Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K202366

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-01-29

    (163 days)

    Product Code
    Regulation Number
    870.1210
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MAGIC Infusion Catheter is intended for regional infusion of contrast materials into selected vessels in the neurovasculature. The MAGIC Infusion Catheter may be used for controlled, regional infusion into selected vessels and is not intended for use in the coronary vasculature.

    Device Description

    The MAGIC Infusion Catheters are microcatheters designed with progressive suppleness and a rigid proximal shaft to allow control and navigability in the vascular system. By their diameter and progressive suppleness features, these catheters are specifically designed for catheterization of small diameter, sinuous distal vessels. The catheter body and its distal tip (ring) are radiopaque to provide visibility under fluoroscopy. The MAGIC Infusion Catheter has an external hydrophilic coating which provides a lubricious surface during use. A PTFE coated mandrel is included inside the MAGIC Infusion Catheter to provide support during product preparation and insertion through the guide catheter. The MAGIC Infusion Catheters and accompanying support mandrels are provided sterile, non-pyrogenic, and intended for single use only.

    AI/ML Overview

    The provided document describes the MAGIC Infusion Catheter and its 510(k) summary. The study is a bench and biocompatibility testing designed to demonstrate substantial equivalence to a predicate device.

    Here's the breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Bench TestAcceptance CriteriaResults
    Dimensional and Physical AttributesShall meet the required dimensional measurements and physical attributes per established specifications.PASS
    Visual InspectionShall meet the established acceptance criteria for surface defects and contamination.PASS
    Torque Strength TestCatheters shall withstand the established acceptance criteria for torque strength without breaking.PASS (Device integrity suitable for intended clinical use)
    Catheter Tensile TestShall meet ISO 105551-1, Annex BPASS
    Corrosion ResistanceShall meet ISO 105551-1, Annex APASS
    Flow RateReference DataN/A (Note: Not an acceptance criterion for passing/failing)
    Dynamic PressureCatheters shall withstand the maximum infusion pressure (as indicated on the product labeling) under dynamic conditions. (Adopted from ISO 105551-1)PASS (Device met maximum labeled infusion pressure)
    Static Burst PressureCatheters shall withstand the maximum infusion pressure (as indicated on the product labeling) under static conditions. (Adopted from ISO 105551-1)PASS (Device integrity suitable for intended clinical use)
    Pressure and Hub Test (Fluid Leakage)Catheters shall not leak when injected with water at the maximum infusion pressure (as indicated on the product labeling) for a 30 second duration.PASS (Device integrity suitable for intended clinical use)
    Simulated Use / Coating Lubricity and DurabilityCatheters shall be tracked using a benchtop model under simulated use conditions per product IFU, and meet established acceptance criteria for product performance, trackability, coating lubricity and durability.PASS (Results comparable to predicate device)
    Kink ResistanceCatheter body shall be wrapped around different pin gauges at various locations and meet the expected acceptance criteria for kink resistance. No kinks shall be noted during Simulated Use.PASS (Results comparable to predicate device)
    RadiopacityMarker band shall be detectable under radiographic techniques. (Per ASTM F640-12)PASS
    ParticulatesShall meet USP <788> criteria. No particulates shall be noted pre and post Simulated Use.PASS (Results comparable to predicate device)
    Biocompatibility: Cytotoxicity – MEM Elution TestISO 10993-5 (implicitly: PASSED)PASSED
    Biocompatibility: Cytotoxicity – Agar Overlay AssayISO 10993-5 (implicitly: PASSED)PASSED
    Biocompatibility: Sensitization – Guinea Pig Maximization TestISO 10993-10 (implicitly: PASSED)PASSED
    Biocompatibility: Irritation – Intracutaneous ReactivityISO 10993-10 (implicitly: PASSED)PASSED
    Biocompatibility: Acute Systemic Toxicity – Systemic Injection TestISO 10993-11 (implicitly: PASSED)PASSED
    Biocompatibility: Systemic Toxicity – Rabbit Pyrogen TestISO 10993-11 (implicitly: PASSED)PASSED
    Biocompatibility: Hemocompatibility – HemolysisISO 10993-4 / ASTM F756 (implicitly: PASSED)PASSED
    Biocompatibility: Hemocompatibility – Complement ActivationISO 10993-4 (implicitly: PASSED)PASSED
    Biocompatibility: Hemocompatibility – Thrombogenicity: Canine ModelISO 10993-4 (implicitly: PASSED)PASSED
    Biocompatibility: Hemocompatibility – Thrombogenicity: Partial Thromboplastin Time AssayISO 10993-4 / ASTM F2382-18 (implicitly: PASSED)PASSED
    Biocompatibility: Hemocompatibility – Thrombogenicity: Platelet/Leukocyte Count AssayISO 10993-4 / ASTM F2888-19 (implicitly: PASSED)PASSED

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the numerical sample sizes for each specific bench test. It generally refers to "the 1.2F MAGIC Infusion Catheter (subject device)" as being tested. For biocompatibility, it states "The 1.2F MAGIC Infusion Catheter was assessed for biocompatibility."

    The provenance of the data is retrospective, as it refers to testing conducted on the subject device to demonstrate substantial equivalence to a previously cleared predicate device. The document does not specify the country of origin of the data beyond the applicant "Balt USA, LLC" being located in Irvine, CA, USA.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable to this type of device submission. The study describes physical and biological bench testing of a medical device, not a diagnostic or AI-powered device that requires expert-established ground truth from a test set of clinical images or data. The "ground truth" for these tests is defined by established engineering standards and biological testing protocols (e.g., ISO standards, ASTM standards, USP criteria).

    4. Adjudication Method for the Test Set

    This information is not applicable. Adjudication methods like 2+1 or 3+1 are used in studies involving expert readers for diagnostic accuracy. This submission relies on objective physical and biological measurements against predefined acceptance criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    This information is not applicable. This submission is for a physical medical device (infusion catheter) and does not involve AI or human readers for diagnostic interpretation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This information is not applicable. This is not an AI-powered device.

    7. The Type of Ground Truth Used

    The ground truth for the bench tests is based on established engineering specifications, international standards (ISO 105551-1, ASTM F640-12, USP <788>), and biological testing protocols (ISO 10993-series, ASTM F756, ASTM F2382-18, ASTM F2888-19). These standards define the acceptable performance characteristics and biological response for such medical devices.

    8. The Sample Size for the Training Set

    This information is not applicable. This is not an AI/machine learning device, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K023351

    Validate with FDA (Live)

    Date Cleared
    2003-02-20

    (136 days)

    Product Code
    Regulation Number
    870.1210
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Magic Infusion Catheter is intended for regional infusion of contrast materials into selected vessels in the neurovasculature. The Magic Infusion Catheter may be used for controlled, regional infusion into selected vessels and is not intended for use in the coronary vasculature.

    Device Description

    The subject device is intended to operate for regional infusion of contrast materials into selected vessels in the neurovasculature. Magic Infusion Catheters are intended to facilitate access through distant, tortuous vasculature. Progressive suppleness ranging from a highly flexible tip to a semi-rigid proximal section allows the catheter to be advanced by the physician. It is not intended for use in the coronary vasculature. Magic catheters are available in 1.8 and 1.5Fr and 165 and 155 cm in length and are available with an "olive" tip configuration.

    AI/ML Overview

    This document is a 510(k) summary for the Magic® Infusion Catheter, indicating its substantial equivalence to a predicate device. It primarily focuses on demonstrating that changes to materials and coating do not affect safety and effectiveness. As such, the information typically found in studies for AI/ML devices or diagnostic tools (like acceptance criteria tables, sample sizes for test sets, expert ground truth, MRMC studies, standalone performance, and training set details) is not applicable here.

    Here's a breakdown of the available information based on your request, highlighting what is present and what is not:

    1. A table of acceptance criteria and the reported device performance:

    The document describes several tests performed on the Magic Infusion Catheter, particularly regarding the changes to its materials and coating. It states that these tests demonstrate biocompatibility and material integrity.

    Test TypeAcceptance Criteria (Implied)Reported Device Performance
    Biocompatibility (Hydrophilic Coating)Non-cytotoxic, non-hemolytic, non-pyrogenic, non-mutagenic, non-sensitizing, non-irritatingTesting demonstrated that the Magic catheter with the hydrophilic coating was non-cytotoxic, non-hemolytic, non-pyrogenic, non-mutagenic, non-sensitizing and non-irritating, "attesting to the catheter with the new coating's biocompatibility."
    Biocompatibility (Polyamid Shaft)Non-cytotoxic, non-hemolytic, non-pyrogenic, non-allergenic, non-toxic, non-irritatingTesting on the catheter including the Polyamid section of the shaft demonstrated that the catheter was non-cytotoxic, non-hemolytic, non-pyrogenic, non-allergenic, non-toxic, and non-irritating.
    Material Change (Tip Marker - Platinum)Material properties are well understood and change is not significant.Change of the material in the catheter's tip marker is not a significant change as the properties of the new material (platinum) is well understood.
    Other Physical Tests (General Catheter)Not explicitly stated, but implied to meet performance similar to predicate.Testing on the Magic catheter including biocompatibility, joint strength, shaft tensile strength, tip tensile strength, tip flexibility, static rupture strength, dynamic rupture strength, and flow rate. (Specific performance values or direct comparison to criteria for these are not provided in the summary.)

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    • Sample Size: Not specified. The document refers to "Testing" but does not detail the number of units or samples used for each test.
    • Data Provenance: Not specified. This is a technical performance study, not a clinical data study, so "country of origin" of data or "retrospective/prospective" does not apply in the typical sense for medical imaging or diagnostic devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    • Not Applicable. This is a physical device testing summary, not a study involving expert readers or ground truth in the context of diagnoses or image interpretation. The "ground truth" here is based on laboratory test results and material properties.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not Applicable. No human adjudication method is described as this is not a study assessing diagnostic performance or subjective interpretations.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable. This is not an AI/ML device or a diagnostic device where MRMC studies would be relevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. This is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Laboratory Test Results and Material Properties. The "ground truth" for the device's performance is derived from standardized laboratory tests for biocompatibility, material strength, flexibility, rupture strength, and flow rate, as well as the known properties of the materials used (e.g., platinum).

    8. The sample size for the training set:

    • Not Applicable. This is not an AI/ML device, so there is no "training set."

    9. How the ground truth for the training set was established:

    • Not Applicable. As there is no training set, this question is not relevant.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1