510(k) Data Aggregation

    K Number
    K243978

    Validate with FDA (Live)

    Device Name
    LED Facial Mask
    Date Cleared
    2025-06-20

    (179 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LED Facial Mask is an over-the-counter (OTC) device intended for the following uses:

    For Red Mode(s): The device emits energy in the red spectrum and is intended for the treatment of full-face wrinkles.

    For Red + NIR Mode: The device emits energy in the red and infrared spectrum and is intended for the treatment of full-face wrinkles.

    For Blue Mode: The device emits light in the blue region of the spectrum and is specifically indicated for the treatment of mild to moderate acne on the face.

    For NIR Mode: The device is intended to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.

    Device Description

    The LED Facial Mask is a device used for treatment of full-face wrinkles, mild to moderate acne on the face, and providing topical heating. LEDs housed inside the device emit light onto the face. These LEDs generate blue, red, and infrared wavelengths.

    This product is composed of the main device, controller, fastening straps, storage bag, blindfold, and charging cable.

    AI/ML Overview

    This FDA 510(k) clearance letter pertains to an LED Facial Mask, which is categorized as a Class II medical device. The information provided outlines the manufacturing, intended use, and comparison to predicate devices, but it does not contain the specific performance data one would expect for a study proving the device meets acceptance criteria.

    The document primarily focuses on demonstrating substantial equivalence to already cleared predicate devices based on technological characteristics and intended use. The "Performance Data" section lists various recognized consensus standards to which the device was tested (e.g., electrical safety, electromagnetic compatibility, usability, biological evaluation), rather than clinical study results establishing efficacy.

    Therefore, I cannot create the detailed table of acceptance criteria and reported device performance, nor can I answer questions about sample size, ground truth establishment, expert adjudication, or MRMC studies, as this information is not present in the provided FDA clearance letter.

    It explicitly states: "Clinical/animal testing was not performed for the proposed device as part of the submission." This confirms the absence of efficacy studies that would typically define acceptance criteria for performance outcomes.

    The acceptance criteria for this device, as implied by the FDA 510(k) process for a Class II device, are largely related to safety, performance against engineering standards, and substantial equivalence to legally marketed predicate devices. The clearance is based on the premise that if the new device is sufficiently similar to existing cleared devices in terms of its technology and intended use, and meets relevant safety standards, it does not require new clinical efficacy data.

    Here's a breakdown of what can be inferred from the document regarding acceptance criteria and the "study" (non-clinical testing) that proves the device meets them:


    Inferred Acceptance Criteria and Device "Performance" (Non-Clinical)

    Since no clinical efficacy studies were performed, the "acceptance criteria" are not related to a specific clinical outcome (e.g., X% reduction in wrinkles, Y% clearance of acne). Instead, they relate to safety, electrical performance, and biocompatibility, as demonstrated through testing against consensus standards.

    Acceptance Criterion (Inferred from Standards)Reported Device Performance (Implied by Clearance)
    Electrical Safety (IEC 60601-1, -2-57, -2-83)Device meets general requirements for basic safety and essential performance, particular requirements for non-laser light source equipment, and home light therapy equipment. Implies safe power delivery, insulation, and fault protection.
    Electromagnetic Compatibility (IEC 60601-1-2)Device meets requirements for electromagnetic disturbances. Implies suitable electromagnetic emissions and immunity to interference.
    Usability (IEC 60601-1-6)Device meets usability requirements. Implies the device is designed for safe and effective user interaction.
    Home Healthcare Environment Suitability (IEC 60601-1-11)Device meets requirements for use in the home healthcare environment. Implies robustness and safety for lay users.
    Photobiological Safety (IEC 62471)Device meets photobiological safety standards. Implies emitted light is safe for the eyes and skin under specified use conditions.
    Software Life Cycle Processes (IEC 62304)Device software development followed medical device software life cycle processes. Implies software is engineered to be safe and reliable.
    Biocompatibility (ISO 10993-1, -5, -10, -23)Materials in contact with the user are biologically evaluated against standards for cytotoxicity, skin sensitization, and irritation. Implies materials are safe for human contact.
    Wavelength AccuracyBlue: 410 nm, 460 nm; Red: 610 nm, 630 nm, 660 nm; NIR: 830 nm, 850 nm, 880 nm. (Measured against specified ranges)
    Intensity/Irradiance (mW/cm²)Red1 (Y) Mode: 36.0 mW/cm²; Red2 (R) Mode: 19.0 mW/cm²; Blue (B) Mode: 6.1 mW/cm²; NIR Mode: 45.0 mW/cm²; Red1 (Y)+NIR Mode: 81.0 mW/cm²; Red2 (R)+NIR Mode: 64.0 mW/cm². (Measured against specified output)
    Substantial Equivalence to Predicate DevicesDevice has similar intended use, technological characteristics (light source, anatomical location, wavelength, intensity/irradiance), and safety profile to previously cleared devices.

    Study Details (Based on the Provided Document)

    1. Sample size used for the test set and the data provenance:

      • Test Set Sample Size: Not applicable in the context of clinical data for efficacy. The "test set" here refers to the device itself being tested against engineering and safety standards. There is no patient sample size.
      • Data Provenance: The tests conducted are in vitro (bench testing) and in silico (software analysis) or conducted on the physical device itself, not on human subjects. Therefore, there's no country of origin for patient data, nor is it retrospective or prospective clinical data.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable. The "ground truth" for the engineering and safety tests is defined by the consensus standards themselves (e.g., a certain voltage tolerance, a specific emission limit). The "experts" would be the engineers and technicians performing the tests and verifying compliance with the standards, not clinical experts establishing ground truth for medical conditions.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable. This concept pertains to clinical studies where human interpretation of medical data needs to be harmonized. For engineering tests, results are typically objective measurements against a standard.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This type of study is for AI/CADe devices and involves human readers reviewing cases. This device is an LED facial mask, not an AI diagnostic tool.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • No. This applies to AI/CADe devices.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for this clearance is adherence to recognized consensus standards for safety, electrical performance, photobiological safety, and biocompatibility, as well as demonstrated technical similarity and intended use alignment with predicate devices. There is no clinical ground truth (e.g., biopsy results, clinician's diagnosis) involved in this substantial equivalence determination.
    7. The sample size for the training set:

      • Not applicable. There is no AI model or "training set" for this device.
    8. How the ground truth for the training set was established:

      • Not applicable.

    In summary, the provided FDA 510(k) clearance document for the LED Facial Mask demonstrates that the device was cleared based on its substantial equivalence to existing devices and its adherence to a series of non-clinical safety and performance standards. No clinical efficacy studies were conducted or required for this particular clearance, as explicitly stated in the document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230351

    Validate with FDA (Live)

    Device Name
    LED Facial Mask
    Date Cleared
    2023-05-05

    (85 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    LED Facial Mask is an over the counter device that is intended to use LED light for the treatment of wrinkles and mild to moderate acne.

    Device Description

    The LED Facial Mask is an over-the-counter, facemask-shaped design light emitting diode (LED) device that emits light energy in the red, blue, and amber spectrum for the treatment of wrinkles or mild to moderate inflammatory acne on the face. The device consists of a mask body unit, controller, and AC adapter. The mask body unit contains light emitting diodes (LEDs) that emit visible red light (650nm+/-5nm), amber light (605nm+/-5nm), or blue light (465nm+/-5nm) to help reduce the appearance of wrinkles or mild to moderate acne. A controller is connected to the mask body unit to control the device, such as turn on/off the device, switch LED color output. An AC adapter is used to power the device. To use the device, user should place the mask body unit over the face and use the controller to operate. The device will automatically turn off after each treatment.

    The LED Facial Mask includes MZ-01, NEWKEY-01, and SP-FM-01 models. Their intended use, performance, structure design, and operation are identical, with the difference being the product appearance, but these differences do not affect or change the intended use of the device.

    AI/ML Overview

    The provided text is a 510(k) summary for an LED Facial Mask, a medical device for over-the-counter use in treating wrinkles and mild to moderate acne. However, it does not contain information about studies that prove the device meets acceptance criteria related to its clinical efficacy.

    Instead, the document focuses on demonstrating substantial equivalence to predicate devices through technical specifications, safety data (biocompatibility, electrical safety, eye safety), and software verification and validation. There's no mention of clinical trials or performance studies evaluating the device's effectiveness in treating wrinkles or acne in accordance with specific acceptance criteria beyond general safety and design similarity to other cleared devices.

    Therefore, I cannot provide the requested information regarding acceptance criteria and the study proving the device meets them for its intended clinical effects because such information is not present in the provided text. The document primarily details technical and safety compliance, not clinical performance metrics.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213056

    Validate with FDA (Live)

    Device Name
    LED Facial Mask
    Date Cleared
    2022-07-07

    (288 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    LED Facial Mask is an over the counter device. The Red light mode is intended for the treatment of periorbital wrinkles. The Red and Infrared light combination mode is intented for the treatment of full-face wrinkles. The blue light is intended for the treatment of mild to moderate inflammatory acne. The device is indicated for adults only.

    Device Description

    LED Facial Mask (Model: AST-804, AST-804A, AST-804B, AST-805, AST-805A, AST-805B)

    AI/ML Overview

    The provided text is an FDA 510(k) clearance letter for an LED Facial Mask. It does not contain information about acceptance criteria, device performance studies, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth establishment, or training set details.

    The letter only states that the device is "substantially equivalent" to a legally marketed predicate device for the following indications for use:

    • Red light mode: Treatment of periorbital wrinkles.
    • Red and Infrared light combination mode: Treatment of full-face wrinkles.
    • Blue light: Treatment of mild to moderate inflammatory acne.
    • Target population: Adults only.

    Therefore, I cannot fulfill your request to describe the acceptance criteria and the study that proves the device meets them based on the provided text. This information would typically be found in the 510(k) submission itself, not in the clearance letter.