Found 1 results

510(k) Data Aggregation

    K Number
    K963988

    Validate with FDA (Live)

    Date Cleared
    1996-12-12

    (69 days)

    Product Code
    Regulation Number
    870.1210
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Intravascular Infusion Device is indicated for the infusion of diagnostic agents into the general vasculature including the peripheral, coronary and neurovasculature.

    Device Description

    The Intravascular Infusion Device is a single lumen device constructed with progressively softer characteristics from proximal to distal end to aid in selective placement in the vasculature.

    AI/ML Overview

    This 510(k) premarket notification describes an Intravascular Infusion Device. It's important to note that this document is for a medical device that predates the widespread use of AI/ML in medical diagnostics, and thus, many of the questions related to AI/ML specific aspects (like multi-reader multi-case studies, standalone algorithm performance, training data sets, and ground truth establishment for AI) are not directly applicable.

    Here's an analysis based on the provided text, focusing on the device's functional and biocompatibility testing:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document lists various tests performed but does not explicitly state specific acceptance criteria (e.g., "Infusion rate must be X mL/min with Y% accuracy") or quantitative reported performance values (e.g., "Infusion rate achieved Z mL/min"). It only lists the types of tests conducted to demonstrate safety and performance.

    Acceptance Criteria (Not explicitly stated in the document)Reported Device Performance (Implied by successful testing)
    Adequate Infusion RateDevice passed Infusion Rate Testing
    Adequate Dynamic InfusionDevice passed Dynamic Infusion Testing
    Resistance to Static Burst PressureDevice passed Static Burst Testing
    Appropriate Tip FlexibilityDevice passed Tip Flexibility Testing
    Responsive TorqueDevice passed Torque Response Testing
    Sufficient Torsion StrengthDevice passed Torsion Strength Testing
    Sufficient Tensile StrengthDevice passed Tensile Strength Testing
    Adequate Coating IntegrityDevice passed Coatings Testing
    Biocompatibility in vivoDevice passed In vivo Testing
    Non-cytotoxicDevice passed Cytotoxicity testing
    Non-hemolyticDevice passed Hemolysis testing
    Non-acutely toxic systemicallyDevice passed Acute Systemic Toxicity testing
    Non-intracutaneous toxicDevice passed Intracutaneous Toxicity testing
    Non-sensitizingDevice passed Sensitization testing
    Non-pyrogenicDevice passed Pyrogenicity testing
    Compliant with USP Physiochemical standardsDevice passed USP Physiochemical Testing

    The conclusion states: "Based on the Indications for Use, technological characteristics and safety and performance testing, the Intravascular Infusion Device has been shown to be sate and effective for its intended use." This implies that the device met all unstated acceptance criteria for these tests.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify the sample size for any of the functional or biocompatibility tests.
    • Data Provenance: The tests are described as "Functional Testing" and "Biocompatibility Testing," implying lab-based, pre-clinical testing conducted by the manufacturer (Boston Scientific Corporation / SciMed Life Systems). The data is therefore prospective in nature, generated specifically for this premarket notification. The country of origin for the data generation would be where the manufacturer is located, which is USA (Maple Grove, MN).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This question is not applicable to this type of device and testing. Functional and biocompatibility tests for an intravascular catheter do not typically involve human experts establishing "ground truth" in the same way an AI diagnostic algorithm for image analysis would. The "ground truth" for these tests would be the measured physical properties or biological responses against predefined engineering specifications or recognized biological safety standards.

    4. Adjudication Method for the Test Set

    This question is not applicable. Adjudication methods like "2+1" or "3+1" are relevant for expert review in clinical studies or AI algorithm validation, not for functional engineering or biocompatibility testing of a physical medical device.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    This question is not applicable. MRMC studies are specific to evaluating human reader performance, particularly in diagnostic imaging, often in the context of AI tools. This 510(k) is for a physical intravascular infusion device manufactured in 1996 and does not involve AI.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This question is not applicable. This 510(k) describes a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used

    For the functional tests (e.g., Infusion Rate, Burst Testing, Tensile Strength), the "ground truth" would be established by engineering specifications and established test methodologies. For biocompatibility tests (e.g., Cytotoxicity, Hemolysis, Pyrogenicity), the "ground truth" is based on recognized biological safety standards and validated laboratory assays (e.g., ISO 10993 series and USP standards), where specific thresholds for toxicity, hemolytic activity, or pyrogenicity define "safe" or "acceptable" outcomes.

    8. The Sample Size for the Training Set

    This question is not applicable. This 510(k) is for a physical medical device and does not involve an AI algorithm with a training set.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable, as there is no AI algorithm training set for this device.