance of the new device and the predicate device are substantially equivalent."
    Pressure-flow characteristics and flow impedance substantially equivalent to predicate device."The pressure-flow characteristics and flow impedance of both the new device and the predicate device are substantially equivalent."
    Operation on the same standard flow generator setting as predicate device."Both the new device and the predicate are designed to operate on the same standard flow generator setting."
    Provision of seal via dual wall silicone interface."The new device and the predicate mask, provide seal via dual wall silicone interface."
    Incorporation of vent holes for continuous air leak."Both the masks incorporate vent holes to provide continuous air leak to flush out the dead space within the mask and minimize the amount of CO2 rebreathed by the patient."
    Connection to conventional air delivery hose via standard conical connectors."Both the masks connect to conventional air delivery hose between the mask and the positive airway-pressure source via standard conical connectors (ref: ISO 5356-1:2004)."
    Provisions for connecting oxygen and pressure sensing tubing."Both the masks have provisions for connecting oxygen and pressure sensing tubing via luer ports."
    Construction using molded plastic components and fabric headgear with safe materials."Both the masks are constructed using molded plastic components and fabric headgear. All the components of both the masks are fabricated using materials deemed safe. (ref: ISO 10993-1)."
    Intended use matching predicate devices."it has the same intended use;"
    Technological characteristics similar to predicate devices."it has similar technological characteristics to both predicates:"
    Does not raise new questions of safety and effectiveness."it does not raise new questions of safety and effectiveness:"
    At least as safe and effective as predicate devices."it is at least as safe and effective as the predicate devices Mirage Micro and Mirage Quattro."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample size for the test set: Not applicable. The submission refers to "bench testing" and "comparison with predicate" rather than a clinical trial or a test set of data points in the context of an AI/ML device.
    • Data provenance: Not applicable. Bench testing is typically performed in a laboratory setting. No country of origin for clinical data is mentioned as none was used.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Number of experts: Not applicable. Ground truth establishment by experts is not described as part of this bench testing for a mechanical device.
    • Qualifications of experts: Not applicable.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Adjudication method: Not applicable. This concept is relevant for expert disagreement resolution in diagnostic studies, not for mechanical device bench testing.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC study: No. This is a medical device, not an AI/ML diagnostic.
    • Effect size: Not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone performance: Not applicable. This is a medical device, not an AI/ML diagnostic.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Type of ground truth: Not explicitly stated as "ground truth" in the AI/ML sense. The performance of the new device was compared against the performance of the predicate devices (Mirage Micro K071808 and Mirage Quattro K063122) through bench testing parameters like CO2 performance, pressure-flow characteristics, and flow impedance. The "truth" is based on the established safety and efficacy of the predicate devices.

    8. The sample size for the training set

    • Sample size for training set: Not applicable. This is a medical device, not an AI/ML system.

    9. How the ground truth for the training set was established

    • Ground truth for training set establishment: Not applicable. This is a medical device, not an AI/ML system.

    Summary of the Study (as described in the 510(k) Summary):

    The submission for the Mirage Micro Traditional nasal mask relies on a bench testing study to demonstrate substantial equivalence to its predicate devices (Mirage Micro Mask K071808 and Mirage Quattro Mask K063122).

    The core of the "study" involved:

    • Comparison of Technological Characteristics: Detail how the new device's design, materials, and features are similar to the predicates. This includes aspects like the dual-wall silicone interface, vent hole design, standard connectors, luer ports, and biocompatible materials.
    • Performance Data Comparison (Bench Testing): Direct comparison of the new device's performance against the predicate Mirage Micro regarding:
      • CO2 performance: Stated as "substantially equivalent."
      • Pressure-flow characteristics: Stated as "substantially equivalent."
      • Flow impedance: Stated as "substantially equivalent."
      • Operation on standard flow generator settings: Stated as designed to operate on the same settings.

    The conclusion drawn from this comparison and bench testing is that the new Mirage Micro is substantially equivalent because it shares the same intended use, similar technological characteristics, does not raise new questions of safety and effectiveness, and is at least as safe and effective as the predicate devices. Clinical data was deemed unnecessary as "Use of vented nasal masks with CPAP or Bilevel therapy is proven technology and is well accepted by the medical community. Bench testing is sufficient to demonstrate safety and efficacy..."

    Ask a Question

    Ask a specific question about this device

    K Number
    K071808

    Validate with FDA (Live)

    Device Name
    MIRAGE MICRO
    Manufacturer
    Date Cleared
    2007-10-10

    (100 days)

    Product Code
    Regulation Number
    868.5905
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Mirage Micro channels airflow non-invasively to a patient from a positive airway pressure device such as a continuous positive airway pressure (CPAP) or bilevel system. The Mirage Micro is: - to be used by adult patients (>66lb / >30kg) for whom positive airway pressure has been prescribed. - intended for single patient re-use in the home and the hospital/institutional environment.

    Device Description

    The Mirage Micro provides seal such that airflow from a positive pressure source is directed to the patient's nose. The mask is held in place with adjustable headgear that straps the mask to the face. Mirage Micro is safe when used under the conditions and purposes intended as indicated in the labeling provided with the product. Mirage Micro is a prescription device supplied nonsterile.

    AI/ML Overview

    The provided text describes the 510(k) summary for the ResMed Mirage Micro™ nasal mask. This document outlines the device's characteristics and its substantial equivalence to predicate devices, rather than a clinical study demonstrating the device meets specific acceptance criteria through comprehensive performance metrics.

    Therefore, many of the requested details about acceptance criteria, study design, sample sizes, and ground truth establishment are not present in this regulatory submission for a nasal mask. The submission emphasizes "bench testing" and "proven technology" for substantial equivalence.

    Here's an analysis based on the information available and not available in the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not present a formal table of acceptance criteria with specific numerical targets. Instead, it relies on a comparison to predicate devices, asserting "substantial equivalence" in key performance characteristics relevant to its function as a nasal mask for PAP therapy.

    Performance Characteristic (Implied "Acceptance Criteria")Reported Device Performance (Comparison to Predicate)
    Seal (adequate for airflow from positive pressure source)"The Mirage Micro provides seal such that airflow from a positive pressure source is directed to the patient's nose."
    Fit (over extended patient population)"Both masks are offered in various sizes to ensure adequate fit over the extended patient population." (Compared to Ultra Mirage II)
    CO2 Rebreathing / Vent Holes (minimize CO2 rebreathing and flush dead space)"The design of the mask components is such that the incorporation of these vent holes do not interfere with the intended performance of the masks." (Compared to Ultra Mirage II) "The CO2 performance the new device and the predicate device are substantially equivalent."
    Connection to Air Delivery Hose (standard conical connectors)"Both the masks connect to conventional air delivery hose between the mask and the positive airway-pressure source via standard conical connectors (ref: ISO 5356-1:2004)" (Compared to Ultra Mirage II)
    Oxygen/Pressure Sensing Ports (luer ports)"Both the masks have provisions for connecting oxygen and pressure sensing tubing via luer ports." (Compared to Ultra Mirage II)
    Biocompatibility of Materials (fabricated using safe materials)"All the components of both the masks are fabricated using materials deemed safe. (ref: ISO 10993-1)." (Compared to Ultra Mirage II)
    Pressure-Flow Characteristics / Flow Impedance"The pressure-flow characteristics and flow impedance of both the new device and the predicate device are substantially equivalent." (Compared to Meridian)
    Safety and Efficacy"Bench testing is sufficient to demonstrate safety and efficacy of the Mirage Micro, as was the case with the predicate devices." (Clinical data section explicitly states this)
    Intended Use (adult patients, single patient reuse, home/hospital environment)"Same intended use with reduced scope being for single patient reuse" (compared to predicate Ultra Mirage II for multi-patient use in hospital/institutional environment)

    2. Sample size used for the test set and the data provenance

    The document explicitly states: "Bench testing is sufficient to demonstrate safety and efficacy of the Mirage Micro, as was the case with the predicate devices." This indicates that the primary performance evaluation was conducted via bench testing, meaning in a laboratory setting, rather than with human subjects in a clinical trial.

    • Sample size: Not specified for bench tests. For physical component characteristics, it would likely involve a sample of manufactured masks.
    • Data provenance: Bench test data, not explicitly described as from a specific country, but linked to internal testing by ResMed Ltd (Australia). The data provenance is retrospective in the sense that it's based on internal validation for product development.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. As the evaluation was based on bench testing and comparison to predicate devices, there was no "ground truth" derived from expert interpretation of clinical data in the traditional sense. The "ground truth" for performance would be the specifications and validated behavior of the predicate devices and general knowledge of CPAP mask function.

    4. Adjudication method for the test set

    Not applicable. No expert adjudication method is described, as it was bench testing.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a medical mask, not an AI-powered diagnostic or assistive technology for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a medical mask, not an algorithm. Its "standalone" performance refers to its physical and functional characteristics on a bench.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for this device's performance is based on engineering specifications, material safety standards (ISO 10993-1), recognized international standards for connectors (ISO 5356-1:2004), and the established performance characteristics of the predicate devices. It’s not clinical ground truth like pathology or expert consensus on patient data.

    8. The sample size for the training set

    Not applicable. This is a physical device, not an algorithm that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. No training set is involved for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1