510(k) Data Aggregation

    K Number
    K153125

    Validate with FDA (Live)

    Date Cleared
    2016-01-13

    (76 days)

    Product Code
    Regulation Number
    870.1200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IntraNovo Microcatheter is intended for the infusion of contrast media into all peripheral vessels up to and including the cervical vessels, all vessels in the lower and upper extremittes, and all coronary vessels. The Microcatheter is also intended for drug infusion in intra-arterial therapy and infusion of embolic materials for hemostasis. The Microcatheter should not be used in cerebral vessels.

    Device Description

    The IntraNovo Microcatheter (the "IntraNovo 21" or the "Microcatheter") is an infusion catheter intended for intravascular use. The Microcatheter consists of a single lumen. The shaft consists of a lubricous inner liner made from polytetrafluoroethylene (PTFE) with a stainless steel coil over the inner liner. The outer liner consists of different lengths of colored polyether block amide with varying durometers hardness. A lubricious hydrophilic coating covers the distal end of the IntraNovo Microcatheter. One radiopaque marker band is placed at the distal end of the Microcatheter and above the stainless steel coil. A polycarbonate hub is attached to the proximal end of the Microcatheter with a strain relief placed over the distal end of the hub. The device shaft has an outer diameter of 2.9 French size on the proximal end and 2.4 French size on the distal end with an inner diameter of 0.021" throughout the shaft. The device is available in lengths of 110 cm and 130 cm. The following accessories are included with the Microcatheter: a stainless steel steam shaping mandrel to allow for manual shaping of the distal tip and a 3.0 mL syringe.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the IntraNovo Microcatheter, a medical device. This document focuses on demonstrating substantial equivalence to a predicate device through bench testing. It does not contain information about clinical studies with human participants or a multi-reader, multi-case comparative effectiveness study. Therefore, several aspects of the requested information, such as sample sizes for test/training sets, data provenance, number/qualifications of experts, adjudication methods, MRMC study details, and standalone algorithm performance, are not applicable or cannot be extracted from this document.

    Here's an analysis of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document states that the IntraNovo Microcatheter "met the predetermined acceptance criteria" for all tests. However, the specific quantitative acceptance criteria for each test are not provided in the table. Only the "Test Results (T=0) and (T=2)" which uniformly indicate "Pass" are given.

    Test #Test NameApplicable Standard or Internal Test MethodAcceptance Criteria (Not Explicitly Stated in Document, only "Pass" is reported for all)Reported Device Performance (T=0 and T=2)
    1Guidewire & Guide Catheter CompatibilityInternal Test MethodPredetermined acceptance criteria (not detailed)T=0 Pass, T=2 Pass
    2In-Vitro Track ForceInternal Test MethodPredetermined acceptance criteria (not detailed)T=0 Pass, T=2 Pass
    3Durability of Hydrophilic CoatingInternal Test MethodPredetermined acceptance criteria (not detailed)T=0 Pass, T=2 Pass
    4Lubricity of Hydrophilic CoatingInternal Test MethodPredetermined acceptance criteria (not detailed)T=0 Pass, T=2 Pass
    5Tip Shape RetentionInternal Test MethodPredetermined acceptance criteria (not detailed)T=0 Pass, T=2 Pass
    6Static Burst PressureISO 10555Predetermined acceptance criteria (not detailed)T=0 Pass, T=2 Pass
    7Dimensional & Physical AttributesISO 10555Predetermined acceptance criteria (not detailed)T=0 Pass, T=2 Pass
    8Corrosion ResistanceISO 10555Predetermined acceptance criteria (not detailed)T=0 Pass, T=2 Pass
    9Dynamic Burst PressureInternal Test MethodPredetermined acceptance criteria (not detailed)T=0 Pass, T=2 Pass
    10Air LeakInternal Test MethodPredetermined acceptance criteria (not detailed)T=0 Pass, T=2 Pass
    11Liquid LeakageInternal Test MethodPredetermined acceptance criteria (not detailed)T=0 Pass, T=2 Pass
    12Tensile StrengthISO 10555Predetermined acceptance criteria (not detailed)T=0 Pass, T=2 Pass
    13Flow RateInternal Test MethodPredetermined acceptance criteria (not detailed)T=0 Pass, T=2 Pass
    14Kink ResistanceInternal Test MethodPredetermined acceptance criteria (not detailed)T=0 Pass, T=2 Pass
    15RadiopacityASTM-F640-12Predetermined acceptance criteria (not detailed)T=0 Pass, T=2 Pass
    16Torque to FailureInternal Test MethodPredetermined acceptance criteria (not detailed)T=0 Pass, T=2 Pass
    17Catheter StiffnessInternal Test MethodPredetermined acceptance criteria (not detailed)T=0 Pass, T=2 Pass
    18Packaging IntegrityASTM F-88-09, ASTM-1929-98Predetermined acceptance criteria (not detailed)T=0 Pass, T=2 Pass
    19Therapeutic AgentsInternal Test MethodPredetermined acceptance criteria (not detailed)T=0 Pass, T=2 Pass
    20Female Luer Hub VerificationISO 594Predetermined acceptance criteria (not detailed)T=0 Pass, T=2 Pass
    21Shipping and Transportation SimulationISTA 3PAPredetermined acceptance criteria (not detailed)T=0 Pass, T=2 Pass
    22Coating IntegrityInternal Test MethodPredetermined acceptance criteria (not detailed)T=0 Pass, T=2 Pass
    23Torque ResponseInternal Test MethodPredetermined acceptance criteria (not detailed)T=0 Pass, T=2 Pass
    24Particulate EvaluationUSP <788>Predetermined acceptance criteria (not detailed)T=0 Pass, T=2 Pass

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    This document describes bench testing of a physical device, not a study involving a test set of data. The "test sets" here refer to batches of manufactured devices used for physical and mechanical evaluations. The document does not specify the sample size for devices used in each bench test, nor does it refer to data provenance in terms of country of origin or retrospective/prospective nature, as these are concepts typically associated with clinical or imaging data sets. The testing was performed on "non-aged devices (T=0) as well as on devices subject to 2 years of accelerated aging (T=2)."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    Not applicable. This is not a study assessing diagnostic performance or requiring expert ground truth. The "ground truth" for these tests would be the measured physical and mechanical properties of the device against engineering specifications.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. As noted above, this is not a study requiring adjudication of interpretations.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document pertains to the 510(k) clearance of a physical medical device (microcatheter) and does not involve AI or human readers.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This document pertains to the 510(k) clearance of a physical medical device (microcatheter) and does not involve an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the bench tests would be the established engineering specifications and performance requirements for the device, often derived from relevant ISO standards (e.g., ISO 10555) or internal test methods. These are objective measurements rather than subjective expert consensus or pathology.

    8. The sample size for the training set

    Not applicable. This document describes bench testing for device clearance, not the development or evaluation of an algorithm based on a training set.

    9. How the ground truth for the training set was established

    Not applicable.