510(k) Data Aggregation

    K Number
    K211153

    Validate with FDA (Live)

    Device Name
    Inomax DSIR
    Date Cleared
    2023-04-20

    (731 days)

    Product Code
    Regulation Number
    868.5165
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The INOmax® DSIR Plus delivers INOMAX® (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant on of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with ventilators and respiratory care devices that the INOmax DSIR Plus has been validated with.

    The INOmax® DSIR Plus provides continuous integrated monitoring of inspired O2, NO2, and NO, and a comprehensive alarm system.

    The INOmax® DSIR Plus incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source.

    The INOmax® DSIR Plus includes a backup NO delivery capability that provides a fixed flow of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breath. It may also use the INOblender® for backup.

    The INOmax® DSIR Plus must only be used in accordance with the indications, warnings and precautions described in the nitric oxide drug packaging inserts and is indicated for use in term and near-term (>34weeks gestation) neonates with hypoxic respiratory failure associated with clinical or echocardiographic evidence of pulmonary hypertension. INOmax DSIR Plus is indicated for a maximum of 14 days of use. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.

    Device Description

    The INOmax DSR® Plus uses a "dual-channel" design to ensure the safe delivery of INOMAX®. The first channel has the delivery CPU, the flow controller and the injector module to ensure the accurate delivery of NO. The second channel is the monitoring system, which includes a separate monitor CPU, the gas cells (NO, NO2, and O2 cells) and the user interface including the display and alarms. The dual-channel approach to delivery and monitoring permits INOMAX® delivery independent of monitoring but also allows the monitoring system to shutdown INOMAX® delivery if it detects a fault in the delivery system such that the NO concentration could become greater than 100 ppm.

    All revisions of INOmax DSm® Plus utilize component technology to deliver Nitric Oxide gas to the patient. The components consist of the Delivery System unit, the blender, a stand/cart and the NO gas tanks. In this revision of the INOmax DSIR® Plus, the significant changes to the device include the labeling and main circuit board.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the INOmax DSIR® Plus device, which is a nitric oxide administration apparatus. It focuses on demonstrating substantial equivalence to a predicate device (K200389) rather than an AI/ML-based device requiring a study to prove meeting acceptance criteria based on performance metrics like sensitivity, specificity, or AUC.

    Therefore, the requested information regarding acceptance criteria, study design for proving device performance, sample sizes for test/training sets, expert involvement, adjudication methods, MRMC studies, standalone performance, and ground truth establishment (which are typical for AI/ML device evaluations) are not applicable to this submission.

    The document explicitly states: "The subject of this premarket submission, INOmax DSm® Plus, did not require clinical studies to support substantial equivalence." This reinforces that the evaluation was based on non-clinical tests demonstrating design changes and continued safety/performance relative to the predicate, not on a clinical performance study with human subjects or AI-based diagnostic/prognostic output.

    The "acceptance criteria" for this device, as implied by the submission, are largely related to engineering, safety, and performance as compared to the predicate device, and these were met through non-clinical testing.

    Here's a breakdown of the relevant information provided, mapping it to your request where applicable, and noting where information is not present due to the nature of the submission:


    Acceptance Criteria and Device Performance (Not Applicable in the AI/ML sense):

    Acceptance Criterion (Implicit based on device type)Reported Device Performance/Testing
    Safety and Essential Performance (Electrica)Demonstrated conformity to IEC 60601-1:2005 via testing.
    Electromagnetic Compatibility (EMC)Demonstrated conformity to IEC 60601-1-2:2014 via testing.
    Biocompatibility of New MaterialsTested in accordance with ISO 18562 and ISO 10993 series.
    Accurate Delivery of NOVerified through integration, performance, and safety testing (module verification, system verification).
    Continuous Integrated Monitoring (O2, NO2, NO)Verified through integration, performance, and safety testing (module verification, system verification).
    Backup NO Delivery CapabilityVerified through integration, performance, and safety testing (module verification, system verification).
    Software FunctionalityVerified through software tests, including minor modifications to troubleshooting help and resolving anomalies.
    Risk ManagementRisk Analysis conducted.
    Requirements Review & Design ReviewsPerformed.
    Substantial Equivalence to PredicateConcluded based on non-clinical testing and comparison of features and intended use.

    Detailed Breakdown per your request:

    1. A table of acceptance criteria and the reported device performance:

      • As shown in the table above, the "acceptance criteria" are implied by the non-clinical tests performed (e.g., meeting IEC standards, successful risk analysis, verification of functionality). The "reported performance" is that these tests were passed, supporting substantial equivalence. There are no quantitative performance metrics like sensitivity/specificity for a diagnostic AI.
    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

      • Not applicable. This was not a data-driven performance study in the context of AI/ML. "Testing" refers to hardware, software, and system verification/validation against engineering specifications and recognized standards, not a clinical test set of patient data.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

      • Not applicable. There was no "ground truth" to establish in the context of an AI/ML diagnostic or prognostic system. The device's function is gas delivery and monitoring, not diagnosis.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable. No expert review or adjudication process was described or required for this type of device submission.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. This device is not an AI-assisted diagnostic tool.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. The device itself is a medical apparatus, not an algorithm, and its performance is assessed via engineering and system validation, not standalone algorithmic evaluations.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • Not applicable. No "ground truth" in the AI/ML sense was used. Device functionality and safety were verified against engineering specifications, simulated physiological conditions (e.g., gas flow and concentration measurements), and recognized standards.
    8. The sample size for the training set:

      • Not applicable. There was no AI/ML training set.
    9. How the ground truth for the training set was established:

      • Not applicable. There was no AI/ML training set.

    In summary, the provided document is a 510(k) summary for a medical device that delivers and monitors nitric oxide. The submission focuses on demonstrating substantial equivalence to a predicate device through non-clinical testing (e.g., electrical safety, performance testing, software verification, biocompatibility), rather than clinical performance studies involving a test set with established ground truth, which would be typical for AI/ML-based diagnostic or prognostic devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200389

    Validate with FDA (Live)

    Device Name
    INOmax DSIR Plus
    Date Cleared
    2020-06-17

    (120 days)

    Product Code
    Regulation Number
    868.5165
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The INOmax® DSIR Plus delivery system delivers INOMAX® (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators.

    The INOmax® DSIR Plus provides continuous integrated monitoring of inspired O2, NO2, and NO, and a comprehensive alarm system.

    The INOmax® DSIR Plus incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source.

    The INOmax® DSIR Plus includes a backup NO delivery capability that provides a fixed flow of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breath. It may also use the INOblender® for backup.

    The target patient population is controlled by the drug labeling for INOMAX® and is currently neonates. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.

    Device Description

    The INOmax DSR® Plus uses a "dual-channel" design to ensure the safe delivery of INOMAX®. The first channel has the delivery CPU, the flow controller and the injector module to ensure the accurate delivery of NO. The second channel is the monitoring system, which includes a separate monitor CPU, the gas cells (NO, NO2, and O2 cells) and the user interface including the display and alarms. The dual-channel approach to delivery and monitoring permits INOMAX® delivery independent of monitoring but also allows the monitoring system to shutdown INOMAX® delivery if it detects a fault in the delivery system such that the NO concentration could become greater than 100 ppm.

    The INOmax® DSIR Plus delivery system delivers INOMAX® (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators.

    The INOmax® DSIS Plus provides continuous integrated monitoring of inspired O2, NO2, and NO, and a comprehensive alarm system.

    The INOmax® DSIR Plus incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source.

    The INOmax® DSIR Plus includes a backup NO delivery capability that provides a fixed flow of 250 mL/min of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breathing circuit. It may also use the INOblender® for backup.

    All revisions of INOmax DSm® Plus utilize component technology to deliver Nitric Oxide gas to the patient. The components consist of the Delivery System unit, the blender, a stand/cart and the NO gas tanks. In this revision of the INOmax DSm® Plus, the only changes to the device includes the labeling for compatibility with respiratory care device.

    AI/ML Overview

    This document, K200389, is a 510(k) premarket notification for the INOmax DSIR Plus, a nitric oxide administration apparatus. It focuses on demonstrating substantial equivalence to a predicate device (K131686), specifically by adding compatibility with new ventilator and breathing devices.

    Based on the provided text, the device performance assessment relies entirely on nonclinical (bench) testing and comparison to a previously cleared predicate device. There is no evidence of clinical studies involving human subjects or AI algorithms in this document. Therefore, many of the requested points regarding AI/MRMC studies, expert ground truth adjudication, and training/test set sample sizes are not applicable to the information presented.

    Here's the breakdown of what can be extracted from the document:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a specific table of acceptance criteria with corresponding performance metrics like "accuracy > X%". Instead, it refers to a "Ventilator/Gas Delivery System Validation Test Protocol" used for the predicate device (K131686) and states that this same protocol, with "insignificant differences," was used for the INOmax DSIR Plus. The general acceptance criterion implied is that the device "performs within published specifications" and that "the hazards were mitigated" based on this protocol.

    • Acceptance Criteria (Implied): Performance according to "published specifications" and mitigation of identified hazards, as demonstrated through the "Ventilator/Gas Delivery System Validation Test Protocol."
    • Reported Device Performance: "Ultimately, the requirements necessary for the operation of the INOmax DSIR passed." and "This Bench Testing was conducted across all platforms to demonstrate that the INOmax DSIR® Plus performs within published specifications."

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not explicitly stated as a number of patient cases or images. The "test set" in this context refers to the bench testing conducted on the device's compatibility with a list of new ventilator and breathing devices. The document lists 11 specific additional ventilator devices that were tested for compatibility:
      • Covidien PB 980 (K131252)
      • GE Healthcare Carescape R860 (K142679)
      • Fisher & Paykel Healthcare RT330 Breathing Circuit and Optiflow Jr (Class I, 510(k) Exempt under 21 CFR 868.5340)
      • Bunnell Inc Life Pulse 204 (P850064)
      • Drager Perseus A500 (K133886)
      • Fisher & Paykel Healthcare Airvo 2 (K131895)
      • Drager Carina (K072885)
      • Maquet Servo u/n (K151814)
      • Hamilton C3 (K161450)
      • IMT Medical Bellavista (K163127)
      • Maquet Flow-i (K160665)
      • Bio-Med TV-100 (K173973)
      • Phillips V60 (K102985)
    • Data Provenance: The data originates from bench testing (laboratory) rather than clinical patient data. Country of origin for the testing is not specified but is presumed to be associated with the manufacturer (Mallinckrodt Manufacturing, LLC, based in Madison, Wisconsin, USA). The testing is prospective in the sense of being conducted specifically for this submission, although it leverages a protocol from a previous clearance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    Not applicable. This device is a hardware apparatus for administering nitric oxide, not an AI algorithm requiring expert human interpretation of medical images or data for ground truth. The "ground truth" for the nonclinical testing would be the engineering specifications and expected performance, verified through the validation protocol.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable, as no human expert adjudication of data (like medical images or clinical outcomes) was performed. The "adjudication" of the bench test results would be whether the device passed or failed the predefined engineering/performance criteria in the validation protocol.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This document explicitly states, "The subject of this premarket submission... did not require clinical studies to support substantial equivalence." This means no human-in-the-loop performance study, MRMC study, or AI assistance was involved.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    No. This device is a physical medical device, not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" used for this device's validation was engineering specifications and performance criteria established in the "Ventilator/Gas Delivery System Validation Test Protocol," likely determined by design requirements and regulatory standards for medical devices of this type.

    8. The sample size for the training set

    Not applicable. There is no AI component or training set mentioned in this submission.

    9. How the ground truth for the training set was established

    Not applicable. There is no AI component or training set mentioned in this submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K143213

    Validate with FDA (Live)

    Date Cleared
    2015-07-10

    (242 days)

    Product Code
    Regulation Number
    868.5165
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The INOmax DSIR® Plus MRI delivery system is indicated for delivery of INOMAX® (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant on of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. The INOmax DSIR® Plus MRI is indicated for use only with MR Conditional ventilators validated to be compatible, as identified in the device labeling.

    The INOmax DSIR® Plus MRI is indicated for continuous integrated monitoring of inspired 02, NO2, and NO.

    The INOmax DSIR® Plus MRI is considered MR Conditional with the use of 1.5 Tesla static magnetic field scanners ONLY in areas where the field strength is less than 100 gauss.

    The target patient population is controlled by the drug labeling for INOMAX® and is currently neonates. The primary targeted clinical setting is a clinical 1.5 Tesla and 3.0 Tesla diagnostic imaging environment.

    Device Description

    The INOmax DSIR® uses a "dual-channel" design to provide delivery of INOMAX®. The first channel has the delivery CPU, the flow controller and the injector module to ensure the accurate delivery of NO. The specially designed injector module enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. The second channel is the monitoring system, which includes a separate monitor CPU, the gas cells (NO, NO2, and O2 cells) and the user interface including the display and a comprehensive alarm system. The dual-channel approach to delivery and monitoring permits INOMAX® delivery independent of monitoring but also allows the monitoring system to shut down INOMAX® delivery if the monitored NO concentration exceeds 100 ppm for 12 consecutive seconds. The delivery system can also shut down delivery if it detects certain serious problems with the monitoring system.

    The INOmax DSIR Plus MRI incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source.

    The INOmax DSIR Plus MRI includes a backup NO delivery capability that provides a fixed flow of 250 mL/min of NO, which along with user supplied 10 L/min of oxygen, provides 20 ppm of NO to a patient breathing circuit. The INOblender® can also be used for backup.

    AI/ML Overview

    The provided document is a 510(k) summary for the INOmax DSIR® Plus MRI, a nitric oxide administration apparatus. It outlines the device description, intended use, technological characteristics, and a comparison to a predicate device, as well as a summary of nonclinical tests performed.

    However, the document does not contain the specific information requested about acceptance criteria and a study proving the device meets those criteria in the context of typical AI/ML medical device submission requirements.

    Here's why and what information can be extracted or deduced:

    This document describes a hardware medical device, not an AI/ML powered software or diagnostic device. Therefore, the concepts of "acceptance criteria" and "device performance" as typically discussed for AI/ML models (e.g., sensitivity, specificity, AUC) are not directly applicable in the same way. The "performance" here refers to the device's functional operation meeting engineering specifications.

    Let's address each point based on the available information:


    1. A table of acceptance criteria and the reported device performance

    • Acceptance Criteria (Deduced from testing goals): The document implies acceptance criteria are related to the device's functional performance:

      • O2 dilution: The device should not unacceptably dilute oxygen concentration.
      • Effect on delivered pressures: The device should not adversely affect ventilator delivered pressures.
      • INOmax DSIR® delivery accuracy: The device should accurately deliver Nitric Oxide.
      • NO2 generation: The device should not generate unacceptable levels of Nitrogen Dioxide.
      • MRI Compatibility: Device classified as MR Conditional for 1.5 Tesla and 3.0 Tesla static magnetic field scanners in areas where field strength is less than 100 gauss (via ASTM standards and other tests).
      • Electrical Safety & EMC: Conformity to IEC 60601-1 and IEC 60601-1-2.
      • Alarm Systems: Conformity to IEC 60601-1-8.
      • Software Functionality: Software version 3.1.2 meets specified requirements.
      • Usability: Meets human factors engineering requirements (ANSI/AAMI HE75).
      • Backup delivery: Provides fixed flow of 250 mL/min of NO.
    • Reported Device Performance:

      • "The INOmax DSIR® performed within published specifications when used with each of the ventilators in both primary and backup delivery."
      • "The INOmax DSIR® Plus MRI meets its system level requirements and that the new/modified features function as specified."
      • Compliance with specific FDA recognized consensus standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, ASTM F2052-06, ASTM F2119-07, ASTM F2503-13).

    Table of (Deduced) Acceptance Criteria and Performance:

    Acceptance Criteria (Inferred)Reported Device Performance
    Functional Safety & Compatibility:
    O2 dilution acceptablePerformed within published specifications when used with ventilators.
    No adverse effect on delivered pressuresPerformed within published specifications when used with ventilators.
    INOmax DSIR® delivery accuracyPerformed within published specifications when used with ventilators; measured values on the INOmax DSIR® were recorded and anomalies noted.
    Acceptable NO2 generationPerformed within published specifications when used with ventilators.
    MR Conditional for 1.5T/3.0T static fields (<100 gauss)Classification as MR Conditional with the use of 1.5 Tesla and 3.0 Tesla static magnetic field scanners ONLY in areas where the field strength is less than 100 gauss. Compliance with ASTM F2052-06, ASTM F2119-07, ASTM F2503-13. Cart-mounted Gauss Alarms (100 Gauss) included.
    Electrical Safety & Quality:
    Electrical safety (Basic Safety and Essential Performance)Conformity to IEC 60601-1:2005.
    Electromagnetic Compatibility (EMC)Conformity to IEC 60601-1-2:2007.
    Alarm System performanceConformity to IEC 60601-1-8:2006.
    Software meets requirements and functions as specifiedVerification confirmed the INOmax DSIR® Plus MRI meets its system level requirements and that new/modified features function as specified.
    Usability (Human Factors Engineering)Formative usability study conducted per ANSI/AAMI HE75. Meets system level requirements.
    Backup NO delivery capability (250 mL/min NO, 20 ppm with O2)Device includes a backup NO delivery capability providing a fixed flow of 250 mL/min of NO; performed within specifications in backup delivery.
    6 hours uninterrupted NO delivery on batteryIntegrates a battery that provides up to 6 hours of uninterrupted NO delivery. (Performance implies this function was confirmed).

    Regarding AI/ML specific questions (2-9), the following applies since this is not an AI/ML device:

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not applicable as this is a hardware device. Testing involved physical device performance and compatibility with other medical equipment, not analysis of a data set. The document mentions "Five INOmax DSIR® settings were used [0 (baseline), 5. 10. 20. 40. and 80 ppm] for each mode of ventilation, as well as the Backup mode." This describes test parameters, not a patient sample size or data provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. Ground truth for a hardware device's functional performance is established through engineering and medical device standards (e.g., accuracy of gas delivery, safety limits). Not expert interpretation of data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. This concept pertains to resolving discrepancies in expert labeling or diagnoses for AI/ML ground truth.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is not an AI/ML device intended to interpret data for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a standalone medical device, but not in the context of an algorithm's performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For a hardware device, "ground truth" relates to its adherence to established engineering specifications, safety standards, and functional requirements. For example, the ground truth for "NO delivery accuracy" would be calibration standards and expected physical/chemical properties.

    8. The sample size for the training set

    • Not applicable. This device does not have a "training set" in the context of AI/ML.

    9. How the ground truth for the training set was established

    • Not applicable. This device does not have a "training set" for AI/ML.
    Ask a Question

    Ask a specific question about this device

    K Number
    K131686

    Validate with FDA (Live)

    Date Cleared
    2013-11-29

    (172 days)

    Product Code
    Regulation Number
    868.5165
    Age Range
    All
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The INOmax® DS delivery system delivers INOMAX® (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators.

    The INOmax® DS provides continuous integrated monitoring of inspired O₂, NO₂, and NO, and a comprehensive alarm system.

    The INOmax® DS incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source.

    The INOmax® DS includes a backup NO delivery capability that provides a fixed flow of 250 mL/min of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breathing circuit. It may also use the INOblender® for backup.

    The target patient population is controlled by the drug labeling for INOMAX® and is currently neonates. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.

    Device Description

    The INOmax DSIR® uses a "dual-channel" design to ensure the safe delivery of INOMAX®. The first channel has the delivery CPU, the flow controller and the injector module to ensure the accurate delivery of NO. The second channel is the monitoring system, which includes a separate monitor CPU, the gas cells (NO, NO₂, and O₂ cells) and the user interface including the display and alarms. The dual-channel approach to delivery and monitoring permits INOMAX® delivery independent of monitoring but also allows the monitoring system to shutdown INOMAX® delivery if it detects a fault in the delivery system such that the NO concentration could become greater than 100 ppm. The delivery system can also shut down delivery if it detects certain serious problems with the monitoring system.

    AI/ML Overview

    The provided document describes the INOmax DSIR® (Delivery System), a device for delivering nitric oxide to patients. The submission is a 510(k) for a software update (version 3.0) and compatibility with two additional respiratory care devices (Hamilton C1 and T1 Ventilators).

    Here's an analysis of the acceptance criteria and the study conducted:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria / RequirementReported Device Performance
    02 dilutionTested and confirmed compatibility with the new respiratory care devices. (Implied that 02 dilution was within acceptable limits, as the "Testing Conclusion" states the device performed within published specifications.)
    Effect on delivered pressuresTested and confirmed compatibility with the new respiratory care devices. (Implied that delivered pressures were within acceptable limits, as the "Testing Conclusion" states the device performed within published specifications.)
    INOmax DSIR® delivery accuracyTested and confirmed compatibility with the new respiratory care devices. (Implied that delivery accuracy was maintained, as the "Testing Conclusion" states the device performed within published specifications.)
    NO2 generationTested and confirmed compatibility with the new respiratory care devices. (Implied that NO2 generation was within acceptable limits, as the "Testing Conclusion" states the device performed within published specifications.)
    Software Functionality (new/modified user convenience features and alarms)Software verification confirmed the INOmax DSIR® is compliant with its system level requirements and that the new/modified user convenience features and alarms function as specified.
    Electrical Safety (IEC 60601-1:2005)Support for substantial equivalence was provided as a result of risk management and testing which included electrical safety tests. (Implied compliance with the standard).
    Electromagnetic Compatibility (IEC 60601-1-2:2007)Support for substantial equivalence was provided as a result of risk management and testing which included performance tests. (Implied compliance with the standard).
    Alarm Systems (IEC 60601-1-8:2006)Support for substantial equivalence was provided as a result of risk management and testing which included safety tests. (Implied compliance with the standard, and new alarm functionality was verified for the "Low Calibration Failed" alarm).
    Usability (Human Factors Engineering) (ANSI/AAMI HE75)A formative usability study was conducted to aid in the development of requirements. (While a formative study is mentioned, the text doesn't explicitly state the acceptance criteria derived from this or the outcome in terms of meeting specific usability performance metrics, only that it aided requirement development.)

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes testing for the new software version and compatibility with two new ventilators. The "test set" consisted of:

    • Two additional respiratory care devices: Hamilton C1 Ventilator (K120574) and Hamilton T1 Ventilator (K120670).
    • INOmax DSIR® settings: Five settings were used: 0 (baseline), 1, 5, 20, 40, and 80 ppm, for each setting and mode of ventilation, as well as the Backup mode.

    The data provenance is non-clinical testing, performed in a controlled laboratory setting (likely within the company or a certified testing facility). There is no indication of country of origin of the data, but the company is based in Madison, Wisconsin, USA. The testing is prospective in the sense that it was conducted specifically to support this 510(k) submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    There is no mention of "experts" being used to establish ground truth in the context of the device's technical performance. The "ground truth" for this engineering validation would be the physical measurements taken by calibrated instruments, and the functionality verification against predetermined specifications. The text does not refer to human experts evaluating the "ground truth" of the device's performance characteristics.

    4. Adjudication Method for the Test Set

    Not applicable. This was a technical performance verification study, not a study involving human interpretation where adjudication would typically be used to resolve discrepancies in expert opinions.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The document explicitly states: "The subject of this premarket submission, INOmax DSix®, with updated software and interfaced to each of the selected respiratory care devices, did not require clinical studies to support substantial equivalence."

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, this was effectively a standalone performance evaluation of the device. The testing described assesses the device's ability to accurately deliver nitric oxide, monitor gases, and function with specific ventilators, without human intervention being part of the performance measurement itself (though human users operate the system). The tests focused on the device's intrinsic mechanical and software performance.

    7. The Type of Ground Truth Used

    The ground truth used for the performance testing was instrumental measurements of various parameters (e.g., O2 dilution, delivered pressures, NO delivery accuracy, NO2 generation) and functional verification against predetermined specifications for software features and alarms.

    8. The Sample Size for the Training Set

    Not applicable. This device is a hardware/software system, not an AI/ML algorithm that requires a "training set" in the conventional sense. The "software update" refers to deterministic code changes and feature enhancements, not a learned model.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" in the context of AI/ML. The "ground truth" for the device's design and functionality would be established through engineering specifications, previous predicate device performance, and compliance with recognized standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K130605

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-05-02

    (56 days)

    Product Code
    Regulation Number
    868.5165
    Age Range
    All
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The INOmax® DS delivery system delivers INOmax® (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators.

    The INOmax® DS provides continuous integrated monitoring of inspired O2, NO2, and NO, and a comprehensive alarm system.

    The INOmax® DS incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source.

    The INOmax® DS includes a backup NO delivery capability that provides a fixed flow of 250 mL/min of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breathing circuit. It may also use the INOblender® for backup.

    The target patient population is controlled by the drug labeling for INOmax® and is currently neonates. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.

    Device Description

    The INOmax DSIR® uses a "dual-channel" design to ensure the safe delivery of INOmax®. The first channel has the delivery CPU, the flow controller and the injector module to ensure the accurate delivery of NO. The second channel is the monitoring system, which includes a separate monitor CPU, the gas cells (NO, NO2, and O2 cells) and the user interface including the display and alarms. The dual-channel approach to delivery and monitoring permits INOmax® delivery independent of monitoring but also allows the monitoring system to shutdown INOmax® delivery if it detects a fault in the delivery system such that the NO concentration could become greater than 100 ppm. The delivery system can also shut down delivery if it detects certain serious problems with the monitoring system.

    AI/ML Overview

    Here's an analysis of the provided text regarding the INOmax DSIR device, focusing on acceptance criteria and the supporting study:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided text describes a compatibility study rather than a traditional performance study with explicit numerical acceptance criteria for accuracy metrics. The study aims to demonstrate substantial equivalence by confirming compatibility with additional respiratory care devices. Therefore, the "acceptance criteria" are implied to be the device performing within its published specifications and meeting the four necessary requirements for compatibility.

    Acceptance Criteria CategorySpecific Criteria (Implied)Reported Device Performance
    Compatibility Requirements1. O2 dilutionPerformed within published specifications.
    2. Effect on delivered pressuresPerformed within published specifications.
    3. INOmax DSIR® delivery accuracyPerformed within published specifications.
    4. NO2 generationPerformed within published specifications.
    Overall PerformanceDevice performs within its own established published specifications with new ventilators.The INOmax DSIR® performed within published specifications when used with each of the ventilators in both primary and backup delivery.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The text states, "Five INOmax DSIR® settings were used [0 (baseline), 1, 5, 20, 40, and 80 ppm] for each setting and mode of ventilation, as well as the Backup mode." It also mentions "both devices were set up and calibrated... and tested using the settings established for each respiratory care device test." The "new respiratory care devices" are the Drager Apollo Anesthesia Ventilator and the CareFusion ReVel Ventilator.
      • This implies testing across different modes of ventilation (which are not specifically enumerated, but would be several for each ventilator) and these 6 concentrations for each mode, for both primary and backup delivery, across two different new ventilators.
      • While specific case numbers aren't given in a medical imaging sense, the "sample size" here refers to the extensive set of configurations and parameters tested: 2 ventilators * (multiple modes) * 6 NO concentrations * 2 delivery types (primary/backup).
    • Data Provenance: The study is nonclinical (laboratory testing) and was conducted by the submitter (INO Therapeutics doing business as Ikaria). The country of origin is not explicitly stated but can be inferred to be the USA, where the company is based and where the 510(k) submission was made. The study is prospective as it involves active testing of the device under specific conditions.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of nonclinical, engineering-focused study does not typically involve human experts establishing "ground truth" in the way a clinical diagnostic study would. The "ground truth" in this context is the objectively measured performance of the device against its own published specifications and the expected behavior of the ventilators. The measurements would be taken by trained technicians or engineers following established protocols. No information is provided about expert qualifications or numbers beyond "manufacturer's recommendations" for setup and calibration.

    4. Adjudication Method (for the test set)

    No formal adjudication method (like 2+1 or 3+1 consensus) is described, as this is not a study involving human interpretation of clinical data. The "adjudication" is based on objective measurements and comparison against established specifications. "Any anomalies found" were recorded, implying a review of results by the testing personnel.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This study is a nonclinical, engineering compatibility test, not a clinical study involving human readers or cases.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, this was a standalone performance study in the sense that it evaluated the performance of the device (INOmax DSIR®) itself when interfaced with two specific ventilators. There was no human-in-the-loop component being evaluated for its diagnostic or therapeutic effectiveness; rather, the device's ability to maintain its intended performance characteristics in a new configuration was assessed.

    7. The Type of Ground Truth Used

    The ground truth used was objective performance measurements of the INOmax DSIR® against its published specifications and the expected operational parameters of the ventilators (e.g., delivered oxygen concentration, pressure, NO concentration, NO2 generation).

    8. The Sample Size for the Training Set

    No training set is mentioned or applicable. This is a conformance and compatibility test, not a machine learning study.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there was no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K110344

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-06-24

    (140 days)

    Product Code
    Regulation Number
    868.5165
    Age Range
    All
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The INOmax DS delivery system delivers INOmax® (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators.

    The INOmax DS provides continuous integrated monitoring of inspired O2, NO2 and NO, and a comprehensive alarm system.

    The INOmax DS incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source.

    The INOmax DS includes a backup NO delivery capability that provides a fixed flow of 250 mL/min of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breathing circuit. It may also use the INOblender for backup.

    The target patient population is controlled by the drug labeling for INOmax and is currently neonates. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.

    Device Description

    The INOmax DSIR uses a "dual-channel" design to ensure the safe delivery of INOmax. The first channel has the delivery CPU the flow controller and the injector module to ensure the accurate delivery of NO. The second channel is the monitoring system, which includes a separate monitor CPU, the gas cells (NO, NO2, and O2 cells) and the user interface including the display and alarms. The dual-channel approach to delivery and monitoring permits INOmax delivery independent of monitoring but also allows the monitoring system to shutdown INOmax delivery if it detects a fault in the delivery system such that the NO concentration could become greater than 100 ppm.

    AI/ML Overview

    The provided text describes the INOmax DS delivery system and its acceptance criteria, primarily focusing on its compatibility with the Vapotherm Precision Flow system. It does not contain information about a study proving the device meets the acceptance criteria in the typical sense of a clinical statistical study for AI/machine learning devices. Instead, it describes non-clinical engineering tests.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance MetricAcceptance CriteriaReported Device Performance
    O2 Dilution (with Vapotherm)Measured O2 within ± 4% v/v of the expected value (does not include Precision Flow tolerance)The measured O2 was within ± 4% v/v of the expected value.
    INOmax DS Delivery AccuracyMeasured values +15% of setting or 1.3 ppm, whichever is greater (within INOmax DSIR specification)The measured values were +15% of setting or 1.3 ppm whichever is greater, which is within specification of the INOmax DSIR.
    NO2 Generation (Max at 80 ppm NO, 100% O2)Expected behavior (no specific numerical limit stated, but within label limits)The maximum generated NO2 was at 80 ppm and 100% oxygen, as would be expected.
    NO2 Levels (general, with Vapotherm)Did not exceed 1.3 ppm on any flow settings (per INOmax label < 3 ppm on 100% FiO2 and any dose setting)NO2 levels did not exceed 1.3 ppm on any of the flow settings.
    NO2 Levels (at 40 ppm NO, 60% O2)< 1 ppm (per FDA guideline)The maximum generated NO2 at 40 ppm and 60% oxygen was 0.4 ppm.
    Backup Delivery (NO Accuracy)Within 5% of the INOmax DSIR specification at flow rates above 5 L/minThe delivered NO during backup delivery was within 5% of the INOmax DSIR specification at flow rates above 5 L/min.
    Backup Delivery (NO2 Levels)Did not exceed 0.2 ppm on any settings tested and 100% oxygen (per INOmax label < 3 ppm on 100% FiO2 and any dose setting)Measured NO2 levels did not exceed 0.2 ppm on any settings tested and 100% oxygen.
    Backup Delivery (O2 Accuracy)Measured O2 within + 3% v/v of the expected valueThe measured O2 was within + 3% v/v of the expected value.

    2. Sample Size Used for the Test Set and Data Provenance
    The document describes non-clinical laboratory testing ("The testing concluded four requirements necessary for the operation of the INOmax DSIR and Vapotherm Precision Flow system to be compatible"). It does not specify a "test set" in terms of patient data or a sample size of cases. The tests relate to the device's compatibility and performance characteristics. Therefore, there is no mention of data provenance (e.g., country of origin, retrospective/prospective).

    3. Number of Experts Used to Establish Ground Truth and Qualifications
    Not applicable. This was a non-clinical engineering validation, not a study requiring expert readers to establish ground truth from patient data.

    4. Adjudication Method for the Test Set
    Not applicable. There was no "test set" in the context of human-reviewed data, and thus no adjudication method as described.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The submission explicitly states: "The subject of this premarket submission, INOmax DSJR interfaced to Vapotherm Precision Flow did not require clinical studies to support substantial equivalence."

    6. Standalone Performance Study
    Yes, in a sense. The described tests ("Summary of Non-Clinical Tests") represent standalone performance evaluations of the INOmax DSIR's functional parameters and its compatibility when interfaced with the Vapotherm Precision Flow system. It's not "algorithm only without human-in-the-loop performance" in the AI context, but rather device-only performance in an engineering context. The results validated that "the additional interface to the Vapotherm Precision Flow system did not alter the safety or effectiveness of the INOmax DS operation."

    7. Type of Ground Truth Used
    The "ground truth" was established based on engineering specifications, expected values, and regulatory guidelines (e.g., "within specification of the INOmax DSIR," "per INOmax label," "per FDA guideline"). It was not derived from expert consensus, pathology, or outcomes data.

    8. Sample Size for the Training Set
    Not applicable. This is not an AI/machine learning device; therefore, there is no concept of a training set as used in that context. The device's operation is based on predefined mechanical and electronic controls and sensors.

    9. How Ground Truth for the Training Set Was Established
    Not applicable, as there is no training set.