510(k) Data Aggregation

    K Number
    K182977

    Validate with FDA (Live)

    Date Cleared
    2018-11-21

    (26 days)

    Product Code
    Regulation Number
    880.5400
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Isolette® 8000 plus is indicated for thermoregulation and controlling oxygen (optional), and humidity (optional) for both premature and full-term infants up to a maximum of 10 kg (22lbs).

    The Isolette® 8000 plus is not intended for home use.

    Device Description

    The Isolette® 8000 plus is a neonatal incubator that provides a controlled environment for both premature and full-term infants up to a maximum of 10 kg (22lbs). It controls temperature, oxygen and humidity (both are optional).

    The predicate device as cleared in K172154 is offered with the hood and shell assembly of the incubator mounted on a variable high adjustable (VHA) trolley. All components and accessories for the device were cleared as part of the system.

    The modification of the device is the replacement of the variable height adjustable (VHA) trolley with a fixed-height cabinet stand trolley. All other components and accessories of the device are identical to those cleared under K172154.

    The modification does not change the intended use or alter the fundamental scientific technology of the currently approved predicate device.

    AI/ML Overview

    This document is a 510(k) premarket notification for a modification to an existing medical device, the Isolette® 8000 Plus Neonatal Incubator. The modification involves replacing the variable height adjustable (VHA) trolley with a fixed-height cabinet stand trolley. No AI/ML components are involved in this device, therefore, many of the requested fields are not applicable.

    Here's a summary based on the provided text:

    1. Table of acceptance criteria and the reported device performance:

    The document focuses on demonstrating that the modified device remains substantially equivalent to its predicate device (Isolette® 8000 plus, K172154) despite the change in the trolley. The acceptance criteria essentially align with the predicate device's established performance and safety standards. The performance of the modified device is reported as "Same" for nearly all specifications, with "Different – No impact on effect on safety and effectiveness" for the few physical attribute changes.

    SpecificationAcceptance Criteria (Predicate Device K172154)Reported Device Performance (Modified Device K182977)Comments
    General Performance
    Air temperature mode set point range20 to 39 C (68-102.2 F)SameN/A
    Warm-up time at 22C (72F) ambient<35minSameN/A
    Variability<0.5C (<0.9F)SameN/A
    Uniformity with a level mattress<0.8C (<1.4F)SameN/A
    Skin Temperature Mode set point range34.0 to 37.0C (93.2 to 98.6F)SameN/A
    Accuracy of incubator temp indication≤0.8CSameN/A
    Noise level within the hood environment<47 dB(A) w/ 37 dB(A) or less ambientSameN/A
    Physical Attributes
    Height133.3 to 153.7 cm (52.5 to 60.5 in)140 cm ± 1.2 cm (55 in ± 0.5 in)Different – No impact on effect on safety and effectiveness.
    Weight<98.5 kg (217.1 lb.) without options/accessories<95.3 kg (210 lb.) without options/accessoriesDifferent – No impact on effect on safety and effectiveness.
    AppearanceVariable Height Stand (VHA) with Isolette 8000 plus incubator mounted on topFixed Height / Cabinet Stand with Isolette 8000 plus incubator mounted on topDifferent – No impact on effect on safety and effectiveness.
    MobilitySupport stand able to raise and lower the incubator on VHA. Storage with swivel drawer storage. Three (3) dual total lock castors, one (1) dual steering castorSupport stand without the ability to raise and lower incubator. Storage ability permitted within cabinet stand. Four (4) Single castors with total lock functionDifferent – No impact on effect on safety and effectiveness.
    Power Routing and Convenience OutletsImage: Incubator and VHAImage: Incubator and Cab StandDifferent – Convenience outlet and actuator/electronics controlling the VHA removed. No impact on effect on safety and effectiveness.
    Safety and EMC
    Instability (mechanical)Per IEC60601-1Verified through non-clinical testingN/A
    Electrical SafetyPer IEC60601-1Verified through non-clinical testingN/A
    Electromagnetic Compatibility (EMC)Per IEC60601-1-2Verified through non-clinical testingN/A

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This document describes a modification to an existing device, not a study involving human or animal subjects that would typically have a "test set" in the context of clinical data. The "test set" here refers to the engineering verification and validation activities. The document summarizes these activities as having been "carried out under well established methods." There is no information regarding the sample size of devices tested for Instability, Electrical Safety, and EMC, nor is there information on data provenance beyond being "non-clinical testing."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable to an engineering verification of a medical device modification. The "ground truth" for the tests performed (Instability, Electrical Safety, EMC) is established by the relevant international standards (IEC60601-1, IEC60601-1-2) which define the pass/fail criteria. Expertise would be in the form of qualified engineers and technicians performing the tests and reviewing the results against these standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is not a study involving human reader interpretation or consensus on data. The verification activities are against predefined technical standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a neonatal incubator, not an AI-powered diagnostic tool, and its modification does not involve human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    The "ground truth" for the non-clinical testing refers to the established standards for medical device safety and performance, specifically IEC60601-1 (general medical electrical equipment safety), IEC60601-1-2 (electromagnetic compatibility), and IEC60601-2-19 (basic safety and essential performance of infant incubators). These standards define the acceptable limits and performance characteristics.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI/ML device that requires a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172154

    Validate with FDA (Live)

    Date Cleared
    2018-04-19

    (276 days)

    Product Code
    Regulation Number
    880.5400
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Isolette® 8000 plus incubator is indicated for thermoregulation and controlling oxygen (optional), and humidity (optional) for both premature and full-term infants up to a maximum of 10 kg (22lbs).

    The Isolette® 8000 plus is not intended for home use.

    Device Description

    The Isolette 8000 plus is a device used to maintain environmental conditions suitable for neonates. The Isolette 8000 plus incubator provides a controlled environment for both premature and full-term infants up to a maximum of 10 kg (22lbs). It controls temperature, oxygen (optional) and humidity (optional). It is a controller-based incubator featuring the ability to enable simultaneous control of temperature, and optional oxygen and humidity. The system operates on AC power. Data is provided on a liquid crystal display. It is operated in closed care therapy as an incubator according to IEC60601-2-19.

    Components include hand ports or iris ports, access panel, serial port, trolley with height adjustment, castor wheels with brakes, power receptacle, a Trendelenburg bed-tilt mechanism (0° to 12°), softbed mattress, skin temperature probes, probe covers, oxygen inlet, sensor module, hose grommets, accessoryrail and cable wrap. Optional components include IV pole, drawers, water reservoir, collection bottler mount litter bag holder, basket for gloves, breathing hose holder kit, tray, high frequency ventilation (HFV) door with grommets, positioning aids, incubator cover, and monitor shelf.

    AI/ML Overview

    The provided text describes the regulatory clearance of the Isolette® 8000 plus neonatal incubator and its comparison to a predicate device (Isolette® Infant Incubator, Model C2HS, K001242) for showing substantial equivalence. It does not contain information about an AI-powered device or a study involving human experts or a training set for machine learning. Therefore, many of the requested details are not applicable as this is a traditional medical device submission without AI.

    However, based on the provided text, I can extract the acceptance criteria and a summary of the non-clinical studies performed to demonstrate that the device meets those criteria.


    Acceptance Criteria and Device Performance for Isolette® 8000 plus (K172154)

    This submission is for a traditional medical device (neonatal incubator) and does not involve AI. Therefore, numerous requested sections related to AI performance, human expert evaluation, and training sets are not applicable.

    1. A table of acceptance criteria and the reported device performance

    The acceptance criteria are generally implied by the stated specifications and compliance with recognized standards. The "Device Under Review" column effectively serves as the reported device performance compared to the predicate.

    Specification (Acceptance Criterion implied by predicate/standards)Predicate (Isolette® Infant Incubator, Model C2HS) PerformanceDevice Under Review (Isolette® 8000 plus) PerformanceComments (Device Meets or is Similar to Predicate)
    Intended UseCare for premature and full-term babies, controlled environment in NICU/Special Care Baby Unit etc.Controlled environment for premature and full-term infants up to 10 kg (22lbs). Controls temperature, oxygen (optional), and humidity (optional). Not for home use.Same - Wording different, but intended use unchanged.
    Indications for UseEffective temperature management, reduces radiant heat loss, minimizes evaporative heat loss, monitors/maintains O2 levels.Thermoregulation and controlling oxygen (optional), and humidity (optional) for premature and full-term infants up to 10 kg (22lbs). Not for home use.Similar - Simplified wording; functional indications unchanged; similar technology.
    Target PopulationPremature and full term infantsNewly born infants up to 10 kg (22lbs)Same - Weight added as information.
    Environment of UseNICU/Special Care Baby Unit, Step Down Nursery, Newborn Nursery and pediatrics.Any department of the hospital that provides neonatal and infant care, including NICU/Special Care Baby Unit, Step Down Nursery, Newborn Nursery and pediatrics.Similar - Broader phrasing for hospital areas.
    Principle of OperationController-based incubator, simultaneous control of temp, O2, humidity.Same as C2HSSame
    Protection ClassClass I, Type BF, continuous, not APSame as C2HSSame
    Ingress of liquids and particulate matter (IEC 60601-1)IPXOSame as C2HSSame
    Height133.4-152.4 cm (52.5-60 in)133.3 to 153.7 cm (52.5 to 60.5 in)Similar - No effect on function.
    Width99 cm (38.0 in)<104 cm (41 in)Similar - No effect on function.
    Depth59.7 cm (23.5in)<76.2 cm (30 in)Similar - No effect on function.
    Weight99 kg (198 lb.)<98.5 kg (217.1 lb.) without options/accessoriesSimilar - No effect on function.
    Mattress size40.6 x 78.7cm(16 x 31.5 in)>38 x 74 x3 cm (15 x 29.1 x 1.2 in)Similar - No effect on function.
    Trendelenburg/Reverse TrendelenburgContinuously variable to 12 deg. +1 deg.Same as C2HSSame
    Infant WeightNot published10 kg (22 lbs.) maximumSame - Specification not previously published.
    Operating temperature20-30°C (68-86°F)Same as C2HSSame
    Operating humidity5-99% RH non-condensing5-95% RH non-condensingSimilar
    Operating altitude3048m (10,000 ft.)Up to 3000 m (9800 ft.)Similar - Meets IEC 60601-1.
    Operating Ambient air pressureNot available110-70 kPaSame - Design elements unchanged.
    Storage temperature-30 to 70°C (-22 to 158°F)-20 to 60°C (-4 to 140°F)Similar - Device packaging marked with requirements.
    Storage humidity0-99% RH, non-condensing5-95% RH non-condensingSimilar
    Storage Ambient air pressureNot available110-50 kPaSame - Design elements unchanged.
    Power req. 100/120V100/120V ± 10%, 50/60 Hz, 1900 W max.100/120V, 50/60 Hz, 1900 W max, 9.9 A maxSimilar - Amperage added; not clinically relevant.
    Auxiliary power socketsAll 50/60 Hz, 100, 120, 220, & 240V ± 10% 500W max.All 50/60 Hz, 100 V, 100W max; 120 & 230 V 300W maxSimilar - Both meet current safety standards.
    Earth Leakage< 300µA 100 & 120V; < 500µA 220 & 240V< 500 µASimilar - Both meet current safety standards.
    Air temperature mode - Set point range20-37°C (68-98.6°F)20 to 39°C (68-102.2°F)Same - I8000 plus publishes entire range.
    Control override range37-39°C (98.6-102.2 F)Same as C2HSSame
    Set point display range15 to 45°C (59-113 F)Same as C2HSSame
    Upwards deviation limitFixed +1.5°C+1.5 to +2.5°C (+2.7 to 4.5°F)Similar - I8000 plus allows adjustment.
    Upwards deviation limit defaultFixed +1.5°C+1.5°C (+2.7°F)Similar - Default value is same.
    Downwards deviation limit rangeFixed -2.5°C-1.5 to -2.5°C (-2.7 to-4.5F)Similar - I8000 plus allows adjustment.
    Downwards deviation limit defaultFixed -2.5°C-2.5°C (-4.5F)Similar - Default value is same.
    Warm-up time at 22°C (72°F) ambient<35minSame as C2HSSame
    Variability<0.5°C (<0.9°F)Same as C2HSSame
    Overshoot<0.5°C (<0.9°F)Same as C2HSSame
    Uniformity with a level mattress<0.8°C (<1.4°F)Same as C2HSSame
    Skin Temperature Mode - Set point range34.0 to 37.0°C (93.2 to 98.6°F)Same as C2HSSame
    Control override temp range37.0 to 38.0°C (98.6 to 100.4°F)Same as C2HSSame
    Set point display range (skin temp)15.0 to 45.0°C (59 to 113°F)Same as C2HSSame
    Accuracy of incubator temp indication≤0.8°CSame as C2HSSame
    Deviation limit range (skin temp)0.5 to 1.0°C0.3 to 1.0°C (0.5 to 1.8°F)Similar - I8000 plus has tighter limit.
    Deviation limit default (skin temp)1.0°C (1.8°F)Same as C2HSSame
    Kangaroo ModeNoYesDifferent - New feature, details tested.
    Noise level within the hood environment≤47 dB(A) w/ 37 dB(A) or less ambientSame as C2HSSame
    Air velocity over the mattress<10 cm/second (4 in/second); average of 5 points at 10cm (4 in) above the mattressSame as C2HSSame
    Carbon Dioxide (CO2) level (per IEC60601-2-19, clause 105)<0.8%Same as C2HSSame
    Oxygen deviation limit range3% fixed3% to 5%Similar - User-adjustable to reduce nuisance alarms.
    Oxygen deviation limit default3%Same as C2HSSame
    Set point data retention (power failure)<10 minSame as C2HSSame
    Acoustical level (all alarms on)75 dB(A) maximumSame as C2HSSame
    External Communication - Type9-pin Sub-D (female), electrically isolated ProtocolSame as C2HSSame
    ConfigurationsSerial Data OutputSerial Data Output (default) or MEDIBUS.XSimilar - MEDIBUS.X addition not clinically relevant.
    Baud rate (Serial Data Output)2400Same as C2HSSame
    Humidification systemManual adjustment of humidity setpointsAutomatic based on user setpoint (Autohumidity)Similar - Both provide humidity control.
    Humidity control duration of operation after refilling>24 hours @ 85% RH and 37°C, in air temp modeSame as C2HSSame
    Humidity control reservoir capacity1000 mL1500 mLSimilar - Difference not clinically relevant.
    Humidity control range30% to 95% in 1% incrementsSame as C2HSSame
    Humidity control accuracy (10-90% @ 20-40°C)±5% RH±6% RHSimilar - 1% difference not clinically relevant.
    Oxygen control range21% to 65%Same as C2HSSame
    Oxygen display accuracy (100% calibration)±3%Same as C2HSSame
    Material used for direct patient contactTextile Polyurethane, PVC, HydrogelTextile Polyurethane, TPE, HydrogelSimilar - TPE material tested for biocompatibility.
    BiocompatibilityAccording to ISO 10993Same as C2HSSame
    ThermonitoringN/A - only displays skin temperatureTrend display of difference between central and peripheral skin temperaturesSimilar. Isolette 8000 plus provides additional trend info.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The studies were non-clinical bench testing, not clinical studies involving patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This is a non-AI medical device submission, and the ground truth was established through engineering specifications, regulatory standards compliance, and bench testing, not human expert interpretation of data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is a non-AI medical device submission.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a non-AI medical device submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a non-AI medical device submission. The device performs as a standalone incubator without human-in-the-loop performance in the context of an AI algorithm; however, human operators are inherently part of its use.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device is based on engineering specifications, recognized international standards (e.g., IEC 60601-1, IEC 60601-2-19, IEC 60601-1-2, ISO-10993), and internal design control procedures. Performance benchmarks are set by these standards and the predicate device's capabilities.

    8. The sample size for the training set

    Not applicable. This is a non-AI medical device submission; there is no "training set" in the machine learning sense.

    9. How the ground truth for the training set was established

    Not applicable.


    Summary of Studies to Prove Acceptance Criteria are Met (Non-Clinical Testing):

    The Isolette® 8000 plus underwent a series of non-clinical bench tests and evaluations to demonstrate its safety and effectiveness and substantial equivalence to the predicate device. These include:

    • Performance Testing: Conducted in accordance with applicable standards and technical system requirements, as well as the device's Instructions for Use (IFU) Technical Data Section. This covers all the operational specifications listed in the table above (e.g., temperature control ranges, humidity accuracy, oxygen control, noise levels, warm-up times).
    • Electrical Safety and Electromagnetic Compatibility (EMC): Compliance with IEC 60601-1, IEC 60601-2-19 (safety and performance), and IEC 60601-1-2 (EMC).
    • Software Verification and Validation Testing: Conducted following FDA guidance for medical device software. The software was considered a "major" level of concern.
    • Biocompatibility Testing: Conducted in accordance with FDA guidance on ISO 10993 standards. Tests included:
      • Integral Test for volatile organic compounds
      • Photogenic bacteria test
      • Emission of particles
      • Material characterization (ISO 10993-18)
      • Toxicological Evaluation (ISO 10993-17)
      • Cytotoxicity
      • Irritation
      • Sensitization
    • Human Factors: Usability evaluation conducted in accordance with FDA guidance on Human Factors and Usability Engineering.
    • Reprocessing: Assessed following FDA guidance ("Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling").
    • Cleaning/Disinfection: Macroscopic and microbiological validation performed with Oxycide by an accredited laboratory (DIN EN ISO/IEC 17025).
    • Kangaroo Mode Performance Testing: Performed according to the technical requirements for this new feature.
    • Autohumidity Performance Testing: Performed according to technical requirements for the optional humidification system.
    • Thermomonitoring Performance Testing: Performed according to technical requirements.
    • External Communication (Serial data output or Medibus X) Performance Testing: Performed according to technical requirements.
    • Trendelenburg (Tilt Mechanism) Regression Testing: Performed for mechanical changes compared to the predicate device.

    No animal or clinical studies were conducted for this submission. The conclusion drawn from these non-clinical studies was that the Isolette® 8000 plus meets performance requirements and standards and is substantially equivalent to the predicate device K001242.