Search Filters

Search Results

Found 8 results

510(k) Data Aggregation

    K Number
    K242434

    Validate with FDA (Live)

    Device Name
    Insulin Syringes
    Manufacturer
    Date Cleared
    2024-12-24

    (130 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Insulin Syringes are intended for subcutaneous injection of U100 insulin or U40 insulin.

    Device Description

    The Insulin Syringes are 0.3 mL, 0.5 mL, or 1.0 mL synnges designed for subcutaneous injection of a desired dose of insulin. The Insulin Syringes consist of a graduated barrel, plunger rod and needle/hub assembly. It is available in various needle gauge sizes (25G, 26G, 27G, 28G, 29G, 30G and 31G) and various needle lengths (6mm). The Insulin Syringes are sterile, single use, and non-toxic. These devices operate on the principles of a piston syringe.

    AI/ML Overview

    The provided document is an FDA 510(k) premarket notification for "Insulin Syringes," which are medical devices, not AI/ML-driven software with specific acceptance criteria related to a test set, expert ground truth, or MRMC studies. The document describes a traditional medical device (insulin syringes) and its substantial equivalence to previously cleared predicate devices.

    Therefore, the requested information regarding acceptance criteria, study design (test set, ground truth, expert adjudication, MRMC, standalone performance, training set details) for an AI/ML device is not available in this document. The document primarily focuses on:

    • Device Description: Physical characteristics (volume, needle gauge/length, materials), intended use (subcutaneous injection of U100/U40 insulin).
    • Modifications: Updates to packaging options and syringe sizes compared to the predicate.
    • Substantial Equivalence Justification: Comparison of indications for use, technological characteristics (design, materials, sterilization) with predicate devices.
    • Performance Testing: References to ISO standards for sterile single-use syringes, hypodermic syringes, stainless steel needle tubing, hypodermic needles, and sterilization validation (e.g., ISO 8537, ISO 7886-1, ISO 9626, ISO 7864, ISO 11135, ASTM F1980-16). These are standard tests for physical and functional properties of syringes, not AI/ML performance metrics.

    The "Performance Testing" section refers to engineering and manufacturing standards for physical device performance, not for the performance of an algorithm or software. There are no mentions of AI/ML, image data, diagnostic accuracy, or human reader studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233794

    Validate with FDA (Live)

    Device Name
    Insulin Syringe
    Date Cleared
    2024-03-06

    (99 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The sterile Insulin Syringe is for single use, with the calibrated unit of insulin for U-40 & U-100. The Insulin Syringe is a device intended for medical purposes for the manual aspiration of insulin, and for the injection of insulin into parts of the body below the surface of the skin. The device is intended to be used for adults only.

    Device Description

    The proposed device, Insulin Syringe, is a sterile device consisting of a calibrated hollow barrel, needle, needle cover and end cap. The Insulin syringe is intended for medical purpose for the manual aspiration of insulin and for the injection of insulin into parts of the body below the surface skin. The device is intended to be used for adults only. The needle is fixed on the syringe. The syringes barrel and plunger is made up of Polypropylene, the gasket is of Thermo Plastic Elastomer (TPE) and the needle cover (Top Cap) and end cap (Bottom Cap) is made of polypropylene.

    AI/ML Overview

    The provided document describes the FDA 510(k) clearance for an Insulin Syringe (K233794). This document focuses on demonstrating substantial equivalence to a predicate device, rather than proving that a device meets specific performance criteria for an AI/ML algorithm.

    Therefore, the information typically required to answer your request regarding "acceptance criteria" and "study that proves the device meets the acceptance criteria" for an AI/ML-powered device is not present in this document.

    This document describes a medical device, an insulin syringe, which is a physical product, not an AI/ML software. The performance data presented refers to standard physical and biological tests for such a device (e.g., sterilization, biocompatibility, needle strength, etc.), not the performance metrics of an AI algorithm.

    Here's a breakdown of why this document doesn't fit your request and what is actually discussed:

    • Type of Device: The device is an "Insulin Syringe," a physical medical instrument for injecting insulin. It does not incorporate AI or machine learning.
    • Purpose of the Document: This is an FDA 510(k) clearance letter. Its primary purpose is to establish "substantial equivalence" of a new device to a legally marketed predicate device, allowing it to be marketed. This is a common regulatory pathway for non-novel medical devices.
    • "Acceptance Criteria" discussed: The "acceptance criteria" for a physical device like an insulin syringe are generally defined by recognized standards (e.g., ISO standards for sterile hypodermic needles, biocompatibility standards). The document states that performance testing was conducted "to ensure the substantial equivalence...and verify conformity to the applicable parts of ISO standards."
    • "Study" discussed: The "study" refers to a series of performance tests (e.g., ISO 7864:2016, ISO 9626:2016, ISO 8537:2016 for physical characteristics; Biocompatibility testing as per ISO 10993-1; Sterilization, shipping, and shelf-life validation). These are laboratory and bench tests, not clinical studies involving AI models.

    In summary, as the provided text is for a physical medical device (Insulin Syringe) and not an AI/ML-powered device, it does not contain the information you requested regarding acceptance criteria and study data for AI/ML performance.

    To answer your prompt for an AI/ML device, you would need a document detailing the validation of an AI algorithm, which would include information on metrics like sensitivity, specificity, AUC, human-in-the-loop studies (MRMC), ground truth establishment, training data, and more. This document does not pertain to such a device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230061

    Validate with FDA (Live)

    Device Name
    Insulin Syringe
    Date Cleared
    2023-11-30

    (325 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    2 - 120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Insulin syringes are intended for subcutaneous injection of U-100 insulin

    Device Description

    The insulin syringe consists of needle cap, barrel, plunger stopper, needle tube and plunger cap. It is sterile disposable, non-pyrogenic, easy to use and meets all of the required standards. The plunger cap is optional. The single insulin syringe in unit package lacks the plunger cap compared with the product in the multi-unit package. The needle cap is used to protect physical protection of the needle prior to use. The needle tube penetrates the patient's skin to inject fluid into the body. The plunger stopper is the sealing assembly that encloses mandrel component and is used to inject fluid into the body. The barrel is a tube-like and cylindrical structure that holds insulin before the injection. The plunger is the rod pushed down or up depending on clinical use. The plunger cap is used to protect plunger before use. The capacity of the insulin device is 0.3 mL, 0.5 mL and 1 mL. The specification of needle tube is 26G, 27G, 28G, 29G, 30G, 31G, 32G and 33G. The needle length is 5 mm, 6 mm, 8 mm and 10 mm.

    AI/ML Overview

    The provided text describes the acceptance criteria and the study that proves the device meets the acceptance criteria for an Insulin Syringe (K230061).

    1. A table of acceptance criteria and the reported device performance

      The document lists the following performance tests and states that the proposed device was tested and demonstrated to be in conformance with the specified FDA recognized standards, and that the performance testing results met the standards' requirements to demonstrate the device's safety and effectiveness. No specific numerical performance values are provided, only conformance to standards.

      Acceptance Criteria (Standard)Reported Device Performance
      ISO 8537-2016: Sterile single-use syringes, with or without needle, for insulinConforms to requirements for: Limits for acidity or alkalinity, Limits for extractable metals, Tolerance on the graduated capacity, Graduation lines, Numbering of scale, Scale interval units, Overall length of scale, Barrel and plunger stopper, Finger grips, Fit of plunger stopper in barrel, Dead space, Freedom from leakage past plunger stopper, Bond between hub and needle tube, Needle cap.
      ISO 7886:1-2017: Sterile hypodermic syringes for single use – Part 1: Syringes for manual useConforms to requirements for Lubricant of syringe.
      ISO 9626-2016: Stainless steel needle tubing for the manufacture of medical devices – Requirements and test methodsConforms to requirements for: Stiffness, Resistance to Breakage, Resistance to Corrosion, Needle O.D., Needle inner diameter.
      ISO 7864-2016: Sterile hypodermic needles for single use - Requirements and test methodsConforms to requirements for: Needle point, Patency of lumen, Cleanliness, Needle length, Lubricant of needle tube, Penetration force and drag force for needles.
      ISO 10993-1 (Biocompatibility)Conforms to requirements for: Cytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity, Hemocompatibility, Pyrogen, Subacute/Subchronic Toxicity, Genotoxicity. (For Externally Communicating Device, Blood Path Indirect, prolonged Contact (>24hrs to 30 days)).
      USP <787> and USP<790> (Particulate testing)Met USP acceptance criteria.
      ISO 11135 (Sterilization)Sterilization process validated, EO and ECH residual does not exceed limit according to ISO 10993-7.
      ASTM F1980-16 (Shelf-Life)Verified claimed shelf-life of 5 years under accelerated aging conditions.
      ASTM D4169-16 (Package integrity - shipping)All packaging deemed acceptable for protection of product and sterility maintenance.
      ASTM D3078-02 (Vacuum Leak Test)Conforms to requirements.
      ASTM F 1929-15 (Dye Penetration Test)Conforms to requirements.
      DIN 58953-6: 2016 (Microbial Barrier Properties Test)Conforms to requirements.
      ASTM F88/F88M-21 (Seal Strength Test)Conforms to requirements.
      FDA Guidance for Medical Devices with Sharps Injury Prevention Features, August 9, 2005 (Simulated Clinical Use Testing)Conducted and demonstrated conformance.
    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

      The document does not specify the exact sample sizes used for each test. For instance, for biocompatibility, it states testing was done in compliance with ISO 10993-1, but not the number of samples. The data provenance is not explicitly stated in terms of country of origin or retrospective/prospective nature, but the manufacturer is Anhui Hongyu Wuzhou Medical Manufacturer Co., Ltd. in China.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

      This information is not applicable and therefore not provided in the document. The device is a physical medical device (insulin syringe), not an AI/software device that requires expert adjudication for ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

      This information is not applicable and therefore not provided in the document, as it pertains to AI/software device assessments.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      No MRMC study was done, as this is a physical medical device (insulin syringe) and not an AI-assisted diagnostic or imaging device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

      Not applicable, as this is a physical medical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

      The "ground truth" for this device is conformance to established international and FDA-recognized standards for sterile single-use syringes and needles, verified through various physical, chemical, and biological tests. This includes mechanical properties, material safety, sterility, packaging integrity, and simulated use. There is no "ground truth" derived from expert consensus or pathology in the context of an AI/software device.

    8. The sample size for the training set

      Not applicable, as this is a physical medical device, not an AI/software device that requires a training set.

    9. How the ground truth for the training set was established

      Not applicable, as this is a physical medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223453

    Validate with FDA (Live)

    Device Name
    Insulin Syringe
    Date Cleared
    2023-03-02

    (106 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Insulin Syringe is intended for subcutaneous injection of U-40 and U-100 insulin in the treatment of diabetes.

    Device Description

    The proposed device Insulin Syringe, a sterile device consisting of a calibrated barrel with plunger, is intended to be used to administer an injection of insulin to a patient subcutaneously. A non-retractable integrated needle is included. The syringe is made of plastic and silicone materials and allowing smooth plunger movement. This is a single-use device. The device is the same as K193273. The purpose of the submission is to add OTC labeling.

    AI/ML Overview

    The provided document is an FDA 510(k) premarket notification for an Insulin Syringe. It describes the device, its intended use, and compares it to a predicate device. However, it does not contain information about acceptance criteria and a study proving the device meets those criteria in the context of an AI/ML medical device, which is implied by your detailed request.

    The document describes typical non-AI medical device testing, such as:

    • Bench Testing: Verifies physical performance characteristics (e.g., needle tube length, bond strength, leakage, fit of plunger, dead space, penetration force, particulate contamination).
    • Biocompatibility Testing: Assesses biological safety (cytotoxicity, irritation, sensitization, systemic toxicity, hemolysis, pyrogens).
    • Sterilization and Shelf-life Testing: Validates sterilization method and confirms shelf life.

    Therefore, I cannot fulfill your request for information related to AI/ML device acceptance criteria, performance metrics (like sensitivity, specificity, AUC), sample sizes for AI test sets, expert ground truth establishment, MRMC studies, or training set details because this information is not present in the provided text. The device in question is a physical insulin syringe, not an AI/ML-based diagnostic or therapeutic device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203211

    Validate with FDA (Live)

    Date Cleared
    2023-02-09

    (832 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Disposable syringe for insulin administration, for the patient with diabetes.

    Device Description

    Insulin syringe with integrated extra thin stainless-steel needle and dead space less than 0.005 ml, made of medical grade plastic, sterile, disposable, pyrogen free and non-toxic. Sterilized by ethylene oxide. The product is composed of the following parts: Cylinder or barrel, Piston, Plunger, Needle, Needle cover or protector.

    AI/ML Overview

    This document describes the acceptance criteria and study results for the "Insulin syringe with integrated needle DL®" (proposed device), demonstrating its substantial equivalence to a predicate device (Disposable Insulin Syringe, K162180).

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are primarily derived from international standards (ISO series) and national standards (MGA, NOM). The "Specifications" column represents the acceptance criteria, and the "Insulin syringe with integrated needle DLPJECT®" column represents the reported device performance.

    #TestAcceptance Criteria (Specifications)Reported Device Performance ("Insulin syringe with integrated needle DLPJECT®")
    1Product designationInsulin syringe with integrated needle. Syringe type 7. Capacity: 0.3 mL and 1 mL. Sterile and disposable.Insulin syringe with integrated needle. Syringe type 7. Capacity: 0.3 mL and 1 mL. Sterile and disposable.
    2Product descriptionMedical use article, disposable, sterile, pyrogen-free, non-reactive tissue materials. Must be sterile, non-toxic, pyrogen-free and non-reactive tissue.Compliant (detailed description provided)
    Cylinder or barrel (clarity, lubrication, length, scale)Enough clarity to see dosage and identify bubbles. Interior lubricated with medical grade silicone (no drops). Minimum length 79 mm. Insulin unit scale (min 57 mm). Graduation lines on longitudinal axis.Compliant (detailed description provided)
    Graduation lines (length, numbering, appearance)Longer lines for zero and every five lines (approx half length of zero). Zero starts at perimeter. Clear numbering (10, 20...100 U.). "U.I." and "U-100" at end. "mL" or "ml" for total capacity. "Not reusing" symbol. Numbers >= 3mm. Lines 0.2-0.4mm thick. Clear contrast.Compliant (detailed description provided)
    PlungerOperated inside cylinder/barrel. Protrusion at distal end prevents slipping. Piston assembled at opposite end. Plunger + piston length 88 x 2 mm.Compliant (detailed description provided)
    NeedlePuncture device attached to syringe for liquid introduction. May have medical grade silicone coating.Compliant (detailed description provided)
    Needle cover or protectorMedical grade plastic, perfectly covers needle, protects edge, avoids accidental punctures.Compliant (detailed description provided)
    3FinishFree of defects (burrs, bubbles, perforations, fractures, roughness, deformations, sharp parts, non-uniform thickness). Needle assembly firm and not separated by normal use.Compliant (detailed description provided)
    4DimensionsVolume (mL/cc): 0.3, 0.5, 1.0Scale division (ml): 5 U.I.Scale subdivision (ml): 1 U.I. (0.3/0.5ml), 2 U.I. (1.0ml)Min length of scale (mm): 41.0 (0.3ml), 43.0 (0.5ml), 57.0 (1.0ml)Max silicone mass (mg): Lubricant will not form fluid droplets on inner surface.Compliant (values match specification)Compliant (values match specification)Compliant (values match specification)Compliant (values match specification)Compliant (matches specification)
    Scale tolerance (when ref line > 50% nominal capacity)± 5% (0.3ml), ± 4% (0.5ml), ± 4% (1.0ml)Compliant (values match specification)
    5Scale numbering0.5 units (0.3ml), 1.0 units (0.5ml), 2.0 units (1.0ml)Compliant (values match specification)
    6Scale positionEnds of similarly long graduation lines align vertically with barrel axis and each other (tolerance +0.5mm) when held vertically.Compliant (matches specification)
    7Characteristics of the cylinder or barrelEyebrow/flange for finger support. Enough clarity. Lubricated with medical grade silicone (no drops). Useful capacity no less than 10% more than nominal capacity or 3mm plunger travel beyond scale mark (whichever is less).Compliant (detailed description provided)
    8FlangeSyringe does not rotate more than 180° when placed on a flat surface with scale upwards at 10° from horizontal.Compliant (matches specification)
    9Plunger and piston featuresPlunger can be pushed by thumb when barrel is held with one hand. Plunger head has grooves to prevent slipping.Compliant (matches specification)
    10Reference lineDefined, visible edge at piston end for capacity reference. In contact with inner surface of barrel.Compliant (matches specification)
    11Dead spaceVolume in barrel/cylinder and pivot when piston fully inserted: max 0.005 mL.Compliant (matches specification)
    12Hypodermic Needle (Dimensions)Gauge (G): 30, 31, 32Nominal external diameter (mm): 0.298-0.320 (30G), 0.254-0.266 (31G), 0.229-0.241 (32G)Minimum nominal inner diameter (mm): 0.133 (30G), 0.114 (31G), 0.089 (32G)Useful length: Needle length tolerance within ± 1.25 mm.Primary angle: 9° to 11°Color code: Orange for U-100Compliant (values match specification)Compliant (values match specification)Compliant (values match specification)Compliant (matches specification)Compliant (matches specification)Compliant (matches specification)
    13Cannula adhesion (N)Minimum joint resistance of needle tube with nominal external diameter < 0.33 should be 11 N.Complies (matches specification)
    14HermeticityNone of the syringes should leak.Complies (matches specification)
    15Systemic injectionMGA-DM 3083. Passes the Test.Passes the Test.
    16Intracutaneous reactivityMGA-DM 3071. Passes the Test.Passes the Test.
    17PyrogensMGA 0711 or MGA 0316 (Bacterial Endotoxins). Satisfies the test method.Satisfies the test method.
    18SterilityMGA 0381. Passes the Test.Passes the Test.
    19Ethylene oxide residuesComplies with ISO 10993-7:2008, 4 mg maximum/24h.Complies with ISO 10993-7:2008, 4 mg maximum/24h.
    20Acidity or alkalinityMGA-DM 0001, Method II, Test compliance.Test compliance.
    21Removable metal limitSample extract in total < 5 mg/L of lead, tin, zinc, iron. Cadmium content < 0.1 mg/L.Complies (matches specification)
    22Product markingClear, legible, permanent characters: name, company name/symbol, nominal capacity (cm3 or ml), single graduated scale.Complies (matches specification)
    23Label and counter-label NOM-137-SSA1-2008Complies with NOM-137-SSA1-2008, Medical Device Labeling.Complies with NOM-137-SSA1-2008.
    24Labeling of the primary Packaging / RIS health supplies regulationComplies with the provisions of the Regulation of Inputs for Health. Second section. Labeling and packaging.Complies with the provisions.
    Biocompatibility Tests
    Cytotoxicity (elution)Reactivity grade: 0 - 2Reactivity grade: 0 (None, Not cytotoxic)
    Cytotoxicity (agar diffusion)Reactivity grade: 0 - 2Reactivity grade: 0 (None, Not cytotoxic)
    Irritability0 to 0.4: Not measurable0.0: Not measurable
    SensitizationGrade 0: No visible change; Grade 1: Slight or irregular erythema; Grade 2: Moderate and confluent erythema; Grade 3: Intense erythema or swellingGrade 0: No visible change
    Systemic ToxicityNo significantly greater biological reaction than blank in treated animals.No significantly greater biological reaction than blank in treated animals.
    PyrogensFor information only (sum of temperature increments).Sum of temperature increments 0.26 °C; considered apyrogenic.
    Acute toxicitySaline solution extract of medical device must not show adverse clinical effects.Saline solution extract did not show adverse clinical effects.
    Subacute toxicitySaline solution extract of medical device must not show adverse clinical effects.Saline solution extract did not show adverse clinical effects.
    Subchronic toxicitySaline solution extract of medical device must not show adverse clinical effects.Saline solution extract did not show adverse clinical effects.
    Sterility
    Sterilization methodEthylene oxide (validated per ISO 11135-1.2015).Ethylene oxide. Validated per ISO 11135-1.2015.
    Pyrogens (ISO 10993-12)Compliant with ISO 10993-12.Compliant with ISO 10993-12.
    Sterility Assurance Level (SAL)10-6Minimum SAL of 10-6.
    Microbial Ingress (Sterile barrier)Passes analytical test procedures.Passes analytical test procedures.
    Syringe air bubble leak (Sterile barrier)Passes analytical test procedures.Passes analytical test procedures.
    Packaging IntegrityASTM D4109-22: acceptable for protection and sterility.Acceptable per ASTM D4109-22.
    Shelf-lifeSupported by real-time stability testing.Real time stability testing supports shelf-life of years.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document extensively references international ISO standards (e.g., ISO 9626:2016, ISO 7864-1:2016, ISO 8537:2016, ISO 10993 series) and national standards (MGA, NOM-137-SSA1-2008, ASTM D4109-22). These standards typically specify the number of samples required for each test. For instance:

    • ISO 10993 (Biocompatibility): Tests like cytotoxicity, irritation, sensitization, acute, subacute, and subchronic toxicity, and pyrogenicity often involve in-vitro (cell cultures) or in-vivo (animal models like Wistar rats) testing with specific animal group sizes outlined in the respective ISO parts. For example, Pyrogens in Table 3 mentions "The sum of the temperature increments of the test animals is 0.26 °C so the sample is considered apyrogenic," implying in-vivo testing.
    • Physical/Performance Tests (e.g., Dimensions, Cannula adhesion, Hermeticity): These are typically performed on a statistically significant number of production samples according to the specific test methods outlined in the cited ISO standards. While exact numbers are not explicitly stated in this summary, the compliance with these standards implies that the required sample sizes were met.

    The data provenance is not explicitly stated as "country of origin." However, the tests are conducted in accordance with international (ISO) and potentially national (MGA, NOM from Mexico, given the submitter's address) standards, suggesting these tests were performed in certified laboratories adhering to these guidelines. The studies appear to be prospective as they are testing the newly developed "Insulin syringe with integrated needle DL®" against established criteria.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This type of medical device (insulin syringe) does not typically involve human expert "ground truth" establishment in the same way an AI diagnostic algorithm might. The "ground truth" for the performance and biocompatibility tests is based on objective measurements and established scientific protocols as defined by the international standards (ISO) and regulatory guidelines (MGA, NOM, ASTM). The "experts" involved would be the certified laboratory technicians and scientists performing these tests and interpreting the results against the defined specifications. Their qualifications would be in analytical chemistry, microbiology, toxicology, materials science, and medical device testing, following Good Laboratory Practice (GLP) and Good Manufacturing Practice (GMP) principles. The number of such professionals is generally defined by the laboratory's quality system and standard operating procedures for each test.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Adjudication methods like "2+1" or "3+1" are typically used for establishing ground truth in subjective assessments for diagnostic devices (e.g., radiologists reviewing images). For this type of device, which relies heavily on objective physical, chemical, and biological testing, traditional adjudication methods are not applicable. The "adjudication" is inherent in the test methodology itself:

    • Compliance with numerical specifications: If a measurement falls within the specified range, it passes.
    • Qualitative observations: For tests like "Finish" (absence of defects) or "Cytotoxicity" (reactivity grade), trained personnel make observations and record results based on predefined criteria in the standard. If a result is deemed "Non-cytotoxic" or "Passes the Test" according to the standard's interpretation guidelines, it meets the criterion.
      The reporting of results as "Complies," "Passes the Test," or specific measurement within tolerance indicates that the results were directly compared to the acceptance criteria without a separate adjudication panel.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic tools where human readers (e.g., radiologists) interpret cases with and without AI assistance. The "Insulin syringe with integrated needle DL®" is a physical medical device, not a diagnostic imaging or AI software product, so such a study would not apply.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    No, a standalone (algorithm-only) performance study was not done. This concept is also specific to AI/software as a medical device (SaMD) where an algorithm makes a determination independently. The device here is a physical syringe. Its "standalone" performance refers to its intrinsic physical and biological properties as tested against the standards.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth for this device's evaluation is primarily established through:

    • International Standards and Regulatory Specifications: The core ground truth is defined by the detailed requirements and test methods of standards like ISO 9626, ISO 7864, ISO 8537, ISO 10993 series, MGA, NOM, and ASTM. These standards define acceptable material properties, dimensions, performance characteristics, and biological compatibility limits.
    • Objective Measurement and Analytical Testing: Laboratory tests provide objective data (e.g., dimensional measurements, force required for cannula adhesion, chemical residue levels, cell reactivity grades, animal physiological responses) which are quantitatively or qualitatively compared to the predefined acceptance criteria from the standards.
    • Biological Endpoints: For biocompatibility, endpoints like cell viability (cytotoxicity), skin reactions (irritation/sensitization), and animal physiological responses (pyrogenicity, acute/subchronic toxicity) serve as the ground truth indicators as interpreted against the ISO 10993 series.

    8. The sample size for the training set

    This device does not involve machine learning algorithms, thus there is no "training set" in the context of AI. The "training" in the manufacturing process refers to process validation and quality control, ensuring that the manufacturing process consistently produces devices meeting specifications. The exact numbers of units produced and tested during process validation are not provided, but they would be governed by internal quality systems and relevant manufacturing standards.

    9. How the ground truth for the training set was established

    As there is no AI training set for this physical device, this question is not applicable. The "ground truth" for the device's design and manufacturing process would refer to the adherence to verified design specifications, material inputs, and validated manufacturing procedures, all established through engineering principles, regulatory requirements, and quality management systems.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221045

    Validate with FDA (Live)

    Date Cleared
    2022-11-03

    (209 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Insulin Syringes without needle is intended for medical purpose in conjunction with sterile needle for the manual aspiration and injection of U100 insulin into subcutaneous tissues.

    Device Description

    The Insulin Syringes without needle is a sterile and single use syringe intended for injection of U-100 insulin into the body, in conjunction with a sterile needle. The proposed device is available in 1 ml. The proposed device is sterilized by Ethylene Oxide to achieve a SAL of 10-6 and supplied in immediate package to maintain the sterility of the device during the shelf life of 5 years.

    AI/ML Overview

    This document is a 510(k) Summary for a medical device called "Insulin Syringes without needle" (K221045) from Berpu Medical Technology Co., Ltd. It compares the proposed device to a predicate device (K162180) to demonstrate substantial equivalence.

    Here's an analysis of the acceptance criteria and supporting studies as presented in the document:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present acceptance criteria in a table format with corresponding reported device performance for a specific study. Instead, it describes adherence to recognized standards and successful completion of various tests. For the purpose of this request, I will synthesize the information into a table based on the non-clinical testing performed and the standards referenced.

    Acceptance Criteria (Inferred from Standards)Reported Device Performance
    Material Performance
    Compliance with ISO 8537:2016 for insulin syringes (performance, dimensions, etc.)Complied with ISO 8537:2016
    Compliance with ISO 80369-7:2021 for small-bore connectors (Luer lock performance)Complied with ISO 80369-7:2021 (performance test conducted with proposed device)
    Biocompatibility (as per ISO 10993-1 for prolonged contact)
    Absence of CytotoxicityAddressed
    Absence of SensitizationAddressed
    Absence of Intracutaneous ReactivityAddressed
    Absence of Acute Systemic ToxicityAddressed
    Absence of Material-Mediated PyrogenicityAddressed
    Absence of Subchronic/Subacute ToxicityAddressed
    HemocompatibilityAddressed
    Particulate Matter in Injections (USP <788>)Met the USP acceptance criteria
    Sterility & Shelf Life
    Sterilization validated per ISO 11135:2014 (Ethylene Oxide)Validated to achieve a SAL of 10^-6^
    Shelf life of 5 years (validated per ASTM F1980-16)5 years shelf life determined based on stability studies (accelerated aging and simulated shipping)
    EO and ECH residuals within limits (ISO 10993-7:2008)Complied
    Bacterial Endotoxin Limit (USP <85>)Complied
    Packaging Integrity (after accelerated aging, environmental conditioning, transport)
    Visual Inspection (ASTM F1886 / F1886M-16)Performed
    Seal Strength (ASTM F88/F88M-15)Performed
    Dye Penetration (ASTM F1929-15)Performed. "All packaging deemed acceptable for protection of product and sterility maintenance." This implies passing criteria for these tests, although specific numerical results/criteria are not provided in this summary.

    2. Sample size used for the test set and the data provenance

    The document does not specify sample sizes for "test sets" in the context of clinical studies, as it explicitly states "No clinical study is included in this submission."

    For non-clinical tests (e.g., performance, biocompatibility, sterilization validation, shelf-life, packaging), specific sample sizes used for testing are not provided in this summary. The data provenance would be from the manufacturer's internal testing or contracted laboratories performing tests according to the cited ISO and ASTM standards. This testing is typically prospective, performed on manufactured devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. No clinical studies were conducted, and therefore, no expert-established ground truth for a clinical test set is relevant to this submission.

    4. Adjudication method for the test set

    Not applicable. No clinical studies were conducted, and therefore, no adjudication method for a clinical test set is relevant to this submission.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an insulin syringe, not an AI-powered diagnostic or assistive tool. No MRMC study was mentioned or would be relevant.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This device is an insulin syringe, not an algorithm or AI system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the non-clinical tests, the "ground truth" is established by the specified international (ISO) and national (ASTM, USP) standards. The device is tested against these standards, and its performance is compared to the requirements outlined in those standards. For example, for particulate matter, the ground truth is the maximum particle count allowed by USP <788>. For sterility, the ground truth is a SAL of 10^-6^ as per ISO 11135.

    8. The sample size for the training set

    Not applicable. No AI/machine learning component is involved, so there is no training set in this context.

    9. How the ground truth for the training set was established

    Not applicable. No AI/machine learning component is involved, so there is no training set or corresponding ground truth establishment in this context.

    Ask a Question

    Ask a specific question about this device

    K Number
    K202570

    Validate with FDA (Live)

    Date Cleared
    2021-07-02

    (301 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Insulin Syringe is a sterile, single use and non-reusable, insulin syringe intended for injection of U-100 insulin into the body.
    The Insulin Syringe with Safety Retractable is a sterile, single use and non-reusable, manual retractable safety insulin syringe intended for injection of U-100 insulin into the body, while reducing the risk of sharps injuries and the potential for insulin syringe reuse.

    Device Description

    (1) Insulin Syringe
    This product uses PP, Polyisoprene rubber and SUS304 as the main materials. It consists of barrel, plunger, plunger stopper, needle tube and needle cap.
    The product is for single use and provided sterile (EtO). The shelf-life of the product is five-years.
    Product Contact Classification: Externally Communicating, Blood Path and Tissue Contact with prolonged contact duration (>24 h to 30 d).
    (2) Insulin Syringe with Safety Retractable
    This product uses PP, Polyisoprene rubber and SUS304 as the main materials. It consists of barrel, plunger, plunger stopper, needle tube, safety shield and needle cap.
    The product is for single use and provided sterile (EtO). The shelf-life of the product is five-years.
    Product Contact Classification: Externally Communicating, Blood Path and Tissue Contact with prolonged contact duration (>24 h to 30 d).

    AI/ML Overview

    1. Acceptance Criteria and Reported Device Performance:

    The document outlines the acceptance criteria by stating that the device "conforms with the requirements of ISO 7864 and ISO 8537" for performance, safety, and effectiveness, and "conforms to the requirement of ISO 10993 series Standards" for biocompatibility.

    For the "Insulin Syringe with Safety Retractable," additional performance criteria specifically related to the safety feature are provided:

    Acceptance Criteria (Safety Feature)Reported Device Performance
    In-Safe mode force: not be more than 5NTest value: 3.45-4.45N
    Resist force: 60s with 20N weights, and the protective device shall not be opened.Test value: >20N/60s

    For the general Insulin Syringe and Insulin Syringe with Safety Retractable, the following criteria are implicitly met by conformance to the cited ISO standards. The "Reported Device Performance" column indicates that the device did conform to these standards.

    Acceptance Criteria (General)Reported Device Performance
    Biocompatibility: No cytotoxicity, no irritation to skin, no significant evidence of sterilization, no systemic toxicity, no hemolysis, no pyrogen (based on ISO 10993 series)Conforms to ISO 10993 series Standards
    Performance, Safety & Effectiveness (general syringe functionality): Cleanliness, limits for acidity or alkalinity, limits for extractable metals, needle cap, tolerances on length, freedom from defects, lubricant, needle point, bond between hub and needle tube, patency of lumen, sharps injury protection (based on ISO 7864)Conforms with ISO 7864
    Performance, Safety & Effectiveness (insulin syringe specific): Color coding, limits for acidity or alkalinity, limits for extractable metals, lubrication of syringes, lubrication of needle tube, barrel and plunger stopper, finger grips, fit of plunger stopper in barrel, position of nozzle on end of barrel, needle tubing for specific syringe types, bond between hub and needle tube, dead space, freedom from leakage at needle, freedom from leakage past plunger stopper (based on ISO 8537)Conforms with ISO 8537
    Stainless Steel Needle Tubing: Dimensions, Stiffness, Resistance to breakage, Resistance to corrosion (based on ISO 9626)Conforms with ISO 9626
    Particulate matter testingPerformed per USP <788>, Method 1
    ETO residualsPerformed per ISO 10993-7:2008
    Packaging and Shelf life testingPerformed per ASTM D4169-16, ASTM F1929-15:1998 ASTM F88:2009
    Risk Management ReportAvailable (implied conformance)
    Sharps Injury ProtectionPerformed per ISO 23908

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the sample sizes used for the non-clinical tests. It refers to standards like ISO 7864, ISO 8537, ISO 10993, ISO 9626, USP <788>, ASTM D4169, ASTM F1929, ASTM F88, and ISO 23908. These standards likely define the minimum sample sizes for relevant tests. The data provenance is implied to be from testing conducted by Jiangsu Micsafe Medical Technology Co., Ltd in China, as it is the submitting company. The studies are prospective in nature, as they involve testing the device to ensure it meets established standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    The document does not mention the use of experts to establish a "ground truth" in the context of clinical or diagnostic studies. The evaluation is focused on demonstrating conformance to established international and national standards for medical devices. Therefore, the "ground truth" is defined by the technical specifications and performance requirements outlined in these standards. The experts involved would be the engineers, scientists, and technicians conducting the tests and interpreting the results in accordance with these standards. Their qualifications would align with the specific tests performed (e.g., biocompatibility specialists, mechanical testing engineers).

    4. Adjudication Method for the Test Set:

    Not applicable. The tests performed are objective measurements against predefined criteria in recognized standards, not subjective interpretations requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is a medical device (insulin syringe) that does not involve "readers" or Artificial Intelligence (AI) for diagnostic or interpretive tasks. The studies are related to the physical performance, safety, and biocompatibility of the syringe itself.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Not applicable. This device is not an algorithm or AI system. It is a physical medical device.

    7. The Type of Ground Truth Used:

    The ground truth used for these studies is defined by the established technical specifications and performance requirements detailed in the referenced international (ISO) and national (ANSI AAMI, ASTM, USP) standards. These standards set the benchmarks for characteristics like biocompatibility, sterility, physical dimensions, material properties, and safety features.

    8. The Sample Size for the Training Set:

    Not applicable. There is no "training set" in the context of a physical medical device undergoing performance and safety testing. This document describes the testing of a finished product against established standards, not the training of a machine learning model.

    9. How the Ground Truth for the Training Set was Established:

    Not applicable, as there is no training set for this type of device submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K193273

    Validate with FDA (Live)

    Device Name
    Insulin Syringe
    Date Cleared
    2021-01-05

    (406 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Insulin Syringe is intended for subcutaneous injection of U-40 and U-100 insulin in the treatment of diabetes.

    Device Description

    The proposed device Insulin Syringe, a sterile device consisting of a calibrated barrel with plunger, is intended to be used to administer an injection of insulin to a patient subcutaneously. A non-retractable integrated needle is included. The syringe is made of plastic and silicone materials and allowing smooth plunger movement. This is a single-use device.

    AI/ML Overview

    The provided document is a 510(k) summary for an Insulin Syringe and does not contain information about acceptance criteria or a study that proves the device meets specific performance criteria in the context of an AI/ML medical device.

    Key points from the document regarding studies and testing:

    • No AI/ML Component: The device described is an "Insulin Syringe," a physical medical device, not an AI/ML algorithm. Therefore, none of the requested information regarding AI/ML performance, sample sizes for test/training sets, ground truth establishment, expert adjudication, or MRMC studies are applicable or present in this document.
    • Non-Clinical Studies: The document details non-clinical studies performed to demonstrate substantial equivalence to a predicate device. These include:
      • Performance Testing: Conducted according to ISO 8537:2016 for sterile single-use syringes for insulin.
      • Biocompatibility Testing: Conducted according to ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 10993-11, ISO 10993-4, ASTM F 756-17, and USP <151> (Pyrogen test), USP <788> (Particulate matters).
      • Sterilization and Shelf-life Testing: Validated according to ISO 11135, with residue levels meeting ISO 10993-7. Shelf-life testing supported a 5-year shelf life.
    • No Clinical Studies: The document explicitly states: "No prospective clinical trials were conducted in support of this 510(K)."

    Since the request pertains to the acceptance criteria and study proving an AI/ML device meets them, and the provided document describes a conventional medical device (insulin syringe), the requested information cannot be extracted.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1