Search Filters

Search Results

Found 25 results

510(k) Data Aggregation

    K Number
    K203210

    Validate with FDA (Live)

    Date Cleared
    2021-04-12

    (164 days)

    Product Code
    Regulation Number
    868.5440
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Invacare® Platinum® 5NXG Oxygen Concentrator is indicated for patients with respiratory disorders requiring supplemental oxygen at flow rates of 1 to 5 liters per minute. For flow rates below 1 liter per minute use with Invacare Pediatric Flowmeter accessory. It is not intended to sustain or support life.

    Device Description

    The subject Invacare® Platinum® 5NXG Oxygen Concentrator device is used by patients who require supplemental oxygen. The device is not intended to sustain or support life. The oxygen concentration level of the output gas ranges from 87% to 96% at an output flow rate range of 1 to 5 liters per minute (LPM). The oxygen is delivered to the user through the use of a nasal cannula. The concentrator provides supplemental oxygen to patients with respiratory disorders, in the home, long term care facility or other non-acute patient care facilities. It is not intended to sustain or support life. It is to be used indoors and is intended to support the patient's lifestyle within the home or care facility. The device operates at a nominal 120VAC/60 Hertz supply.

    AI/ML Overview

    The provided document is a 510(k) Pre-Market Notification for the Invacare Platinum 5NXG Oxygen Concentrator. It describes the device, its intended use, and compares it to a predicate device (Invacare Perfecto2™V Oxygen Concentrator K200890) to demonstrate substantial equivalence.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a "table of acceptance criteria and reported device performance" in the format of pass/fail measurements against specific numerical criteria for all aspects. Instead, it describes general performance categories and states that the device "met the performance requirements" or "demonstrates that the subject device meets performance specifications and requirements."

    However, we can infer some criteria and reported performance from the "Design Characteristics Comparison" table on page 6 and the "Performance Testing – Bench" section on page 9.

    Acceptance Criteria (Inferred from Predicate/Standards)Reported Device Performance (Invacare Platinum 5NXG Oxygen Concentrator)
    Oxygen Output Concentration Levels87% to 96% at all flow rates
    (Predicate was 95.6% to 87% at all flow rates)
    Maximum Flow Rate5.0 L/min
    (Same as predicate)
    Adjustable Flow Range1 to 5 L/min in 0.5 L/min increments
    (Same as predicate)
    Oxygen Purity Fault IndicationIf oxygen purity falls below 82% but above 50%
    (Predicate was if oxygen purity falls below 85% (±2%) but above 73% (±3%))
    Rated Current Input3.0 A
    (Predicate was 3.2 A)
    Electrical Requirements120 VAC ±10%, 60 Hz
    (Same as predicate)
    Weight33 ± 2 lbs.
    (Predicate was 40 lbs.)
    Height24 ± 3/8"
    (Predicate was 23 ± 3/8")
    Width16 ± 3/8"
    (Predicate was 15 ± 3/8")
    Depth12 ± 3/8"
    (Same as predicate)
    Basic Safety and Essential PerformanceDevice meets performance specifications and requirements
    (Assessed against ISO 80601-2-69, AAMI ES60601-1, IEC 60601-1 etc.)
    Risk ManagementRisk profile unchanged, no significant impact on safety/effectiveness.
    (In accordance with ISO 14971:2012)
    Software Functionality (including alarms and indicators)Evaluated and verified
    (In accordance with FDA guidance and IEC 62304:2006, ISO 80601-2-69)
    Electrical Safety and EMCConfirmed post-change functionality, demonstrated equivalence
    (Against various IEC and AAMI standards)
    BiocompatibilityPerformed on components; met requirements for inorganic gases (CO, CO2, O2) and odor.
    (In accordance with ISO 18562 series and USP monograph)
    Power and Oxygen Output PerformanceMet performance specifications and requirements
    Transportation Testing (ISTA 2A)Successfully performed
    Sound TestingSuccessfully performed

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document refers to "design verification testing data" and "bench performance testing." These are laboratory or engineering tests performed on the device itself (or its components) rather than clinical data from a human population. Therefore, the concept of a "sample size" in the clinical sense (e.g., number of patients) is not applicable here. The testing involves physical devices.
    The provenance of the data is internal to the manufacturer (Invacare Corporation) as it describes their internal verification testing. The document does not specify the country of origin of the test data (e.g., if the tests were performed in the US, or elsewhere) nor does it explicitly state whether the testing was prospective (done specifically for this submission) or retrospective (using existing data), though the context of "design verification testing" implies it was specifically conducted for the submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This product is an oxygen concentrator, a medical device for delivering oxygen. It is not an AI/imaging device requiring expert interpretation for ground truth. Therefore, the concept of "experts establishing ground truth for a test set" with qualifications such as radiologists is not applicable to this submission. The ground truth for performance is established by objective measurements and adherence to recognized consensus standards by qualified testing engineers and laboratories.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. As noted above, this device does not involve human interpretation requiring adjudication for establishing ground truth. Performance is measured objectively.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The device is an oxygen concentrator, not an AI-assisted diagnostic tool. No MRMC study was performed, nor is it relevant to the device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. The device is an oxygen concentrator, not an algorithm or software-only device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the device's performance is established by objective measurements against defined engineering specifications and recognized consensus standards. This includes:

    • Measurements of oxygen concentration, flow rates, electrical characteristics, weight, dimensions, and sound levels.
    • Adherence to specific clauses and requirements within international standards like ISO 80601-2-69 (specific for oxygen concentrators), AAMI ES60601-1, IEC 60601-1 (general medical electrical equipment safety), IEC 62304 (software life cycle), ISO 14971 (risk management), and ISO 18562 series (biocompatibility).

    8. The sample size for the training set

    Not applicable. This is a hardware medical device, not an AI/machine learning model that requires a "training set" of data.

    9. How the ground truth for the training set was established

    Not applicable. There is no training set for this type of medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K202379

    Validate with FDA (Live)

    Date Cleared
    2021-03-19

    (211 days)

    Product Code
    Regulation Number
    890.3860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Invacare AVIVA Storm RX Power Wheelchair is indicated to provide mobility and positioning to persons limited to a sitting position.

    Device Description

    The subject device is a rearward drive wheel version substantially equivalent to the previously cleared forward drive wheel Invacare® AVIVA FX Power Wheelchair (K192216) with LiNX Electronics and Ultra Low Maxx Seating System. The subject device version consists of the following changes:

    • . Rearward drive wheel placement
    • . Modification to suspension system
    • . Modification to battery box design and seat interface brackets
    • Minor changes to mechanical components such as new rims and colors .
      The Invacare® AVIVA Storm RX Power Wheelchair is a 24V DC system, motor-driven wheelchair, utilizing the predicate device LiNX® control system. The subject device consists of a rigid or "non-folding" type power wheelchair base with rearward drive wheel placement with two (2) casters in the front and two anti-tippers in the rear. It is powered by two 12-volt DC batteries and two 4-pole single stage drive motors.
    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a powered wheelchair, the Invacare AVIVA Storm RX Power Wheelchair. It focuses on demonstrating substantial equivalence to a predicate device (Invacare AVIVA FX Power Wheelchair) through non-clinical testing against various international standards.

    Here's an analysis of the acceptance criteria and study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally established by compliance with a comprehensive set of international (ISO), American National Standards Institute (ANSI), and European (EN) standards. The reported device performance is that the "subject device components met the performance requirements."

    Acceptance Criteria (Test Standard)Reported Device Performance
    ISO 7176-1:2014: Wheelchairs Part 1: Determination of Static StabilitySubject device met performance requirements.
    ISO 7176-2:2017: Wheelchairs Part 2: Determination of Dynamic Stability of Electrically Powered WheelchairsSubject device met performance requirements.
    ISO 7176-3:2012: Wheelchairs Part 3: Determination of Effectiveness of BrakesSubject device met performance requirements.
    ISO 7176-4:2008: Wheelchairs Part 4: Energy Consumption of Electrical Wheelchairs and Scooters for Determination of Theoretical Distance RangeSubject device met performance requirements.
    ISO 7176-5:2008: Wheelchairs Part 5: Determination of Dimensions, Mass and Maneuvering SpaceSubject device met performance requirements.
    ISO 7176-6:2018: Wheelchairs Part 6: Determination of Maximum Speed, Acceleration and Deceleration of Electric WheelchairsSubject device met performance requirements.
    ISO 7176-7:1998: Wheelchairs Part 7: Measurement of Seating and Wheel DimensionsSubject device met performance requirements.
    ISO 7176-8:2014: Wheelchairs Part 8: Requirements and Test Methods for Static, Impact and Fatigue StrengthsSubject device met performance requirements.
    ISO 7176-9:2009: Wheelchairs Part 9: Climatic Tests for Electric WheelchairsSubject device met performance requirements.
    ISO 7176-10:2008: Wheelchairs Part 10: Determination of Obstacle Climbing Ability of Electrically Powered WheelchairsSubject device met performance requirements.
    ISO 7176-11:2012: Wheelchairs Part 11: Test DummiesSubject device met performance requirements.
    ISO 7176-13:1989: Wheelchairs Part 13: Determination of Coefficient of Friction of Test SurfaceSubject device met performance requirements.
    ISO 7176-14:2008: Wheelchairs Part 14: Power and Control Systems for Electrically Powered Wheelchairs and Scooters – Requirements and Test MethodsSubject device met performance requirements.
    ISO 7176-15:1996: Wheelchairs Part 15: Requirements for Information Disclosure, Documentation and LabelingSubject device met performance requirements.
    ISO 7176-16:2012: Wheelchairs Part 16: Resistance to Ignition of Postural Support DevicesSubject device met performance requirements.
    ISO 7176-19: Wheelchairs Part 19: Wheeled Mobility Devices for use as Seats in Motor VehiclesSubject device met performance requirements.
    ISO 7176-21:2008: Wheelchairs Part 21: Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and scooters, and battery chargersSubject device met performance requirements.
    ISO 7176-22:2014: Wheelchairs Part 22: Set-up ProceduresSubject device met performance requirements.
    ISO 7176-25:2013: Wheelchairs Part 25: Batteries and chargers for powered wheelchairsSubject device met performance requirements.
    IEC 62304:2006: Medical Device Software – Software Life CycleSubject device met performance requirements.
    ANSI C63.27:2017: American National Standard for Evaluation of Wireless CoexistenceSubject device met performance requirements.
    Other tests mentioned (Flammability, Ignitability of Upholstered Furniture)"Subject device components and features are the same or substantially equivalent to the predicate device."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a "test set" in terms of a patient cohort or a specific number of devices tested individually. Instead, it refers to "non-clinical test reports" and "design verification test reports" that support the finding of substantial equivalence. The testing is based on compliance with established international standards for wheelchairs. Therefore:

    • Sample size: Not explicitly stated as a numerical sample of devices or patients. The assumption is that sufficiently representative samples were used to conduct the required tests under each standard.
    • Data provenance: Not explicitly stated, as the focus is on compliance with standards rather than clinical data from a specific geographical region or retrospective/prospective study. The testing is implied to be laboratory-based non-clinical testing.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This document does not describe the use of human experts to establish "ground truth" for a performance test set. The evaluation is based on objective, standardized non-clinical engineering and performance tests carried out by the manufacturer to demonstrate compliance with recognized standards.

    4. Adjudication Method

    There is no mention of an adjudication method as the evaluation relies on compliance with technical standards and objective measurements, not subjective expert reviews.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for AI or diagnostic devices where human reader performance is a key metric. This submission is for a physical medical device (power wheelchair) where performance is evaluated against engineering standards.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This question is not applicable. The device is a power wheelchair, not an algorithm, and its performance is evaluated against engineering and safety standards, not an algorithm's diagnostic output.

    7. The Type of Ground Truth Used

    The "ground truth" in this context is implicitly defined by the pass/fail criteria of the various international and national standards (ISO, IEC, ANSI) listed in the document. These standards define objective performance thresholds and test methods for aspects like stability, braking, speed, safety, and electromagnetic compatibility.

    8. The Sample Size for the Training Set

    This question is not applicable. There is no mention of a "training set" as this is a physical medical device, not a machine learning algorithm.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200890

    Validate with FDA (Live)

    Date Cleared
    2020-08-06

    (125 days)

    Product Code
    Regulation Number
    868.5440
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Invacare® Perfector™V Oxygen Concentrator is indicated for patients with respiratory disorders requiring supplemental oxygen at flow rates of 1 to 5 liters per minute. For flow rates below 1 liter per minute use with Invacare Pediatric Flowmeter accessory. It is not intended to sustain or support life.

    Device Description

    The purpose of this this Premarket Notification [510(k)] submission is to obtain commercial clearance for Invacare® Perfecto2™V Oxygen Concentrator. The subject device is used by patients who require supplemental oxygen concentration level of the output gas ranges from 87% to 95.6% at an output flow rate range of 1/2 to 5 liters per minute. The oxygen is delivered to the user through the use of a nasal cannula.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for the Invacare Perfecto2V Oxygen Concentrator. It asserts substantial equivalence to a predicate device, the Invacare Platinum 5 Oxygen Concentrator (K020386), rather than providing detailed acceptance criteria and a study to prove meeting those criteria in the context of a new functional claim (e.g., related to AI performance).

    Therefore, I cannot extract the information required for the requested table and study details (acceptance criteria, performance, sample sizes, expert involvement, adjudication, MRMC, standalone, ground truth, training set specifics) because the provided text describes a submission for a medical device (oxygen concentrator), not a software/AI device that would involve the assessment of algorithmic performance against a ground truth.

    The document primarily focuses on:

    • Substantial Equivalence: Demonstrating that the new device has the same intended use, technological characteristics, and performs similarly to a previously cleared predicate device.
    • Bench Testing: Performance testing (e.g., oxygen concentration levels, flow rates, electrical safety, biocompatibility) to show the device meets its defined specifications and is equivalent to the predicate.
    • Risk Management and Software Verification: Ensuring the device's design and software meet relevant standards.

    There is no mention of an AI component, image analysis, human reader studies, or any of the elements typically associated with the kind of "acceptance criteria" and "study" described in your prompt, which are relevant to diagnostic or AI-driven medical devices.

    In summary, the provided text does not contain the information requested in your prompt regarding acceptance criteria and a study proving a device meets them in the context of AI or sophisticated diagnostic performance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192216

    Validate with FDA (Live)

    Date Cleared
    2020-01-02

    (140 days)

    Product Code
    Regulation Number
    890.3860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Invacare® AVIVA FX Power Wheelchair is indicated to provide mobility and positioning to persons limited to a sitting position.

    Device Description

    The subject device is a front wheel drive version of the existing previously cleared Invacare TDX SP2 (Center Wheel Drive) Power Wheelchair (K170507) with LiNX Electronics and Ultra Low Maxx Seating System. The subject version of the Invacare TDX SP2 Power Wheelchair consists of the following changes: Alternative Front Wheel Drive Base, Seating Brackets to connect the Ultra Low Maxx Seating System to the base, Wiring Harness to connect the LiNX controller, the AVIVA FX base and the Ultra Low Maxx Seating System. The Invacare® AVIVA FX Power Wheelchair is a 24V DC battery-powered, motor-driven wheelchair, utilizing the predicate device LiNX® Control System and Ultra Low Maxx Seating System. The subject device consists of a rigid or "non-folding" type power wheelchair base with front wheel drive with 2 casters in the rear and two anti-tippers in the front. It is powered by two 12-volt DC batteries and two 4-pole single stage drive motors.

    AI/ML Overview

    This document describes the Invacare® AVIVA FX Power Wheelchair (Model: IFX-20MP), a front-wheel-drive version of the previously cleared Invacare TDX SP2 Power Wheelchair, with LiNX Electronics and Ultra Low Maxx Seating System.

    Here's an analysis of the acceptance criteria and the study proving the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document primarily relies on demonstrating substantial equivalence to a predicate device (Invacare TDX SP2 Power Wheelchair, K170507) and adherence to recognized international and national standards for safety and performance. The acceptance criteria are implicitly derived from these standards and the characteristics of the predicate device.

    Acceptance Criteria CategorySpecific Criteria (from standards/predicate)Reported Device Performance (as demonstrated by testing)
    Indications for UseProvide mobility and positioning to persons limited to a sitting position (same as predicate).Met (subject device shares the same Indications for Use as the predicate).
    Material CompositionSame as predicate device.Met (subject device utilizes the same material composition).
    Technological CharacteristicsSimilar to predicate device, with differences not raising new questions of safety/effectiveness.Met (differences, primarily the front-wheel-drive base, were evaluated and found not to raise new safety/effectiveness concerns).
    Static StabilityAdherence to ISO 7176-1:2014.Demonstrated compliance through non-clinical testing.
    Dynamic StabilityAdherence to ISO 7176-2:2017.Demonstrated compliance through non-clinical testing.
    Effectiveness of BrakesAdherence to ISO 7176-3:2012.Demonstrated compliance through non-clinical testing.
    Energy Consumption (Theoretical Distance Range)Adherence to ISO 7176-4:2008.Demonstrated compliance through non-clinical testing.
    Dimensions, Mass, Maneuvering SpaceAdherence to ISO 7176-5:2008 / -7:1998.Demonstrated compliance through non-clinical testing.
    Maximum Speed, Acceleration, DecelerationAdherence to ISO 7176-6:2018.Demonstrated compliance through non-clinical testing (Maximum Speed 6.25mph, predicate 5mph/5.8mph).
    Static, Impact, Fatigue StrengthsAdherence to ISO 7176-8:2014.Demonstrated compliance through non-clinical testing.
    Climatic TestsAdherence to ISO 7176-9:2009.Demonstrated compliance through non-clinical testing.
    Obstacle Climbing AbilityAdherence to ISO 7176-10:2008.Demonstrated compliance through non-clinical testing.
    Power and Control SystemsAdherence to ISO 7176-14:2008.Demonstrated compliance through non-clinical testing.
    Information Disclosure, Documentation, LabelingAdherence to ISO 7176-15:1996.Demonstrated compliance through non-clinical testing.
    Resistance to Ignition of Postural Support DevicesAdherence to ISO 7176-16:2012.Demonstrated compliance through non-clinical testing.
    Wheeled Mobility Devices for Use as Seats in Motor VehiclesAdherence to ISO 7176-19:2008.Demonstrated compliance through non-clinical testing.
    Electromagnetic Compatibility (EMC)Adherence to ISO 7176-21:2008.Demonstrated compliance through non-clinical testing.
    Batteries and ChargersAdherence to ISO 7176-25:2013.Demonstrated compliance through non-clinical testing.
    Software Life CycleAdherence to IEC 62304:2006.Software Verification Testing performed.
    FlammabilityAdherence to CAL117:2013, Section 1 and EN 1021-2:2014.Demonstrated compliance through non-clinical testing.
    Wireless CoexistenceAdherence to ANSI C63.27.Demonstrated compliance through non-clinical testing.
    BiocompatibilityAdherence to AAMI/ANSI/ISO 10993-5:2009 (Cytotoxicity) and AAMI/ANSI/ISO 10993-10:2010 (Irritation).Demonstrated compliance through non-clinical testing.

    2. Sample size used for the test set and the data provenance:

    The document describes non-clinical laboratory testing. For such tests, the "test set" typically refers to the devices or components manufactured and tested according to the specified standards. The document doesn't explicitly state the exact number of units tested for each standard, but it implies that tests were conducted on the subject device (Invacare® AVIVA FX Power Wheelchair).

    • Sample Size: Not explicitly stated as a number of devices, but testing was performed on the "subject Invacare® AVIVA FX Power Wheelchair" (singular/representative unit for engineering testing).
    • Data Provenance: Non-clinical laboratory data generated to specifically evaluate the subject device against recognized standards. It is prospective testing designed to confirm compliance. The country of origin of the data is not specified but would typically be from accredited testing laboratories, likely within the manufacturer's region or designated by them.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This device is a powered wheelchair and its evaluation primarily relies on engineering and performance standards, not on interpretation of images or clinical diagnoses by human experts to establish "ground truth" in the medical imaging sense. Therefore, the concept of "experts establishing ground truth" as typically seen in AI/diagnostic device submissions using a test set of patient data, is not directly applicable here.

    Instead, the "ground truth" for the performance criteria is derived from:

    • The requirements set forth in the international and national standards (e.g., ISO, IEC, CAL, ANSI).
    • The published specifications and validated performance of the predicate device.

    The "experts" involved would be:

    • Engineers and quality assurance personnel from Invacare Corporation responsible for design, manufacturing, and internal testing.
    • Personnel at accredited testing laboratories who perform the tests according to the standards and provide certified results. These individuals would have expertise in mechanical engineering, electronics, software validation, biocompatibility, etc., relevant to the specific test being conducted. Their qualifications are inherent in their accreditation to perform such tests.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Adjudication methods like 2+1 or 3+1 are typically used in clinical studies where human readers interpret data (e.g., medical images) and their interpretations need to be reconciled to establish a consensus "ground truth." This is not relevant for the non-clinical performance and engineering tests described for this powered wheelchair.

    For the non-clinical tests, the "adjudication" is essentially:

    • Direct measurement and observation: The device either meets the quantitative specification (e.g., a stability angle, a speed, a dimension) or it does not.
    • Compliance to a specified test protocol: The testing laboratory follows a defined protocol from the standard, and the results are reported as compliant or non-compliant. Deviations and failures would be investigated and resolved by the manufacturer to achieve compliance.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No. An MRMC comparative effectiveness study is not mentioned and would not be applicable to a powered wheelchair. Such studies are typically conducted for diagnostic devices (e.g., AI in medical imaging) to assess how AI assistance impacts human reader performance. This submission focuses on the mechanical and electrical safety and performance of the wheelchair as a standalone mobility device.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    The concept of "standalone performance" for an "algorithm only" device is not directly applicable here in the same way it would be for an AI diagnostic algorithm. While the wheelchair does incorporate the LiNX Control System (which has software and control algorithms), its performance is evaluated as an integrated system (device + software) through the various ISO, IEC, CAL, and ANSI standards. The software verification testing (IEC 62304:2006) ensures the software itself functions correctly and safely within the device's overall design. The overall device's performance is tested, essentially, "without human-in-the-loop" in the sense that the tests are designed to objectively measure the device's inherent capabilities against a standard, rather than evaluating a human operator's improved performance with the device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for the evaluation of this powered wheelchair is primarily:

    • Industry and regulatory performance standards: International (ISO, IEC) and national (CAL, ANSI) standards define acceptable performance limits, test methods, and safety requirements.
    • Predicate device characteristics: The specifications and previously cleared performance of the Invacare TDX SP2 Power Wheelchair (K170507) serve as a benchmark for substantial equivalence.
    • Engineering specifications and design verification: Internal technical requirements define the expected operating parameters and safety features.

    8. The sample size for the training set:

    Not applicable. This submission is for a physical medical device (powered wheelchair), not an AI algorithm that requires a training set of data.

    9. How the ground truth for the training set was established:

    Not applicable. As there is no "training set," there is no "ground truth for the training set."

    Ask a Question

    Ask a specific question about this device

    K Number
    K182946

    Validate with FDA (Live)

    Date Cleared
    2018-12-11

    (49 days)

    Product Code
    Regulation Number
    890.3860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The indication for use of the Invacare® TDX® SP2-HD with Captain's Seat is to provide mobility and positioning to persons limited to a sitting position.

    Device Description

    The subject device is an update to the existing previously cleared Invacare® TDX® SP2 Power Wheelchair (K170507) with LiNX Electronics and Ultra Low Maxx Seating System. The updated subject version of the Invacare® TDX® SP2 Power Wheelchair has the following changes:

    • Increase in the weight capacity of the previously cleared device from 300 lb to 450 lb.
      Invacare® TDX® SP2 Power Wheelchair is a battery-powered, motor-driven powered wheelchair, controlled by the LiNX® control system with enhanced suspension and additional back, arm and leg rest types. The subject device is a rigid or "non-folding" type power wheelchair base with center-wheel drive capability, two casters in the rear and two casters in the front. It is powered by two 12-volt DC batteries and two 4-pole single stage drive motors. No new accessories were added to the new device.
    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a powered wheelchair, the Invacare® TDX® SP2-HD with Captain's Seat. This document focuses on demonstrating substantial equivalence to a previously cleared predicate device (Invacare® TDX® SP2 Power Wheelchair, K170507), primarily by performing non-clinical testing to meet recognized consensus standards.

    Here's an analysis of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document states that "The acceptance criteria for the full verification of the design and acceptance criteria for each section of the Rehabilitation Engineering and Assistive Technology Society of North America (RESNA) testing standard was met." However, it does not provide a specific table detailing the acceptance criteria for each test and the corresponding reported performance of the device.

    It lists the following design characteristics and notes some differences compared to the predicate device, which could be interpreted as performance metrics:

    DescriptionSubject Device Invacare® TDX® SP2-HD with Captain's SeatPredicate Device Invacare® TDX® SP2 Power WheelchairMet Acceptance Criteria (Implied)
    Weight Capacity450lb.300lb.Yes (improved)
    Speed5mph5mph or 5.8mphYes
    Range (GP24 Batteries)> 20.7 miles> 20.7 milesYes
    Stopping Distance45.7" to 69.3" (depending on chair configuration)45.7" to 69.3" (depending on chair configuration)Yes
    Incline CapabilityYes
    Ground Clearance> 2.5"> 2.5"Yes
    Obstacle Climbing (Forward)2.95"2.95"Yes
    Obstacle Climbing (Reverse)0.98"0.98"Yes
    Motor Gearbox Sound Level54dBa54dBaYes
    Base Width25.5"24" or 25.5" (depending on battery type)Yes
    Turning Diameter50.4" to 65.4" (depending on seat configuration)50.4" to 65.4" (depending on seat configuration)Yes

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document mentions "Design verification and validation testing was performed on the subject device," but it does not specify the sample size used for the test set. It describes non-clinical testing, implying a prospective testing approach on physical units of the device. There is no information provided about the country of origin of the data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This product is a powered wheelchair and the testing described is non-clinical, focusing on engineering and performance standards (RESNA). Therefore, the concept of "ground truth established by experts" in the context of medical image analysis or diagnostics does not apply to this type of device and study. The "ground truth" here is the defined performance parameters and limits specified by the RESNA standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. As described in point 3, this is non-clinical performance testing against established engineering standards, so there is no expert adjudication process for ground truth.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a powered wheelchair; it is not an AI-assisted diagnostic tool or system that would involve human readers or a MRMC study.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical product (a powered wheelchair), not an algorithm or AI system. The "standalone" performance here relates to the wheelchair's physical performance characteristics according to engineering standards.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" in this context is the performance requirements and limits specified by recognized consensus standards, specifically ANSI/RESNA WC-1 and WC-2 series. The device's performance was compared against these established engineering standards.

    8. The sample size for the training set

    Not applicable. This device is a physical powered wheelchair, not a machine learning or AI model that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. As described in point 8, there is no training set for a physical device like a powered wheelchair.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181090

    Validate with FDA (Live)

    Date Cleared
    2018-11-14

    (203 days)

    Product Code
    Regulation Number
    890.3850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Invacare® Solara® 3G Manual Wheelchair and Invacare® Solara® 3G Spree Manual Wheelchair are intended to provide mobility to adults limited to a sitting position.

    Device Description

    The Invacare® Solara® 3G Manual Wheelchair is an attendant operated and self-propelled, mechanical wheelchair for adults. The indication for use of the Invacare® Solara® 3G Manual Wheelchair is to provide mobility to adults limited to a sitting position. The indication for use of the Invacare® Solara® 3G Spree Manual Wheelchair is to provide mobility to person limited to a sitting position. These devices are rigid, "non-folding" type wheelchairs that incorporate a solid seating surface (seat pan).

    AI/ML Overview

    The Invacare® Solara® 3G and Solara® 3G Spree Manual Wheelchairs were evaluated for substantial equivalence to their predicate devices through non-clinical performance testing.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly list "acceptance criteria" against which performance is measured for each test, but rather states that design verification testing demonstrated the device met the requirements and was substantially equivalent. The performance data focuses on demonstrating equivalence to the predicate devices through adherence to ANSI/RESNA WC-1:2009 standards and other relevant tests.

    Test ConductedReported Device Performance
    ANSI/RESNA WC-1:2009 Section 1: Determination of Static StabilityDesign Verification Testing demonstrated that the subject Invacare® Solara® 3G Manual Wheelchairs (300/400 lbs) and Solara® 3G Spree Manual Wheelchair (200 lbs) "are substantially equivalent to the marketed predicate devices" regarding static stability.
    ANSI/RESNA WC-1:2009 Section 3: Determination of Effectiveness of BrakesDesign Verification Testing demonstrated that the subject devices "are substantially equivalent to the marketed predicate devices" regarding effectiveness of brakes.
    ANSI/RESNA WC-1:2009 Section 5: Determination of Dimensions, Mass and Maneuvering SpaceDesign Verification Testing demonstrated that the subject devices "are substantially equivalent to the marketed predicate devices" regarding dimensions, mass, and maneuvering space.
    ANSI/RESNA WC-1:2009 Section 7: Method of Measurement of Seating and Wheel DimensionsDesign Verification Testing demonstrated that the subject devices "are substantially equivalent to the marketed predicate devices" regarding measurement of seating and wheel dimensions.
    ANSI/RESNA WC-1:2009 Section 8: Requirements and Test Methods for Static, Impact and Fatigue StrengthsDesign Verification Testing demonstrated that the subject devices "are substantially equivalent to the marketed predicate devices" regarding static, impact, and fatigue strengths.
    ANSI/RESNA WC-1:2009 Section 15: Requirements for Information Disclosure, Documentation and LabelingDesign Verification Testing demonstrated that the subject devices "are substantially equivalent to the marketed predicate devices" regarding information disclosure, documentation, and labeling.
    ANSI/RESNA WC-1:2009 Section 13: Determination of Coefficient of Friction of Test SurfacesDesign Verification Testing demonstrated that the subject devices "are substantially equivalent to the marketed predicate devices" regarding coefficient of friction of test surfaces.
    ANSI/RESNA WC-1:2009 Section 11: Test DummiesUsed as part of the testing methodology for other sections where dummies are relevant (e.g., static stability, impact, fatigue strengths). Performance is implicitly covered by the "substantially equivalent" statements for those sections.
    CAL 117:2013, Section 1: Flammability TestBiocompatibility evaluation was conducted. (Assumed to meet standard if listed, no specific performance pass/fail provided).
    ISO 8191-1:1987 & 8191-2:1988: Flammability TestingBiocompatibility evaluation was conducted. (Assumed to meet standard if listed, no specific performance pass/fail provided).
    AAMI / ANSI / ISO 10993-5:2009: Tests for in vitro CytotoxicityBiocompatibility evaluation was conducted in accordance with FDA Blue Book Memorandum #G95 – 1 and ISO 10993-1. Specific results are not provided, but the statement implies the device passed these tests to meet "Biocompatibility Testing" requirements.
    AAMI / ANSI / ISO 10993-10:2010: Tests for Skin IrritationBiocompatibility evaluation was conducted in accordance with FDA Blue Book Memorandum #G95 – 1 and ISO 10993-1. Specific results are not provided, but the statement implies the device passed these tests to meet "Biocompatibility Testing" requirements.
    Risk Management (ISO 14971:2012)A Risk Assessment Device Hazard Analysis (DHA) was performed. "The subject device's risk profile remained unchanged and there is no significant impact on the safety or effectiveness of the subject device."

    2. Sample size used for the test set and the data provenance:

    • Sample Size: The document does not specify the exact number of devices or components tested. It generally refers to "the subject device" or "the subject Invacare® Solara® 3G Manual Wheelchair with weight capacities of 300 and 400 lbs. and Invacare® Solara® 3G Spree Manual Wheelchair with capacity of 200 lbs."
    • Data Provenance: The tests are non-clinical laboratory tests, likely conducted at an Invacare facility or a contracted testing lab. The location/country of origin for the testing is not specified, but the manufacturer is listed as INVAMEX S.DE R.L. DE C.V. in Reynosa Tamaulipas, Mexico. The nature of the tests (e.g., static stability, impact tests) indicates they are typically prospective, conducted specifically for the submission.

    3. Number of experts used to establish the ground truth for the test set and their qualifications:

    Not applicable. This is a medical device for mobility (manual wheelchair) and the testing is based on engineering and safety standards (e.g., ANSI/RESNA, ISO), not expert interpretation of medical images or conditions. The "ground truth" for these tests is the performance against specified physical and mechanical standards.

    4. Adjudication method for the test set:

    Not applicable. As this involves objective non-clinical performance testing against established standards, an adjudication method for expert consensus is not relevant. The pass/fail criteria are defined by the standards themselves.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a manual wheelchair, not an AI-powered diagnostic or assistive tool for human interpretation.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used:

    The "ground truth" for the non-clinical testing of the manual wheelchairs is the adherence to the objective performance specifications and safety requirements outlined in recognized national and international standards, such as ANSI/RESNA WC-1:2009, CAL 117:2013, ISO 8191, and ISO 10993. This is best described as objective engineering and safety standards conformance.

    8. The sample size for the training set:

    Not applicable. There is no AI model or algorithm being trained for this device. The "training set" concept does not apply to non-clinical performance testing of a physical medical device.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set for this type of device submission.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The indication for use of the Invacare® TDX® SP2 Power Wheelchair is to provide mobility and positioning to persons limited to a sitting position.

    Device Description

    The subject device is an update to the existing previously cleared Invacare® TDX® SP2 Power Wheelchair (K141783) with the MK6i™ control system and Formula™ CG Seating System. The updated subject version of the Invacare® TDX® SP2 Power Wheelchair has the following changes:

    • . The LiNX® control system, which incorporates updated software, wireless technology and a touch user interface,
    • The Ultra-Low Maxx™ Seating System, and ●
    • Minor changes to mechanical components such as rim inserts and new colours. .

    The TDX® SP2 Power Wheelchair is a battery-powered, motor-driven powered wheelchair, controlled by the LiNX® control system with enhanced suspension and additional back, arm and leg rest types. The subject device is a rigid or "non-folding" type power wheelchair base with centre-wheel drive capability, two casters in the rear and two casters in the front. It is powered by two 12-volt DC batteries and two 4-pole single stage drive motors.

    AI/ML Overview

    This document describes acceptance criteria and testing for the Invacare® TDX® SP2 Power Wheelchair.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document highlights conformity to various ISO, IEC, ANSI, CAL, and EN standards for different aspects of the wheelchair's hardware, software, and materials. Due to the nature of the device (a configurable power wheelchair) and the submission type (510(k) for substantial equivalence), the acceptance criteria are primarily described as meeting the requirements and passing the tests of these recognized standards. The reported device performance is generally stated as meeting these requirements and demonstrating substantial equivalence.

    Here's a summary derived from the document, focusing on categories rather than specific numerical values where not explicitly provided for the subject device in comparison tables:

    Test CategoryAcceptance Criteria (based on standards)Reported Device Performance
    Hardware Performance (Physical Characteristics)Conformity to relevant ISO 7176 standards (e.g., static/dynamic stability, braking, energy consumption, dimensions, speed, impact/fatigue strengths, climatic tests, obstacle climbing, power/control systems)Met all acceptance criteria for each section of ISO 7176
    Material PerformanceConformity to flammability standards (CAL117, EN 1021-1, EN 1021-2)Met relevant flammability standards
    Electromagnetic Compatibility (EMC)Conformity to ISO 7176-21, ANSI/RESNA WC-2:2009 Section 21 EMC requirementsMet relevant EMC requirements
    Wireless CoexistenceConformity to ANSIC63.27 Wireless Coexistence standardMet relevant wireless coexistence standard requirements
    Software Life CycleConformity to IEC 62304:2006 for medical device software life cycleEvaluated against IEC 62304:2006, passed verification
    Software FunctionalityFunctionality as designed, demonstrating safety and effectivenessSoftware verification testing confirmed functionality
    BiocompatibilityConformity to ISO 10993-5 (cytotoxicity) and ISO 10993-10 (skin irritation)Passed cytotoxicity and skin irritation tests
    Information Disclosure/LabelingConformity to ISO 7176-15 requirementsMet requirements for information disclosure/labeling
    Risk ManagementConducted in accordance with ISO 14971:2012Risk management conducted per ISO 14971:2012

    2. Sample Size Used for the Test Set and Data Provenance:

    The document states: "Because the subject Invacare® TDX® SP2 Power Wheelchair is highly configurable Power Wheelchair (PWC), intended to serve a diverse population with unique individual needs, a wide variety of options and accessories have been designed. However, due to the highly customizable nature of the subject device, testing to every possible combination would not be practical. However, every design feature of the subject device was verified. This was accomplished by identifying the specific wheelchair configurations that represent the full range of product permutations."

    • Sample Size: The exact numerical sample size for "test sets" (e.g., number of individual wheelchairs or components tested) is not explicitly provided. Instead, it indicates that "specific wheelchair configurations that represent the full range of product permutations" were identified and tested for each design feature. This implies a comprehensive, representative testing approach rather than a fixed "sample size" in the sense of clinical trials.
    • Data Provenance: The testing was "Non-clinical Test" and "Software Verification Testing" performed on the "subject Invacare® TDX® SP2 Power Wheelchair." This indicates that the data was generated through laboratory testing by the manufacturer (Invacare Corporation) as part of their design verification process. There is no mention of country of origin for the data or whether it was retrospective or prospective in the clinical sense, as it was non-clinical laboratory testing.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    This information is not applicable to this submission. The device is a physical product (power wheelchair) and the testing performed is non-clinical laboratory testing against established engineering and safety standards (ISO, IEC, etc.). The "ground truth" is defined by the technical specifications and performance limits set by these international standards, not by expert consensus on clinical diagnoses or interpretations.

    4. Adjudication Method for the Test Set:

    This information is not applicable. As the testing is non-clinical laboratory testing against defined standards, there is no ambiguity or subjective interpretation requiring adjudication by multiple parties or a specific method like 2+1 or 3+1. The tests yield objective results (e.g., pass/fail for a specified load, speed, or range).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for diagnostic devices where human interpretation (e.g., radiologists reading images) is involved, and the AI's assistance to human readers is being evaluated. This submission is for a physical medical device (power wheelchair) and involves non-clinical performance and safety testing.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

    While not an "algorithm only" in the sense of an AI diagnostic tool, the non-clinical tests performed on the physical wheelchair and its software are effectively "standalone" tests in that they evaluate the device's performance against defined criteria without active human intervention in the loop of the specific test itself (beyond setting up and observing the test). For example, stability tests, braking tests, and energy consumption tests are direct measurements of the device's inherent capabilities. The "Software Verification Testing" also falls into this category, assessing the software's functionality independently.

    7. The Type of Ground Truth Used:

    The ground truth used for these tests is objective performance specifications and safety requirements defined by internationally recognized standards (e.g., ISO 7176 series for wheelchairs, IEC 62304 for software, ISO 10993 for biocompatibility, flammability standards). The device's performance characteristics are measured and compared against the limits and criteria set forth in these standards.

    8. The Sample Size for the Training Set:

    This information is not applicable. This is a 510(k) submission for a physical medical device based on engineering and safety testing, not a machine learning or AI-driven diagnostic device that would typically involve a "training set" of data.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162621

    Validate with FDA (Live)

    Date Cleared
    2017-04-17

    (209 days)

    Product Code
    Regulation Number
    890.3850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Invacare® Tracer Manual Wheelchair, Invacare® Tracer® SX5 Manual Wheelchair, Invacare Tracer SX5 Recliner Manual Wheelchair, and Invacare® Tracer IV® Heavy Duty/Extra Wide Manual Wheelchair are intended to provide mobility to persons limited to a sitting position.

    Device Description

    The Invacare® Tracer® SX5 Manual Wheelchair is intended to provide mobility to persons who may be restricted to a sitting position. The manual wheelchair is intended for ongoing daily use. The models offer a wide range of customization which allows for a better fit for the end user resulting in comfort, lightweight, easier propelling, transferring, and smooth ride. The manual wheelchair is intended for indoor and outdoor use on firm surfaces free of climbing obstacles.

    The Invacare® Tracer® SX5 Recliner Manual Wheelchair is a Tracer SX5 model with the addition of a recline feature. This recliner model incorporates a mechanism that allows the backrest to recline in angles from 90° to 180°. Both models of the chair have a weight capacity of 250 to 300 lbs. depending on the seat width. Both models also come with an anti-tipper device to ensure the chair remains stable while in use.

    The Invacare® Tracer IV® Heavy Duty/Extra Wide Manual Wheelchair has the same intended use as the Invacare® Tracer® SX5 Manual Wheelchair. However, the Tracer IV does not come with a recliner model and has a higher weight capacity of 350 lbs. depending on the configuration. Both models of Tracer IV chairs come with a push-to-lock or pull-to-lock feature to engage and disengage the wheel locks. Both models also come with anti-tippers that allow clearance of 1.5 to 2 inches.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for three manual wheelchair models by Invacare Corporation: Invacare Tracer SX5 Manual Wheelchair, Invacare Tracer SX5 Recline Manual Wheelchair, and Invacare Tracer IV Heavy Duty/Extra Wide Manual Wheelchair. The document focuses on demonstrating substantial equivalence to predicate and reference devices, rather than establishing acceptance criteria and conducting a study to prove performance against those criteria in the context of an AI/algorithm-based medical device.

    Therefore, most of the requested information regarding acceptance criteria, study design for AI devices, sample sizes, expert involvement, ground truth establishment, MRMC studies, and standalone performance is not applicable to this document. This is because the submission is for a mechanical medical device (manual wheelchairs), not a software or AI-driven device.

    However, I can extract the relevant information from the document that pertains to the types of "performance testing" conducted to demonstrate substantial equivalence for these mechanical devices.

    Here's a breakdown based on the document's content:

    1. A table of acceptance criteria and the reported device performance:

    The document doesn't explicitly state "acceptance criteria" in a quantitative, pass/fail manner for each test. Instead, it states that "Verification Testing demonstrated that the subject Invacare® Tracer® SX5 Manual and Invacare® Tracer IV® Heavy Duty/Extra Wide Manual are substantially equivalent to the marketed predicate and reference devices." This implies that the devices met the performance standards outlined in the referenced ANSI/RESNA and ISO standards. The performance reporting is implicit; the tests were successfully passed, indicating conformance with the standards.

    Test Performed (Standard Referenced)Implied Acceptance Criteria (Based on passing the standard)Reported Device Performance
    ANSI / RESNA WC/Volume 1 2009, Section 1: Determination of Static StabilityMust meet the static stability requirements defined in the standard.Demonstrated substantial equivalence; tests were successful.
    ANSI / RESNA WC/Volume 1 2009, Section 3: Determination of Effectiveness of BrakesBrakes must meet the effectiveness requirements defined in the standard.Demonstrated substantial equivalence; tests were successful.
    ANSI / RESNA WC/Volume 1 2009, Section 5: Determination of Dimensions, Mass and Maneuvering SpaceDimensions, mass, and maneuvering space must conform to the standard's specifications or be within acceptable variance from predicate.Demonstrated substantial equivalence; tests were successful (implicitly, also supported by "design characteristics comparison").
    ANSI / RESNA WC/Volume 1 - 2009, Section 7: Measurement of Seating and Wheel DimensionsSeating and wheel dimensions must conform to the standard's specifications or be within acceptable variance from predicate.Demonstrated substantial equivalence; tests were successful (implicitly).
    ANSI / RESNA WC/Volume 1 2009, Section 8: Static, Impact and Fatigue Strengths TestsMust withstand static, impact, and fatigue stresses as defined in the standard.Demonstrated substantial equivalence; tests were successful.
    RESNA WC-1:2009 Section 11: Test Dummies(This section concerns the test apparatus, not a device performance metric itself. Implies proper use of test dummies.)N/A (Methodology, not a direct performance outcome)
    RESNA WC-1:2009 Section 13: Determination of Coefficient of Friction of Test Surfaces(This section concerns the test environment, not a device performance metric itself. Implies proper control of test surfaces.)N/A (Methodology, not a direct performance outcome)
    ANSI / RESNA WC/Volume 1 2009, Section 15: Requirements for Information Disclosure, Documentation and LabelingInformation disclosure, documentation, and labeling must comply with the standard.Demonstrated substantial equivalence; tests were successful.
    CAL117:2013, Section 1: Flammability TestingMust meet flammability resistance requirements defined in the standard.Tests were successful.
    ISO 8191-1:1987 & 8191-2:1988: Flammability TestingMust meet flammability resistance requirements defined in the standard.Tests were successful.
    Biocompatibility (AAMI / ANSI / ISO 10993-5:2009 & 10993-10:2010)Must demonstrate acceptable levels of cytotoxicity and skin irritation.Tests were successful.

    2. Sample sizes used for the test set and the data provenance:

    • Test Set Sample Size: The document does not specify the number of individual wheelchairs tested. It refers to "the subject Invacare® Tracer Manual Wheelchairs" and "the subject Invacare® Tracer® SX5 Manual and Invacare® Tracer IV® Heavy Duty/Extra Wide Manual," suggesting representative samples were tested as per the standards. Specific quantities are not provided.
    • Data Provenance: The testing was "Non-clinical laboratory testing" conducted on the subject devices. Location of testing and whether it was retrospective or prospective is not explicitly stated, but for a mechanical device undergoing performance testing, it would inherently be prospective testing of new models.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This is not applicable. The "ground truth" for mechanical performance tests like these is established by adherence to recognized international and national standards (e.g., ANSI/RESNA, ISO) and the physical properties and performance characteristics of the devices themselves, measured by accredited laboratories and engineers. It does not involve expert consensus in the way an AI diagnostic tool's ground truth would.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    This is not applicable. Adjudication methods like 2+1 or 3+1 are used for establishing ground truth in subjective assessments (e.g., medical image interpretation). Mechanical performance testing relies on objective measurements against established engineering standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable. MRMC studies are specific to AI-assisted diagnostic or interpretative tasks involving human readers. This submission is for mechanical manual wheelchairs.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    This is not applicable. This refers to an AI algorithm's performance independent of human input. The device is a mechanical wheelchair.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for these mechanical devices is established through adherence to recognized national and international performance standards (ANSI/RESNA, ISO, CAL117) via physical and material testing. For biocompatibility, the ground truth is established by the results of cytotoxicity and skin irritation tests conforming to ISO standards.

    8. The sample size for the training set:

    This is not applicable. "Training set" refers to data used to train machine learning models. This is a mechanical device.

    9. How the ground truth for the training set was established:

    This is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160630

    Validate with FDA (Live)

    Date Cleared
    2016-09-27

    (204 days)

    Product Code
    Regulation Number
    868.5440
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Invacare® Platinum™ Mobile Oxygen Concentrator is intended to provide supplemental oxygen to patients with respiratory disorders. The device can be used in a home, institution, vehicle, or other environments outside the home.

    Device Description

    The Invacare® Platinum™ Mobile Oxygen Concentrator is a portable oxygen concentrator that provides oxygen in pulsed demand flow dosages at settings of P1 through P4. The oxygen concentration level of the output gas ranges from 87% to 95.6%. The device is used with a nasal cannula to direct oxygen from the device to the patient. The Invacare® Platinum™ Mobile Oxygen Concentrator can be used in a home, institution, vehicle, or other environments outside the home. The device is not intended to be life supporting or life sustaining.

    AI/ML Overview

    This document is a 510(k) premarket notification for the "Invacare® Platinum™ Mobile Oxygen Concentrator." It asserts that the device is substantially equivalent to a legally marketed predicate device. The information provided heavily focuses on non-clinical performance and comparison to a predicate, rather than an AI/ML-driven device with detailed acceptance criteria for diagnostic performance outcomes.

    Given the nature of the document, the questions about acceptance criteria and study details for an AI/ML device are not directly applicable. This is a traditional medical device submission, not specifically one for software as a medical device (SaMD) or an AI/ML diagnostic tool. Therefore, many of the requested fields about AI/ML specific criteria (like "multi-reader multi-case comparative effectiveness study," "standalone algorithm performance," "sample size for training set," "number of experts," etc.) cannot be answered from this document.

    However, I can extract the closest analogous information from the provided text regarding performance testing and acceptance.

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of "acceptance criteria" against which specific numerical performance metrics are reported in the way demanded by an AI/ML diagnostic device (e.g., sensitivity, specificity, AUC). Instead, it lists various performance tests that were conducted to ensure the device's functionality, safety, and equivalence to the predicate. The implication is that successful completion of these tests, demonstrating that the device operates within expected parameters and similar to the predicate, serves as the acceptance basis for this type of device.

    Test CategorySpecific Tests PerformedAcceptance Criteria (Implied)Reported Device Performance (Implied)
    Respiratory Testing- Product Weight Test- Pulse Volume/Pulse Time/Trigger Sensitivity / Oxygen Purity Test- External Battery Charger Charge Time- Oxygen Sensor Accuracy- Oxygen Purity at AC Supply Mains Extremes- Oxygen Purity at DC Power Extremes- Oxygen Purity after Storage Thermal Cycling- Oxygen Purity at Low and High Temperature- Hot – Swapping Batteries- Battery Discharge Times- Battery Charge Times- Oxygen Purity at Low and High Altitude- Software Functional Test- Oxygen Alarms Test- Operational Alarms Test- Full Assembly Sound Performance Test- POC1 Battery Charge Time When POC1 Unit Not Running- Device consistently provides oxygen within specified ranges (87% to 95.6% purity).- Alarms function correctly (e.g., low oxygen purity, oxygen sensor failure, no breath detect indication).- Battery performance (charge/discharge) meets specifications.- Device maintains performance under various environmental conditions (temperature, altitude, power extremes).- Product weight is as specified.- Software functions as intended.- Pulse volume, pulse time, and trigger sensitivity are within specifications.- Output gas ranges from 87% to 95.6%.- Alarms for low oxygen purity and oxygen sensor failure are present. (The predicate device had "No breath detect" with specific actions, while the subject device lists "No breath detect" as an alarm, but without detailed actions, noting "No 'Auto-Pulse' feature").- Overall Conclusion: "The non-clinical laboratory data support the substantial equivalence of Invacare® Platinum™ Mobile Oxygen Concentrator and demonstrates that the subject device performs as intended in the specified use conditions." Specific numerical results for each test are not provided in this summary.
    Electrical Safety and EMC- AAMI ES60601-1- IEC 60601-1- ISO 80601-2-67- ISO 80601-2-69- IEC 60601-1-2- IEC 60601-1-6- IEC 62366- IEC 60601-1-8- IEC 60601-1-11- Compliance with all listed consensus standards for electrical safety, electromagnetic compatibility, usability, and alarm systems.- Overall Conclusion: "The non-clinical laboratory data support the substantial equivalence of Invacare® Platinum™ Mobile Oxygen Concentrator and demonstrates that the subject device performs as intended in the specified use conditions." Implies successful compliance with all standards.
    Software Verification Testing- Testing conducted as recommended by "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."- Software functions correctly and reliably, consistent with a "minor level of concern" where failure will not result in injury (as the device is not life-supporting).- Overall Conclusion: "The non-clinical laboratory data support the substantial equivalence of Invacare® Platinum™ Mobile Oxygen Concentrator and demonstrates that the subject device performs as intended in the specified use conditions." Implies successful verification.
    Biocompatibility Testing- AAMI / ANSI / ISO 10993-5 (in vitro Cytotoxicity)- AAMI / ANSI / ISO 10993-10 (Skin Irritation)- ISO 10993-11 (Systemic Toxicity)- Volatile Organic Chemicals Using the EPA TO-17 Method- Particulate Matter (PM)- Raw materials making contact with the patient meet biocompatibility standards as per ISO 10993 series and FDA guidance.- "The test results demonstrated that the subject device is biocompatible."

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in a detailed format within this 510(k) summary. For device testing (e.g., respiratory, electrical safety, biocompatibility), the "sample size" would typically refer to the number of device units tested. This is usually a small number (e.g., 3-5 units) sufficient to demonstrate manufacturing consistency and design performance. The document only states that "Respiratory testing was performed," "Electrical Safety and Electromagnetic Compatibility testing was performed," etc., without specifying the number of units or any "data provenance" like country of origin for the data. This type of testing is typically prospective, performed in a lab setting by the manufacturer, rather than on patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This device is a portable oxygen concentrator, not an AI/ML diagnostic device requiring expert interpretation for ground truth. The "ground truth" for this device's performance is established through physical measurements, adherence to engineering specifications, and compliance with recognized standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. No expert adjudication method is mentioned or required for the type of testing described (physical performance, electrical safety, biocompatibility).

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. No MRMC study was conducted. This is not an AI/ML diagnostic device for human reader assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm-only device. It's a hardware device with embedded software for control and monitoring. Its performance is inherent in the device's physical operation.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For this device, the "ground truth" is based on:

    • Physical Measurements and Engineering Specifications: Regarding oxygen purity, flow rates, battery life, weight, dimensions, alarm thresholds, etc.
    • Compliance with Recognized Standards: Such as those for electrical safety (IEC 60601 series), biocompatibility (ISO 10993 series), and software validation.
    • Functional Verification: Ensuring the device operates as designed under various conditions.

    8. The sample size for the training set

    Not applicable. This device does not use a "training set" in the context of machine learning or AI models. It's not an AI/ML device.

    9. How the ground truth for the training set was established

    Not applicable. As there is no training set for AI/ML, there is no ground truth established in that context.

    Ask a Question

    Ask a specific question about this device

    K Number
    K083751

    Validate with FDA (Live)

    Date Cleared
    2009-05-21

    (155 days)

    Product Code
    Regulation Number
    868.5440
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Invacare TPOC Portable Oxygen Concentrator is intended to be used by patients with respiratory disorders who require supplemental oxygen. It supplies a high concentration of supplemental oxygen and is used with a nasal cannula to channel oxygen from the concentrator to the patient. The Invacare TPOC Portable Oxygen Concentrator can be used in a home, institution, vehicle, and various mobile environments. The Invacare TPOC Portable Oxygen Concentrator does not nor is it intended to sustain or support life.

    Device Description

    The Invacare TPOC is to be used by patients with respiratory disorders who require supplemental oxygen. The device can be used in the home, an institution, vehicle and various mobile and outdoor environments. The device is not intended to sustain or support life. The device is typically used with a nasal cannula to direct oxygen from the device to the patient.

    When set for Continuous Flow Mode, the TPOC can deliver up to 3 LPM continuous flow in 0.5 LPM increments. While in Pulse Delivery Mode, when the demand for oxygen is detected, the oxygen is delivered through pulsed flow with pulse flow settings of 1 through 5. The oxygen concentration level of the output gas ranges from 87% to 95.6% in all modes.

    The modified TPOC contains an integrated oxygen sensor to provide added monitoring of the output gas purity, as well as additional patient warnings and alarms based on those levels. The oxygen sensor process uses a heated Zirconia element as the method of measurement. This is the exact same sensor process used in previously cleared oxygen concentrators such as the Invacare Platinum 5 Oxygen Concentrator, K020386.

    With the addition of continuous flow, the TPOC is now capable of supporting two features commonly with the addition of ocultified no ne nextrators; usc with a humidification and use in conjunction with PAP, Bi-Level and other such devices as an oxygen bleed-in. The use of this equipment is only during continuous flow mode, as indicated in the User's Manual. odulphileit is only aaning othium battery pack, standard power options include switching power Anong with a rents from AC power outlet (120 VAC/ 60 Hertz nominal) and from accessory DC outlets typically found in a mobile vehicle type environment (12 VDC nominal).

    The Invacare TPOC uses a molecular sieve and pressure swing adsorption methodology to produce the The in theater Tr C abber a nair enters the device, is filtered and then compressed. This compressed air is then directed toward one of two nitrogen adsorbing sieve beds. Concentrated oxygen exits the opposite end of the active sieve bed and is directed into an oxygen reservoir where it is delivered to the patient in specific volumes during the inhalation portion of a detected breath in pulse mode. In continuous flow mode, the operation is the same with the exception that the oxygen is delivered continuously, without the need for breath detection.

    The basic technology of the Invacare TPOC is equivalent to its predicate device.

    AI/ML Overview

    This document describes the Invacare Transportable Oxygen Concentrator (TPOC, Model TPOC100) and its 510(k) summary, which essentially acts as a claim for substantial equivalence to a predicate device, the Invacare Flyer Oxygen Concentrator (K0719028).

    There is no detailed study provided within the given text that explicitly describes acceptance criteria and then presents study results to prove those criteria were met for the Invacare TPOC.

    The document states: "Performance Testing: The performance testing conducted as a result of the modifications to the Invacare Flyer, as required by the risk analysis, was performed and the results demonstrated that the predetermined acceptance criteria were met." However, it does not provide the specifics of these acceptance criteria or the study details.

    Based only on the provided text, here's what can be extracted and what is missing:


    1. Table of Acceptance Criteria and Reported Device Performance

    Feature/ParameterAcceptance Criteria (Inferred/Missing Details)Reported Device Performance (Invacare TPOC100)
    Intended UseMatches predicate device.Matches predicate device (Invacare Flyer).
    Oxygen Purity"87% minimum at all flow rates""87% minimum at all flow rates"
    Flow RatesContinuous: 0.5 - 3.0 LPM (0.5 LPM resolution)Continuous: 0.5 - 3.0 LPM (0.5 LPM resolution)
    Pulse Flow: 1, 2, 3, 4, 5Pulse Flow: 1, 2, 3, 4, 5
    Method of Oxygen ReleaseMolecular sieve (mechanical)Molecular sieve (mechanical)
    Process of Oxygen ReleasePressure swing adsorptionPressure swing adsorption
    Sieve Bed MaterialSynthetic zeoliteSynthetic zeolite
    Software/HardwareAnalog and digital electronics with microprocessorAnalog and digital electronics with microprocessor
    Flow ControlMicroprocessor controlled valvesMicroprocessor controlled valves
    Oxygen SensorYes (integrated for added monitoring)Yes (integrated)
    Weight≤ 18.0 lbs. (without battery module or cart)≤ 18.0 lbs. (without battery module or cart)
    Power OptionsAC (100-240 V, 50/60 Hz), DC (11-16 V)AC (100-240 V, 50/60 Hz), DC (11-16 V)
    BatteryLi-ion, 16.8 Volt, 15.6 AhLi-ion, 16.8 Volt, 15.6 Ah

    Explanation: The "acceptance criteria" are implied by the performance characteristics listed for the device itself and against its predicate. The text explicitly states that "Performance Testing... was performed and the results demonstrated that the predetermined acceptance criteria were met," but it does not list those specific criteria or the detailed results of the tests. The table above largely compares the TPOC's specifications to the predicate and notes its own specifications as the "reported device performance."


    Regarding the study proving the device meets the acceptance criteria:

    The provided text only makes a general statement about performance testing, rather than describing a specific study with detailed methodology. Therefore, most of the following information is missing from the input text:

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Missing. The document does not specify any sample size for a test set or the provenance of data. The "performance testing" mentioned is generic.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable / Missing. This device is an oxygen concentrator, not an imaging or diagnostic device that would typically rely on expert human interpretation for ground truth establishment. Therefore, there would not be "experts" establishing ground truth in this context, but rather engineering/technical verification against specifications.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable / Missing. As above, this concept typically applies to studies where human reviewers are assessing data (e.g., medical images) and their agreement is critical for ground truth. It's not relevant to the performance testing of a physical medical device like an oxygen concentrator against its specifications.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This is not an AI-assisted diagnostic or interpretive device. An MRMC study would not be applicable to an oxygen concentrator.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not applicable. This isn't an algorithm; it's a hardware device with embedded software. Its performance would inherently be "standalone" in terms of its mechanical and electronic functions. The "standalone" concept typically applies to pure AI/algorithm performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Inferred: The "ground truth" for oxygen concentrators meeting performance criteria is based on direct physical measurements against engineering specifications and validated test methods (e.g., oxygen concentration using a sensor, flow rate measurement, weight measurement, battery life tests, safety tests). The document mentions an integrated oxygen sensor and "predetermined acceptance criteria" from risk analysis, which implies such direct measurements.

    8. The sample size for the training set

    • Not applicable/Missing. This device does not use a "training set" in the context of machine learning or AI. Performance testing involves demonstrating the device meets its specifications, not training an algorithm.

    9. How the ground truth for the training set was established

    • Not applicable/Missing. No training set is involved for this type of device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 3