510(k) Data Aggregation

    K Number
    K163077

    Validate with FDA (Live)

    Date Cleared
    2016-11-17

    (14 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended for use by a qualified physician for ultrasound evaluation of Fetal; Abdominal; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transrectal; Transvaginal; Transesophageal and Intraoperative (abdominal, thoracic and vascular).

    Device Description

    The LOGIQ E9 is a full featured, general purpose diagnostic ultrasound system which consists of a mobile console approximately 58 cm wide, 87cm deep and 130 cm high that provides digital acquisition, processing and display capability. The user interface includes a computer keyboard, specialized controls, 10-inch LCD touch screen and color wide screen image display.

    AI/ML Overview

    Here's a breakdown of the requested information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document (a 510(k) summary) does not detail specific quantitative acceptance criteria or corresponding reported device performance metrics in the way one might expect for a clinical study comparing an AI algorithm's diagnostic accuracy to a threshold. Instead, the document focuses on demonstrating substantial equivalence to a predicate device.

    The "acceptance criteria" here are implicitly related to meeting safety and performance standards, and having comparable intended use, imaging capabilities, and technological characteristics to the predicate device. The "reported device performance" is the conclusion that the device is "as safe, as effective, and performance is substantially equivalent."

    Here's an interpretation based on the document's content:

    Acceptance Criteria (Implicit)Reported Device Performance
    Equivalent intended use to predicate deviceThe LOGIQ E9 and predicate LOGIQ E9 systems have the same clinical intended use.
    Equivalent imaging capabilities to predicate deviceThe LOGIQ E9 and predicate LOGIQ E9 systems have the same imaging modes.
    Equivalent technological characteristics to predicate deviceThe LOGIQ E9 employs the same fundamental scientific technology as its predicate device(s).
    Safe materialsThe systems are manufactured with materials which have been evaluated and found to be safe for the intended use of the device.
    Acoustic power levels below FDA limitsThe systems have acoustic power levels which are below the applicable FDA limits.
    Similar capability in measurements, image capture, review, reportingThe LOGIQ E9 and predicate LOGIQ E9 systems have similar capability in terms of performing measurements, capturing digital images, reviewing and reporting studies.
    Compliance with electrical and physical safety standardsThe LOGIQ E9 and predicate systems have been designed in compliance with approved electrical and physical safety standards.
    Transducer equivalenceThe LOGIQ E9 and predicate LOGIQ E9 systems transducers are identical.
    Software feature functionality (Parametric Imaging)Parametric Imaging is an evolution of the previously cleared TIC feature on LOGIQ E9.
    Compliance with specific voluntary standards (e.g., AAMI/ANSI ES60601-1, IEC60601-1-2, etc.)The device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic and mechanical safety, and has been found to conform with applicable medical device safety standards. The LOGIQ E9 and its applications comply with voluntary standards (listed).

    2. Sample Size Used for the Test Set and Data Provenance

    The document states: "The subject of this premarket submission, LOGIQ E9, did not require clinical studies to support substantial equivalence." This indicates that no specific test set of patient data was used to evaluate the substantial equivalence of this device in a clinical context. The evaluation was based on non-clinical tests and comparison to a predicate device.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    As no clinical studies were required, there was no test set for which expert ground truth would have been established.

    4. Adjudication Method for the Test Set

    Not applicable, as no clinical test set was used.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was done, as the document explicitly states that clinical studies were not required to support substantial equivalence. Therefore, there is no effect size of human readers improving with or without AI assistance to report.

    6. Standalone (Algorithm Only) Performance Study

    Not applicable. The document describes an ultrasound system (hardware and integrated software features), not a standalone AI algorithm. The evaluation was for the entire system's substantial equivalence.

    7. Type of Ground Truth Used

    Not applicable, as no clinical studies were required and no ground truth from patient data was established for the purpose of demonstrating substantial equivalence. The "ground truth" in this context would be adherence to safety and performance standards and demonstrable equivalence to the predicate device through non-clinical testing.

    8. Sample Size for the Training Set

    Not applicable. This document describes a medical device (ultrasound system), not an AI algorithm that requires a training set in the conventional sense. The "training" for the device's functionality would be embedded in its design, engineering, and adherence to established physical principles of ultrasound imaging.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As noted above, this summary does not pertain to an AI algorithm that was "trained" on data to learn a task. Its "ground truth" is its compliance with a predefined design and manufacturing process that ensures safety and performance in line with a predicate device and industry standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K152309

    Validate with FDA (Live)

    Date Cleared
    2015-09-14

    (31 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended for use by a qualified physician for ultrasound evaluation of Fetal; Abdominal; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transrectal; Transvaginal; Transesophageal and Intraoperative (abdominal, thoracic and vascular).

    Device Description

    The LOGIQ E9 is a full featured, general purpose diagnostic ultrasound system which consists of a mobile console approximately 58 cm wide, 87cm deep and 130 cm high that provides digital acquisition, processing and display capability. The user interface includes a computer keyboard, specialized controls, 10-inch LCD touch screen and color wide screen image display.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification Submission for the GE LOGIQ E9, LOGIQ E9 XDclear 2.0 Diagnostic Ultrasound System. This document focuses on establishing substantial equivalence to previously cleared devices rather than presenting a new clinical study with acceptance criteria and device performance. Therefore, much of the requested information regarding acceptance criteria and performance studies in the traditional sense of a clinical trial for a novel device is not directly available in this document.

    However, I can extract information related to the device's intended use, comparison to predicates, and the non-clinical tests performed.

    Important Note: The acceptance criteria and performance data requested in the prompt typically refer to a clinical study demonstrating the efficacy and safety of a novel device. As this submission is for substantial equivalence to existing devices (K142160 GE LOGIQ E9, K131527 GE LOGIQ S8, K131514 GE Vivid E9), the "acceptance criteria" here are met by demonstrating that the new device shares the same intended use, imaging capabilities, technological characteristics, and safety and effectiveness as the predicate devices, and complies with applicable safety standards. There is no new clinical study presented that would define novel acceptance criteria and then report performance against them.

    Here's the information based on the provided document:


    1. Table of Acceptance Criteria and Reported Device Performance

    As this is a substantial equivalence submission, the "acceptance criteria" are implied by the features and performance of the predicate devices and compliance with relevant standards. The "reported device performance" is the demonstration that the LOGIQ E9, LOGIQ E9 XDclear 2.0 system performs equivalently to the predicates for its intended uses and complies with safety standards.

    Acceptance Criteria (Implied by Substantial Equivalence and Standards)Reported Device Performance (Claimed in Submission)
    Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body for specified clinical applications (Fetal, Abdominal, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Cardiac, Peripheral Vascular, Musculo-skeletal, Urology, Transrectal, Transvaginal, Transesophageal, Intraoperative).The device is stated to have the same clinical intended use as the predicate LOGIQ E9. The individual Intended Use Forms for various transducers (e.g., 3CRF, 3CRF-D, 6S-D, 6Tc, 9L-D, 11L-D, C1-5-D, IC5-9-D, L8-18i-D, M4S-D, M5S-D, M6C-D, ML6-15-D, P2D, P6D, RAB2-5-D, RAB4-8-D, RIC5-9-D, RNA5-9-D, RSP6-16-D, S1-5, S4-10, S4-10-D, C2-9-D, RAB6-D, S1-5-D, C1-6-D, C2-9VN-D, C1-6VN-D, C2-7-D, C2-7VN-D, L3-9i-D, C3-10-D, C2-6b-D) detail the specific applications and modes (B, M, PW Doppler, CW Doppler, Color Doppler, Color M Doppler, Power Doppler, Combined Modes, Harmonic Imaging, Coded Pulse, 3D/4D Imaging, Needle Guidance, Volume Navigation, Elastography Imaging, Shear Wave Elastography).
    Imaging Capabilities: Similar imaging modes and capabilities as predicate devices.The LOGIQ E9 and predicate LOGIQ E9 systems have the same imaging modes. The LOGIQ E9 and predicate LOGIQ E9 systems have similar capability in terms of performing measurements, capturing digital images, reviewing, and reporting studies.
    Technological Characteristics: Employ the same fundamental scientific technology.The LOGIQ E9 employs the same fundamental scientific technology as its predicate device(s). New transducers (C2-7-D, C2-7VN-D, C3-10-D, L3-9i-D) are described as "incremental improvements to the existing transducers and are equivalent." Two new software features (AutoEF and AFI) have been migrated from a predicate (Vivid E9 K131514).
    Safety and Effectiveness: Compliance with safety standards and acoustic power limits.The device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness, as well as thermal, electrical, electromagnetic, and mechanical safety. It has been found to conform with applicable medical device safety standards (AAMI/ANSI ES60601-1, IEC60601-1-2, IEC60601-2-37, NEMA UD 3, ISO10993-1, NEMA UD 2, ISO14971). Acoustic power levels are below applicable FDA limits. The system has been designed in compliance with approved electrical and physical safety standards.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This document explicitly states: "The subject of this premarket submission, LOGIQ E9, did not require clinical studies to support substantial equivalence." Therefore, there is no test set, sample size, or data provenance from a new clinical study provided in this submission for device performance evaluation. The substantial equivalence relies on demonstrating that the new device operates similarly and safely as the predicate devices, which were previously cleared.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable, as no new clinical study (and thus no "test set" and "ground truth" to establish) was required for this substantial equivalence submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as no new clinical study was conducted.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a diagnostic ultrasound system, not an AI-assisted diagnostic tool in the sense of image interpretation. The "AutoEF and AFI" mentioned are software features but there's no MRMC study detailed here to show improved human reader performance.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable, as this is a diagnostic ultrasound imaging system, not a standalone algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable, as no new clinical study was required for this substantial equivalence submission. The "ground truth" for the predicate devices' clearance would have been established through methods appropriate for diagnostic ultrasound systems at the time of their original clearance (e.g., comparison to other imaging modalities, clinical correlation).

    8. The sample size for the training set

    Not applicable. This document describes a medical device, not a machine learning algorithm that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for an algorithm described in this submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K142160

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-10-10

    (65 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended for use by a qualified physician for ultrasound evaluation of Fetal; Abdominal; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transrectal; Transvaginal; Transesophageal and Intraoperative (abdominal, thoracic and vascular).

    Device Description

    The LOGIQ E9 is a full featured, general purpose diagnostic ultrasound system which consists of a mobile console approximately 58 cm wide, 86cm deep and 141 cm high that provides digital acquisition, processing and display capability. The user interface includes a computer keyboard, specialized controls, 10-inch LCD touch screen and color 19-inch LCD image display.

    AI/ML Overview

    The provided text describes the GE LOGIQ E9 Diagnostic Ultrasound System and its various transducers. It highlights the device's indications for use and a summary of non-clinical tests conducted to prove its safety and substantial equivalence to predicate devices. However, it explicitly states that no clinical studies were required or performed to support substantial equivalence.

    Therefore, I cannot provide details on the acceptance criteria and a study that proves the device meets those criteria, as no such clinical study is reported in the provided document.

    Here's a breakdown of the information that can be extracted or inferred based on your prompt and the document:


    1. A table of acceptance criteria and the reported device performance:

    Since no clinical studies were performed, there are no specific performance metrics derived from such studies to present in a table. The "acceptance criteria" are implicitly met by demonstrating substantial equivalence to predicate devices through non-clinical testing and adherence to recognized standards.

    Acceptance Criteria (Inferred from Substantial Equivalence and Standards Adherence)Reported Device Performance (as stated in document)
    Safety:Adherence to medical device safety standards.
    Acoustic Output below FDA limitsAcoustic power levels are below applicable FDA limits.
    BiocompatibilityBiocompatibility evaluated and found safe.
    Cleaning and Disinfection EffectivenessCleaning and disinfection effectiveness evaluated.
    Thermal, Electrical, Electromagnetic, Mechanical SafetyConforms with applicable medical device safety standards.
    Effectiveness (Substantial Equivalence):Same fundamental scientific technology as predicate devices.
    Intended UseSame clinical intended use as predicate LOGIQ E9.
    Imaging CapabilitiesSame imaging modes as predicate LOGIQ E9 (with Shear Wave Elastography for Aixplorer).
    Transducer Equivalence/ImprovementTransducers are identical to predicate LOGIQ E9 versions, or incremental improvements (C1-6VN-D, C2-9-VN-D).
    Capability for Measurements, Digital Images, Reviewing, ReportingSimilar to predicate LOGIQ E9 systems.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    • Sample Size for Test Set: Not applicable, as no clinical studies were conducted that would involve a "test set" of patients for performance evaluation. Non-clinical tests typically involve phantom, bench, or ex-vivo testing.
    • Data Provenance: Not applicable for patient data, as no clinical studies were conducted. The document details adherence to international and national standards (AAMI/ANSI, IEC, NEMA, ISO), indicating these standards and tests were likely performed according to established protocols in manufacturing or testing facilities.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    Not applicable, as no clinical studies were conducted requiring ground truth established by experts for a patient test set.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    Not applicable, as no clinical studies were conducted requiring adjudication for a patient test set.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. The GE LOGIQ E9 is a diagnostic ultrasound system and the document does not suggest or describe any AI components or MRMC studies for improved human reader performance.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Not applicable. The device is a diagnostic ultrasound system, not an algorithm, and the document does not describe standalone algorithm performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    Not applicable for clinical ground truth, as no clinical studies were described. For non-clinical tests (acoustic output, safety, etc.), the "ground truth" would be defined by the specifications, standards, and established methodologies for each test (e.g., calibrated measurement devices for acoustic output).

    8. The sample size for the training set:

    Not applicable, as no clinical studies involving machine learning or AI models with a "training set" were described.

    9. How the ground truth for the training set was established:

    Not applicable, as no clinical studies involving machine learning or AI models with a "training set" were described.

    Ask a Question

    Ask a specific question about this device

    K Number
    K123564

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-12-18

    (28 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended for use by a qualified physician for ultrasound evaluation of Fetal; Abdominal; Pediatric; Small . Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transrectal; Transvaginal; Transesophageal and Intraoperative (abdominal, thoracic, vascular and neurosurgical).

    Device Description

    The LOGIQ E9 is a full featured, general purpose diagnostic ultrasound system which consists of a mobile console approximately 58 cm wide, 86cm deep and 141 cm high that provides digital acquisition, processing and display capability. The user interface includes a computer keyboard, specialized controls, 10-inch LCD touch screen and color 19-inch LCD image display.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification Submission for the GE LOGIQ E9 Diagnostic Ultrasound System. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving safety and effectiveness through novel clinical studies against acceptance criteria.

    Therefore, the document explicitly states: "The subject of this premarket submission, LOGIQ E9, did not require clinical studies to support substantial equivalence."

    Because no clinical studies were performed, it is not possible to extract the requested information regarding acceptance criteria, device performance, sample sizes, ground truth establishment, or multi-reader multi-case studies.

    The document instead focuses on:

    • Summary of Non-Clinical Tests: Evaluation for acoustic output, biocompatibility, cleaning and disinfection effectiveness, and thermal, electrical, electromagnetic, and mechanical safety.
    • Compliance with Voluntary Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, NEMA UD 3, ISO 10993-1, NEMA UD 2, ISO 14971, and NEMA DICOM Set.
    • Quality Assurance Measures: Risk Analysis, Requirements Reviews, Design Reviews, Testing on unit level (Module verification), Integration testing (System verification), Performance testing (Verification), Safety testing (Verification), and Final Acceptance testing (Validation).
    • Intended Use: A broad range of diagnostic ultrasound evaluations for various anatomical regions and clinical applications, with specific modes of operation (B, M, PW Doppler, CW Doppler, Color Doppler, Color M Doppler, Power Doppler, Combined Modes, Harmonic Imaging, Coded Pulse, and other features like 3D/4D imaging and needle guidance) listed for each supported transducer.

    In summary, the provided text does not contain the information required to populate the table or answer the questions as it pertains to clinical acceptance criteria and a study demonstrating performance against those criteria. The device's clearance was based on demonstrating equivalence, not on de novo clinical performance studies with acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K110943

    Validate with FDA (Live)

    Device Name
    LOGIQ E9
    Manufacturer
    Date Cleared
    2011-07-22

    (109 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended for use by a qualified physician for ultrasound evaluation of Fetal; Abdominal; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transrectal; Transvaginal; Transesophageal and Intraoperative (abdominal, thoracic, vascular and neurosurgical).

    Device Description

    The LOGIQ E9 is a full featured, general purpose diagnostic ultrasound system which consists of a mobile console approximately 58 cm wide, 86cm deep and 141 cm high that provides digital acquisition, processing and display capability. The user interface includes a computer keyboard, specialized controls, 10-inch LCD touch screen and color 19-inch LCD image display.

    AI/ML Overview

    The GE LOGIQ E9 Diagnostic Ultrasound System is a full-featured, general-purpose diagnostic ultrasound system. It consists of a mobile console and provides digital acquisition, processing, and display capabilities. The user interface includes a computer keyboard, specialized controls, a 10-inch LCD touch screen, and a 19-inch LCD image display.

    Here's an analysis of the provided information regarding acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document does not explicitly state specific quantitative acceptance criteria or numerical performance metrics for the LOGIQ E9 system's image quality or diagnostic accuracy. Instead, the "acceptance criteria" are implied through the statement of substantial equivalence to its predicate device(s).

    The reported device performance is that it is considered "as safe, as effective, and performance is substantially equivalent to the predicate device(s)."

    Acceptance Criteria (Implied)Reported Device Performance
    Safety (electrical, thermal, mechanical, biocompatibility etc.)Conforms with applicable medical device safety standards.
    Effectiveness (general ultrasound imaging capabilities)Substantially equivalent to predicate device(s).
    Performance (image quality, functionality across indications)Substantially equivalent to predicate device(s).
    Conformance to voluntary standardsComplies with voluntary standards detailed in sections 9, 11, and 17.

    2. Sample Size Used for the Test Set and Data Provenance

    The document explicitly states: "The subject of this premarket submission, LOGIQ E9, did not require clinical studies to support substantial equivalence."

    Therefore, there is no test set of clinical data from human subjects mentioned in this 510(k) summary. The evaluation focused on non-clinical tests.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    Since no clinical studies were performed on a test set, no experts were used to establish ground truth for a test set in the context of clinical performance evaluation. The evaluation was primarily engineering and performance testing against standards and the predicate device.

    4. Adjudication Method for the Test Set

    As no clinical test set was utilized, there was no adjudication method described for clinical data.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was done. The submission explicitly states no clinical studies were required.

    6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop Performance)

    Since the LOGIQ E9 is a diagnostic ultrasound system and not an AI-powered algorithm for image interpretation, the concept of "standalone (algorithm only without human-in-the-loop performance)" as typically applied to AI/ML devices is not directly applicable.

    The "performance" evaluation for this device was based on engineering verification and validation, demonstrating that the system's technical specifications and imaging capabilities were comparable to its predicate devices, rather than a standalone AI performance study.

    7. Type of Ground Truth Used

    For the non-clinical tests, the "ground truth" would be established by:

    • Engineering specifications and test standards: For acoustic output, thermal, electrical, electromagnetic, and mechanical safety.
    • Biocompatibility testing results: Conformance to established biocompatibility protocols.
    • Cleaning and disinfection effectiveness testing results: Verification against established protocols.
    • Comparison to predicate device's established performance: The predicate device serves as a benchmark for "substantial equivalence."

    No pathology, expert consensus on images, or outcomes data were used as ground truth for clinical performance, as clinical studies were not performed.

    8. Sample Size for the Training Set

    The LOGIQ E9 is a traditional diagnostic ultrasound system, not an AI/ML device that requires a "training set" of data in the same way. Therefore, there is no mention of a training set sample size.

    9. How the Ground Truth for the Training Set Was Established

    As there is no mention of an AI/ML training set, this question is not applicable to this 510(k) submission.

    Summary of Device Evaluation Approach:

    The GE LOGIQ E9 Diagnostic Ultrasound System received 510(k) clearance based on its substantial equivalence to previously cleared predicate devices (K092271 GE LOGIQ E9 Diagnostic Ultrasound System and K052441 GE LOGIQ 7 Diagnostic Ultrasound System). This determination was made primarily through non-clinical tests and verification activities, rather than clinical studies. The non-clinical tests included evaluations for acoustic output, biocompatibility, cleaning and disinfection effectiveness, and various safety aspects (thermal, electrical, electromagnetic, mechanical). The device also underwent a series of internal quality assurance measures such as risk analysis, requirements reviews, design reviews, and various levels of testing (unit, integration, performance, safety, and final acceptance). The core argument for substantial equivalence rested on the device employing the "same fundamental scientific technology as its predicate device(s)" and conforming to applicable safety standards and voluntary standards.