Found 4 results

510(k) Data Aggregation

    K Number
    K223403

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-12-09

    (30 days)

    Product Code
    Regulation Number
    866.3310
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LIAISON® Anti-HAV assay is an in vitro chemiluminescent immunoassay intended for the qualitative detection of total antibodies to hepatitis A (anti-HAV) in human serum and sodium heparin plasma samples using the LIAISON® Analyzer family*. The assay is indicated as an aid in the laboratory diagnosis of current or previous HAV infections in conjunction with other serological and clinical information and to determine the presence of an antibody response to HAV in vaccine recipients.

    This assay is not intended for screening blood or solid or soft tissue donors.

    Device Description

    The DiaSorin LIAISON® XS Analyzer is a fully automated, closed, continuous loading of samples and reagents in vitro diagnostic immunoassay system utilizing chemiluminescent technology to provide rapid sample results. The analyzer uses DiaSorin proprietary reagents in which chemiluminescence of an analyte is measured in a sample by the reaction of a magnetic particle solid phase coated with antigen or antibody and a chemiluminescent tracer. The LIAISON® XS Analyzer is intended for use in professional clinical laboratories only.

    The method for qualitative determination of anti-HAV is a competitive sandwich chemiluminescence immunoassay (CLIA) based on neutralization. The assay uses magnetic particles (solid phase) coated with IgG antibodies to HAV (mouse monoclonal), and a mouse monoclonal anti-HAV antibody conjugate linked to an isoluminol derivative (isoluminolantibody conjugate).

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a modified medical device, the LIAISON® XS Analyzer, used with the LIAISON® Anti-HAV assay. However, the document does not contain specific details about acceptance criteria, reported device performance (in terms of sensitivity, specificity, etc.), sample sizes for test sets, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance, or ground truth details for either test or training sets.

    The submission is for a device modification (moving fluid canisters onboard) to an already cleared device (K210272). The focus of the provided text is on demonstrating that these modifications do not negatively impact the device's performance or safety/effectiveness, rather than a full de novo performance study of the Anti-HAV assay itself.

    Therefore, most of the requested information cannot be extracted from this document. The "Summary of Performance Data" section states that "Non-clinical verification and validation activities conducted with the LIAISON® XS Analyzer demonstrate that the modified device met predetermined acceptance criteria," but it does not specify what those criteria were or quantitatively report the performance. It merely lists the types of studies conducted.

    Here is what can be inferred or stated based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly list the acceptance criteria or quantitative performance results (e.g., sensitivity, specificity, accuracy) for the LIAISON Anti-HAV assay after the modifications. It broadly states: "Non-clinical verification and validation activities conducted with the LIAISON® XS Analyzer demonstrate that the modified device met predetermined acceptance criteria, supporting equivalency of the modified device to the cleared device." And "Testing verified all acceptance criteria were met."

    The primary goal of this 510(k) is to demonstrate that the modifications to the analyzer (moving fluid canisters onboard) do not alter the safety and effectiveness of the existing cleared device. The previous clearance (K210272) would have contained the detailed performance data for the LIAISON® Anti-HAV assay itself.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not provided in this document. The document refers to "non-clinical verification and validation activities" which are typically internal testing, not necessarily clinical studies with patient test sets.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable and not provided. This information would be relevant for a de novo clinical study with expert ground truth, which is not the focus of this modification submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable and not provided.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The LIAISON® XS Analyzer is an in vitro diagnostic immunoassay system, not an AI-assisted diagnostic tool that requires human reader interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The LIAISON® Anti-HAV assay on the LIAISON® XS Analyzer is a standalone diagnostic test. Its performance is evaluated based on its accuracy in detecting antibodies, as indicated by the chemiluminescence signal, and does not involve human interpretation of complex images or signals in the same way an AI algorithm might. The document does not provide the specific performance metrics (e.g., sensitivity, specificity, NPV, PPV) for this standalone device in the context of this specific 510(k) submission, as it refers to these having been established in the previous clearance (K210272).

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not explicitly stated in this specific document. For an immunoassay like this, the ground truth for clinical studies would typically be established through a combination of:

    • Established reference methods: Usually another FDA-cleared or gold standard HAV antibody test.
    • Clinical diagnosis: Based on patient symptoms, epidemiological information, and other laboratory markers.
    • Seroconversion panels: Well-characterized samples from individuals demonstrating progression of infection or immune response.

    Since this 510(k) is for a modification to an existing device, it relies on the ground truth established during the original clearance of the LIAISON® Anti-HAV assay.

    8. The sample size for the training set

    Not applicable and not provided. Immunoassays are not "trained" in the same way machine learning models are. Performance characteristics are established through various analytical and clinical studies.

    9. How the ground truth for the training set was established

    Not applicable and not provided (see point 8).

    Ask a Question

    Ask a specific question about this device

    K Number
    K210272

    Validate with FDA (Live)

    Device Name
    LIAISON Anti-HAV
    Manufacturer
    Date Cleared
    2021-02-09

    (8 days)

    Product Code
    Regulation Number
    866.3310
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LIAISON® Anti-HAV assay is an in vitro chemiluminescent immunoassay intended for the qualitative detection of total antibodies to hepatitis A (anti-HAV) in human serum and sodium heparin plasma samples using the LIAISON® Analyzer family. The assay is indicated as an aid in the laboratory diagnosis of current or previous HAV infections in conjunction with other serological and clinical information and to determine the presence of an antibody response to HAV in vaccine recipients.

    This assay is not intended for screening blood or solid or soft tissue donors.

    Device Description

    The DiaSorin LIAISON® XS Analyzer is a fully automated, closed, continuous loading of samples and reagents in vitro diagnostic immunoassay system utilizing chemiluminescent technology to provide rapid sample results. The analyzer uses DiaSorin proprietary reagents in which chemiluminescence of an analyte is measured in a sample by the reaction of a magnetic particle solid phase coated with antigen or antibody and a chemiluminescent tracer. The LIAISON® XS Analyzer is intended for use in professional clinical laboratories only.

    The method for qualitative determination of anti-HAV is a competitive sandwich chemiluminescence immunoassay (CLIA) based on neutralization. The assay uses magnetic particles (solid phase) coated with IgG antibodies to HAV (mouse monoclonal), and a mouse monoclonal anti-HAV antibody conjugate linked to an isoluminol derivative (isoluminol-antibody conjugate).

    AI/ML Overview

    The information provided pertains to the DiaSorin LIAISON® Anti-HAV assay running on the LIAISON® XS Analyzer. This premarket notification is a "Special 510(k)" for device modifications to the existing LIAISON® XS analyzer (K193532), primarily addressing improvements in reliability related to the reagent pipettor. The LIAISON® Anti-HAV assay component and procedures themselves remain unchanged.

    Here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    ParameterAcceptance criteriaReported Device PerformanceAcceptance criteria met?
    Analytical Sensitivity, as concentration at cut off threshold vs WHO standard preparationAnalytical sensitivity in the range 15.5 - 21.5 mIU/mlRun 1: 21 mIU/mLRun 2: 20 mIU/mLYes
    Total precision, as value of the percentage coefficient of variation (CV)≤14.5%3.3 - 7.2%Yes
    Positive agreement≥95%97.0%Yes
    Negative agreement≥95%98.2%Yes

    2. Sample size used for the test set and the data provenance

    The document does not explicitly state the sample size used for the test set for the immunometrical performance assessment.

    • Data Provenance: Not specified. It's unclear if the data is retrospective or prospective, or the country of origin.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable for this type of in vitro diagnostic device (immunoassay). Ground truth for these assays is typically established by reference methods or clinical diagnosis, not by experts reviewing images or other data.

    4. Adjudication method for the test set

    Not applicable for this type of in vitro diagnostic device. Result determination is quantitative or qualitative based on the assay's output measurements against a defined cutoff, not through expert adjudication of individual cases.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

    No, an MRMC comparative effectiveness study was not done. This type of study is typically associated with imaging devices or AI-assisted diagnostic tools where human readers interpret results. The LIAISON® Anti-HAV assay is an automated chemiluminescent immunoassay.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the performance presented in Table 6-3 represents the standalone performance of the LIAISON® Anti-HAV assay on the modified LIAISON® XS Analyzer. This device is an automated immunoassay system, and its output is directly interpreted as a qualitative detection of antibodies.

    7. The type of ground truth used

    The ground truth for the immunometrical performance assessment:

    • Analytical Sensitivity: Established against a "WHO standard preparation."
    • Positive/Negative Agreement: Implied to be established against a reference method or clinical diagnosis for Hepatitis A infection, as is standard for serological assays. The document does not explicitly name the specific reference method used for establishing positive and negative agreement.

    8. The sample size for the training set

    The document does not provide information regarding a separate "training set" or its sample size. For an immunoassay like this, the development likely involves optimization and validation steps, but not a distinct "training set" in the machine learning sense. The performance data presented is for the evaluation of the validated device.

    9. How the ground truth for the training set was established

    As there is no explicit mention of a training set in the context of machine learning, there is no information on how its ground truth was established. The development of such an assay involves careful analytical and clinical validation, ensuring its performance aligns with established diagnostic standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K193532

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-03-02

    (73 days)

    Product Code
    Regulation Number
    866.3310
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LIAISON® Anti-HAV assay is an in vitro chemiluminescent immunoassay intended for the qualitative detection of total antibodies to hepatitis A (anti-HAV) in human serum and sodium heparin plasma samples using the LIAISON® Analyzer family. The assay is indicated as an aid in the laboratory diagnosis of current or previous HAV infections in conjunction with other serological and clinical information and to determine the presponse to HAV in vaccine recipients.

    The assay is not intended for screening blood or solid or soft tissue donors.

    Device Description

    The method for qualitative determination of anti-HAV is a competitive sandwich chemiluminescence immunoassay (CLIA) based on neutralization. The assay uses magnetic particles (solid phase) coated with IqG antibodies to HAV (mouse monoclonal), and a mouse monoclonal anti-HAV antibody conjugate linked to an isoluminol derivative (isoluminol-antibody conjuqate). The first incubation step consists of adding the HAV antigen to calibrators, samples or controls, during which anti-HAV present in calibrators, samples or controls binds to a fixed and limited amount of HAV. thus forming an HAV-anti-HAV immune complex. After this step the second incubation follows and it involves addition of magnetic microparticles and conjugate into the cuvette, during which the antibody conjugate and the solid-phase antibody compete with anti-HAV present in the specimen for HAV. This allows the conjugate to bind to the solid phase and to form a sandwich. If all HAV added is sequestered in an HAV-anti-HAV immune complex during the first incubation, no sandwich is formed during the second incubation. After the second incubation, the unbound material is removed with a wash cycle. Subsequently, the starter reagents are added and a flash chemiluminescence reaction is thus induced. The light signal, and hence the amount of isoluminol-antibody conjugate, is measured by a photomultiplier as relative light units (RLU) and is inversely indicative of anti-HAV present in calibrators, samples or controls.

    AI/ML Overview

    Here's an analysis of the provided text regarding the DiaSorin Inc. LIAISON® Anti-HAV device, outlining acceptance criteria and study details:

    Acceptance Criteria and Device Performance

    The provided document describes two main performance studies: a Method Comparison study and a Reproducibility study. The "acceptance criteria" are implied by the reported agreement percentages and coefficient of variation (%CV) values, which are generally expected to be high for agreement and low for variation in diagnostic assays.

    Table of Acceptance Criteria and Reported Device Performance

    Performance MetricAcceptance Criteria (Implied)Reported Device Performance
    Method Comparison
    Negative AgreementHigh agreement (e.g., >85-90%) with predicate.97.4% (38/39) 95% Cl: 86.8% to 99.5%
    Positive AgreementHigh agreement (e.g., >85-90%) with predicate.96.7% (58/60) 95% Cl: 88.6% to 99.1%
    Overall AgreementHigh agreement (e.g., >90%) with predicate.97.0% (96/99) 95% Cl: 91.5% to 99.0%
    Reproducibility
    Repeatability (within Day)Low %CV (e.g., <10-15%) for various sample concentrations.Ranged from 1.3% to 2.4% for samples; 1.7% for Negative Control, 2.4% for Positive Control.
    Between Day %CVLow %CV (e.g., <10-15%) for various sample concentrations.Ranged from 3.5% to 6.2% for samples; 3.5% for Negative Control, 4.7% for Positive Control.
    Between Laboratory %CVLow %CV (e.g., <10-15%) across different testing sites.Ranged from 0.0% to 5.0% for samples; 2.1% for Negative Control, 4.6% for Positive Control.
    Reproducibility (Total) %CVLow %CV (e.g., <15-20%) representing overall precision.Ranged from 4.4% to 7.1% for samples; 4.5% for Negative Control, 7.0% for Positive Control.

    Study Details for LIAISON® Anti-HAV

    Here's a breakdown of the specific information requested, based on the provided text:

    2. Sample Size Used for the Test Set and Data Provenance

    • Method Comparison Test Set: 100 frozen serum samples.

    • Data Provenance: Samples were "either selected or prepared by DiaSorin Inc." The text does not explicitly state the country of origin or if the samples were retrospective or prospective, but the selection/preparation by the manufacturer suggests a controlled, potentially retrospective or spiked sample set rather than a purely prospective real-world patient cohort. The testing was performed at "2 external sites and at DiaSorin Inc."

    • Reproducibility Study Test Set: A coded precision panel consisting of seven (7) serum specimens manufactured by DiaSorin S.p.A. and two (2) kit controls (a positive and negative from a single control lot). Each sample was run with 4 replicates per day for 12 days at each of the 3 sites, totaling 144 replicates per sample (4 replicates/day * 12 days * 3 sites) for both samples and controls.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    The studies described are for an in vitro diagnostic (IVD) immunoassay, not an imaging device or AI-driven diagnostic that typically involves human expert interpretation for ground truth. Therefore, no human experts were used to establish ground truth in the traditional sense for these performance studies.

    • In the method comparison study, the ground truth for samples was based on the results from a predicate device (LIAISON® Anti-HAV, Reference K082049).
    • In the reproducibility study, the "ground truth" for the precision panel samples would be their known or assigned concentration/status based on their manufacturing and characterization, against which the assay's consistency is measured. The text doesn't specify how these concentrations were established, but it would typically be through reference methods or rigorous characterization during manufacturing.

    4. Adjudication Method for the Test Set

    Given that these are performance studies for an in vitro diagnostic where comparison is primarily against a predicate device or known sample characteristics, no human adjudication method (e.g., 2+1, 3+1) was mentioned or would typically be applicable. The "adjudication" is effectively the direct comparison of results between the new device and the predicate device or the statistical analysis of quantitative measurements against expected values (for reproducibility).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. MRMC studies are typically performed for medical imaging or AI diagnostics where multiple human readers interpret cases and their performance is compared with and without AI assistance. This document describes an in vitro diagnostic assay that determines the presence of antibodies in serum/plasma, not a technology that human readers interpret.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    The LIAISON® Anti-HAV assay itself is a standalone in vitro diagnostic device, meaning its performance is evaluated based on its own output (chemiluminescence signal leading to qualitative detection of antibodies) without direct human interpretation of complex patterns or images. The "algorithm" in this context refers to the assay's chemical and instrumental process, which produces results independently. The provided studies directly assess this standalone performance (e.g., agreement with a predicate, and reproducibility).

    7. The Type of Ground Truth Used

    • Method Comparison Study: The ground truth for the test set was established by the predicate device (LIAISON® Anti-HAV, Reference K082049) results. This is a common practice for demonstrating substantial equivalence for IVDs.
    • Reproducibility Study: The ground truth, in terms of expected values for precision, was based on the known characteristics of the manufactured serum specimens and kit controls. This is not "expert consensus" or "pathology" but rather scientifically characterized samples.

    8. The Sample Size for the Training Set

    The document does not specify a training set sample size. This is expected because the LIAISON® Anti-HAV assay is a chemiluminescent immunoassay, which is a traditional laboratory diagnostic method based on biochemical reactions, not a machine learning or AI algorithm that requires a "training set" in the context of model development. The development of such an assay involves extensive R&D, reagent optimization, and analytical validation but not a "training set" in the computational sense.

    9. How the Ground Truth for the Training Set Was Established

    Since there is no "training set" in the AI/ML sense, this question is not applicable based on the provided information. The development of the assay would have relied on established scientific principles, method validation, and characterization of reagents and controls.

    Ask a Question

    Ask a specific question about this device

    K Number
    K082049

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2008-12-05

    (140 days)

    Product Code
    Regulation Number
    866.3310
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LIAISON® Anti-HAV assay is an in vitro chemiluminescent immunoassay intended for the qualitative detection of total antibodies to hepatitis A (anti-HAV) in human serum and sodium heparinized plasma samples using the automated LIAISON® Analyzer. The assay is indicated as an aid in the laboratory diagnosis of current or previous HAV Infections in conjunction with other serological and clinical information and to determine the presence of an antibody response to HAV in vaccine recipients.

    This assay is not intended for screening blood or solid or soft tissue donors. Assay performance characteristics have not been established for immunocompromised or immunosuppressed patients. The user is responsible for establishing their own assay performance characteristics in these populations.

    The LIAISON® Control Anti-HAV (negative and positive) is intended for use as assayed quality control samples to monitor the performance of the LIAISON® Anti-HAV assay.

    Device Description

    The method for qualitative determination of anti-HAV is a competitive sandwich chemiluminescence immunoassay (CLIA) based on neutralization. The assay uses magnetic particles (solid phase) coated with IgG antibodies to HAV (mouse monoclonal), and a mouse monoclonal anti- HAV antibody conjugate linked to an isoluminol derivative (isoluminol-antibody conjugate).

    The first incubation step consists of adding the HAV antigen to calibrators, samples or controls, during which anti-HAV present in calibrators, samples or controls binds to a fixed and limited amount of HAV, thus forming an HAV-anti-HAV immune complex.

    After this step the second incubation follows and it involves addition of magnetic microparticles and conjugate into the reaction module, during which the antibody conjugate and the solid-phase antibody compete with anti-HAV present in the specimen for HAV. This allows the conjugate to bind to the solid phase and to form a sandwich. If all HAV added is sequestered in an HAV-anti-HAV immune complex during the first incubation, no sandwich is formed during the second incubation. After the second incubation, the unbound material is removed with a wash cycle.

    Subsequently, the starter reagents are added and a flash chemiluminescence reaction is thus induced. The light signal, and hence the amount of isoluminolantibody conjugate, is measured by a photomultiplier as relative light units (RLU) and is inversely indicative of anti-HAV present in calibrators, samples or controls.

    AI/ML Overview

    The provided text describes the performance data for the DiaSorin LIAISON® Anti-HAV assay. However, it does not explicitly state specific acceptance criteria (e.g., "Positive agreement must be >= 95%"). Instead, it reports the observed performance (percent agreement and confidence intervals) which implies these results were deemed acceptable by the regulatory body for 510(k) clearance based on substantial equivalence to a predicate device.

    Therefore, the table below will present the reported device performance, and the description will elaborate on the various studies conducted to demonstrate this performance.


    Acceptance Criteria and Study Details for DiaSorin LIAISON® Anti-HAV Assay

    1. Table of Acceptance Criteria and Reported Device Performance

    As explicit quantitative acceptance criteria (e.g., "The positive agreement must be at least X%") are not stated in the provided document, the table will reflect the reported performance that was found to be sufficient for substantial equivalence. The document presents "Percent Agreement" values along with their 95% Confidence Intervals as the primary performance metrics. The underlying acceptance for these values is implicit through the 510(k) clearance process, which confirms the device is safe and effective as a predicate device.

    Performance Metric CategoryReported Device Performance (Percent Agreement)95% Confidence Interval
    Prospective Population
    At Risk / HAV Testing (Positive)96.6%93.5% - 98.0%
    At Risk / HAV Testing (Negative)99.0%97.9% - 99.6%
    Pediatric Population (Positive)78.6%49.2% - 95.3%
    Pediatric Population (Negative)96.8%92.0% - 99.1%
    Retrospective Population
    Current/Previous HAV Infection100.0%98.0% - 100%
    Pediatric Current/Previous HAV Inf.100.0%98.0% - 100%

    Note: The low positive agreement for the Pediatric Population (78.6%) with a wide confidence interval (49.2 – 95.3%) indicates a smaller number of positive samples in that specific cohort, but was still accepted for clearance.

    2. Sample Sizes and Data Provenance

    • Prospective Studies:
      • HAV Testing and At Risk Populations: 739 samples.
        • 500 excess serum samples from individuals in the Northeastern U.S. sent for HAV testing.
        • 239 samples from individuals at risk for viral hepatitis (homosexual males, healthcare workers, commercial sex workers, drug users, prison inmates, dialysis patients, hemophiliacs).
      • Pediatric Population: 108 samples from children in the United States.
      • Vaccine Study: 73 individuals (9 TWINRIX sets, 32 HAVRIX sets, 32 VAQTA sets of pre- and post-vaccine samples).
    • Retrospective Studies:
      • Current/Previous HAV Infection (Adults): 109 samples from adults in Eastern U.S. and Egypt.
      • Current HAV Infection (Pediatric): 42 samples from pediatric patients in Egypt.
    • Expected Values (Prevalence Study): 802 apparently healthy adults (301 from Western U.S., 501 from Eastern U.S.).

    Data Provenance: The data is a mix of prospective and retrospective collections. Geographically, samples were from the Northeastern U.S., other unspecified regions of the U.S. (for at-risk groups and pediatric prospective), Eastern U.S., Western U.S., and Egypt.

    3. Number of Experts and Qualifications for Ground Truth

    The document does not explicitly state the number or qualifications of experts used to establish the ground truth for the test sets. Instead, the "Comparator ELISA" assay is consistently referred to as the reference method for determining the true positive, negative, or equivocal status of samples. This implies that the accepted clinical standard for Hepatitis A testing at the time (the predicate ELISA assay) served as the ground truth.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method involving multiple experts. The comparator ELISA assay results were the reference. For samples that yielded "Equivocal" or "Borderline" results by the comparator ELISA, these were sometimes retested per the Instructions for Use of the comparator method. Samples that remained equivocal or borderline after retesting were handled as distinct categories in the comparison tables.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    A multi-reader multi-case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance was not done. This device is an in vitro diagnostic (IVD) assay designed to be read by automated equipment (LIAISON® Analyzer) and interpreted by laboratory professionals, not primarily by human "readers" interpreting images or complex data in the same way an AI for medical imaging would. The performance is compared to a predicate assay, not human interpretation.

    6. Standalone (Algorithm Only) Performance

    Yes, a standalone performance study was done. The entire performance evaluation section ("Performance Data") describes the performance of the LIAISON® Anti-HAV assay as a standalone algorithm/device, comparing its results directly against those of a predicate ELISA assay on various sample populations. There is no human-in-the-loop component described for these performance evaluations.

    7. Type of Ground Truth Used

    The ground truth was established by a comparator ELISA assay, which is referred to throughout the document as the reference method. Specifically, the predicate device, DiaSorin Inc. ETI-AB-HAVK Plus assay (PMA #P890019/S05), served as the standard for comparison.

    8. Sample Size for the Training Set

    The document does not specify a separate "training set" size. For in vitro diagnostic (IVD) assays, the development process typically involves internal method validation and optimization by the manufacturer, rather than a distinct "training set" in the machine learning sense. The samples described in the "Performance Data" section are effectively the "test set" or clinical validation set used to demonstrate performance for regulatory submission.

    9. How Ground Truth for the Training Set Was Established

    As no explicit training set is mentioned in the context of machine learning, this question is not directly applicable. For the overall assay development and validation, the ground truth was established by comparison to existing, clinically accepted methods, presumably the predicate ELISA assay during earlier stages of development and optimization.