Found 2 results

510(k) Data Aggregation

    K Number
    K140141

    Validate with FDA (Live)

    Date Cleared
    2014-07-03

    (162 days)

    Product Code
    Regulation Number
    862.1345
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LabonaCheck® Gluppy Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip. The LabonaCheck® Gluppy Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.

    The LabonaCheck® Gluppy Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The LabonaCheck® Gluppy Blood Glucose Monitoring System should not be used for the diagnosis or screening of diabetes or for neonatal use.

    The LabonaCheck® Gluppy Blood Glucose Test Strips are for use with the LabonaCheck® Gluppy Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip.

    Device Description

    The system consists of Test Meter, Test Strips, Lancing Device, lancets, 3 V battery and Carrying Case. The system measures the amount of glucose (sugar) in whole blood. Blood is applied to the absorbent hole of the test strip and automatically drawn into the reaction zone where reaction between reagent and glucose occurs.

    AI/ML Overview

    The provided text is a 510(k) summary for the LabonaCheck® Gluppy Blood Glucose Monitoring System. It does not contain information about the specific acceptance criteria, a study proving the device meets those criteria, or details regarding sample sizes for training/test sets, ground truth establishment, or expert involvement as requested.

    The document primarily focuses on establishing substantial equivalence to a predicate device (K102751) by comparing various features and performance characteristics. It states that the "Performance Characteristics" are the same as the predicate and refers to "N. Performance Characteristics on 6 page" which is not provided in the input text.

    Therefore, I cannot fulfill your request for the specific details regarding acceptance criteria, study design, and performance results. This document is a regulatory submission for premarket notification, not a detailed scientific study report.

    Ask a Question

    Ask a specific question about this device

    K Number
    K102751

    Validate with FDA (Live)

    Date Cleared
    2013-02-11

    (872 days)

    Product Code
    Regulation Number
    862.1345
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LabonaCheck Gluppy Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip. The LabonaCheck Gluppy Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.

    The LabonaCheck Gluppy Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The LabonaCheck Gluppy Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes or for neonatal use.

    The LabonaCheck Gluppy Blood Glucose Test Strips are for use with the LabonaCheck Gluppy Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip.

    The LabonaCheck Gluppy Control Solution is for use with the LabonaCheck Gluppy Blood Glucose Monitoring System as a quality control check to verify that the meter and test strips are working together properly.

    Device Description

    The system consists of Test Meter, Test Strips, Code key, Lancing Device, lancets, 3V battery and Carrying Case.

    The system measures the amount of glucose (sugar) in whole blood. Blood is applied to the absorbent hole of the test strip and automatically drawn into the reaction zone where reaction between reagent and glucose occurs.

    AI/ML Overview

    The provided text is a 510(k) summary for the LabonaCheck Gluppy Blood Glucose Monitoring System. It describes the device, its intended use, and compares it to a predicate device to establish substantial equivalence. However, it does not contain any information about acceptance criteria, a specific study proving the device meets those criteria, sample sizes, expert involvement, adjudication methods, MRMC studies, standalone performance, or training set details.

    The document is a regulatory submission for premarket notification, which focuses on demonstrating substantial equivalence to a legally marketed device rather than providing detailed performance study results with specific acceptance criteria and statistical analysis.

    Therefore, I cannot fulfill your request for the specific points about acceptance criteria and detailed study information based on the provided text. The document primarily focuses on:

    • Device Name: LabonaCheck Gluppy Blood Glucose Monitoring System
    • Manufacturer: CERAGEM Medisys Inc.
    • Submission Date: February 11, 2013
    • Predicate Device: RIGHTEST BLOOD GLUCOSE MONITORING SYSTEM, Model GM 100 (K081451)
    • Indication for Use: Quantitative measurement of glucose in fresh capillary whole blood samples from the fingertip for self-testing by people with diabetes at home (not for diagnosis, screening, or neonatal use).
    • Similarities to Predicate: Amperometry detection method, Glucose Oxidase enzyme, Potassium ferricyanide mediator, test range (20-600 mg/dL), temperature range (10-40°C), noble metal electrode, 3-year warranty, CR2032 battery.
    • Differences from Predicate: Sample volume (1.0 uL vs 1.4 uL), humidity range (Below 80% vs 10-90%), open use time for test strips (4 months vs 3 months), coding (Code Key vs No coding), test time (5 seconds vs 8 seconds), hematocrit range (20-60% vs 30-55%), memory (500 results vs 10 results).

    To answer your questions, a different type of document, such as a validation study report or a more detailed performance specification document, would be required.