510(k) Data Aggregation

    K Number
    K242775

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-11-12

    (60 days)

    Product Code
    Regulation Number
    868.1890
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The InPen System is a home-use reusable pen for single-patient use by people with diabetes under the supervision of an adult caregiver, or by a patient aged 7 and older for the self-injection of a desired dose of insulin. The pen injector is compatible with Lilly Humalog® U-100 3.0 mL cartridges, Novo Nordisk Novolog® U-100 3.0 mL cartridges, and Novo Nordisk Fiasp® U-100 3.0 mL cartridges and single-use detachable and disposable pen needles (not included). The pen Smart Insulin pen allows the user to dial the desired dose from 0.5 to 30 units in one-half (1/2) unit increments.

    The InPen dose calculator, a component of the InPen app, is indicated for the management of diabetes by people with diabetes under the supervision of an adult caregiver, or by a patient aged 7 and older for calculating an insulin dose or carbohydrate intake based on user entered data.

    For an insulin dose based on amount of carbohydrates, a healthcare provide patient-specific target blood glucose, insulin-to-carbohydrate ratio, and insulin sensitivity parameters to be programmed into the software prior to use.

    For an insulin dose based on fixed/variable meal sizes, a healthcare provide patient-specific fixed doses/ meal sizes to be programmed into the software prior to use.

    Device Description

    The InPen App is a software application with versions that are compatible with mobile phones running the iOS or Android operating system. The App is a component of the InPen system and is used with the InPen Smart Insulin pen for the management of insulin-requiring diabetes. The InPen App communicates with the InPen Smart Insulin pen to communicate doses that are delivered by the user. The InPen App is also compatible for use with blood glucose (BG) meters, Medtronic Continuous Glucose Monitors (CGMs), and the Dexcom CGMs. The InPen App includes a dose calculator that can calculate and recommend a dose for the user to review and consider as part of following the treatment plan prescribed by the healthcare provider. The dose calculator features in the App require that a healthcare professional provide patient-specific values for various therapy settings for programming into the App prior to use by the patient. These therapy settings include glucose target(s), duration of insulin action time, insulin sensitivity factor(s), and insulin-to-carbohydrate ratio(s) or fixed insulin doses for meal types and sizes. The dose calculator feature is unavailable to the user until these patient-specific values, provided by the healthcare professional, are programmed and an InPen has been paired to the App. A healthcare provider may also provide long-acting insulin settings to be programmed into the InPen App. The App includes a logbook feature that displays the patient's recent activity related to BG values, meal types and sizes, dose calculations, doses by insulin type (rapid- or long-acting), cartridge replacement and priming. The App also provides reminders and alerts that can notify the user to check their glucose, dose insulin (for potential missed meals, correction doses, and longacting insulin doses) and log doses according to schedule, replace a cartridge, or if the insulin pen has been exposed to very low or very high temperatures. The App can generate a supplemental summary report of recent therapy information for review by the patient or healthcare professional (HCP).

    The InPen Cloud includes a therapy report component that the user and the health care provider (HCP) can view and print to assess the overall diabetes control and treatment plan. The report displays data based upon user and HCP-defined inputs, such as glucose, insulin and carb trended information, during the defined period, as well as dose calculator usage and alerts and reminders. The Insulin Notification Service (INS) is a subcomponent of the InPen Cloud that can receive Medtronic CGM sensor glucose measurements from the CareLink Cloud. The INS includes two algorithms that assess "real time" sensor glucose measurements to identify whether a user has missed a dose or if their glucose is rising and a correction dose is needed. If either of these conditions exist, a silent notification is sent by the INS to the InPen App. The InPen App confirms the data and can provide and visual alert to the user. The user can act on the alert by assessing their glucose levels followed by calculating a dose utilizing the InPen App dose calculator.

    AI/ML Overview

    The provided text is a 510(k) summary for the Medtronic MiniMed InPen System. It describes the device, its indications for use, and a comparison to a predicate device to establish substantial equivalence. However, the document does not contain the specific details about acceptance criteria, the study design (including sample size, data provenance, expert ground truth establishment, or adjudication methods), or performance data in the way requested in the prompt.

    The "Performance Data" section (Page 10) broadly states that "Software verification and validation testing was performed in accordance with the FDA's Guidance..." and mentions "unit level testing, integration level testing, and systems testing." It also notes "Cybersecurity Testing" and "risk management activities." Finally, it mentions an "additional assessment of the changes to the InPen Cloud occurred through a summative usability evaluation" where "patients...used the InPen Cloud to perform a series of critical tasks involving the use of the InPen App and Cloud, including the additional alerts and reminder."

    Therefore, it is not possible to fill in most of the requested information based on the provided text. The document focuses on regulatory arguments for substantial equivalence rather than detailed clinical or technical study results that would typically include such specifics.

    Here's a breakdown of what can be extracted and what cannot:

    Information that CANNOT be extracted from the provided text:

    1. A table of acceptance criteria and the reported device performance: This detail is not present. The document focuses on showing the device is substantially equivalent to a predicate, not on specific performance metrics against pre-defined acceptance criteria.
    2. Sample sizes used for the test set and the data provenance: No specific sample sizes for particular tests (e.g., test sets for algorithms) are mentioned, nor is the data provenance (country, retrospective/prospective). The "summative usability evaluation" is mentioned, but without sample size or details about the data used.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not mentioned.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not mentioned.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not mentioned. This type of study is more common for diagnostic imaging AI, whereas this device is an insulin management system with a dose calculator.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: While the InPen App has an algorithm for dose calculation and the InPen Cloud has algorithms for notifications, the document doesn't provide performance data for these algorithms in isolation. The "summative usability evaluation" implies human-in-the-loop testing.
    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc): Not explicitly stated how the accuracy of the dose calculations or alert effectiveness was verified as "ground truth." The "summative usability evaluation" likely involved checking if users performed tasks correctly, but the underlying ground truth for algorithm accuracy isn't detailed.
    8. The sample size for the training set: Not mentioned.
    9. How the ground truth for the training set was established: Not mentioned.

    Information that CAN be extracted or inferred from the provided text:

    • Device Type: Insulin management system (reusable pen, app with dose calculator).
    • Regulatory Class: Class II (Product Code: NDC).
    • Software Level of Concern: Major.
    • Key Software Features: Dose calculator (insulin dose based on carbohydrates or fixed/variable meal sizes), logging, reminders, alerts (missed dose, high glucose).
    • New Features (compared to predicate):
      • Two new algorithms in the InPen Cloud (Insulin Notification Service) to identify and send "silent notifications" to the InPen App for missed insulin doses or rising glucose.
      • Two new visual and audible alerts in the InPen App for missed insulin doses or high glucose, for users with a Medtronic CGM.
      • Improved existing Long-Acting Reminder to include both audible and visual alerts.
    • Performance Data (General Statement): "Software verification and validation testing was performed in accordance with the FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions'', and FDA's guidance "General Principles of Software Validation". Verification and validation activities included software testing consisting of unit level testing, integration level testing, and systems testing of the InPen App and InPen Cloud. Cybersecurity Testing was performed... Risk management activities... were undertaken... Additional assessment of the changes to the InPen Cloud occurred through a summative usability evaluation."

    In summary, the provided document is a regulatory submission focused on substantial equivalence, not a detailed technical report of study results with specific performance metrics and study methodologies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201337

    Validate with FDA (Live)

    Device Name
    InPen System
    Date Cleared
    2020-06-18

    (29 days)

    Product Code
    Regulation Number
    868.1890
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The InPen is a home-use reusable pen injector for single-patient use by people with diabetes under the supervision of an adult caregiver, or by a patient age 7 and older for the self-injection of a desired dose of insulin. The pen injector is compatible with Lilly Humalog® U-100 3.0 mL cartridges, Novo Nordisk Novolog® U-100 3.0 mL cartridges, and Novo Nordisk Fiasp® U-100 3.0 mL cartridges and single-use detachable and disposable pen needles (not included). The pen injector allows the user to dial the desired dose from 0.5 to 30 units in one-half (1/2) unit increments.

    The InPen dose calculator, a component of the InPen app, is indicated for the management of diabetes by people with diabetes under the supervision of an adult caregiver, or by a patient age 7 and older for calculating an insulin dose or carbohydrate intake based on user entered data.

    For an insulin dose based on amount of carbohydrates, a healthcare professional must provide patient-specific target blood glucose, insulin-to-carbohydrate ratio, and insulin sensitivity parameters to be programmed into the software prior to use.

    For an insulin dose based on fixed/variable meal sizes, a healthcare professional must provide patient-specific fixed doses/meal sizes to be programmed into the software prior to use.

    Device Description

    The InPen System consists of a manually-controlled pen injector and an app containing a logbook and a dose (bolus) calculator.

    The InPen is a manual pen injector containing a non-replaceable battery and electronics to communicate via Bluetooth® with the app on an iOS® mobile device. The intended dose is manually set by the user by rotating a dose knob. The insulin is injected by manually depressing the dose knob which causes the piston in the insulin cartridge to expel the intended dose. The InPen is provided in two different models for compatibility with the available U-100 insulin cartridges, i.e. Humalog®, Novolog®, and Fiasp®. The device is provided with Instructions For Use and a Quick Start Guide. The device is used with sterile needles and U-100 insulin cartridges (supplied separately).

    The app is designed to manage the wireless transfer of insulin dose data from the InPen, log insulin dose data, and provide a dose calculator to aid mealtime insulin dose calculations. The insulin dose calculations provided by the app are meant for patients undergoing multiple daily injection (MDI) therapy. The InPen app is not intended to serve as an accessory to an insulin pump.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the InPen System, based on the provided FDA 510(k) summary:

    Important Note: The provided document is a 510(k) summary for a substantial equivalence determination. This means the device is being compared to a previously cleared predicate device (K190487). The performance data presented often focuses on demonstrating that the new device meets the same standards as the predicate and that any changes (like expanded indications for use) do not raise new questions of safety or effectiveness. It does not typically include detailed, de novo clinical trial results for the predicate device.


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (What the device must achieve)Reported Device Performance (How the device performed)
    InPen Pen Injector:
    Dose Accuracy (Meets ISO 11608-1 requirements)Meets ISO 11608-1 requirements
    Biocompatibility (Meets ISO 10993-1 requirements)Meets ISO 10993-1 requirements
    Compliance with regulations for a Piston Syringe (21 CFR 880.5860; Class II)Classified as Class II, Product Codes FMF, NDC
    InPen App Dose Calculator:
    Ability to calculate insulin dose or carbohydrate intake based on user-entered data for people with diabetes (including pediatric users age 7 and older)Indicated for calculating an insulin dose or carbohydrate intake based on user-entered data for patients age 7 and older (with adult caregiver supervision or self-injection). Specific parameters (target blood glucose, insulin-to-carbohydrate ratio, insulin sensitivity, fixed/variable meal sizes) must be programmed by a healthcare professional.
    Compliance with regulations for a Medical Device Data System (21 CFR 868.1890)Classified as Class II, Product Codes FMF, NDC
    Appropriate for pediatric patients and their healthcare providersHuman Factors evaluation and literature review determined appropriateness for pediatric patients. Updated labeling provides further information regarding potential risks for pediatric users.
    Ensure safety and effectiveness when expanding indications for Fiasp compatibility.Drug compatibility was updated to reflect the addition of Fiasp compatibility in a previous submission. (Implies previous demonstration of safety/effectiveness with Fiasp).
    Ensure safety and effectiveness despite changes in Indications For Use (specifically, the expanded age range for pediatric users and Fiasp compatibility).The document states: "The differences in Indications For Use do not raise new questions of safety or effectiveness." This is supported by human factors evaluations and the existing technological characteristics being the same as the predicate.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state a numerical sample size for a "test set" in the context of an algorithm's performance evaluation.

    • Provenance: The primary basis for comparison is a previously cleared predicate device (K190487). The "study" referenced for the current submission primarily relates to human factors evaluation and literature review to support the expanded indications (pediatric use). This suggests the data provenance is a combination of:
      • Previously established performance of the predicate device (likely from earlier submissions).
      • New human factors data (prospective) related to the pediatric use case.
      • Literature review (retrospective).

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not explicitly provided in the summary. Since the evaluation primarily relies on demonstrating substantial equivalence to a predicate and conducting human factors testing for expanded use, it's unlikely a separate "ground truth" establishment by a panel of experts, as might be done for an AI diagnostic device, occurred for this specific 510(k) submission. Human factors may involve usability experts, but their role is different from establishing a diagnostic ground truth.

    4. Adjudication Method for the Test Set

    This information is not explicitly provided. Given the nature of the device (insulin pen and dose calculator) and the type of evaluation described (human factors, substantial equivalence), an adjudication method for a "test set" in the context of expert review for diagnostic ground truth is not typically applicable or detailed in this type of submission.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study is not mentioned or explicitly described. This type of study is more common for diagnostic imaging AI systems where human readers interpret cases with and without AI assistance. The InPen system is an insulin delivery and dose calculation device, not a diagnostic imaging tool.

    6. Standalone (Algorithm Only) Performance Study

    While the device includes a "dose calculator" component (an algorithm), the document does not describe a standalone "algorithm only" performance study in the way one might for a diagnostic AI. The substantial equivalence argument relies on the overall system (pen injector + app) meeting the same performance standards as the predicate, and the clinical parameters used by the dose calculator are provided by a healthcare professional. Human factors testing focuses on the user interface and interaction, rather than an isolated algorithm's numerical output accuracy.

    7. Type of Ground Truth Used

    For the InPen pen injector:

    • Engineering standards: Adherence to ISO 11608-1 (dose accuracy) and ISO 10993-1 (biocompatibility). These standards define the ground truth for mechanical and material performance.

    For the InPen App dose calculator (algorithm):

    • Clinical parameters/guidelines: The algorithm's calculations rely on patient-specific parameters (target blood glucose, insulin-to-carbohydrate ratio, insulin sensitivity, fixed/variable meal sizes) programmed by a healthcare professional. The "ground truth" for the dose calculation would be its accurate application of these parameters (as per established medical guidelines for insulin dosing) to user-entered data.
    • Human Factors Evaluation: For the expanded pediatric indication, human factors testing serves as the ground truth for user interface and usability safety/effectiveness in the intended population.

    8. Sample Size for the Training Set

    This information is not provided. The document is concerned with the current device's performance relative to a predicate, not the development or training of a de novo AI algorithm. The "dose calculator" is likely based on established physiological formulas and clinical guidelines rather than machine learning trained on a large dataset in the conventional AI sense.

    9. How the Ground Truth for the Training Set Was Established

    Since a "training set" in the context of de novo machine learning is not implied or described, the method for establishing ground truth for such a set is not applicable or provided. The dose calculation logic is based on well-established medical algorithms and clinical parameters set by healthcare professionals.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192841

    Validate with FDA (Live)

    Device Name
    InPen System
    Date Cleared
    2020-02-10

    (130 days)

    Product Code
    Regulation Number
    868.1890
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The InPen is a home-use reusable pen injector for single-patient use by people with diabetes age 12 and older for the self-injection of a desired dose of insulin. The pen injector is compatible with Lilly Humalog® U-100 3.0 mL cartridges, Novo Nordisk Novolog® U-100 3.0 mL cartridges, and Novo Nordisk Fiasp® U-100 3.0 mL cartridges and single-use detachable and disposable pen needles (not included). The pen injector allows the user to dial the desired dose from 0.5 to 30 units in one-half (1/2) unit increments.

    The InPen dose calculator, a component of the InPen app, is indicated for the management of diabetes by people with diabetes age 12 and older by calculating an insulin dose or carbohydrate intake based on user entered data. Prior to use, a healthcare professional must provide the patient-specific target blood glucose, insulin-to-carbohydrate ratio, and insulin sensitivity parameters to be programmed into the software.

    Device Description

    The InPen System consists of a manually-controlled pen injector and an app containing a logbook and a dose (bolus) calculator.

    The InPen is a manual pen injector containing a non-replaceable battery and electronics to communicate via Bluetooth® with the app on an iOS® mobile device. The intended dose is manually set by the user by rotating a dose knob. The insulin is injected by manually depressing the dose knob which causes the piston in the insulin cartridge to expel the intended dose. The InPen is provided in two different models for compatibility with the available U-100 insulin cartridges, i.e. Humalog®, Novolog®, and Fiasp®. The device is provided with Instructions For Use and a Quick Start Guide. The device is used with sterile needles and U-100 insulin cartridges (supplied separately).

    The app is designed to manage the wireless transfer of insulin dose data from the InPen, log insulin dose data, and provide a dose calculator to aid mealtime insulin dose calculations. The insulin dose calculations provided by the app are meant for patients undergoing multiple daily injection (MDI) therapy. The InPen app is not intended to serve as an accessory to an insulin pump.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the InPen System, focusing on its substantial equivalence to a previously cleared predicate device (K160629). It primarily addresses the device's compatibility with an additional insulin cartridge (Fiasp®) and the InPen app's dose calculator.

    Based on the provided information, I can extract the following details regarding acceptance criteria and performance study:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't explicitly present a table of "acceptance criteria" for the entire InPen System in the typical sense of quantitative targets. Instead, it demonstrates continued conformance to existing standards and principles for both the pen injector and the app.

    Here's a summary of the attributes compared and the reported performance, indicating that the new device meets the same standards as the predicate:

    AttributeAcceptance Criteria (Implied / Standard)Reported Device Performance
    InPen Pen Injector
    ClassificationClass II - FMF - 21 CFR §880.5860Same (Subject Device is Class II - FMF - 21 CFR §880.5860)
    Indications For UseSame as predicate, with added Fiasp® compatibilityFiasp® compatibility does not raise new safety/effectiveness questions. Labeling updated to mitigate errors.
    Cartridge Volume3 ml (300 units)Same
    DrugU-100 insulinSame
    Syringe TypePiston SyringeSame
    Single Patient UseYesSame
    Reusable DeviceYesSame
    Dose AccuracyMeets ISO 11608-1 requirementsMeets ISO 11608-1 requirements
    BiocompatibilityMeets ISO 10993-1 requirementsMeets ISO 10993-1 requirements
    Maximum Dose30 UnitsSame
    User FeedbackAudible and tactile clicks per incrementSame
    Dose DialingTwo-waySame
    BatteryNon-rechargeableSame
    ElectronicsFolded Flex CircuitSame
    SoftwareYesSame
    Dose DeliveryMechanicalSame
    Unit IncrementsHalf-Unit increments after 0.5 UnitSame
    Dimensions6.5″ x ø0.6″Same
    Weight35 gramsSame
    Fluid Pathway ContactNoneSame
    Dose Calculator CommunicationYesSame
    InPen App
    ClassificationClass II - NDC - 21 CFR §868.1890Same (Subject Device is Class II - NDC - 21 CFR §868.1890)
    Indications For UseSame as predicateSame (No changes to app)
    Prescription UseYesSame
    User GroupDiabetes patients with MDI therapySame
    Communication with insulin pumpsNoSame
    Software Level of ConcernMajorSame
    Wireless ConnectivityBluetooth Low Energy (BLE)Same
    Control or affect blood glucose measurementsNoSame
    Control or affect insulin deliveryNoSame
    Reports, graphs, and Electronic Log BookYesSame
    Meal Size EntryGrams of carbohydratesSame
    Insulin Dose CalculatorCalculates doses for meals & corrections, accounts for IOBSame
    Carbohydrate CalculatorCalculates carbohydrate intake based on user dataSame
    Manual Dose EntryYesSame
    InPen Dose EntryYesSame
    Tracking of residual bolus insulin to mitigate stackingYesSame
    Operating platformAndroid and iOS platformsSame
    UI StandardsAndroid and iOS standardsSame

    2. Sample size used for the test set and the data provenance:

    • InPen Pen Injector (Fiasp® compatibility):
      • Sample Size: Not explicitly stated as a number of devices or injections. The compliance is demonstrated by "demonstrating conformance to ISO 11608-1...and Part 2" and "hazard analysis according to ISO 14971". This typically involves a defined number of test units and test conditions as specified by the ISO standards.
      • Data Provenance: Not specified, but generally, ISO standard testing is conducted in controlled laboratory environments. No indication of retrospective or prospective human clinical data.
    • Labeling Changes Validation:
      • Sample Size: Not explicitly stated but mentions "human factors testing." Human factors testing involves a representative user sample, but the specific number is not provided.
      • Data Provenance: Not specified. This would typically be prospective data collected during human factors studies.
    • InPen Dose Calculator (Fiasp® compatibility):
      • Sample Size: Not applicable. The compatibility was shown through "clinical literature review." No new test set was used for this aspect of the calculator.
      • Data Provenance: Clinical literature (specifically referencing "Mudaliar et al (1999)").

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • For the InPen Pen Injector and Labeling Changes: Not explicitly stated. For ISO standard compliance, the "ground truth" is typically the standard itself, validated by accredited testing labs or internal quality assurance. For human factors, experts review the results, but their specific number and qualifications are not detailed.
    • For the InPen Dose Calculator: No new "ground truth" was established based on expert consensus for the Fiasp® compatibility beyond the existing clinical literature review. The "ground truth" for the original algorithm would have been established during the predicate device's clearance, likely through clinical trials and medical expert validation, but this information is not in the provided text.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • The document does not describe any adjudication method (like 2+1 or 3+1 refereeing) for establishing ground truth or evaluating performance. The assessment relies on compliance with ISO standards, hazard analysis, human factors testing, and clinical literature review.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC comparative effectiveness study was done or is mentioned in the provided text. The InPen System is not an AI-assisted diagnostic tool where "human readers" (e.g., radiologists) would be evaluating cases with or without AI assistance. The dose calculator is a direct calculation tool, not an AI interpretation system.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, for the InPen Pen Injector's compatibility with Fiasp®: Conformance to ISO 11608-1 (Part 1 and Part 2) is a standalone performance test of the device's mechanical and functional aspects, independent of a human user in a clinical setting at the time of testing.
    • Yes, for the InPen Dose Calculator's compatibility with Fiasp®: The assessment was based on "clinical literature review," which is an algorithm-only (or calculation-principle-only) evaluation against existing clinical understanding, without direct human-in-the-loop testing for this specific change. The app's calculator algorithm itself performs independently once parameters are set.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • InPen Pen Injector: Conformance to ISO 11608-1 standards (which define precision, accuracy, and other physical characteristics of injection systems) and ISO 14971 (risk management). These are engineering and safety standards, representing an accepted "ground truth" for device performance characteristics.
    • Labeling Changes: Likely human factors engineering principles and results of user testing, representing the "ground truth" for usability and clarity.
    • InPen Dose Calculator: Clinical literature review (specifically mentioning Mudaliar et al (1999) for duration of insulin action). This implies the "ground truth" for the dose calculation algorithm's underlying physiological models and parameters is derived from established medical knowledge and clinical research.

    8. The sample size for the training set:

    • The document describes a 510(k) submission for a device change (Fiasp® compatibility) and the app remaining unchanged. There is no mention of a "training set" in the context of machine learning or AI algorithm development for the current submission. The InPen dose calculator, as described, performs calculations based on user-entered data and pre-programmed parameters (target BG, I:C ratio, ISF) provided by a healthcare professional, rather than being an adaptive AI model trained on a dataset. The original algorithm for the predicate device would have been developed and validated, but details about its training set are not in this document.

    9. How the ground truth for the training set was established:

    • As noted above, no "training set" for a machine learning algorithm is discussed in this submission. For the underlying principles of the dose calculator, the "ground truth" for its rules and parameters would have been established through established medical guidelines, clinical trials, and medical expertise during the development of the original algorithm for the predicate device. The current submission relies on the existing algorithm and its previous validation, plus a literature review for Fiasp® compatibility, not a new training process.