Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K223037

    Validate with FDA (Live)

    Date Cleared
    2023-07-21

    (295 days)

    Product Code
    Regulation Number
    880.2910
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Infrared Thermometer (Model: IR8807) is intended for the intermittent measurement of human body temperature from the forehead. It is a non-contact clinical thermometer and intended for use on people of all ages except for neonates/newborns.

    Device Description

    The Infrared Thermometer (Model: IR8807) is an electronic thermometer using an infrared sensor to measure infrared energy radiated from the forehead. This energy is collected through the lens and converted to a temperature value. The Infrared Thermometer (Model: IR8807) has a forehead measurement mode and an object measurement mode.

    The Infrared Thermometer (Model: IR8807), consists of the following parts:
    a) Thermopile Sensor
    b) Application-Specific Integrated Circuitry
    c) Erasable Programmable Read-Only Memory Integrated Circuit
    d) Capacitance-touch Integrated Circuit
    e) LCD display
    f) 2 buttons (Scan button, Setting button,)
    q) 2×1.5V AAA dry batteries

    The operation principle is based on infrared sensor technology. The sensor in the thermometer detects the infrared energy, and the device processes the data to body temperature result with its internal microcontroller unit. It operates in an adjusted mode.

    The main functions of Infrared Thermometer (Model: IR8807) are as followings:
    Forehead temperature measure functions
    Wide range of temperature readings: from 32.0ºCto 42.9ºC (89.6ºF ~ 109.2ºF)
    The measurement results can be displayed in either °Cor °F units
    The LCD display showing clear measurement result.
    Memory data reading/removal function: The device can store a total of 32 groups of memorable values. All memory values can be removed after the battery is removed
    Low battery indicator
    Buzzer reminder function
    Users can set temperature alarm value
    Environment or object temperature measurement function
    The reference body site of the output temperature is Oral.

    AI/ML Overview

    The provided document describes the Infrared Thermometer, Model: IR8807, and its compliance with relevant standards and performance criteria. Here's a breakdown of the requested information based on the text:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (from ASTM E 1965-98)Reported Device Performance (Model: IR8807)
    Measurement Range32°C ~ 42.9°C (89.6°F ~ 109.2°F)
    Accuracy (Forehead mode)±0.2°C for 35.0°C ~ 42.0°C
    ±0.3°C outside this temperature range
    Measurement Time≤3 seconds
    Measurement Distance2-5 cm

    Note: The document states that the device's accuracy requirements meet ASTM E1965-98. The specific ASTM E1965-98 requirements for measurement range, accuracy, measurement time, and measurement distance are not explicitly detailed in the provided text beyond what's shown for the device's performance. The table above uses the device's reported performance as the "reported device performance" and implies these meet the ASTM E 1965-98 criteria because the document explicitly states compliance.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample size for clinical test set: 169 subjects.
    • Data Provenance: The document does not specify the country of origin. The test was a "clinical test" which implies a prospective study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided in the document. The document only mentions "clinical tests were conducted" and that the "clinical performance of the subject device complied with the requirement of ASTM E1965-98 (2016)." It does not describe the method or personnel used to establish the ground truth (reference temperature).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided in the document.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is an Infrared Thermometer. The document describes clinical tests for device accuracy, not a comparative effectiveness study involving human readers or AI assistance. Therefore, no MRMC comparative effectiveness study was done in the context of human readers improving with or without AI.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Yes, a standalone performance study was done for the device. The "clinical tests" described in section 8 were conducted on the Infrared Thermometer (Model: IR8807) itself, evaluating its direct measurement performance against a ground truth (implied by compliance with ASTM E1965-98 for clinical accuracy). This assesses the algorithm/device's performance without human interpretation or intervention in the measurement process beyond standard use.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The document does not explicitly state the type of ground truth used. However, given that it's a clinical thermometer study and states compliance with ASTM E1965-98, the ground truth would most likely be a reference clinical thermometer (e.g., a highly accurate rectal, oral, or tympanic thermometer, depending on the standard's protocol for correlating to forehead temperature) used simultaneously or sequentially with the test device.

    8. The sample size for the training set

    This device is a physical thermometer, not an AI/ML model that typically requires a "training set." The concept of a training set is not applicable to this device.

    9. How the ground truth for the training set was established

    As the concept of a training set is not applicable to this device, this question is not relevant.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1