Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K081147

    Validate with FDA (Live)

    Date Cleared
    2009-05-29

    (402 days)

    Product Code
    Regulation Number
    870.1210
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IsoFlow™ Infusion Catheter is a multi-lumen dual balloon catheter designed to isolate a specific treatment region from blood flow while allowing infusion of fluids into the region and perfusion of blood past the region.

    The device also has the ability to deliver physician specified fluids out the distal tip with or without inflation of the balloons.

    Device Description

    The IsoFlow™ Infusion Catheter is a multi-lumen dual balloon catheter designed to isolate a specific treatment region from blood flow while allowing infusion of fluids into the region and perfusion of blood past the region. The exterior surface of the distal 120cm catheter length is treated with a hydrophilic coating.

    The 2.4F IsoFlow™ Infusion Catheter System is intended to be used with guide catheters 4F (0.050") and larger, along with a 0.010" guide wire for positioning the catheter in the desired region. A standard Y-adaptor RHV provides for guide wire entry and saline flush into the main guide wire / bypass lumen of the catheter. Radiopaque markers at the distal tip and between the two balloons allow for final position adjustment under fluoroscopy guidance.

    The two compliant balloons are inflated simultaneously using radiopaque fluid delivered via a single inflation lumen.

    Physician specified infusion fluid through either the labeled infusion lumen or the labeled guidewire lumen is delivered via the 1-way stopcock connection. The mixture of infusion and radiopaque agents is delivered directly to the target region between the balloons, or out the distal tip respectively.

    For infusion out of the sideport holes, retracting the guide wire to the radiopaque marker band proximal to both balloons allows blood to bypass the isolated tarqet region via holes connecting the guide wire / bypass lumen with the catheter exterior. Complete removal of the guidewire allows delivery from the distal tip.

    All components of the catheter system are provided sterile. Each device is intended for single use. Do not reuse or attempt to resterilize any component of the system.

    AI/ML Overview

    The provided text does not contain detailed information about acceptance criteria, device performance,
    or a study comparing the device against specific criteria in the way typically found for AI/ML-based
    medical devices. The document is a 510(k) summary for a physical medical device—the IsoFlow™
    Infusion Catheter—seeking substantial equivalence to predicate devices.

    Therefore, many of the requested categories (such as sample size for test set, number of experts,
    adjudication method, MRMC study, standalone performance, training set details) are not applicable
    to this type of submission and are not present in the provided text.

    However, I can extract the information that is present concerning the device's characteristics and its
    comparison to predicate devices, which forms the basis for its acceptance (i.e., substantial
    equivalence) by the FDA.

    Here's a summary based on the provided text, addressing the applicable points:

    1. A table of acceptance criteria and the reported device performance

    For a traditional medical device like the IsoFlow™ Infusion Catheter, "acceptance criteria" are
    typically demonstrated through substantial equivalence to legally marketed predicate devices, along
    with adherence to performance standards and safety testing (which are summarized, but not detailed
    in a quantitative study in this specific 510(k) summary). The performance is shown by demonstrating
    that the device's characteristics are "Substantially Equivalent" to the predicates.

    CategoryAcceptance Criteria (Demonstrated by Substantial Equivalence to Predicate Devices)Reported Device Performance (IsoFlow™ Infusion Catheter)
    Intended UseSubstantially Equivalent to predicate devices (K932616, K913517)Substantially Equivalent
    Physical DescriptionSubstantially Equivalent to predicate devices (K932616, K913517)Substantially Equivalent
    Anatomical SitesSubstantially Equivalent to predicate devices (K932616, K913517)Substantially Equivalent
    DesignSubstantially Equivalent to predicate devices (K932616, K913517)Substantially Equivalent
    MaterialsSubstantially Equivalent to predicate devices (K932616, K913517)Substantially Equivalent
    Primary IndicationIsolate a specific treatment region from blood flow while allowing infusion/perfusion.Functionally meets this indication as stated.
    Secondary IndicationDeliver physician specified fluids out the distal tip with or without inflation.Functionally meets this indication as stated.

    2. Sample size used for the test set and the data provenance

    Not applicable. The document describes a 510(k) submission for a physical medical device (infusion
    catheter), not an AI/ML device relying on a test set of data. The "study" here is the comparison of
    the device's characteristics to predicate devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of
    those experts

    Not applicable. This information is typically relevant for AI/ML device evaluations. For this traditional
    medical device, "ground truth" mainly refers to the functional and material specifications meeting
    established medical device standards and being comparable to legally marketed devices.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is not relevant for the type of device and submission described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was
    the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This type of study is for AI/ML-assisted diagnostic or therapeutic devices, not for
    a physical infusion catheter.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is for AI/ML devices. The IsoFlow™ Infusion Catheter is a physical device
    used by a physician.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable in the context of an AI/ML device. For this physical catheter, the "ground truth" for
    its acceptance is established through:

    • Engineering design specifications and material science: Ensuring the catheter is constructed
      from biocompatible materials and designed to safely perform its intended functions.
    • Comparison to predicate devices: Demonstration of substantial equivalence in intended use,
      technological characteristics (physical description, anatomical sites, design, materials), and
      performance. This implies that the predicate devices themselves have a history of safe and
      effective use.
    • Performance testing: While not detailed in this summary, a 510(k) submission typically
      includes bench testing (e.g., burst pressure, tensile strength, flow rates, balloon integrity),
      biocompatibility testing, and sterilization validation to ensure safety and performance.

    8. The sample size for the training set

    Not applicable. There is no AI/ML model to train for this device.

    9. How the ground truth for the training set was established

    Not applicable. There is no AI/ML model to train for this device.